ISO 1135-3:2016
(Main)Transfusion equipment for medical use — Part 3: Blood-taking sets for single use
Transfusion equipment for medical use — Part 3: Blood-taking sets for single use
ISO 1135-3:2016 specifies requirements for types of blood-taking sets for medical use in order to ensure functional interchangeability of transfusion equipment. It is applicable to sterilized blood-taking sets intended for single use only. ISO 1135-3:2016 also aims to provide a) specifications relating to the quality and performance of materials used in transfusion equipment, and b) a unified presentation of terms for such equipment. In some countries, the national pharmacopoeia or other national regulations are legally binding and take precedence over ISO 1135-3:2016.
Matériel de transfusion à usage médical — Partie 3: Appareils non réutilisables pour prélèvement sanguin
ISO 1135-3:2016 fixe les spécifications des divers types d'appareils pour prélèvement sanguin à usage médical afin d'assurer I'interchangeabilité fonctionnelle du matériel de transfusion. Elle s'applique aux appareils stériles non réutilisables pour prélèvement sanguin. ISO 1135-3:2016 a également pour but de fournir a) des spécifications relatives à la qualité et aux performances des matériaux utilisés pour le matériel de transfusion; et b) une terminologie uniforme concernant ce type de matériel. Dans certains pays, la pharmacopée nationale ou d'autres règlements nationaux sont juridiquement contraignants et prévalent sur les dispositions de l'ISO 1135-3:2016.
General Information
- Status
- Published
- Publication Date
- 20-Oct-2016
- Current Stage
- 9093 - International Standard confirmed
- Start Date
- 11-Apr-2022
- Completion Date
- 12-Feb-2026
Relations
- Effective Date
- 12-Feb-2026
- Effective Date
- 14-Jun-2014
Overview
ISO 1135-3:2016 - "Transfusion equipment for medical use - Part 3: Blood-taking sets for single use" defines requirements for sterilized, single‑use blood‑taking sets to ensure functional interchangeability of transfusion equipment. The standard covers design, materials, sterilization, physical/chemical/biological performance, labelling, packaging and disposal. It is intended to harmonize terminology and define testable criteria so manufacturers, regulators and users can trust compatibility and safety of blood‑collection sets.
Key topics and technical requirements
- Scope and intent
- Applies to sterilized blood‑taking sets intended for single use only.
- Aims to specify material quality/performance and unify terms; national regulations or pharmacopoeias may take precedence.
- Component definitions
- Two main assemblies: blood‑taking assembly (blood‑taking needle + bottle needle + tubing) and air‑outlet assembly (air filter housing + air‑outlet needle).
- Protective caps required to preserve sterility of needles and internal surfaces.
- Physical requirements
- Minimize particulate contamination; surfaces of fluid pathways must be smooth and clean (particulate limits and test procedures in Annex A).
- No air leakage; connections (excluding caps) must withstand a static tensile force ≥ 15 N for 15 s.
- Dimensional and performance limits:
- Bottle needle: length ≥ 35 mm, external diameter ≥ 1.8 mm, internal diameter ≥ 70% of external.
- Blood‑taking needle: length ≥ 35 mm, external diameter ≤ 2 mm, internal diameter ≥ 70% of external.
- Air‑outlet needle: internal diameter ≥ 0.7 mm, external diameter ≤ 1.9 mm, length ≤ 25 mm.
- Tubing: internal diameter ≥ 2.7 mm, length ≥ 600 mm, flexible and kink‑resistant.
- Flow regulator must adjust flow from zero to maximum and withstand continuous use without damaging tubing.
- Chemical and biological
- Chemical extractables limits (e.g., reducing matter and metal ion limits - see Annex B).
- Biological requirements include sterility, pyrogenicity, haemolysis and toxicity testing (Annex C).
- Sterilization & packaging
- Sets must be sterile in the unit container with evidence of sterilization process effectiveness; packaging guidance aligns with ISO 11607.
- Labelling & disposal
- Requirements for unit and multi‑unit labelling; Clause added on disposal of single‑use sets.
Applications and users
- Manufacturers of blood‑taking and transfusion equipment - for design, testing and regulatory compliance.
- Quality assurance / regulatory affairs teams - to implement normative tests and documentation.
- Hospitals, blood banks and transfusion services - to select compatible, safe single‑use collection sets.
- Test laboratories - to execute the physical, chemical and biological test methods specified in the annexes.
- Procurement specialists - for specification of interoperable blood‑collection products.
Related standards (selected)
- ISO 7864 - Sterile hypodermic needles for single use
- ISO 11607‑1 - Packaging for terminally sterilized medical devices
- ISO 14644‑1 - Cleanroom classification
- ISO 15223‑1 - Symbols on medical device labelling
Keywords: ISO 1135-3:2016, blood‑taking sets, transfusion equipment, single‑use, sterilization, tubing, needles, flow regulator, biocompatibility, medical device standards.
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
ISO 1135-3:2016 is a standard published by the International Organization for Standardization (ISO). Its full title is "Transfusion equipment for medical use — Part 3: Blood-taking sets for single use". This standard covers: ISO 1135-3:2016 specifies requirements for types of blood-taking sets for medical use in order to ensure functional interchangeability of transfusion equipment. It is applicable to sterilized blood-taking sets intended for single use only. ISO 1135-3:2016 also aims to provide a) specifications relating to the quality and performance of materials used in transfusion equipment, and b) a unified presentation of terms for such equipment. In some countries, the national pharmacopoeia or other national regulations are legally binding and take precedence over ISO 1135-3:2016.
ISO 1135-3:2016 specifies requirements for types of blood-taking sets for medical use in order to ensure functional interchangeability of transfusion equipment. It is applicable to sterilized blood-taking sets intended for single use only. ISO 1135-3:2016 also aims to provide a) specifications relating to the quality and performance of materials used in transfusion equipment, and b) a unified presentation of terms for such equipment. In some countries, the national pharmacopoeia or other national regulations are legally binding and take precedence over ISO 1135-3:2016.
ISO 1135-3:2016 is classified under the following ICS (International Classification for Standards) categories: 11.040.20 - Transfusion, infusion and injection equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
ISO 1135-3:2016 has the following relationships with other standards: It is inter standard links to EN ISO 1135-3:2017, ISO 1135-3:1986. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ISO 1135-3:2016 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
INTERNATIONAL ISO
STANDARD 1135-3
Second edition
2016-11-01
Transfusion equipment for medical
use —
Part 3:
Blood-taking sets for single use
Matériel de transfusion à usage médical —
Partie 3: Appareils non réutilisables pour prélèvement sanguin
Reference number
©
ISO 2016
© ISO 2016, Published in Switzerland
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Ch. de Blandonnet 8 • CP 401
CH-1214 Vernier, Geneva, Switzerland
Tel. +41 22 749 01 11
Fax +41 22 749 09 47
copyright@iso.org
www.iso.org
ii © ISO 2016 – All rights reserved
Contents Page
Foreword .iv
1 Scope . 1
2 Normative references . 1
3 General requirements . 1
3.1 Types of sets . 1
3.2 Blood taking assembly. 2
3.3 Air-outlet assembly . 2
3.4 Sterilization . 3
3.5 Maintenance of sterility. 3
4 Materials . 3
5 Physical requirements . 3
5.1 Particulate contamination . 3
5.2 Leakage . 3
5.3 Tensile strength . 3
5.4 Bottle needle . 3
5.5 Air-outlet needle . 3
5.6 Blood-taking needle . . 3
5.7 Tubing . 4
5.8 Flow regulator . 4
5.9 Protective caps . 4
6 Chemical requirements . 4
6.1 Reducing (oxidizable) matter . 4
6.2 Metal ions . 4
6.3 Titration acidity or alkalinity. 4
6.4 Residue on evaporation . 4
6.5 UV absorption of extract solution . 5
7 Biological requirements . 5
7.1 General . 5
7.2 Sterility . 5
7.3 Pyrogenicity . 5
7.4 Haemolysis . 5
7.5 Toxicity . 5
8 Labelling . 5
8.1 General . 5
8.2 Unit container . 5
8.3 Shelf or multi-unit container . 6
9 Packaging . 6
10 Disposal . 6
Annex A (normative) Physical tests . 7
Annex B (normative) Chemical tests . 9
Annex C (normative) Biological tests .11
Bibliography .12
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation on the meaning of ISO specific terms and expressions related to conformity assessment,
as well as information about ISO’s adherence to the World Trade Organization (WTO) principles in the
Technical Barriers to Trade (TBT) see the following URL: www.iso.org/iso/foreword.html.
The committee responsible for this document is ISO/TC 76, Transfusion, infusion and injection, and blood
processing equipment for medical and pharmaceutical use.
This second edition cancels and replaces the first edition (ISO 1135-3:1986), which has been technically
revised with the following changes:
— part title has been amended by “for single use” in alignment with the other parts of ISO 1135;
— figures have been updated;
— subclause 3.6, “Designation examples” has been deleted;
— physical, chemical and biological requirements have been aligned with ISO 1135-4;
— Clause 10, “Disposal” has been added;
— Annexes A, B and C have been aligned with ISO 1135-4;
— all references have been updated.
ISO 1135 consists of the following parts, under the general title Transfusion equipment for medical use:
— Part 3: Blood-taking sets for single use
— Part 4: Transfusion sets for single use, gravity feed
— Part 5: Transfusion sets for single use with pressure infusion apparatus
iv © ISO 2016 – All rights reserved
INTERNATIONAL STANDARD ISO 1135-3:2016(E)
Transfusion equipment for medical use —
Part 3:
Blood-taking sets for single use
1 Scope
This part of ISO 1135 specifies requirements for types of blood-taking sets for medical use in order
to ensure functional interchangeability of transfusion equipment. It is applicable to sterilized blood-
taking sets intended for single use only.
This part of ISO 1135 also aims to provide
a) specifications relating to the quality and performance of materials used in transfusion
equipment, and
b) a unified presentation of terms for such equipment.
In some countries, the national pharmacopoeia or other national regulations are legally binding and
take precedence over this part of ISO 1135.
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.
ISO 3696:1987, Water for analytical laboratory use — Specification and test methods
ISO 7864, Sterile hypodermic needles for single use
ISO 11607-1, Packaging for terminally sterilized medical devices — Part 1: Requirements for materials,
sterile barrier systems and packaging systems
ISO 14644-1:2015, Cleanrooms and associated controlled environments — Part 1: Classification of air
cleanliness by particle concentration
ISO 15223-1, Medical devices — Symbols to be used with medical device labels, labelling and information to
be supplied — Part 1: General requirements
3 General requirements
3.1 Types of sets
The blood-taking set shall consist of the blood-taking assembly and the air-outlet assembly, which may
be separate or combined.
A diagram of a typical blood-taking set is illustrated in Figure 1.
a) Blood-taking assembly b) Air outlet assembly
Key
1 protective cap 5 blood-taking needle
2 bottle needle 6 air filter
3 tubing 7 air filter housing
4 flow regulator (optional) 8 air-outlet needle
Figure 1 — Examples of typical blood-taking sets
3.2 Blood taking assembly
The blood-taking assembly shall consist of a blood-taking needle for vein puncture and of a bottle
needle to be inserted through one of the specified areas provided on the bottle closure. Each needle is
connected to one end of a length of tubing.
3.3 Air-outlet assembly
The air-outlet assembly shall consist of an air filter housing with air filter combined with an air-outlet
needle for piercing the specified area provided on the bottle closure.
The filter shall be capable of preventing microbial ingress.
2 © ISO 2016 – All rights reserved
3.4 Sterilization
The blood-taking set shall be sterile in its unit container. Evidence of the effectiveness of the sterilization
process used shall be provided.
3.5 Maintenance of sterility
The blood-taking set shall be provided with protective caps designed to maintain sterility of the
internal surface of the set and the internal and external surfaces of the needles until the set is used.
4 Materials
The materials from which the blood-taking set is made shall not have undesirable effects on the blood
passing through the set under ordinary conditions of use, or on the fluids used in connection with the
blood. They shall not produce any general toxic effects or any local reaction on the recipient of the blood.
Appropriate type tests for assessing biological compatibility are given in Annex C.
5 Physical requirements
5.1 Particulate contamination
The blood-taking sets shall be manufactured under conditions that minimize particulate contamination.
All parts shall be smooth and clean at the fluid pathway surfaces. When tested as specified in A.1, the
number of particles detected shall not exceed the contamination index limit.
5.2 Leakage
The blood-taking set, when tested in accordance with A.2, shall show no signs of air leakage.
5.3 Tensile strength
Any connections between the components of the blood-taking set, excluding protective caps, shall
withstand a static tensile force of not less than 15 N for 15 s.
5.4 Bottle needle
5.4.1 The bottl
...
NORME ISO
INTERNATIONALE 1135-3
Deuxième édition
2016-11-01
Matériel de transfusion à usage
médical —
Partie 3:
Appareils non réutilisables pour
prélèvement sanguin
Transfusion equipment for medical use —
Part 3: Blood-taking sets for single use
Numéro de référence
©
ISO 2016
DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2016, Publié en Suisse
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni utilisée
sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie, l’affichage sur
l’internet ou sur un Intranet, sans autorisation écrite préalable. Les demandes d’autorisation peuvent être adressées à l’ISO à
l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Ch. de Blandonnet 8 • CP 401
CH-1214 Vernier, Geneva, Switzerland
Tel. +41 22 749 01 11
Fax +41 22 749 09 47
copyright@iso.org
www.iso.org
ii © ISO 2016 – Tous droits réservés
Sommaire Page
Avant-propos .iv
1 Domaine d’application . 1
2 Références normatives . 1
3 Exigences générales . 1
3.1 Types d’appareils . 1
3.2 Dispositif pour prélèvement sanguin complet . 2
3.3 Dispositif de prise d’air complet . 2
3.4 Stérilisation . 3
3.5 Maintien de la stérilité . 3
4 Matériaux . 3
5 Exigences physiques . 3
5.1 Contamination particulaire . 3
5.2 Fuite . 3
5.3 Résistance à la traction . 3
5.4 Aiguille de transfert au flacon . 3
5.5 Aiguille de la prise d’air . 3
5.6 Aiguille pour prélèvement sanguin . 4
5.7 Tubulure . 4
5.8 Régulateur de débit . 4
5.9 Capuchons . 4
6 Exigences chimiques . 4
6.1 Substances réductrices (ou oxydables) . 4
6.2 Ions métalliques . 4
6.3 Acidité ou alcalinité de titrage . 5
6.4 Résidu après évaporation . 5
6.5 Absorption UV de l’extrait . 5
7 Exigences biologiques . 5
7.1 Généralités . 5
7.2 Stérilité . 5
7.3 Pyrogénicité . 5
7.4 Hémolyse . 5
7.5 Toxicité . 5
8 Étiquetage . 5
8.1 Généralités . 5
8.2 Emballage unitaire . 6
8.3 Emballage de protection ou de groupage . 6
9 Emballage. 6
10 Élimination . 7
Annexe A (normative) Essais physiques . 8
Annexe B (normative) Essais chimiques .10
Annexe C (normative) Essais biologiques .12
Bibliographie .13
Avant-propos
L’ISO (Organisation internationale de normalisation) est une fédération mondiale d’organismes
nationaux de normalisation (comités membres de l’ISO). L’élaboration des Normes internationales est
en général confiée aux comités techniques de l’ISO. Chaque comité membre intéressé par une étude
a le droit de faire partie du comité technique créé à cet effet. Les organisations internationales,
gouvernementales et non gouvernementales, en liaison avec l’ISO participent également aux travaux.
L’ISO collabore étroitement avec la Commission électrotechnique internationale (IEC) en ce qui
concerne la normalisation électrotechnique.
Les procédures utilisées pour élaborer le présent document et celles destinées à sa mise à jour sont
décrites dans les Directives ISO/IEC, Partie 1. Il convient, en particulier, de prendre note des différents
critères d’approbation requis pour les différents types de documents ISO. Le présent document a été
rédigé conformément aux règles de rédaction données dans les Directives ISO/IEC, Partie 2 (voir www.
iso.org/directives).
L’attention est appelée sur le fait que certains des éléments du présent document peuvent faire l’objet de
droits de propriété intellectuelle ou de droits analogues. L’ISO ne saurait être tenue pour responsable
de ne pas avoir identifié de tels droits de propriété et averti de leur existence. Les détails concernant
les références aux droits de propriété intellectuelle ou autres droits analogues identifiés lors de
l’élaboration du document sont indiqués dans l’Introduction et/ou dans la liste des déclarations de
brevets reçues par l’ISO (voir www.iso.org/brevets).
Les appellations commerciales éventuellement mentionnées dans le présent document sont données
pour information, par souci de commodité, à l’intention des utilisateurs et ne sauraient constituer un
engagement.
Pour une explication de la signification des termes et expressions spécifiques de l’ISO liés à
l’évaluation de la conformité, ou pour toute information au sujet de l’adhésion de l’ISO aux principes
de l’OMC concernant les obstacles techniques au commerce (OTC), voir le lien suivant: Avant-propos —
Informations supplémentaires.
Le comité chargé de l’élaboration du présent document est l’ISO/TC 76, Appareils de transfusion, de
perfusion et d’injection et appareils destinés au traitement du sang à usage médical et pharmaceutique.
Cette deuxième édition annule et remplace la première édition (ISO 1135-3:1986) qui a fait l’objet d’une
révision technique avec les modifications suivantes:
— modification du titre de la présente partie avec l’ajout de la mention «non réutilisables», à des fins
d’harmonisation par rapport aux autres parties de l’ISO 1135;
— mise à jour des figures;
— suppression du paragraphe 3.6 «Exemples de désignation»;
— alignement des exigences physiques, chimiques et biologiques sur l’ISO 1135-4;
— ajout d’un Article 10 «Élimination»;
— alignement des Annexes A, B et C sur l’ISO 1135-4;
— mise à jour de toutes les références.
L’ISO 1135 comprend les parties suivantes, présentées sous le titre général Matériel de transfusion à
usage médical:
— Partie 3: Appareils non réutilisables pour prélèvement sanguin
— Partie 4: Appareils de transfusion non réutilisables à alimentation par gravité
— Partie 5: Appareils de transfusion non réutilisables avec les appareils de perfusion sous pression
iv © ISO 2016 – Tous droits réservés
NORME INTERNATIONALE ISO 1135-3:2016(F)
Matériel de transfusion à usage médical —
Partie 3:
Appareils non réutilisables pour prélèvement sanguin
1 Domaine d’application
La présente partie de l’ISO 1135 fixe les spécifications des divers types d’appareils pour prélèvement
sanguin à usage médical afin d’assurer I’interchangeabilité fonctionnelle du matériel de transfusion.
Elle s’applique aux appareils stériles non réutilisables pour prélèvement sanguin.
La présente partie de I’ISO 1135 a également pour but de fournir
a) des spécifications relatives à la qualité et aux performances des matériaux utilisés pour le matériel
de transfusion; et
b) une terminologie uniforme concernant ce type de matériel.
Dans certains pays, la pharmacopée nationale ou d’autres règlements nationaux sont juridiquement
contraignants et prévalent sur les dispositions de la présente partie de l’ISO 1135.
2 Références normatives
Les documents ci-après, dans leur intégralité ou non, sont des références normatives indispensables à
l’application du présent document. Pour les références datées, seule l’édition citée s’applique. Pour les
références non datées, la dernière édition du document de référence s’applique (y compris les éventuels
amendements).
ISO 3696:1987, Eau pour laboratoire à usage analytique — Spécification et méthodes d’essai
ISO 7864, Aiguilles hypodermiques stériles, non réutilisables
ISO 11607-1, Emballages des dispositifs médicaux stérilisés au stade terminal — Partie 1: Exigences
relatives aux matériaux, aux systèmes de barrière stérile et aux systèmes d’emballage
ISO 14644-1, Salles propres et environnements maîtrisés apparentés — Partie 1: Classification de la
propreté particulaire de l’air
ISO 15223-1, Dispositifs médicaux — Symboles à utiliser avec les étiquettes, l’étiquetage et les informations
à fournir relatifs aux dispositifs médicaux — Partie 1: Exigences générales
3 Exigences générales
3.1 Types d’appareils
L’appareil pour prélèvement sanguin doit être constitué d’un dispositif pour prélèvement sanguin
complet et d’un dispositif de prise d’air complet, qui peuvent être indépendants ou combinés.
La configuration classique des appareils pour prélèvement sanguin est représentée à la Figure 1.
a) Appareil pour prélèvement sanguin complet b) Dispositif de prise d’air complet
Légende
1 capuchon 5 aiguille pour prélèvement sanguin
2 aiguille de transfert au flacon 6 filtre d’air
3 tubulure 7 étui de filtre d’air
4 régulateur de débit (en option) 8 aiguille de la prise d’air
Figure 1 — Exemples d’appareils pour prélèvement sanguin
3.2 Dispositif pour prélèvement sanguin complet
Le dispositif pour prélèvement sanguin complet doit être composé d’une aiguille pour prélèvement
sanguin pour ponction veineuse et d’une aiguille de transfert au flacon à insérer à travers l’un des points
spécifiés indiqués sur le bouchon du flacon. Chaque aiguille est reliée à une extrémité de la portion de
tubulure.
3.3 Dispositif de prise d’air complet
Le dispositif de prise d’air complet doit être constitué d’un étui de filtre d’air avec filtre d’air, combiné à
une aiguille de prise d’air servant à perforer le point spécifié indiqué sur le bouchon du flacon.
Le filtre doit pouvoir empêcher la pénétration des microbes.
2 © ISO 2016 – Tous droits réservés
3.4 Stérilisation
L’appareil pour prélèvement sanguin doit être stérile dans son emballage unitaire. Des preuves de
l’efficacité du procédé de stérilisation utilisé doivent être fournies.
3.5 Maintien de la stérilité
L’appareil pour prélèvement sanguin doit être pourvu de capuchons conçus pour maintenir la stérilité
de sa surface interne et celle des surfaces interne et externe des aiguilles, jusqu’à l’utilisation de
l’appareil.
4 Matériaux
Les matériaux constitutifs de l’appareil pour prélèvement sanguin ne doivent avoir d’effet indésirable
sur le sang qui traverse l’appareil dans les conditions normales d’emploi, ni sur les fluides utilisés
conjointement avec le sang. Ils ne doivent entraîner aucun effet toxique général, ni aucune réaction
locale sur la personne recevant le sang.
Des essais de type appropriés à l’évaluation de la compatibilité biologique sont mentionnés à l’Annexe C.
5 Exigences physiques
5.1 Contamination particulaire
Les appareils pour prélèvement sanguin doivent être fabriqués dans des conditions permettant de
limiter le plus possible la contamination particulaire. Tous les éléments doivent être lisses et propres au
niveau des surfaces sur lesquelles passe le fluide. Lors de l’essai spécifié en A.1, le nombre de particules
détectées ne doit pas dépasser l’indice limite de contamination.
5.2 Fuite
Lorsqu’il est soumis à l’essai conformément à A.2, l’appareil pour prélèvement sanguin ne doit présenter
aucun signe de fuite d’air.
5.3 Résistance à la traction
Toutes les connexions entre les composants de l’appareil pour prélèvement sanguin (à l’exception des
capuchons) doivent résister à une force de traction statique d’au moins 15 N pendant 15 s.
5.4 Aiguille de transfert au flacon
5.4.1 L’aiguille de transfert au flacon doit avoir une longueur minimale de 35 mm. Le diamètre
extérieur ne doit pas être inférieur à 1,8 mm et le diamètre intérieur ne doit pas être inférieur à 70 % du
diamètre extérieur.
5.4.2 Les surfaces intérieure et extérieure du tube de l’aiguille doivent être propres et lisses.
5.4.3 L’aiguille de transfert au flacon doit être conforme à I’ISO 7864 afin de réduire
...