Health Informatics - Application of clinical risk management to the manufacture of health software

ISO/TS 29321:2008 describes the risk management processes required to ensure patient safety in respect to the manufacture of health software products as defined in this document. It does not apply to software which is: necessary for the proper application of a medical device or which is an accessory to a medical device or which is a medical device in its own right. ISO/TS 29321:2008 applies to any health software product whether or not it is placed on the market as an off the shelf or configurable product and whether or not it is for sale or free of charge. It is addressed to all manufacturers of health software products as defined in this document.

Informatique de santé — Application de la gestion du risque clinique à la fabrication de logiciel de santé

General Information

Status
Withdrawn
Current Stage
5098 - Project deleted
Start Date
11-Jun-2010
Completion Date
13-Dec-2025
Ref Project

Relations

Draft
ISO/PRF TS 29321 - Health Informatics -- Application of clinical risk management to the manufacture of health software
English language
80 pages
sale 15% off
sale 15% off

Frequently Asked Questions

ISO/PRF TS 29321 is a draft published by the International Organization for Standardization (ISO). Its full title is "Health Informatics - Application of clinical risk management to the manufacture of health software". This standard covers: ISO/TS 29321:2008 describes the risk management processes required to ensure patient safety in respect to the manufacture of health software products as defined in this document. It does not apply to software which is: necessary for the proper application of a medical device or which is an accessory to a medical device or which is a medical device in its own right. ISO/TS 29321:2008 applies to any health software product whether or not it is placed on the market as an off the shelf or configurable product and whether or not it is for sale or free of charge. It is addressed to all manufacturers of health software products as defined in this document.

ISO/TS 29321:2008 describes the risk management processes required to ensure patient safety in respect to the manufacture of health software products as defined in this document. It does not apply to software which is: necessary for the proper application of a medical device or which is an accessory to a medical device or which is a medical device in its own right. ISO/TS 29321:2008 applies to any health software product whether or not it is placed on the market as an off the shelf or configurable product and whether or not it is for sale or free of charge. It is addressed to all manufacturers of health software products as defined in this document.

ISO/PRF TS 29321 is classified under the following ICS (International Classification for Standards) categories: 35.240.80 - IT applications in health care technology. The ICS classification helps identify the subject area and facilitates finding related standards.

ISO/PRF TS 29321 has the following relationships with other standards: It is inter standard links to ISO 10362-1:1999. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

You can purchase ISO/PRF TS 29321 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of ISO standards.

Standards Content (Sample)


TECHNICAL ISO/TS
SPECIFICATION 29321
First edition
2008-##-##
Health Informatics — Application of
clinical risk management to the
manufacture of health software
Informatique de santé — Application de la gestion du risque clinique à
la fabrication de logiciel de santé

PROOF/ÉPREUVE
Reference number
ISO/TS 29321:2008(E)
©
ISO 2008
ISO/TS 29321:2008(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.

©  ISO 2008
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii PROOF/ÉPREUVE © ISO 2008 – All rights reserved

ISO/TS 29321:2008(E)
Contents Page
Foreword. v
Introduction . vi
1 Scope . 1
2 Terms and definitions. 2
3 Abbreviated terms . 6
4 General requirements for effective clinical risk management . 7
4.1 Clinical risk management process. 7
4.2 Management responsibilities . 7
4.3 Competencies of personnel. 8
4.4 Clinical risk management planning . 8
4.5 Clinical risk management file . 9
4.6 Clinical safety case. 9
5 Clinical risk analysis. 10
5.1 Clinical risk analysis process. 10
5.2 Intended use and identification of characteristics related to the clinical safety of the
health software product . 10
5.3 Identification of hazards to patients . 10
5.4 Estimation of the clinical risk(s) to a patient for each hazardous situation. 11
6 Clinical risk evaluation. 11
7 Clinical risk control. 11
7.1 Clinical risk reduction . 11
7.2 Clinical risk control option analysis . 11
7.3 Implementation of clinical risk control measure(s). 12
7.4 Residual clinical risk evaluation . 12
7.5 Clinical risk/benefit analysis. 12
7.6 Clinical risks arising from clinical risk control measures. 13
7.7 Completeness of clinical risk control. 13
7.8 Evaluation of overall residual clinical risk acceptability . 13
8 Clinical safety case report(s). 13
9 Stage reports and pre-release clinical risk management process review. 15
10 Post-deployment monitoring. 15
11 Product modification. 16
12 Regular clinical risk management process review and maintenance. 16
13 Compliance with this Technical Specification. 17
Annex A (informative) Examples of potential harm presented by health software. 18
Annex B (informative) Conclusions of the CEN/TR measures for ensuring patient safety of health
software . 20
Annex C (informative) Decision support software. 22
Annex D (informative) Rationale for this Technical Specification . 24
Annex E (informative) Clinical risk management plan . 33
Annex F (informative) Components of a generic risk management process. 35
ISO/TS 29321:2008(E)
Annex G (informative) Relationship between clinical risk management file, clinical safety case,
clinical safety case reports, stage reports and product life-cycle. 39
Annex H (informative) Clinical risk estimation and evaluation guidance. 43
Annex I (informative) Risk control guidance . 50
Annex J (informative) Content of a clinical safety case report. 54
Annex K (informative) Example structures of reports within a clinical risk management process. 57
Bibliography . 79

iv PROOF/ÉPREUVE © ISO 2008 – All rights reserved

ISO/TS 29321:2008(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
In other circumstances, particularly when there is an urgent market requirement for such documents, a
technical committee may decide to publish other types of document:
⎯ an ISO Publicly Available Specification (ISO/PAS) represents an agreement between technical experts in
an ISO working group and is accepted for publication if it is approved by more than 50 % of the members
of the parent committee casting a vote;
⎯ an ISO Technical Specification (ISO/TS) represents an agreement between the members of a technical
committee and is accepted for publication if it is approved by 2/3 of the members of the committee casting
a vote.
An ISO/PAS or ISO/TS is reviewed after three years in order to decide whether it will be confirmed for a
further three years, revised to become an International Standard, or withdrawn. If the ISO/PAS or ISO/TS is
confirmed, it is reviewed again after a further three years, at which time it must either be transformed into an
International Standard or be withdrawn.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO/TS 29321 was prepared by Technical Committee ISO/TC 215, Health informatics in collaboration with
Technical Committee CEN/TC 251, Health informatics.
ISO/TS 29321:2008(E)
Introduction
The threat to patient safety
There is mounting concern around the world about the substantial number of avoidable clinical incidents which
have an adverse effect on patients of which a significant proportion result in avoidable death or serious
disability. See References [1], [2], [3], [4], [5] and [6]. A number of such avoidable incidents involved poor or
“wrong” diagnoses or other decisions. A contributing factor is often missing or incomplete information or
simply ignorance, e.g. of clinical options in difficult circumstances or of the cross-reaction of treatments.
It is increasingly claimed that information systems such as decision support, protocols, guidelines and
pathways could markedly reduce such adverse effects. If for no other reasons – and there are others – this
will lead, and is leading, to increasing utilization of decision support and disease management systems which
inevitably will increase in sophistication and complexity. It can also be anticipated that, due to pressures on
time and medico-legal aspects, clinicians will increasingly rely on such systems with less questioning of their
“output”. Indeed, as such systems become integrated with medical care any failure to use standard support
facilities may be criticised on legal grounds.
Increased decision support can be anticipated not only in clinical treatment but also in areas just as important
to patient safety, such as referral decision-making, where failure to make a “correct” referral or to make one “in
time” can have serious consequences.
Economic pressures are also leading to more decision support systems. The area of generic and/or economic
prescribing is the most obvious, but economy in number and costs of clinical investigative tests is another.
Systems such as those for decision support have considerable potential for reducing clinical errors and
improving clinical practice, e.g. in the reduction in errors resulting from the deployment of electronic
prescribing.
Thus decision support and IT in general can bring substantial benefit to patients. However, unless they are
safe and fit for purpose they may also present potential for harm.
Annex A provides some examples of the potential for harm of some health software systems.
Harm can of course result from unquestioning and/or non-professional use although designers and suppliers
can mitigate such circumstances through, for example, instructions for use, training and on-screen
presentation techniques, guidance or instruction. The potential for harm may equally lie in the system design
such as:
⎯ flaws in the requirement for the use of the system;
⎯ poor evidence base for design;
⎯ failure in design logic to properly represent design intentions;
⎯ failure in logic to represent good practice or evidence in the design phase;
⎯ poor or confusing presentation of information or poor search facilities;
⎯ failure to update in line with current knowledge.
Some of these system deficiencies are insidious and may be invisible to the user.
vi PROOF/ÉPREUVE © ISO 2008 – All rights reserved

ISO/TS 29321:2008(E)
Failures and deficiencies in health software products can, of course, have adverse impact other than causing
harm to patients. They may, for example, create administrative inconvenience or even administrative chaos,
with a range of impacts on the organization including financial loss. Harm to a patient may also have a
consequent impact on the organization such as financial loss resulting from litigation. Whereas these adverse
organizational impacts will be significant to an organization they are not the subject of this document unless
they result in harm to a patient. It is the potential harm to the patient which is the subject of this document.
Controlling the risks
The safety of medicines and of medical devices is ensured in many countries through a variety of legal and
administrative measures. In the European Union the safety of medical devices is subject to several EU
directives. See References [7], [8], [9] and [37]. These measures are often backed by a range of safety-related
standards from a number of sources, both national and international, including the International Organization
for Standardization (ISO), the International Electrotechnical Commission (IEC), the European Committee for
Standardization (CEN), the European Committee for Electrotechnical Standardization (CENELEC) and the
Global Harmonization Task Force (GHTF). Some software such as that necessary for the proper application
or functioning of a medical device is often encompassed by these legislative controls. Some software may be
considered a medical device in its own right. However other software applied to health of a stand-alone nature
is not usually covered or is encompassed in a less than clear manner or is not currently a primary focus of
some regulatory bodies. Depending on national regulations, examples might be general
practitioners'/physicians' computer systems, electronic health records, patient administrative systems,
applications of bar coding, for example to identify patients or medicinal products, a range of clinical decision
support software, ambulance dispatch systems, call and recall screening software. These matters are
[11]
complex and changing. For a full analysis see EN TR 15640 . This document is concerned with software
applied to health excluding that which is encompassed by medical device controls.
A necessary precursor for determining and implementing appropriate design and production controls to
minimize risks to patients from poor design, product malfunction or inadequate performance is a set of safety
requirements. These should be derived from an initial set of hazards and require a clear understanding of the
risks which a product might present to patients if malfunction or an unintended event were to occur, and the
likelihood of such a malfunction or event causing harm to the patient. Additionally, if guidance is to be given to
designers and producers of health software products as to design and production control (and corresponding
standards produced) then it will need to be recognised that the controls necessary for products presenting low
risks may not be the same, or applied with the same rigour, as for those presenting high risks. The controls
which are selected need to match both the level and types of risk which a product might present to a patient.
For these purposes many standards, legislation and specifications dealing with control of risks in design and
production, group products into a limited number of classes or types according to the risk they might present.
Controls are then tailored to the class or type. For medical devices such groupings are well established. For
[10]
health software in 2006 CEN published EN TS 15260 which, subject to validation of its risk classes in its
Table 4, could provide a process for grouping health software based on risk characteristics.
What control measures might be necessary for the safety of health software has been considered by
[11]
CEN/TC 251 in EN TR 15640 . The latter contains eleven conclusions which are reproduced in Annex B.
Conclusion 8 reads:
“If risk management is to be part of the requirements for ensuring the safety of health software products then:
[12] [13]
⎯ A new standard, consistent at a high level with the results of ISO/TMB JWG , ISO 14971 and
[14], [15]
ISO 61508 , is required specifically for health software products. That standard should embody the
[16] [17]
concepts in GHTF/SG3/NI5R8 and build on the experience of the use of CRAMM with ISO 17799
[18]
(now numbered ISO 27001:2006) .
⎯ The new standard should be backed by an implementation guide specific to health software products.”
In the document “Measures for ensuring patient safety of health software (APSOHIP): Proposed next
[19]
steps” , CEN/TC 251 considered this conclusion a priority. This standard addresses Conclusion 8.
ISO/TS 29321:2008(E)
Relationship to ISO 14971 and other safety related standards for medical devices
[13]
ISO 14971 is widely used throughout the world for compliance with medical device safety regulations.
Such regulations for medical devices in most countries encompass software that is necessary for the proper
application of a medical device or software that is an accessory to a medical device. In some jurisdictions,
medical device regulations also cover some other software. Thus medical device manufacturers have
considerable experience in the application of ISO 14971. Many manufacturers, particularly of electrical
medical devices, are experienced in the incorporation of software in medical devices, in producing software
supporting such medical devices and/or producing software that is a medical device in its own right. A number
of these manufacturers may also produce other health software of a type not encompassed by medical device
regulations. It would be advantageous to such manufacturers and any regulators if the standard for the
application of risk management to health software bore as close a relationship as practicable to ISO 14971.
This may in particular be an advantage in circumstances where software which is part of a medical device
complying with ISO 14971, interacts with software not controlled as a medical device but compliant with this
Technical Specification. Each may contribute a hazard to the other and thus access to the risk analysis for
both may be necessary.
For these reasons this Technical Specification takes as its baseline ISO 14971. As far as practicable and
appropriate the layout and requirements of the main text of ISO 14971 have been retained. However, most of
the annexes to ISO 14971 are clearly not applicable to health software (e.g. deal with biological hazards, in
vitro devices and characteristics of medical devices) and have therefore been replaced or amended as
appropriate.
Wherever appropriate this Technical Specification also takes account of work progressing in IEC/SC62A and
[20] [21]
ISO/TC 210 on a work item TR 80002 which itself has been drafted in the context of IEC 62304 . It also
[34]
takes note of the work item IEC 80001 .

viii PROOF/ÉPREUVE © ISO 2008 – All rights reserved

TECHNICAL SPECIFICATION ISO/TS 29321:2008(E)

Health Informatics — Application of clinical risk management to
the manufacture of health software
1 Scope
This Technical Specification describes the risk management processes required to ensure patient safety in
respect to the manufacture of health software products as defined in 2.17. It does not apply to software which
is:
⎯ necessary for the proper application of a medical device;
⎯ an accessory to a medical device;
⎯ a medical device in its own right.
This Technical Specification applies to any health software product whether or not it is placed on the market
as an off-the-shelf or configurable product and whether or not it is for sale or free of charge. It is addressed to
all manufacturers of health software products as defined in 2.17.
This Technical Specification does not cover the manufacture of non-health software which may be
incorporated in health software, for example OTS products such as operating systems, e.g. UNIX (Windows),
DBMS or SOUP products. However where a non-health software product such as an OTS or SOUP product is
incorporated by a manufacturer into a health software product, this Technical Specification shall apply to the
totality of that engineered product and shall include the non-health software product on which it is based.
NOTE 1 The scope is intended to cover health software products that are not controlled by medical device regulations.
It is acknowledged that, on the boundary, there are health software products that are encompassed by medical device
[11]
regulations in some countries but not in others. This matter is considered in detail in the CEN/TR 15640 .
NOTE 2 The life cycle of a health software product includes:
⎯ requirements capture and concept development;
⎯ detailed design;
⎯ production;
⎯ software release/marketing;
⎯ deployment;
⎯ use;
⎯ decommissioning.
A manufacturer is responsible for requirements capture and concept development, detailed design, production and
software release/marketing and can be responsible for deployments particularly the first “go live” of complex systems.
Where a customer contracts out responsibility for IT services to the manufacturer, the latter may also be responsible for
use of the application and its decommissioning. This Technical Specification applies to all the life-cycle phases for which
the manufacturer is responsible where this will depend on the contractual scope with the customer.
NOTE 3 Failures and deficiencies in software products used in the health environment can, of course, have adverse
impacts other than causing harm to patients. They might, for example, create administrative inconvenience with a range of
impacts on the organization including financial loss. Harm to a patient may also have a consequent impact on the
organization such as loss of reputation and/or financial loss resulting from litigation. Whereas these adverse organizational
ISO/TS 29321:2008(E)
impacts will be significant to an organization, they are not the subject of this Technical Specification unless they can result
in harm to a patient. It is the potential harm to the patient that is the subject of this Technical Specification.
NOTE 4 Whereas this Technical Specification might well be useful to regulators if health software products were to be
regulated or controlled in some formal or informal or voluntary manner whether national, regional or local, it is not the
purpose of this document to recommend whether or not health software products should be regulated.
NOTE 5 Guidance on the proper processes to be used by the user community to ensure the patient safety of health
[35]
software as it is deployed, is given in ISO/TR 29322 .
NOTE 6 Throughout this document the term “clinical” is used to make clear that the scope is limited to matters of risks
to patient safety as distinct from other types of risk such as financial. The use of the term “clinical” is not be taken to mean
that the manufacturer is expected to be involved in clinical decisions affecting the treatment of patients in the direct clinical
settings. This Technical Specification however makes clear that decisions about risks to patients posed by a health
software product in a clinical environment need to involve appropriate, experienced and knowledgeable clinicians.
2 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
2.1
clinical hazard
potential source of harm to a patient
[ISO/IEC Guide 51:1999, definition 3.5]
2.2
clinical risk
combination of the likelihood of occurrence of harm to a patient and the severity of that harm
NOTE Adapted from ISO/IEC Guide 51:1999 (definition 3.2).
2.3
clinical risk analysis
systematic use of available information to identify and estimate a risk
NOTE Adapted from ISO/IEC Guide 51:1999 (definition 3.10).
2.4
clinical risk assessment
overall process comprising a clinical risk analysis and a clinical risk evaluation
[ISO/IEC Guide 51:1999, definition 3.12]
2.5
clinical risk control
process in which decisions are made and measures implemented by which clinical risks are reduced to, or
maintained within, specified levels
2.6
clinical risk estimation
process used to assign values to the likelihood of occurrence of harm to a patient and the severity of that
harm
2.7
clinical risk evaluation
process of comparing the estimated clinical risk against given risk criteria to determine the acceptability of the
clinical risk
2 PROOF/ÉPREUVE © ISO 2008 – All rights reserved

ISO/TS 29321:2008(E)
2.8
clinical risk management
systematic application of management policies, procedures and practices to the tasks of analysing, evaluating
and controlling clinical risk
2.9
clinical risk management file
repository of all records and other documents that are produced by the clinical risk management process
2.10
clinical safety
freedom from unacceptable clinical risk to patients
NOTE Adapted from ISO/IEC Guide 51:1999 (definition 3.1).
2.11
clinical safety case
accumulation, through the life cycle of the health software system, of product and business process
documentation and of evidence structured such as to enable a safety argument to be developed to provide a
compelling, comprehensible and valid case that a system is, as far as the clinical risk management process
can realistically ascertain, free from unacceptable clinical risk for its intended use
2.12
clinical safety case report
report that summarises the arguments and supporting evidence of the clinical safety case at a defined point in
the health software's life cycle
2.13
clinical safety management system
organizational structure, processes, procedures and methodologies that enable the direction and control of the
activities necessary to meet clinical safety requirements and clinical safety policy objectives
2.14
OTS
off-the-shelf software that is not health software
2.15
harm
death, physical injury and/or damage to the health or well-being of a patient
NOTE Adapted from ISO/IEC Guide 51:1999.
2.16
hazardous situation
circumstance in which a patient is exposed to one or more hazard(s)
NOTE Adapted from ISO/IEC Guide 51:1999 (definition 3.6).
2.17
health software product
software product for use in the health sector for health related purposes but excluding software that is:
⎯ necessary for the proper application of a medical device;
⎯ an accessory to a medical device;
⎯ a medical device in its own right.
NOTE 1 This definition is intended for this Technical Specification only.
ISO/TS 29321:2008(E)
NOTE 2 For the purposes of this Technical Specification software includes firmware.
NOTE 3 This definition is intended to cover software products used in the health sector which are not covered by
medical device regulations. It is acknowledged that, on the boundary, there are health software products that are
encompassed by medical device regulations in some countries but not in others. This matter is considered in detail in
[11]
EN/TR 15640 .
2.18
intended use
use of a product, process or service in accordance with the specifications, instructions and information
provided by the manufacturer to customers
NOTE Information provided by the manufacturer includes information relevant to misuse as determined by the risk
management processes laid down in this Technical Specification.
2.19
life cycle
all phases in the life of a health software product, from the initial conception to final decommissioning and
disposal
2.20
manufacturer
natural or legal person with responsibility for the design, manufacture, packaging or labelling of a health
software product, assembling a system, or adapting a health software product before it is placed on the
market and/or put into service, regardless of whether these operations are carried out by that person or on
that person's behalf by a third party
NOTE A manufacturer can be involved in part of or the whole of the software life-cycle including deployment, use and
decommissioning of the software according to the contractual scope with the customer.
2.21
medical device
any instrument, apparatus, implement, machine, appliance, implant, in vitro reagent or calibrator, software,
material or other similar or related article, intended by the manufacturer to be used, alone or in combination,
for human beings for one or more of the specific purpose(s) of:
⎯ diagnosis, prevention, monitoring, treatment or alleviation of disease;
⎯ diagnosis, monitoring, treatment, alleviation of or compensation for an injury;
⎯ investigation, replacement, modification, or support of the anatomy of a physiological process;
⎯ supporting or sustaining life;
⎯ control of conception;
⎯ disinfection of medical devices;
⎯ providing information for medical purposes by means of in vitro examination of specimens derived from
the human body;
and which does not achieve its primary intended action in or on the human body by pharmacological,
immunological or metabolic means, but which may be assisted in its function by such means
NOTE This definition has been developed by the Global Harmonization Task Force (GHTF). However, definitions
vary from country to country and the extent to which software is covered by such definitions varies within regulatory
environments of different countries. This Technical Specification is intended to cover health software products that are not
covered by medical device regulations (see Clause 1). It is acknowledged that, on the boundary, there are health software
products that are encompassed by medical device regulations in some countries but not in others. This matter is
[11]
considered in detail in the EN/TR 15640 .
4 PROOF/ÉPREUVE © ISO 2008 – All rights reserved

ISO/TS 29321:2008(E)
2.22
objective evidence
data supporting the existence or verity of something
NOTE Objective evidence can be obtained through observation, measurement, testing or other means.
[ISO 9000:2005, definition 3.8.1]
2.23
patient
any person who is subject to a health software product
NOTE In this document that shall be taken to include healthy persons where applicable (e.g. a healthy person
accessing a knowledge data base to obtain health-related information).
2.24
pre-release stage report
stage report produced and signed off by top management, before a health software product is released for
distribution or deployment
2.25
post-production
part of the life cycle of the product after the design has been completed and the health software product has
been manufactured, prior to its release for use
2.26
post-deployment
part of the life cycle of the health software product after it has been manufactured, released and deployed
ready for use
2.27
procedure
specified way to carry out an activity or a process
[ISO 9000:2000, definition 3.4.5]
2.28
process
set of interrelated or interacting activities which transforms inputs into outputs
[ISO 9000:2000, definition 3.4.1]
2.29
product
entire entity of software proffered by a manufacturer to a user including instructions for use and, where
applicable, training and other such related services
2.30
product liability
legal liability incurred by a manufacturer, merchant or distributor as a result of injury or damage resulting from
the use of a product
NOTE In many countries there is strict liability namely, in essence, the plaintiff in litigation needs only to prove a link
to his/her injury or damage and a defective product to be successful. In the EU these matters are addressed in the EU
[33].
Directive 85/374/EEC
2.31
record
document stating results achieved or providing evidence of activities performed
[ISO 9000:2000, definition 3.7.6]
ISO/TS 29321:2008(E)
2.32
residual clinical risk
clinical risk remaining after risk control measures have been taken
[30]
NOTE ISO/IEC Guide 51:1999 , definition 3.9 uses the term “protective measures” rather than “risk control
measures”. However, in the context of this Technical Specification, “protective measures” are only one option for
controlling risk as described in 7.2.
2.33
severity
measure of the significance of the possible consequences of a hazard
2.34
stage report
report of a review of the clinical risk management process at a defined stage, to ensure all that is required to
be done at that stage has been done, as defined in the clinical risk management plan
2.35
top management
person or group of people who directs and controls a manufacturer at the highest level
NOTE Adapted from ISO 9000:2000 (definition 3.2.7).
2.36
verification
confirmation, through the provision of objective evidence, that specified requirements have been fulfilled
[ISO 9000:2000, definition 3.8.4]
NOTE 1 The term “verified” is used to designate the corresponding status.
NOTE 2 Confirmation can comprise activities such as:
⎯ performing alternative calculations;
⎯ comparing a new design specification with a similar proven design;
⎯ undertaking tests and demonstrations;
⎯ reviewing documents prior to issue;
⎯ checking that requirements have been addressed.
3 Abbreviated terms
For the purposes of this document, the following abbreviations apply.
EU European Union
DBMS Database Management System (software)
GHTF Global Harmonization Task Force
GP General practitioner
OTS off the shelf
SOUP Software of Unknown Provenance
6 PROOF/ÉPREUVE © ISO 2008 – All rights reserved

ISO/TS 29321:2008(E)
4 General requirements for effective clinical risk management
4.1 Clinical risk management process
The manufacturer shall establish, document and maintain throughout the life cycle an ongoing process for
identifying clinical hazards associated with a health software product, estimating and evaluating the
associated clinical risks, controlling these risks, and monitoring the effectiveness of the controls throughout
the life cycle. This process shall include the following elements:
⎯ context, requirements and scope identification;
⎯ creation of clinical risk management plan;
⎯ setting the requirements for and defining the competencies of personnel;
⎯ clinical hazard identification;
⎯ clinical risk analysis;
⎯ clinical risk evaluation;
⎯ clinical risk control;
⎯ residual clinical risk acceptance;
⎯ creation of clinical safety case report(s);
⎯ post deployment monitoring;
⎯ post-production maintenance of clinical risk management process.
Annex F gives an example of the necessary components of a generic risk management process.
4.2 Management responsibilities
Top management shall provide evidence of its commitment to the clinical risk management process by:
⎯ ensuring the provision of sufficient resources;
⎯ ensuring the assignment of suitably qualified and experienced personnel (see 4.3) for clinical risk
management.
NOTE 1 It is good practice for the top management team to appoint a suitably and sufficiently independent safety
function to oversee the effective operation of risk management practices.
Top management shall:
⎯ define and document the organization's risk management policy including criteria for establishing clinical
risk acceptability; this policy shall ensure, where applicable, that criteria are based upon national or
regional regulations and relevant International Standards, and take into account available information
such as the generally accepted state of the art and known stakeholder concerns;
⎯ ensure that a suitably staged approach is taken for the life cycle of the manufacturer's product such that
the risk management process can be efficiently and effectively applied; at each stage, top management
shall sign off the appropriate stage report (see Clause 9);
ISO/TS 29321:2008(E)
⎯ review the suitability of the clinical risk management process at planned, regular intervals to ensure
continuing effectiveness of the clinical risk management process and document any decisions and
actions taken.
This review should be linked to the manufacturer’s clinical safety management system which in turn may be
part of its quality management system or its enterprise risk management system.
NOTE 2 The documents can be incorporated within the documents produced by the manufacturer’s quality
management system and these documents can be referenced in the clinical risk management file.
4.3 Competencies of personnel
Persons performing risk management tasks shall have the knowledge, experience and competencies
appropriate to the tasks assigned to them. This shall include, where appropriate, knowledge and experience of
the particular health software product (or similar health software products) and all its applications, the
technologies involved or risk management techniques. This should include an appropriate registered clinician.
Appropriate competency and experience records shall be maintained.
NOTE Clinical risk management tasks can, and should, be performed by representatives of several functions, each
contributing their specialist knowledge. Of particular importance will be clinical input from clinicians who are familiar with
the practical realities of the environments within which the product will be used and the clinical processes to which the
product is directed.
4.4 Clinical risk management planning
Clinical risk management activities shall be planned. Therefore, for the particular health software product
being considered, the manufacturer shall establish and document a clinical risk management plan in
accordance with the clinical risk management process. The clinical risk management plan shall be the part of
the clinical risk management file.
This plan shall include at least the following:
⎯ the scope of the planned clinical risk management activities, identifying and describing the health
software product, the clinical context in which it will be used and the life-cycle phases for which each
element of the plan is applicable and the life-cycle phases which the plans covers;
⎯ assignment of responsibilities and authorities;
⎯ requirements for review of clinical risk management activities;
⎯ identification of relevant risk management procedures and processes to be used;
⎯ criteria to be used in analysis of the risks;
⎯ criteria for assessing clinical risk acceptability, based on the manufacturer’s policy for determining
acceptable clinical risk, including criteria for accepting clinical risks when the likelihood of occurrence of
patient harm cannot be estimated;
⎯ verification activities;
⎯ activities related to collection and review of relevant production, post-production and post-deployment
information.
NOTE 1 Refer to Annex E for guidance on developing a clinical risk management plan.
NOTE 2 The clinical risk management plan will be the first component of the clinical risk management file although not
all parts of the plan need to be created at the same time. The plan or parts of it can be developed over time as the relevant
risk management process requirements become better understood.
8 PROOF/ÉPREUVE © ISO 2008 – All rights reserved

ISO/TS 29321:2008(E)
NOTE 3 The criteria for clinical risk acceptability are essential for the ultimate effectiveness of the clinical risk
management process (See Annexes F and H).
NOTE 4 Clinical risk management planning needs to be based on a clear understanding of the context in which the
health software is intended to be used.
If the plan changes during the life cycle of the health software product, a record of the changes shall be
maintained in the clinical risk management file.
4.5 Clinical risk management file
For the particular health software product being considered, the manufacturer shall establish and maintain a
clinical risk management file that shall be the repository of the documentation produced and shall provide
suitable traceability. In addition to the requirements of other clauses of this Technical Specification, the clinical
risk management file shall provide traceability for each identified hazard to:
⎯ the requirements and objectives of the software development;
⎯ the clinical risk analysis;
⎯ the clinical risk evaluation;
⎯ the implementation and verification of the clinical risk control measures;
⎯ the assessment of the acceptability of any residual clinical risk(s).
NOTE 1 The records and other documents that make up the clinical risk management file can form part of other
documents and files required, for example, by a manufacturer’s clinical safety management system and/or quality
management system. The clinical risk management file need not physically contain all the records and other documents;
however, effective compliance will require it to contain at least references or pointers to all required documentation and
version control references. The manufacturer should be able to assemble the information referenced in the risk
management file in a timely fashion.
NOTE 2 The clinical risk management file can be in any form or type of medium.
NOTE 3 For the relationship between clinical risk management file, clinical safety case, clinical safety case report and
clinical risk management close-out report and product life-cycle, see Annex G.
4.6 Clinical safety c
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...