Transfusion equipment for medical use - Part 5: Transfusion sets for single use with pressure infusion apparatus

This document specifies requirements for single use transfusion sets for use with pressure infusion equipment capable of generating pressures. It ensures compatibility with containers for blood and blood components as well as intravenous equipment. This document also provides guidance on specifications relating to the quality and performance of materials used in transfusion sets, to present designations for transfusion set components, and to ensure the compatibility of sets with red cell and plasma blood components. NOTE In some countries, the national pharmacopoeia or other national regulations are legally binding and take precedence over this document.

Matériel de transfusion à usage médical — Partie 5: Transfuseurs non réutilisables avec des appareils de perfusion sous pression

Le présent document spécifie les exigences relatives aux transfuseurs non réutilisables, couplés à des appareils de perfusion sous pression capables de générer des pressions. Il assure la compatibilité avec les conteneurs de sang ou de composants sanguins, ainsi qu'avec le matériel d'injection intraveineuse. Le présent document fournit également des recommandations relatives aux spécifications de qualité et de performance des matériaux utilisés dans les transfuseurs, propose des désignations pour les composants de ces transfuseurs, et assure la compatibilité de ces appareils avec les concentrés de globules rouges et le plasma sanguin. NOTE Dans certains pays, la pharmacopée ou d'autres règlements nationaux sont légalement obligatoires et ont donc préséance sur le présent document.

General Information

Status
Published
Publication Date
08-May-2025
Current Stage
6060 - International Standard published
Start Date
09-May-2025
Due Date
20-Oct-2024
Completion Date
09-May-2025

Relations

Effective Date
06-Jun-2022

Overview

ISO 1135-5:2025 - Transfusion equipment for medical use - Part 5: Transfusion sets for single use with pressure infusion apparatus - defines requirements for single‑use transfusion sets intended to be used with pressure infusion equipment. The standard ensures mechanical and biological compatibility with containers for blood and blood components (red cells and plasma) and with intravenous equipment. It replaces ISO 1135-5:2015 and updates definitions, biological evaluation approaches, labelling and packaging references, and test descriptions for physical and chemical performance.

Key topics and technical requirements

  • Scope and intent: Applies to single‑use transfusion sets designed to operate under pressure (pressure infusion apparatus) and to ensure safe delivery of blood and blood components.
  • Designation and components: Standardizes component terminology (e.g., drip chamber, filter for blood, tubing, closure‑piercing device, injection site, male conical fitting, protective caps) to improve clarity and compatibility.
  • Materials and performance: Specifies material quality requirements and directs that components in contact with blood must meet additional chemical and biological requirements.
  • Physical requirements: Includes tests for particulate contamination, leakage, tensile strength, tubing characteristics, filters, drip chamber and drip tube, flow regulator and flow rates, injection sites, protective caps, and post‑occlusion bolus volume. Physical tests are detailed in Annex A.
  • Chemical requirements: Addresses extractables and residues, metal ions, titration acidity/alkalinity and UV absorption testing; normative chemical test procedures are provided in Annex B.
  • Biological requirements: Requires sterility and a biological risk assessment following ISO 10993‑1, with hemocompatibility considerations under ISO 10993‑4 and sample preparation per ISO 10993‑12.
  • Labelling, packaging and disposal: Labelling is aligned with ISO 15223‑1 symbol requirements; packaging follows ISO 11607‑1 guidance for terminally sterilized devices. The standard includes requirements for unit and multi‑unit containers and disposal instructions.
  • Annexes: Normative Annex A (physical tests), Annex B (chemical tests) and Annex C (determination of tube volumes).

Note: National pharmacopoeias or regulations may take precedence over the standard where legally binding.

Practical applications and users

  • Medical device manufacturers designing single‑use transfusion sets for use with pressure infusion systems - to ensure product safety, compatibility and regulatory alignment.
  • Regulatory bodies and notified bodies evaluating compliance for market approval and conformity assessment.
  • Clinical procurement and hospital biomedical engineering teams selecting transfusion sets compatible with blood components and pressure infusion devices.
  • Test laboratories performing physical, chemical and biological testing specified in the normative annexes.
  • Standards writers and quality managers integrating ISO 1135‑5:2025 requirements into design controls, risk management and labelling.

Related standards (selected)

  • ISO 1135-4 (gravity feed transfusion sets)
  • ISO 10993‑1, ISO 10993‑4, ISO 10993‑12 (biological evaluation)
  • ISO 15223‑1 (medical device symbols)
  • ISO 11607‑1 (packaging for terminally sterilized devices)
  • ISO 80369‑7 / ISO 80369‑20 (small‑bore connectors)

Keywords: ISO 1135-5:2025, transfusion sets, pressure infusion apparatus, single‑use, blood transfusion equipment, sterility, hemocompatibility, device labelling, packaging, ISO standards.

Standard

ISO 1135-5:2025 - Transfusion equipment for medical use — Part 5: Transfusion sets for single use with pressure infusion apparatus Released:9. 05. 2025

English language
20 pages
sale 15% off
Preview
sale 15% off
Preview
Standard

ISO 1135-5:2025 - Matériel de transfusion à usage médical — Partie 5: Transfuseurs non réutilisables avec des appareils de perfusion sous pression Released:9. 05. 2025

French language
21 pages
sale 15% off
Preview
sale 15% off
Preview

Frequently Asked Questions

ISO 1135-5:2025 is a standard published by the International Organization for Standardization (ISO). Its full title is "Transfusion equipment for medical use - Part 5: Transfusion sets for single use with pressure infusion apparatus". This standard covers: This document specifies requirements for single use transfusion sets for use with pressure infusion equipment capable of generating pressures. It ensures compatibility with containers for blood and blood components as well as intravenous equipment. This document also provides guidance on specifications relating to the quality and performance of materials used in transfusion sets, to present designations for transfusion set components, and to ensure the compatibility of sets with red cell and plasma blood components. NOTE In some countries, the national pharmacopoeia or other national regulations are legally binding and take precedence over this document.

This document specifies requirements for single use transfusion sets for use with pressure infusion equipment capable of generating pressures. It ensures compatibility with containers for blood and blood components as well as intravenous equipment. This document also provides guidance on specifications relating to the quality and performance of materials used in transfusion sets, to present designations for transfusion set components, and to ensure the compatibility of sets with red cell and plasma blood components. NOTE In some countries, the national pharmacopoeia or other national regulations are legally binding and take precedence over this document.

ISO 1135-5:2025 is classified under the following ICS (International Classification for Standards) categories: 11.040.20 - Transfusion, infusion and injection equipment. The ICS classification helps identify the subject area and facilitates finding related standards.

ISO 1135-5:2025 has the following relationships with other standards: It is inter standard links to ISO 1135-5:2015. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

ISO 1135-5:2025 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


International
Standard
ISO 1135-5
Second edition
Transfusion equipment for
2025-05
medical use —
Part 5:
Transfusion sets for single use with
pressure infusion apparatus
Matériel de transfusion à usage médical —
Partie 5: Transfuseurs non réutilisables avec des appareils de
perfusion sous pression
Reference number
© ISO 2025
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii
Contents Page
Foreword .iv
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 2
4 General requirements . 3
5 Materials . 4
6 Physical requirements . 4
6.1 General .4
6.2 Particulate contamination .5
6.3 Leakage .5
6.4 Tensile strength .5
6.5 Closure-piercing device .5
6.6 Tubing .5
6.7 Filter for blood and blood components .6
6.8 Drip chamber and drip tube .6
6.9 Flow regulator .6
6.10 Flow rate of blood and blood components .6
6.11 Injection site .6
6.12 Male conical fitting .7
6.13 Protective caps .7
6.14 Post-occlusion bolus volume .7
7 Chemical requirements . 7
7.1 General .7
7.2 Reducing (oxidizable) matter .7
7.3 Metal ions .7
7.4 Titration acidity or alkalinity.7
7.5 Residue on evaporation .7
7.6 UV absorption of extract solution .7
8 Biological requirements . 8
8.1 General .8
8.2 Sterility .8
8.3 Additional device specific requirements .8
9 Labelling . 8
9.1 General .8
9.2 Unit container .8
9.3 Shelf or multi-unit container .9
10 Packaging. 9
11 Disposal . 10
Annex A (normative) Physical tests . .11
Annex B (normative) Chemical tests .15
Annex C (normative) Determination of tube volumes . 17
Bibliography .20

iii
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO document should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent
rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a)
patent(s) which may be required to implement this document. However, implementers are cautioned that
this may not represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO's adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 76, Transfusion, infusion and injection, and blood
processing equipment for medical and pharmaceutical use, in collaboration with the European Committee for
Standardization (CEN) Technical Committee CEN/TC 205, Non-active medical devices, in accordance with the
Agreement on technical cooperation between ISO and CEN (Vienna Agreement).
This second cancels and replaces the first edition (ISO 1135-5:2015), which has been technically revised.
The main changes are as follows:
— the definitions of the different ‘volume’ terms have been amended;
— 6.11 “Injection site” has been amended regarding the use of needle-free injection ports and Luer-activated
devices;
— 6.13 “Protective caps” has been amended to clarify how to prevent contamination;
— 6.14 has been completely revised and renamed to clarify the described volume;
— Clause 8 has been revised to meet state-of-the-art methodology:
— biological risk assessment shall follow ISO 10993-1;
— sterility subclause remains;
— subclause on hemocompatibility assessment has been revised;
— Clause 9 “Labelling” has been updated especially regarding the referenced ISO 15223-1;
— Clause 10 “Packaging” has been amended by adding a reference to ISO 11607-1;
— Annex A “Physical test” has been amended by a general introduction on the pre-conditioning. In addition,
the description of the test for leakage has been extended;
— Annex C “Biological tests” has been deleted;

iv
— the former Annex D (now Annex C) has been revised; it has been renamed from "Storage volume" to
"Determination of tube volumes", and the subclause D.3 "Labelling" has been deleted;
— the Normative references have been updated.
A list of all parts in the ISO 1135 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.

v
International Standard ISO 1135-5:2025(en)
Transfusion equipment for medical use —
Part 5:
Transfusion sets for single use with pressure infusion
apparatus
1 Scope
This document specifies requirements for single use transfusion sets for use with pressure infusion
equipment capable of generating pressures. It ensures compatibility with containers for blood and blood
components as well as intravenous equipment.
This document also provides guidance on specifications relating to the quality and performance of
materials used in transfusion sets, to present designations for transfusion set components, and to ensure
the compatibility of sets with red cell and plasma blood components.
NOTE In some countries, the national pharmacopoeia or other national regulations are legally binding and take
precedence over this document.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes
requirements of this document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
ISO 1135-4, Transfusion equipment for medical use — Part 4: Transfusion sets for single use, gravity feed
ISO 3696, Water for analytical laboratory use — Specification and test methods
1)
ISO 3826-1:2019 , Plastics collapsible containers for human blood and blood components — Part 1: Conventional
containers
ISO 10993-1, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk
management process
ISO 10993-4:2017, Biological evaluation of medical devices — Part 4: Selection of tests for interactions with blood
ISO 10993-12, Biological evaluation of medical devices — Part 12: Sample preparation and reference materials
ISO 11607-1, Packaging for terminally sterilized medical devices — Part 1: Requirements for materials, sterile
barrier systems and packaging systems
ISO 14644-1, Cleanrooms and associated controlled environments — Part 1: Classification of air cleanliness by
particle concentration
ISO 15223-1, Medical devices — Symbols to be used with information to be supplied by the manufacturer — Part
1: General requirements
ISO 80369-7, Small-bore connectors for liquids and gases in healthcare applications — Part 7: Connectors for
intravascular or hypodermic applications
1) As impacted by ISO 3826-1:2019/Amd 1:2023

ISO 80369-20, Small-bore connectors for liquids and gases in healthcare applications — Part 20: Common
test methods
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at https:// www .electropedia .org/
3.1
filling volume
V
F
volume of tube during “pressureless” filling, respectively filling by gravity downstream of the infusion pump
Note 1 to entry: The tube remains unstressed without any additional pressure.
Note 2 to entry: The filling volume is to be equated with the calculated volume of the tube.
3.2
storage volume
V
S
tube volume during pressurization downstream of the infusion pump
3.3
post-occlusion bolus volume
V
B
increased tube volume downstream of the infusion pump during pressurization in comparison with the
unstressed tube
Note 1 to entry: V equals to V – V considering that for V and V the tubes have the same length.
B S F S F
Note 2 to entry: For illustration of the post-occlusion bolus volume, see Figure 1.

Key
1 patient 4 expanded tube with occlusion
2 occlusion 5 syringe pump
3 unstressed tube without occlusion
Figure 1 — Post-occlusion bolus volume
4 General requirements
4.1 The nomenclature for components of transfusion sets is given in Figure 2.

Key
1 protective cap of the closure-piercing device 8 flow regulator
2 closure-piercing device 9 injection site
3 fluid channel 10 male conical fitting
4 drip tube 11 protective cap of the male conical fitting
a
5 drip chamber Indicates alternative locations of the filter for blood
and blood components. Other designs are acceptable if
the same safety aspects are ensured.
b
6 filter for blood and blood components Injection site is optional.
7 tubing
Figure 2 — Example of a transfusion set
4.2 The transfusion set shall be provided with protective caps.
5 Materials
The materials from which the transfusion sets given in Clause 4 are manufactured shall conform with the
requirements specified in Clause 6. If components of the transfusion set come into contact with blood and
blood components, they shall additionally conform with the requirements specified in Clauses 7 and 8.
6 Physical requirements
6.1 General
Testing for physical requirements shall be carried out in accordance with Annex A.

6.2 Particulate contamination
The transfusion sets shall be manufactured under conditions that minimize particulate contamination. All
fluid contacting surfaces shall be smooth and clean. The number of particles detected shall not exceed the
contamination index limit. Test in accordance with A.2.
6.3 Leakage
The transfusion set, when tested in accordance with A.3, shall show no signs of air leakage.
6.4 Tensile strength
Any connections between the components of the transfusion set, excluding protective caps, shall withstand
a static tensile force of not less than 15 N for 15 s.
6.5 Closure-piercing device
6.5.1 The dimensions of the closure-piercing device shall conform to the dimensions shown in Figure 3.
NOTE The dimension of 15 mm in Figure 3 is a reference measurement. The cross-section of the piercing device at
this site is a circle.
Dimensions in millimetres
Figure 3 — Dimensions of the closure-piercing device
6.5.2 The closure-piercing device shall be capable of piercing and penetrating the closure of a container
for blood and blood components without pre-piercing. No coring shall occur during this procedure.
A carefully controlled surface treatment of the closure-piercing device (e.g. siliconization) is recommended
to facilitate its insertion into the blood bag port. The same effect can be achieved by a careful selection
of material for the closure-piercing device. Typical results including test equipment for penetration forces
between spikes and blood bag ports have been published. See References [13] and [14].
A closure-piercing device with a centred tip should be preferred to an asymmetric design as it enables easier
insertion.
6.5.3 The closure-piercing device shall resist a pull force of 15 N for 15 s after insertion into a blood bag
port that conforms with ISO 3826-1:2019.
6.5.4 The connection between the closure-piercing device and the blood bag port shall show no evidence
of leakage after insertion into a blood bag port that conforms with ISO 3826-1:2019, 5.3.
6.6 Tubing
6.6.1 The tubing, made of flexible material, shall be transparent or sufficiently translucent so that the
interface of air and water during the passage of air bubbles can be observed with normal or corrected-to-
normal vision.
6.6.2 The tubing from the distal end to the drip chamber shall be not less than 1 500 mm in length,
including the injection site, when provided, and the male conical fitting.
6.6.3 The tubing from the distal end to the drip chamber shall be capable of resisting (without collapsing)
negative pressures generated by the pressure infusion apparatus.
6.7 Filter for blood and blood components
The transfusion set shall be provided with a filter for blood and blood components. The filter shall have
uniform pores and shall cover a total area of not less than 10 cm . When tested in accordance with A.4 the
mass of solid material retained on the filter shall be not less than 80 % (mass fraction) of that retained on
the reference filter. In countries where human blood is not available for testing, an equivalent test method or
a substitute of human blood shall be established.
If the filter has a confirmed thread diameter of (100 ± 10) µm and a pore size of (200 ± 20) µm, with a single
warp and a single weft, a filtration performance test can be exempted.
Pore size measurement can be performed by microscopic inspection.
6.8 Drip chamber and drip tube
The drip chamber shall permit continuous observation of the fall of drops. The liquid shall enter the drip
chamber through a tube which projects into the chamber. There shall be a distance of not less than 40 mm
between the end of the drip tube and the outlet of the chamber, or a distance of not less than 20 mm between
the drip tube and the filter for blood and blood components. The wall of the drip chamber shall not be
closer than 5 mm to the end of the drip tube. The drip tube shall be such that 20 drops of distilled water at
(23 ± 2) °C and at a flow rate of (50 ± 10) drops/min deliver (1 ± 0,1) ml [(1 ± 0,1) g].
The drip chamber shall permit and facilitate the procedure of priming.
6.9 Flow regulator
The flow regulator shall adjust the flow of the blood and blood components between zero and maximum.
The flow regulator shall be capable of continuous use throughout a transfusion without the tubing being
damaged. There shall be no deleterious reaction between the flow regulator and the tubing when stored in
such a manner that there is contact.
6.10 Flow rate of blood and blood components
The transfusion set shall deliver not less than 1 000 ml of blood at (23 ± 2) °C in 30 min with a pressure
difference of 10 kPa. The transfusion set shall also deliver no less than 500 ml of blood in 2 min under a
pressure of 30 kPa above atmospheric pressure. In countries where human blood is not available for testing,
an equivalent test method or a substitute of human blood shall be established.
The blood shall be collected into a suitable anticoagulant solution and stored for not less than 2 weeks and
be free of large clots.
6.11 Injection site
When provided, the self-sealing injection site shall reseal when tested in accordance with A.5, and there
shall be no leakage of water.
Transfusion sets for use with pressure infusion apparatus shall not be fitted with an elastomeric buffer.
Needle-free injection ports or Luer-activated devices shall be used in accordance with the instructions
for use and tested to a minimum gauge pressure of 200 kPa unless otherwise require
...


Norme
internationale
ISO 1135-5
Deuxième édition
Matériel de transfusion à usage
2025-05
médical —
Partie 5:
Transfuseurs non réutilisables avec
des appareils de perfusion sous
pression
Transfusion equipment for medical use —
Part 5: Transfusion sets for single use with pressure infusion
apparatus
Numéro de référence
DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2025
Tous droits réservés. Sauf prescription différente ou nécessité dans le contexte de sa mise en œuvre, aucune partie de cette
publication ne peut être reproduite ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique,
y compris la photocopie, ou la diffusion sur l’internet ou sur un intranet, sans autorisation écrite préalable. Une autorisation peut
être demandée à l’ISO à l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Case postale 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Genève
Tél.: +41 22 749 01 11
E-mail: copyright@iso.org
Web: www.iso.org
Publié en Suisse
ii
Sommaire Page
Avant-propos .iv
1 Domaine d'application . 1
2 Références normatives . 1
3 Termes et définitions . 2
4 Exigences générales . . 4
5 Matériaux . 4
6 Exigences physiques . 5
6.1 Généralités .5
6.2 Contamination particulaire .5
6.3 Fuite .5
6.4 Résistance à la traction .5
6.5 Perforateur .5
6.6 Tubulure .6
6.7 Filtre pour le sang et ses composants .6
6.8 Chambre compte-gouttes et tube compte-gouttes .6
6.9 Régulateur de débit .6
6.10 Débit du sang et des composants sanguins .6
6.11 Site d'injection .7
6.12 Raccord conique mâle .7
6.13 Protecteurs .7
6.14 Volume de bolus post-occlusion .7
7 Exigences chimiques . 7
7.1 Généralités .7
7.2 Substances réductrices (ou oxydables) .7
7.3 Ions métalliques .7
7.4 Acidité ou alcalinité de titrage .8
7.5 Résidu après évaporation .8
7.6 Absorption UV de l'extrait .8
8 Exigences biologiques. 8
8.1 Généralités .8
8.2 Stérilité .8
8.3 Exigences supplémentaires spécifiques au dispositif .8
9 Étiquetage . 9
9.1 Généralités .9
9.2 Emballage individuel.9
9.3 Emballage multiple ou de protection .9
10 Emballage .10
11 Élimination.10
Annexe A (normative) Essais physiques .11
Annexe B (normative) Essais chimiques .16
Annexe C (normative) Détermination des volumes du tube .18
Bibliographie .21

iii
Avant-propos
L'ISO (Organisation internationale de normalisation) est une fédération mondiale d'organismes nationaux
de normalisation (comités membres de l'ISO). L'élaboration des Normes internationales est en général
confiée aux comités techniques de l'ISO. Chaque comité membre intéressé par une étude a le droit de faire
partie du comité technique créé à cet effet. Les organisations internationales, gouvernementales et non
gouvernementales, en liaison avec l'ISO participent également aux travaux. L'ISO collabore étroitement avec
la Commission électrotechnique internationale (IEC) en ce qui concerne la normalisation électrotechnique.
Les procédures utilisées pour élaborer le présent document et celles destinées à sa mise à jour sont
décrites dans les Directives ISO/IEC, Partie 1. Il convient, en particulier, de prendre note des différents
critères d'approbation requis pour les différents types de documents ISO. Le présent document
a été rédigé conformément aux règles de rédaction données dans les Directives ISO/IEC, Partie 2
(voir www.iso.org/directives).
L'ISO attire l'attention sur le fait que la mise en application du présent document peut entraîner l'utilisation
d'un ou de plusieurs brevets. L'ISO ne prend pas position quant à la preuve, à la validité et à l'applicabilité
de tout droit de propriété revendiqué à cet égard. À la date de publication du présent document, l'ISO
n'avait pas reçu notification qu'un ou plusieurs brevets pouvaient être nécessaires à sa mise en application.
Toutefois, il y a lieu d'avertir les responsables de la mise en application du présent document que des
informations plus récentes sont susceptibles de figurer dans la base de données de brevets, disponible à
l'adresse www.iso.org/brevets. L'ISO ne saurait être tenue pour responsable de ne pas avoir identifié tout ou
partie de tels droits de brevet.
Les appellations commerciales éventuellement mentionnées dans le présent document sont données pour
information, par souci de commodité, à l'intention des utilisateurs et ne sauraient constituer un engagement.
Pour une explication de la nature volontaire des normes, la signification des termes et expressions
spécifiques de l'ISO liés à l'évaluation de la conformité, ou pour toute information au sujet de l'adhésion de
l'ISO aux principes de l'Organisation mondiale du commerce (OMC) concernant les obstacles techniques au
commerce (OTC), voir www.iso.org/avant-propos.
Le présent document a été élaboré par le comité technique ISO/TC 76, Appareils de transfusion, de perfusion
et d'injection et appareils destinés au traitement du sang à usage médical et pharmaceutique, en collaboration
avec le comité technique CEN/TC 205, Dispositifs médicaux non actifs, du Comité européen de normalisation
(CEN), conformément à l'Accord de coopération technique entre l'ISO et le CEN (Accord de Vienne).
Cette deuxième édition annule et remplace la première édition (ISO 1135-5:2015), qui a fait l'objet d'une
révision technique.
Les principales modifications sont les suivantes:
— les définitions des différents termes de «volume» ont été modifiées;
— 6.11 «Site d'injection» a été modifié en ce qui concerne l'utilisation de ports d'injection sans aiguille et de
dispositifs activés par Luer;
— 6.13 «Protecteurs» a été modifié pour clarifier la manière de prévenir la contamination;
— 6.14 a été entièrement révisé et renommé pour clarifier le volume décrit;
— l'Article 8 a été révisé afin de satisfaire à la méthodologie de pointe:
— l'évaluation des risques biologiques doit être conforme à l'ISO 10993-1;
— le paragraphe relatif à la stérilité reste inchangé;
— le paragraphe relatif à l'évaluation de l'hémocompatibilité a été révisé;
— l'Article 9 «Étiquetage» a été mis à jour, notamment en ce qui concerne la norme référencée ISO 15223-1;
— l'Article 10 «Emballage» a été modifié par l'ajout d'une référence à l'ISO 11607-1;

iv
— l'Annexe A «Essai physique» a été modifiée par une introduction générale sur le préconditionnement. En
outre, la description de l'essai de fuite a été étendue;
— l'Annexe C «Essais biologiques» a été supprimée;
— l'ancienne Annexe D (désormais l'Annexe C) a été révisée; elle a été renommée de «Volume stocké» à
«Détermination des volumes de tubes», et l'Article D.3 «Étiquetage» a été supprimé;
— les références normatives ont été mises à jour.
Une liste de toutes les parties de la série ISO 1135 se trouve sur le site web de l'ISO.
Il convient que l'utilisateur adresse tout retour d'information ou toute question concernant le présent
document à l'organisme national de normalisation de son pays. Une liste exhaustive desdits organismes se
trouve à l'adresse www.iso.org/fr/members.html.

v
Norme internationale ISO 1135-5:2025(fr)
Matériel de transfusion à usage médical —
Partie 5:
Transfuseurs non réutilisables avec des appareils de
perfusion sous pression
1 Domaine d'application
Le présent document spécifie les exigences relatives aux transfuseurs non réutilisables, couplés à des
appareils de perfusion sous pression capables de générer des pressions. Il assure la compatibilité avec les
conteneurs de sang ou de composants sanguins, ainsi qu'avec le matériel d'injection intraveineuse.
Le présent document fournit également des recommandations relatives aux spécifications de qualité et de
performance des matériaux utilisés dans les transfuseurs, propose des désignations pour les composants
de ces transfuseurs, et assure la compatibilité de ces appareils avec les concentrés de globules rouges et le
plasma sanguin.
NOTE Dans certains pays, la pharmacopée ou d'autres règlements nationaux sont légalement obligatoires et ont
donc préséance sur le présent document.
2 Références normatives
Les documents suivants sont cités dans le texte de sorte qu'ils constituent, pour tout ou partie de leur
contenu, des exigences du présent document. Pour les références datées, seule l'édition citée s'applique. Pour
les références non datées, la dernière édition du document de référence s'applique (y compris les éventuels
amendements).
ISO 1135-4, Matériel de transfusion à usage médical — Partie 4: Transfuseurs non réutilisables, à alimentation
par gravité
ISO 3696, Eau pour laboratoire à usage analytique — Spécification et méthodes d'essai
1)
ISO 3826-1:2019 , Poches en plastique souple pour le sang et les composants du sang — Partie 1: Poches
conventionnelles
ISO 10993-1, Évaluation biologique des dispositifs médicaux — Partie 1: Évaluation et essais au sein d’un
processus de gestion du risque
ISO 10993-4:2017, Évaluation biologique des dispositifs médicaux — Partie 4: Choix des essais pour les
interactions avec le sang
ISO 10993-12, Évaluation biologique des dispositifs médicaux — Partie 12: Préparation des échantillons et
matériaux de référence
ISO 11607-1, Emballages des dispositifs médicaux stérilisés au stade terminal — Partie 1: Exigences relatives
aux matériaux, aux systèmes de barrière stérile et aux systèmes d'emballage
ISO 14644-1, Salles propres et environnements maîtrisés apparentés — Partie 1: Classification de la propreté
particulaire de l'air
1) Telle qu'impactée par l'ISO 3826-1:2019/Amd 1:2023

ISO 15223-1, Dispositifs médicaux — Symboles à utiliser avec les informations à fournir par le fabricant —
Partie 1: Exigences générales
ISO 80369-7, Raccords de petite taille pour liquides et gaz utilisés dans le domaine de la santé — Partie 7:
Connecteurs pour les applications intravasculaires ou hypodermiques
ISO 80369-20, Raccords de petite taille pour liquides et gaz utilisés dans le domaine de la santé — Partie 20:
Méthodes d'essai communes
3 Termes et définitions
Pour les besoins du présent document, les termes et définitions suivants s'appliquent.
L'ISO et l'IEC tiennent à jour des bases de données terminologiques destinées à être utilisées en normalisation,
consultables aux adresses suivantes:
— ISO Online browsing platform: disponible à l'adresse https:// www .iso .org/ obp
— IEC Electropedia: disponible à l'adresse https:// www .electropedia .org/
3.1
volume de remplissage
V
F
volume du tube durant le remplissage, respectivement « à pression nulle » ou par gravité, en aval de la pompe
à perfusion
Note 1 à l'article: Sans pression supplémentaire, le tube reste non contraint.
Note 2 à l'article: Le volume de remplissage doit être équivalent au volume calculé du tube.
3.2
volume stocké
V
S
volume du tube pendant la pressurisation en aval de la pompe à perfusion
3.3
volume de bolus post-occlusion
V
B
augmentation du volume du tube en aval de la pompe à perfusion lors de la pressurisation par rapport au
tube non contraint
Note 1 à l'article: V est égal à V – V en considérant que pour V et V les tubes ont la même longueur.
B S F S F
Note 2 à l'article: Le volume de bolus post-occlusion est illustré à la Figure 1.

Légende
1 patient 4 tube expansé avec occlusion
2 occlusion 5 pousse-seringue
3 tube non contraint sans occlusion
Figure 1 — Volume de bolus post-occlusion

4 Exigences générales
4.1 La nomenclature des composants des transfuseurs est indiquée à la Figure 2.
Légende
1 protecteur du perforateur 7 tubulure
2 perforateur 8 régulateur de débit
3 conduit de fluide 9 site d'injection
4 tube compte-gouttes 10 raccord conique mâle
5 chambre compte-gouttes 11 protecteur du raccord conique mâle
6 filtre pour le sang et ses composants
a
Indique les autres emplacements du filtre pour le sang et ses composants. D'autres configurations sont acceptables,
si les mêmes considérations de sécurité sont garanties.
b
Le site d'injection est facultatif.
Figure 2 — Exemple de transfuseur
4.2 Le transfuseur doit être fourni avec des protecteurs.
5 Matériaux
Les matériaux constituant les transfuseurs décrits à l'Article 4 doivent être conformes aux exigences
spécifiées à l'Article 6. Si des composants d'un transfuseur entrent en contact avec du sang ou ses composants,
ils doivent en outre être conformes aux exigences spécifiées aux Articles 7 et 8.

6 Exigences physiques
6.1 Généralités
L'essai des exigences physiques doit être réalisé conformément à l'Annexe A.
6.2 Contamination particulaire
Les transfuseurs doivent être fabriqués dans des conditions permettant de réduire au minimum la
contamination particulaire. Toutes les surfaces en contact avec le fluide doivent être lisses et propres.
Le nombre de particules détectées ne doit pas dépasser la limite de l'indice de contamination. Les essais
doivent être réalisés conformément à A.2.
6.3 Fuite
Lorsqu'il est soumis à essai conformément à A.3, le transfuseur ne doit présenter aucun signe de fuite d'air.
6.4 Résistance à la traction
Toutes les connexions entre les composants du transfuseur (à l'exception des protecteurs) doivent résister à
une force de traction statique d'au moins 15 N pendant 15 s.
6.5 Perforateur
6.5.1 Les dimensions du perforateur doivent être conformes celles présentées à la Figure 3.
NOTE La dimension de 15 mm de la Figure 3 est une mesure de référence. À cet endroit, la section transversale du
perforateur est circulaire.
Dimensions en millimètres
Figure 3 — Dimensions du perforateur
6.5.2 Le perforateur doit être capable de percer et de pénétrer le port d'un conteneur de sang et de
composants sanguins sans percement préalable. Aucun poinçonnage ne doit être effectué au cours de cette
procédure.
Il est recommandé de soumettre le perforateur à un traitement de surface minutieux (siliconage, par
exemple) afin de faciliter son insertion dans le port de la poche de sang. On peut également obtenir le même
effet en choisissant soigneusement le matériau du perforateur. Des résultats types, incluant l'équipement
d'essai et concernant les forces de pénétration des perforateurs dans les ports de poche de sang, ont été
publiés. Voir Références [13] et [14].
Il convient de privilégier un perforateur à pointe centrale à un dispositif à configuration asymétrique, car
cela facilite l'insertion.
6.5.3 Le perforateur doit résister à une force de traction de 15 N pendant 15 s après insertion dans un port
de poche de sang qui doit être conforme à l'ISO 3826-1:2019.

6.5.4 La connexion entre le perforateur et le port de la poche de sang ne doit présenter aucun signe de
fuite après insertion dans un port de la poche de sang qui est conforme à l'ISO 3826-1:2019, 5.3.
6.6 Tubulure
6.6.1 La tubulure, composée d'un matériau souple, doit être transparente ou suffisamment translucide
pour qu'il soit possible d'observer, en vision normale ou corrigée, l'interface air/eau pendant le passage de
bulles d'air.
6.6.2 La longueur de la tubulure entre l'extrémité distale et la chambre compte-gouttes ne doit pas être
inférieure à 1 500 mm, site d'injection (si présent) et raccord conique mâle inclus.
6.6.3 La tubulure entre l'extrémité distale et la chambre compte-gouttes doit être capable de résister
(sans s'écraser) aux pressions négatives générées par l'appareil de perfusion sous pression.
6.7 Filtre pour le sang et ses composants
Le transfuseur doit être pourvu d'un filtre pour le sang et les composants sanguins. Ce filtre doit présenter
des pores uniformes et couvrir une surface totale d'au moins 10 cm . Lors de l'essai conformément à A.4, la
masse de matériau solide retenue par le filtre ne doit pas être inférieure à 80 % (fraction massique) de celle
retenue par le filtre de référence. Dans les pays où le sang humain n'est pas disponible pour les essais, une
méthode d'essai équivalente ou un substitut de sang humain doit être établi.
Si le filtre présente un diamètre de fil confirmé de (100 ± 10) µm et une porosité de (200 ± 20) µm, avec une
seule chaîne et une seule trame, l'essai de performance de filtration peut être omis.
La porosité peut être mesurée par examen microscopique.
6.8 Chambre compte-gouttes et tube compte-gouttes
La chambre compte-gouttes doit permettre l'observation continue de la chute des gouttes. Le liquide doit
entrer dans la chambre compte-gouttes par l'intermédiaire d'un tube se prolongeant dans cette dernière.
Il doit y avoir une distance d'au moins 40 mm entre l'extrémité du tube compte-gouttes et la sortie de la
chambre, ou une distance d'au moins 20 mm entre le tube compte-gouttes et le filtre pour le sang et ses
composants. La paroi de la chambre compte-gouttes ne doit pas se trouver à moins de 5 mm de l'extrémité
du tube compte-gouttes. Le tube compte-gouttes doit être tel que 20 gouttes d'eau distillée à (23 ± 2) °C à un
débit de (50 ± 10) gouttes/min fournissent (1 ± 0,1) ml [(1 ± 0,1) g].
La chambre compte-gouttes doit permettre et faciliter la procédure d'amorçage.
6.9 Régulateur de débit
Le régulateur de débit doit permettre de régler le flux du sang et des composants sanguins entre zéro et le
débit maximal.
L'usage continu du régulateur de débit tout au long d'une transfusion ne doit pas endommager la tubulure.
Il ne doit y avoir aucune interaction délétère entre le régulateur de débit et la tubulure du fait d'un contact
prolongé pendant le stockage.
6.10 Débit du sang et des composants sanguins
Le transfuseur ne doit pas délivrer moins de 1 000 ml de sang à (23 ± 2) °C en 30 min avec une différence
de pression de 10 kPa. De même, le transfuseur ne doit pas délivrer moins de 500 ml de sang en 2 min à
une pression de 30 kPa au-dessus de la pression atmosphérique. Dans les pays où le sang humain n'est pas
disponible pour les essais, une méthode d'essai équivalente ou un substitut de sang humain doit être établi.
Le sang doit être recueilli dans une solution anticoagulante adaptée et stocké pendant au moins 2 semaines
sans qu'aucun caillot ne se forme.

6.11 Site d'injection
S'il est présent, le site d'injection auto-obturant doit se resceller lorsqu'il est soumis à essai conformément
à A.5; il ne doit présenter aucune fuite d'eau.
Les transfuseurs utilisés avec des appareils de perfusion sous pression ne doivent pas être dotés d'un
ballonnet en élastomère.
Les ports d'injection sans aiguille ou les dispositifs activés par Luer doivent être utilisés conformément
aux instructions d'utilisation et soumis à essai à une pression manométrique minimale de 200 kPa, sauf
indication contraire. Lorsque l'essai est effectué conformément au A.5, il ne doit pas y avoir de fuite d'eau, à
l'exception de l'utilisation d'un cône Luer mâle au lieu d'une aiguille.
Le site d'injection doit être placé à proximité du raccord conique mâle.
NOTE La co-administration de médicaments par le site d'injection n'est pas admise dans certains pays.
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...