ISO 13914:2026
(Main)Soil, treated biowaste and sludge — Determination of dioxins and furans and dioxin-like polychlorinated biphenyls by gas chromatography with mass selective detection (high resolution mass spectrometry, HRMS, and tandem mass spectrometry, MS/MS)
Soil, treated biowaste and sludge — Determination of dioxins and furans and dioxin-like polychlorinated biphenyls by gas chromatography with mass selective detection (high resolution mass spectrometry, HRMS, and tandem mass spectrometry, MS/MS)
This document specifies a method for quantitative determination of 17 2,3,7,8-chlorine substituted dibenzo-p-dioxins and dibenzofurans and dioxin-like polychlorinated biphenyls in sludge, treated biowaste and soil using liquid column chromatographic clean-up methods and gas chromatography/high resolution mass spectrometry (GC/HRMS). Detection by tandem mass spectrometry (MS/MS) can be used in an equivalent way. The analytes that can be determined with the method specified in this document are listed in Table 1. The limit of detection depends on the kind of sample, the congener, the equipment used and the quality of chemicals used for extraction and clean-up. Under the conditions specified in this document, limits of detection better than 1 ng/kg (expressed as dry matter) can be achieved. This method is “performance based”. The method can be modified if all performance criteria given in this method are met. This document is applicable for several types of matrices and validated for municipal sludge (see also Annex D for the results of the validation). NOTE In principle, this method can also be applied for sediments, mineral wastes and for vegetation. It is the responsibility of the user of this document to validate the application for these matrices. For measurement in complex matrices such as fly ashes adsorbed on vegetation, it can be necessary to further improve the clean-up. This can also apply to sediments and mineral wastes.
Sols, biodéchets traités et boues — Dosage des dioxines, des furanes et des polychlorobiphényles de type dioxine par chromatographie en phase gazeuse avec détection sélective de masse (spectrométrie de masse à haute résolution ou SMHR, et spectrométrie de masse en tandem ou SM/SM)
Le présent document spécifie une méthode de dosage quantitatif de 17 polychlorodibenzo‑p‑dioxines et polychlorodibenzofuranes substitués par des atomes de chlore en 2,3,7,8 ainsi que de polychlorobiphényles de type dioxine dans les boues, les biodéchets traités et les sols à l’aide de méthodes de purification chromatographique sur colonne en phase liquide et d’analyse par chromatographie en phase gazeuse avec spectrométrie de masse à haute résolution (CG/SMHR). La détection par spectrométrie de masse en tandem (SM/SM) peut être utilisée de manière équivalente. Les analytes qui peuvent être dosés avec la méthode spécifiée dans le présent document sont répertoriés dans le Tableau 1. La limite de détection dépend du type d’échantillon, du congénère, de l’équipement utilisé et de la qualité des produits chimiques utilisés pour l’extraction et la purification. Dans les conditions spécifiées dans le présent document, des limites de détection supérieures à 1 ng/kg (exprimées en matière sèche) peuvent être atteintes. Cette méthode est "fondée sur les performances". Elle peut être modifiée si tous les critères de performance donnés dans la méthode sont respectés. Le présent document est applicable pour plusieurs types de matrices et validé pour les boues de stations d’épuration des eaux usées urbaines (voir également l’Annexe D pour les résultats de la validation). NOTE En principe, cette méthode peut être également appliquée pour les sédiments, les déchets minéraux et les végétaux. Il appartient à l’utilisateur du présent document de valider l’application pour ces matrices. Pour les mesures dans les matrices complexes telles que les cendres volantes adsorbées sur des végétaux, une amélioration de la purification peut être nécessaire. Cela peut également s’appliquer aux sédiments et aux déchets minéraux.
General Information
- Status
- Published
- Publication Date
- 18-Mar-2026
- Technical Committee
- ISO/TC 190/SC 3 - Chemical and physical characterization
- Drafting Committee
- ISO/TC 190/SC 3/WG 6 - Organic analysis
- Current Stage
- 6060 - International Standard published
- Start Date
- 19-Mar-2026
- Due Date
- 30-Apr-2026
- Completion Date
- 19-Mar-2026
Relations
- Effective Date
- 07-Jan-2025
Overview
ISO/PRF 13914 (2025) specifies a performance‑based analytical method for the quantitative determination of dioxins, furans and dioxin‑like polychlorinated biphenyls (dl‑PCBs) in soil, treated biowaste and sludge. The method uses liquid column chromatographic clean‑up and gas chromatography with mass selective detection - either high resolution mass spectrometry (GC/HRMS) or tandem mass spectrometry (GC‑MS/MS). It targets the 17 toxicologically relevant 2,3,7,8‑substituted PCDD/PCDF congeners and dl‑PCBs, and under specified conditions can achieve limits of detection better than 1 ng/kg (dry matter). The standard is validated for municipal sludge and can be extended to sediments, mineral wastes and vegetation with user validation and potentially enhanced clean‑up.
Key topics and technical requirements
- Analytes covered: 17 2,3,7,8‑chlorine substituted PCDD/PCDF congeners and dioxin‑like PCBs (listed in Table 1 of the document).
- Analytical principle: Isotope dilution combined with GC separation and mass selective detection (HRMS or MS/MS). Use of multiple 13C‑labelled internal standards is required for accurate quantification.
- Sample preparation: Protocols for sample storage, pretreatment, solvent extraction (e.g., Soxhlet or equivalent), and multi‑step liquid column chromatographic clean‑up to remove matrix interferences.
- Instrument requirements: Minimum performance and identification/quantification criteria for GC/HRMS and GC‑MS/MS, system calibration, recovery standards and quantification procedures.
- Performance criteria: The method is performance based - laboratories may modify procedures provided all specified performance criteria (precision, recovery, LOD) are met.
- Safety and competence: Only trained personnel should implement the method; the document warns about handling highly toxic compounds.
- Supporting material: Informative annexes include TEF information, example extraction/clean‑up workflows, GC/HRMS operation examples and precision data.
Applications
- Environmental monitoring of dioxins, furans and dl‑PCBs in soil, sludge and treated biowaste for regulatory compliance, contamination assessment and remediation verification.
- Routine and investigative laboratory analyses where sensitive detection (sub‑ng/kg) is required.
- Adaptable to related matrices (sediments, mineral wastes, vegetation, fly ash) following matrix‑specific validation and potentially enhanced clean‑up.
Who should use this standard
- Analytical and environmental testing laboratories performing dioxin/dl‑PCB quantification.
- Regulatory agencies setting or enforcing soil and sludge contaminant limits.
- Researchers and consultants involved in contaminated site assessment, waste management and environmental risk assessment.
- Instrument vendors and QA managers ensuring method performance for GC/HRMS or MS/MS workflows.
Related standards
- EN 15934 / ISO 11465 (referenced) - determination of dry matter fraction for sludges, soils and related matrices.
- ISO/TC 190 (Soil quality) guidance and other ISO methods for persistent organic pollutant analysis.
Keywords: ISO/PRF 13914, ISO 13914, dioxins, furans, dioxin‑like PCBs, GC/HRMS, MS/MS, soil testing, sludge analysis, treated biowaste, isotope dilution, environmental monitoring.
Buy Documents
ISO 13914:2026 - Soil, treated biowaste and sludge — Determination of dioxins and furans and dioxin-like polychlorinated biphenyls by gas chromatography with mass selective detection (high resolution mass spectrometry, HRMS, and tandem mass spectrometry, MS/MS)
ISO 13914:2026 - Sols, biodéchets traités et boues — Dosage des dioxines, des furanes et des polychlorobiphényles de type dioxine par chromatographie en phase gazeuse avec détection sélective de masse (spectrométrie de masse à haute résolution ou SMHR, et spectrométrie de masse en tandem ou SM/SM)
Get Certified
Connect with accredited certification bodies for this standard

NSF International
Global independent organization facilitating standards development and certification.
CIS Institut d.o.o.
Personal Protective Equipment (PPE) certification body. Notified Body NB-2890 for EU Regulation 2016/425 PPE.

Kiwa BDA Testing
Building and construction product certification.
Sponsored listings
Frequently Asked Questions
ISO 13914:2026 is a standard published by the International Organization for Standardization (ISO). Its full title is "Soil, treated biowaste and sludge — Determination of dioxins and furans and dioxin-like polychlorinated biphenyls by gas chromatography with mass selective detection (high resolution mass spectrometry, HRMS, and tandem mass spectrometry, MS/MS)". This standard covers: This document specifies a method for quantitative determination of 17 2,3,7,8-chlorine substituted dibenzo-p-dioxins and dibenzofurans and dioxin-like polychlorinated biphenyls in sludge, treated biowaste and soil using liquid column chromatographic clean-up methods and gas chromatography/high resolution mass spectrometry (GC/HRMS). Detection by tandem mass spectrometry (MS/MS) can be used in an equivalent way. The analytes that can be determined with the method specified in this document are listed in Table 1. The limit of detection depends on the kind of sample, the congener, the equipment used and the quality of chemicals used for extraction and clean-up. Under the conditions specified in this document, limits of detection better than 1 ng/kg (expressed as dry matter) can be achieved. This method is “performance based”. The method can be modified if all performance criteria given in this method are met. This document is applicable for several types of matrices and validated for municipal sludge (see also Annex D for the results of the validation). NOTE In principle, this method can also be applied for sediments, mineral wastes and for vegetation. It is the responsibility of the user of this document to validate the application for these matrices. For measurement in complex matrices such as fly ashes adsorbed on vegetation, it can be necessary to further improve the clean-up. This can also apply to sediments and mineral wastes.
This document specifies a method for quantitative determination of 17 2,3,7,8-chlorine substituted dibenzo-p-dioxins and dibenzofurans and dioxin-like polychlorinated biphenyls in sludge, treated biowaste and soil using liquid column chromatographic clean-up methods and gas chromatography/high resolution mass spectrometry (GC/HRMS). Detection by tandem mass spectrometry (MS/MS) can be used in an equivalent way. The analytes that can be determined with the method specified in this document are listed in Table 1. The limit of detection depends on the kind of sample, the congener, the equipment used and the quality of chemicals used for extraction and clean-up. Under the conditions specified in this document, limits of detection better than 1 ng/kg (expressed as dry matter) can be achieved. This method is “performance based”. The method can be modified if all performance criteria given in this method are met. This document is applicable for several types of matrices and validated for municipal sludge (see also Annex D for the results of the validation). NOTE In principle, this method can also be applied for sediments, mineral wastes and for vegetation. It is the responsibility of the user of this document to validate the application for these matrices. For measurement in complex matrices such as fly ashes adsorbed on vegetation, it can be necessary to further improve the clean-up. This can also apply to sediments and mineral wastes.
ISO 13914:2026 is classified under the following ICS (International Classification for Standards) categories: 13.080.10 - Chemical characteristics of soils. The ICS classification helps identify the subject area and facilitates finding related standards.
ISO 13914:2026 has the following relationships with other standards: It is inter standard links to ISO 13914:2023. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ISO 13914:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
International
Standard
ISO 13914
Third edition
Soil, treated biowaste and sludge —
2026-03
Determination of dioxins and furans
and dioxin-like polychlorinated
biphenyls by gas chromatography
with mass selective detection (high
resolution mass spectrometry,
HRMS, and tandem mass
spectrometry, MS/MS)
Sols, biodéchets traités et boues — Dosage des dioxines,
des furanes et des polychlorobiphényles de type dioxine par
chromatographie en phase gazeuse avec détection sélective de
masse (spectrométrie de masse à haute résolution ou SMHR, et
spectrométrie de masse en tandem ou SM/SM)
Reference number
© ISO 2026
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii
Contents Page
Foreword .v
Introduction .vi
1 Scope . 1
2 Normative references . 2
3 Terms and definitions . 2
4 Abbreviated terms . 3
5 Principle . 3
6 Reagents . 3
6.1 Chemicals .3
6.2 Standards .4
7 Apparatus . 4
7.1 General .4
7.2 Equipment for sample preparation .4
7.3 Soxhlet extractor .4
7.4 Clean-up apparatus .4
7.5 Concentration apparatus .5
7.6 Other equipment .5
8 Sample storage and sample pretreatment . 6
8.1 Sample storage .6
8.2 Sample pretreatment .6
9 Extraction and clean-up . 6
9.1 General .6
9.2 Extraction .6
9.3 Clean-up .7
9.3.1 General .7
9.3.2 Gel permeation chromatography .7
9.3.3 Multilayer column .8
9.3.4 Sulphuric acid treatment .8
9.3.5 Activated carbon column .8
9.3.6 Aluminium oxide column .8
9.3.7 Removal of sulphur .8
9.4 Final concentration of cleaned sample extract .8
9.5 Addition of recovery standard .9
10 GC/MS (HRMS or MS/MS) analysis. 9
10.1 General .9
10.2 Gas chromatographic analysis .9
10.3 Mass spectrometric detection .9
10.3.1 General .9
10.3.2 High resolution mass spectrometry (HRMS) .9
10.3.3 Tandem mass spectrometry (MS/MS) .11
10.4 Minimum requirements for identification of PCDF/PCDD and PCB . 12
10.5 Minimum requirements for quantification of PCDF/PCDD and PCB . 13
10.6 Calibration of the GC/MS system (HRMS or MS/MS) .14
10.6.1 General .14
10.6.2 Calibration for the analytes .14
10.6.3 Calibration for sum of homologue groups . 15
10.7 Quantification of GC/MS (HRMS or MS/MS) results .16
10.7.1 Quantification of concentrations of the analytes .16
10.7.2 Quantification of recovery rates of C -labelled standards .16
10.7.3 Quantification of sum of homologue groups .17
10.7.4 Calculation of the toxic equivalent .17
iii
10.7.5 Calculation of the limit of detection and the limit of quantification .17
11 Expression of results .18
12 Precision .18
13 Test report .18
Annex A (informative) Toxic equivalency factor (TEF) . 19
Annex B (informative) Example of extraction and clean-up methods .21
Annex C (informative) Examples of operation of GC/HRMS determination.28
Annex D (informative) Repeatability and reproducibility data .32
Bibliography .38
iv
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO documents should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent
rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a)
patent(s) which may be required to implement this document. However, implementers are cautioned that
this may not represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO’s adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 190, Soil quality, Subcommittee SC 3, Chemical
and physical characterization.
This third edition cancels and replaces the second edition (ISO 13914:2023), which has been technically
revised.
The main changes are as follows:
— tandem mass spectrometry (MS/MS) has been added as an alternative gas chromatographic analysis to
high resolution mass spectrometry (HRMS);
— validation results for MS/MS have been added.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.
v
Introduction
Two groups of related chlorinated aromatic ethers are known as polychlorinated dibenzo-p-dioxins (PCDDs)
and polychlorinated dibenzofurans (PCDFs). They consist of a total of 210 individual substances (congeners):
75 PCDD and 135 PCDF of which 17 have chlorine substitution in the 2,3,7,8-positions.
A group of chlorinated aromatic compounds similar to polychlorinated dibenzo-p-dioxins (PCDDs) and
polychlorinated dibenzofurans (PCDFs) is known as polychlorinated biphenyls (PCBs) which consists of 209
individual substances.
PCDD and PCDF can form in the combustion of organic materials; they also occur as undesirable by-
products in the manufacture or further processing of chlorinated organic chemicals. PCDD/PCDF enter the
environment mainly via these emission paths and through the use of contaminated materials. In fact, they
are universally present at very small concentrations. The 2,3,7,8-substituted congeners are toxicologically
significant. Toxicologically much less significant than the tetrachlorinated to octachlorinated dibenzo-p-
dioxins/dibenzofurans are the 74 monochlorinated to trichlorinated dibenzo-p-dioxins/dibenzofurans.
PCB have been produced over a period of approximately 50 years until the end of the 1990s for the purpose
of different uses in open and closed systems, e.g. as electrical insulators or dielectric fluids in capacitors and
transformers, as specialized hydraulic fluids and as a plasticizer in sealing material. Worldwide more than
one million tons of PCB were produced.
PCDD/PCDF as well as PCB are emitted during thermal processes, for example waste incineration. In 1997 a
group of experts of the World Health Organization (WHO) fixed toxicity equivalent factors (TEF) for PCDD
and twelve PCB, known as dioxin-like PCB (dl-PCB, see Annex A). These twelve dioxin-like PCB consist of
four non-ortho PCB and eight mono-ortho PCB (no or only one chlorine atoms in 2-, 2’-, 6- and 6’-position),
having a planar or mostly planar structure. Dioxin-like PCB can contribute considerably to the total WHO-
TEQ.
Only skilled operators who are trained in handling highly toxic compounds should apply the method
described in this document.
vi
International Standard ISO 13914:2026(en)
Soil, treated biowaste and sludge — Determination of dioxins
and furans and dioxin-like polychlorinated biphenyls by
gas chromatography with mass selective detection (high
resolution mass spectrometry, HRMS, and tandem mass
spectrometry, MS/MS)
WARNING — Persons using this document should be familiar with normal laboratory practice. This
document does not purport to address all of the safety problems, if any, associated with its use. It is
the responsibility of the user to establish appropriate safety and health practices.
IMPORTANT — Tests conducted according to this document shall be carried out by suitably trained
staff.
1 Scope
This document specifies a method for quantitative determination of 17 2,3,7,8-chlorine substituted dibenzo-
p-dioxins and dibenzofurans and dioxin-like polychlorinated biphenyls in sludge, treated biowaste and soil
using liquid column chromatographic clean-up methods and gas chromatography/high resolution mass
spectrometry (GC/HRMS). Detection by tandem mass spectrometry (MS/MS) can be used in an equivalent
way.
The analytes that can be determined with the method specified in this document are listed in Table 1.
Table 1 — Analytes and their abbreviations
Substance Abbreviation
Tetrachlorodibenzo-p-dioxin TCDD
Pentachlorodibenzo-p-dioxin PeCDD
Hexachlorodibenzo-p-dioxin HxCDD
Heptachlorodibenzo-p-dioxin HpCDD
Octachlorodibenzo-p-dioxin OCDD
Tetrachlorodibenzofuran TCDF
Pentachlorodibenzofuran PeCDF
Hexachlorodibenzofuran HxCDF
Heptachlorodibenzofuran HpCDF
Octachlorodibenzofuran OCDF
Polychlorinated biphenyl PCB
a
Trichlorobiphenyl TCB
Tetrachlorobiphenyl TeCB
Pentachlorobiphenyl PeCB
Hexachlorobiphenyl HxCB
Heptachlorobiphenyl HpCB
a
Decachlorobiphenyl DecaCB
a
Groups containing no dl-PCB, given for informative purpose only.
The limit of detection depends on the kind of sample, the congener, the equipment used and the quality
of chemicals used for extraction and clean-up. Under the conditions specified in this document, limits of
detection better than 1 ng/kg (expressed as dry matter) can be achieved.
This method is “performance based”. The method can be modified if all performance criteria given in this
method are met.
This document is applicable for several types of matrices and validated for municipal sludge (see also
Annex D for the results of the validation).
NOTE In principle, this method can also be applied for sediments, mineral wastes and for vegetation. It is the
responsibility of the user of this document to validate the application for these matrices. For measurement in complex
matrices such as fly ashes adsorbed on vegetation, it can be necessary to further improve the clean-up. This can also
apply to sediments and mineral wastes.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes
requirements of this document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
ISO 11465, Sludge and solid environmental matrices — Determination of dry residue or water content and
calculation of the dry matter fraction on a mass basis
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at https:// www .electropedia .org/
3.1
internal standard
C -labelled analyte analogue added to samples prior to extraction against which the concentrations of
native analytes are calculated
[SOURCE: ISO 18073:2004, 3.1.5, modified — “2,3,7,8-PCDD/PCDF” and “PCDDs and PCDFs” were changed to
“analytes”.]
3.2
recovery standard
C -labelled dl-polychlorinated biphenyl (dl-PCB) and polychlorinated dibenzo-p-dioxin/polychlorinated
dibenzofuran (PCDD/PCDF), added before injection into the GC
[SOURCE: ISO 18073:2004, 3.1.12, modified — “2,3,7,8-chloro-substituted PCDD/PCDF” was changed to “dl-
PCB and PCDD/PCDF”.]
4 Abbreviated terms
dl-PCB dioxin-like polychlorinated biphenyls
MRM multiple reaction monitoring
PCDD/PCDF or PCDD/F Polychlorinated dibenzo-p-dioxins/dibenzofurans
I-TEQ International toxic equivalent
WHO-TEF Toxic equivalent factor proposed by WHO in 2005
WHO-TEQ Toxic equivalent obtained by multiplying the mass determined with the corre-
sponding WHO-TEF including PCDD, PCDF and dl-PCB
NOTE 1 The I-TEQ is obtained by multiplying the mass determined with the corresponding I-TEF including PCDD
and PCDF (for detailed description, see Annex A). Should only be used for comparison with older data.
NOTE 2 For detailed description of WHO-TEF, see Annex A.
For detailed description of WHO-TEQ, see Annex A. WHO-TEQPCB and WHO-TEQPCDD/PCDF should be used
to distinguish different compound classes.
5 Principle
This document is based on the use of gas chromatography/mass spectrometry combined with the isotope
dilution technique to enable the separation, detection and quantification of PCDD/PCDF and dioxin-like PCB
in sludge, biowaste and soil. For the isotope dilution method, at least 16 out of 17 labelled PCDD/PCDF and
12 labelled PCB internal standards are used. The extracts for the GC-MS measurements contain at least one
13 13
recovery standard per group, i.e. at least one C -PCDD/PCDF and one C -PCB. The use of more recovery
12 12
standards is recommendable. The gas chromatographic parameters offer information which enables the
identification of congeners (position of chlorine substitutes) whereas the mass spectrometric parameters
enable the differentiation between isomers with different numbers of chlorine substitutes and between
dibenzo-p-dioxins, furans and PCB.
C -labelled PCDD/PCDF and PCB congeners are added to the sample prior to extraction and GC/MS (HRMS
or MS/MS) measurement. Losses during extraction and clean-up are detected and compensated by using
these added congeners as internal standards for quantification together with recovery standards which are
added just before the analysis. For the determination of these substances it is necessary to separate PCB
from PCDD/PCDF and vice versa.
The main purpose of the clean-up procedure of the raw sample extract is the removal of sample matrix
components, which can overload the separation method, disturb the quantification or otherwise severely
impact the performance of the identification and quantification method and the separation of PCDD/PCDF
from dioxin-like PCB. Furthermore, the enrichment of the analytes in the final sample extract is achieved.
Extraction procedures are usually based on Soxhlet or equivalent extraction methods of dried, preferably
freeze-dried, samples. Sample clean-up is usually carried out by multi-column liquid chromatographic
techniques using different adsorbents. The determination of PCDD/PCDF and PCB is based on quantification
by the isotope-dilution technique using GC/MS (HRMS or MS/MS).
6 Reagents
6.1 Chemicals
Solvents used for extraction and clean-up shall be of pesticide grade or equivalent quality and checked for
blanks. Adsorbents such as aluminium oxide, silica gel, diatomaceous earth and others used for clean-up
shall be of analytical grade quality or better and pre-cleaned and activated if necessary.
NOTE See Annex B for a specific list of solvents and chemicals.
6.2 Standards
— C -spiking solution for PCDD/PCDF (internal standard);
— C -spiking solution for PCB (internal standard);
— calibration solutions PCDD/PCDF;
— calibration solutions PCB;
13 13
— C -spiking solution as recovery standard for PCDD/PCDF (remark: typical C-congeners used as
13 13 13
recovery standards are C -123789-HexaCDD, C -1234-TetraCDD, C -1234-TetraCDD or -DF);
12 12 12
— C -spiking solution as recovery standard for dl-PCB.
NOTE See Annex B for examples of concentration of the standard solutions.
7 Apparatus
7.1 General
The apparatus and materials listed below are meant as minimum requirements for “conventional” sample
treatment with Soxhlet extraction and column chromatographic clean-up. Additional apparatus and
materials can be necessary due to different methods of sample extraction and clean-up methods.
7.2 Equipment for sample preparation
7.2.1 Laboratory fume hood, of sufficient size to contain the sample preparation equipment listed below.
7.2.2 Desiccator.
7.2.3 Balances, consisting of an analytical type capable of weighing 0,1 mg and a top-loading type capable
of weighing 10 mg.
7.2.4 Snyder column.
7.3 Soxhlet extractor
Alternatively, other automated Soxhlet-like devices can be used, if applicable. It is the responsibility of the
user of this document to validate the application for these alternative devices.
7.3.1 Soxhlet, 50 mm internal diameter, 150 ml or 250 ml capacity with 500 ml round bottom flask.
7.3.2 Thimble, 43 mm × 123 mm, to fit Soxhlet.
7.3.3 Hemispherical heating mantle, to fit 500 ml round-bottom flask.
7.4 Clean-up apparatus
7.4.1 Disposable pipettes, either disposable Pasteur pipettes, or disposable serological pipettes.
7.4.2 Glass chromatographic columns of the following sizes:
— 150 mm length × 8 mm internal diameter, with coarse-glass frit or glass-wool plug, 250 ml reservoir and
glass or polytetrafluoroethylene (PTFE) stopcock;
— 200 mm length × 15 mm internal diameter, with coarse-glass frit or glass-wool plug, 250 ml reservoir
and glass or PTFE stopcock;
— 300 mm length × 25 mm internal diameter, with coarse-glass frit or glass-wool plug, 300 ml reservoir
and glass or PTFE stopcock.
7.4.3 Oven, capable of maintaining a constant temperature (±5 °C) in the range of 105 °C to 450 °C for
baking and storage of adsorbents.
7.5 Concentration apparatus
Alternatively, other evaporation devices can be used, if applicable. It is the responsibility of the user of this
document to validate the application for these alternative evaporation devices.
7.5.1 Rotary evaporator, equipped with a variable temperature water bath and:
— vacuum source for rotary evaporator equipped with shutoff valve at the evaporator and vacuum gauge;
— recirculating water pump and chiller, providing cooling water of (9 ± 4) °C (use of tap water for cooling
the evaporator wastes large volumes of water and can lead to inconsistent performance as water
temperatures and pressures vary);
— round-bottom flask, 100 ml and 500 ml or larger, with ground-glass fitting compatible with the rotary
evaporator.
7.5.2 Nitrogen blowdown apparatus, equipped with either a water bath controlled in the range of 30 °C
to 60 °C or a heated stream of nitrogen or of another suitable inert gas, installed in a fume hood.
1)
7.5.3 Kuderna-Danish concentrator.
7.5.4 Sample vials, of the following types:
— amber glass, nominated volume 2 ml to 5 ml, with PTFE-lined screw cap;
— glass, 0,3 ml, conical, with PTFE-lined screw or crimp cap.
7.6 Other equipment
7.6.1 Gas chromatograph, equipped with a splitless or on-column or temperature programmed injection
port for the use with capillary columns, and an oven temperature programme which enables isothermal
hold.
7.6.2 GC column for PCDD/PCDF and for isomer specificity for 2,3,7,8-TCDD (e.g. 60 m length × 0,32 mm
internal diameter; 0,25 µm; 5 % phenyl, 94 % methyl, 1 % vinyl silicone bonded-phase fused-silica capillary
column).
7.6.3 Mass spectrometer, 28 eV to 80 eV electron impact ionization, capable of repetitively selectively
monitoring of twelve exact masses minimum during a period of approximately 1 s either at high mass
resolution (≥10 000 at 10 % peak valley or equivalent) or 70 eV electron impact ionization following analysis
utilizing a triple quad MS/MS system on specific multiple reaction monitoring (MRM) (see Table 5).
7.6.4 Data system, capable of collecting, recording, and storing mass spectrometric data.
1) Kuderna Danish is an example of a suitable product available commercially. This information is given for the
convenience of users of this document and does not constitute an endorsement by ISO of this product.
8 Sample storage and sample pretreatment
8.1 Sample storage
Samples should be stored in suitable containers with an appropriate closure material such as
polytetrafluoroethylene (PTFE). Samples to be frozen may be stored in aluminium containers pre-cleaned
by heating to 450 °C for minimum 4 h or by rinsing with a non-chlorinated solvent.
Samples should be kept cool and dark where necessary in order to avoid alteration of sample constitution
e.g. in combination with water content determination.
8.2 Sample pretreatment
Drying and homogenization should be carried out according to EN 16179, if not otherwise specified. Store
the ground material in a desiccator or a tightly closed glass container.
Determination of water content shall be carried out according to ISO 11465 or any equivalent method which
is internationally standardized.
9 Extraction and clean-up
9.1 General
In this document, the minimum requirements for extraction and clean-up to be met are described as well
as examples of operation. The analyst may use any of the procedures given below and in Annex B or any
suitable alternative procedures.
The determination of PCDD/PCDF and dl-PCB is based on quantification by the isotope-dilution technique
using GC/MS (HRMS or MS/MS). C -labelled dl-PCB and 2,3,7,8-chlorine substituted PCDD/PCDF
congeners are added at different stages of the whole method. Losses during extraction and clean-up can
be detected and compensated by using these added congeners as internal standards for quantification
together with recovery standards which are added just before the GC/MS analysis. However, due to possible
differences in the binding and adsorption characteristics between the native analytes and the C -
labelled congeners, which are added during analysis, complete substantiation of the extraction efficiency
and compensation of losses during clean-up is not ensured. Therefore, in addition the applied methods shall
be validated thoroughly. Examples of well-proven extraction and clean-up methods are given in Annex B.
Annex C gives examples of operation of GC/HRMS determination.
The main purpose of the clean-up procedure of the raw sample extract is the removal of sample matrix
components, which can overload the separation method, disturb the quantification or otherwise severely
impact the performance of the identification and quantification method and to separate dioxin-like PCB from
PCDD/PCDF. Furthermore, an enrichment of the analytes in the final sample extract is achieved. Extraction
procedures are normally based on Soxhlet extraction of the < 2 mm fraction of the dry and ground or sieved
solid sample. Sample clean-up is usually carried out by multi-column liquid chromatographic techniques
using different adsorbents.
In principle any clean-up method can be used which recovers the analytes in sufficient quantities.
Furthermore, the final sample extract shall not affect adversely the performance of the analytical system
or the quantification step. However, all applied methods shall be tested thoroughly and shall pass a set of
method validation requirements before they can be employed (see Annex D). In addition, the verification of
the method performance for each single sample shall be part of the applied quality assurance protocol.
9.2 Extraction
The sample amount used for extraction can vary from 5 g to 50 g depending on the expected level of
contamination.
The internal standard consisting of C -labelled congeners listed in Table 2 shall be added directly onto
the sample before extraction.
The extraction procedure is carried out using Soxhlet extraction with toluene. Duration of extraction should
be adjusted according to kind and amount of sample used. The minimum requirement is 50 extraction cycles
(approximately 12 h).
Other solvents or other methods such as pressurized liquid extraction can also be used but shall be of proven
equal performance.
Table 2 — C labelled congeners included in the internal standard
C -spiking solution – internal standard
PCDD/PCDF congeners PCB congeners
13 13
2,3,7,8- C -TCDD C -PCB 77
12 12
13 13
1,2,3,7,8- C -PeCDD C -PCB 81
12 12
13 13
1,2,3,4,7,8- C -HxCDD C -PCB 126
12 12
13 13
1,2,3,6,7,8- C -HxCDD C -PCB 169
12 12
13 a
1,2,3,7,8,9- C -HxCDD
13 13
1,2,3,4,6,7,8- C -HpCDD C -PCB 105
12 12
13 13
C -OCDD C -PCB 114
12 12
C -PCB 118
13 13
2,3,7,8- C -TCDF C -PCB 123
12 12
13 13
1,2,3,7,8- C -PeCDF C -PCB 156
12 12
13 13
2,3,4,7,8- C -PeCDF C -PCB 157
12 12
13 13
1,2,3,4,7,8- C -HxCDF C -PCB 167
12 12
13 13
1,2,3,6,7,8- C -HxCDF C -PCB 189
12 12
2,3,4,6,7,8- C -HxCDF
1,2,3,7,8,9- C -HxCDF
1,2,3,4,6,7,8- C -HpCDF
1,2,3,4,7,8,9- C -HpCDF
C -OCDF
a 13
1,2,3,7,8,9- C -HxCDD is traditionally widely used as a recovery standard. Its use as internal standard is
therefore not mandatory.
9.3 Clean-up
9.3.1 General
Clean-up methods shall prepare the sample extract in an appropriate manner for the subsequent quantitative
determination. Clean-up procedures shall concentrate PCDD/PCDF and dioxin-like PCB in the extracts and
to remove interfering matrix components present in the raw extract.
Proven clean-up procedures shall be used including usually two or more of the following techniques which
can be combined in different orders. A detailed description of some of the procedures is given in Annex B.
Other methods can also be used but shall be of proven equal performance as the techniques described below.
9.3.2 Gel permeation chromatography
The interesting molecular weight range for PCDD/PCDF and dioxin-like PCB of 200 g/mol to 500 g/mol can
be isolated from larger molecules and polymers which might overload other clean-up methods. This method
can also be used for the removal of sulphur.
9.3.3 Multilayer column
Multilayer column liquid chromatography using silica with different activity grades and surface
modifications is used. Compounds with different chemical properties than PCDD/PCDF and dl-PCB can be
removed.
9.3.4 Sulphuric acid treatment
A direct treatment of the sample extract with sulphuric acid is possible but is not recommended due to risk
of accident. Furthermore, this shall be carried out very carefully to avoid losses of PCDD/PCDF and dl-PCB
on the formed carboniferous surfaces.
9.3.5 Activated carbon column
Column adsorption chromatography using activated carbon can be used to separate planar PCDD/PCDF and
coplanar PCB molecules from mono-ortho PCB and other interfering non-planar molecules.
9.3.6 Aluminium oxide column
Column liquid chromatography on aluminium oxide of different activity grade and acidity or basicity is used.
Interfering compounds with small differences in polarity or structure compared to PCDD/PCDF and dioxin-
like PCB can be removed.
Additionally, aluminium oxide columns can be used to separate PCDD/PCDF from dioxin-like PCB.
9.3.7 Removal of sulphur
The removal of sulphur can be achieved by refluxing the extract with powdered copper or by gel permeation
chromatography.
9.4 Final concentration of cleaned sample extract
To achieve sufficient detection limits, the cleaned sample extract shall be concentrated to a volume in the
order of 25 µl to 100 µl before quantification. The final solvent shall be nonane, toluene or another high
boiling solvent.
Though PCDD/PCDF and dl-PCB have rather high boiling points (>320 °C) vapour phase transfer mechanisms
and aerosol formation during solvent evaporation can lead to substantial losses when concentrating volumes
below 10 ml. Depending on the method to be used for solvent volume reduction the following precautions
shall be taken into consideration.
a) Rotary evaporators:
Losses can be substantial when reducing solvent volumes below 10 ml. Counter measures are the use of
controlled vacuum conditions according to the vapour pressure and boiling point of the solvent, addition
of a high-boiling solvent as a keeper as well as the use of specially shaped vessels (e.g. V-shaped).
b) Counter gas flow evaporators:
Volumes should not be reduced to less than 1 ml.
c) Nitrogen flow:
An excessive flow of nitrogen which disturbs the solvent surface should be avoided. The vial shape has
also some influence on possible losses. V-shaped vials or vial inserts shall be used for volume reductions
below around 200 µl.
2)
d) Kuderna Danish :
To avoid initial losses pre-wet the column with about 1 ml of solvent. Boiling chips should be added.
Adjust the vertical position of the apparatus. At the proper rate of distillation, the balls of the column
actively chatter but the chambers do not flood. Adjust the water bath temperature accordingly. When
reaching an extract volume of 1 ml remove the evaporation flask, replace the Snyder column (7.2.4) by a
smaller one and continue the evaporation.
9.5 Addition of recovery standard
The very last step before quantification is the addition of the recovery standards for calculation of the
recovery rates of the internal standards.
Recovery standards shall be added after the final volume reduction. Any further direct volume reduction
shall be avoided. A slow evaporation at room temperature from the open sample vial to a volume of
approximately 25 µl is acceptable. After addition of the recovery standards, sample extracts should be
stored as briefly as possible before measurement and any further uncontrolled solvent evaporation until
dryness shall be avoided.
10 GC/MS (HRMS or MS/MS) analysis
10.1 General
GC-MS analyses of PCDD/PCDF and dioxin-like PCB shall be carried out on a GC-MS instrument equipped
with a high resolution gas chromatograph, an auto sampler, a mass spectrometer (high resolution or MS/
MS) and a data system for instrument control, data acquisition and processing.
10.2 Gas chromatographic analysis
Gas chromatographic separation shall be carried out in such a way that sufficient separation of all PCDD/
PCDF and dioxin-like PCB congeners is achieved, and the quality criteria specified in 10.4 and 10.5 are met.
For PCDD/PCDF there is no capillary column available at present that allows the separation of all
2,3,7,8-substituted congeners from all other non-2,3,7,8-substituted congeners. Complete separation can
only be achieved by analysing a sample on different capillary columns of different polarity.
For dioxin-like PCB analysis similar problems exist for the separation of all coplanar and mono-ortho
congeners. There is no column available at present which is able to completely separate all twelve dioxin-like
PCB congeners from all other non-dioxin-like PCB congeners.
For routine purposes, a state-of-the-art single-GC-column analysis is usually acceptable unless otherwise
demanded.
10.3 Mass spectrometric detection
10.3.1 General
Both a high resolution mass spectrometer or an MS/MS system on specific MRM can be used for
determination of PCDD/PCDF and dioxin-like PCB.
10.3.2 High resolution mass spectrometry (HRMS)
A high resolution mass spectrometer at a minimum resolution of ≥ 10 000 is used for the detection of PCDD/
PCDF and dioxin-like PCB. This allows the use of C -labelled congeners as internal standards for the
2,3,7,8-PCDD/PCDF congeners and dioxin-like PCB congeners of interest.
2) Kuderna Danish is an example of a suitable product available commercially. This information is given for the
convenience of users of this document and does not constitute an endorsement by ISO of this product.
The mass spectrometer is used in the MID mode (multiple ion detection), the GC column is directly coupled
to the mass spectrometer. The ion source temperature should be between 250 °C to 270 °C depending on
type of instrument. To achieve appropriate sensitivity the detection capability should be at least 200 fg for
2,3,7,8-TCDD.
For identification and quantification, the masses given in Table 3 and Table 4 shall be recorded in MID mode.
For each PCDD/PCDF or PCB congener of interest at least two ions of the molecular isotope cluster shall be
recorded for both the native and the added C -labelled congeners.
In addition, masses for quality control of the mass calibration shall be measured depending on the type of
instrument, e.g. lock mass, calibration mass, lock mass check.
The time slots for the MID windows shall be defined by a calibration standard in a way that all congeners
of interest elute within the related MID-window. In the case the sum of the concentrations of isomer groups
is needed, the retention time window for all isomers of an isomer group shall be defined by measuring a
standard mixture containing the first and last eluting isomers of each isomer group corresponding to the
used GC column. As an alternative, a fly ash extract or any other solution containing all native PCDD/PCDF
congeners can be used.
Table 3 — Masses for the detection and quantification of PCDD/PCDF in [amu]
Substance Dibenzofurans Dibenzo-p-dioxins
12 13 12 13
C C C C
12 1
...
Norme
internationale
ISO 13914
Troisième édition
Sols, biodéchets traités et boues —
2026-03
Dosage des dioxines, des furanes
et des polychlorobiphényles de
type dioxine par chromatographie
en phase gazeuse avec détection
sélective de masse (spectrométrie
de masse à haute résolution ou
SMHR, et spectrométrie de masse en
tandem ou SM/SM)
Soil, treated biowaste and sludge — Determination of dioxins
and furans and dioxin-like polychlorinated biphenyls by gas
chromatography with mass selective detection (high resolution
mass spectrometry, HRMS, and tandem mass spectrometry, MS/
MS)
Numéro de référence
DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2026
Tous droits réservés. Sauf prescription différente ou nécessité dans le contexte de sa mise en œuvre, aucune partie de cette
publication ne peut être reproduite ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique,
y compris la photocopie, ou la diffusion sur l’internet ou sur un intranet, sans autorisation écrite préalable. Une autorisation peut
être demandée à l’ISO à l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Case postale 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Genève
Tél.: +41 22 749 01 11
E-mail: copyright@iso.org
Web: www.iso.org
Publié en Suisse
ii
Sommaire Page
Avant‑propos .v
Introduction .vi
1 Domaine d’application . 1
2 Références normatives . 2
3 Termes et définitions . 2
4 Abréviations . 3
5 Principe. 3
6 Réactifs . 4
6.1 Produits chimiques .4
6.2 Étalons .4
7 Appareillage . 4
7.1 Généralités .4
7.2 Matériel pour la préparation des échantillons .4
7.3 Extracteur Soxhlet .5
7.4 Matériel de purification .5
7.5 Matériel de concentration . .5
7.6 Autres équipements .6
8 Conservation et prétraitement des échantillons . 6
8.1 Conservation des échantillons .6
8.2 Prétraitement des échantillons .6
9 Extraction et purification . . 6
9.1 Généralités .6
9.2 Extraction .7
9.3 Purification .8
9.3.1 Généralités .8
9.3.2 Chromatographie par perméation de gel .8
9.3.3 Colonne multicouche .8
9.3.4 Traitement à l’acide sulfurique .8
9.3.5 Colonne de charbon actif .9
9.3.6 Colonne d’oxyde d’aluminium .9
9.3.7 Élimination du soufre .9
9.4 Concentration finale de l’extrait d’échantillon purifié .9
9.5 Ajout des étalons internes de rendement .10
10 Analyse par CG/SM (SMHR ou SM/SM) .10
10.1 Généralités .10
10.2 Analyse par chromatographie en phase gazeuse .10
10.3 Détection par spectrométrie de masse .10
10.3.1 Généralités .10
10.3.2 Spectrométrie de masse haute résolution (SMHR) .10
10.3.3 Spectrométrie de masse en tandem (SM/SM) . 12
10.4 Exigences minimales relatives à l’identification des PCDD/PCDF et des PCB . 13
10.5 Exigences minimales relatives à la quantification des PCDD/PCDF et des PCB .14
10.6 Étalonnage du système CG/SM (SMHR ou SM/SM) . . 15
10.6.1 Généralités . 15
10.6.2 Étalonnage des analytes . 15
10.6.3 Étalonnage pour la somme de groupes homologues . 15
10.7 Quantification des résultats d’analyse par CG/SM (SMHR ou SM/SM) .16
10.7.1 Quantification des concentrations des analytes .16
10.7.2 Détermination des taux de récupération des étalons marqués au C .17
10.7.3 Quantification de la somme de groupes homologues .17
10.7.4 Calcul de l’équivalent toxique .17
iii
10.7.5 Calcul des limites de détection et de quantification .18
11 Expression des résultats .18
12 Fidélité .18
13 Rapport d’essai . 19
Annexe A (informative) Facteur d’équivalence toxique (TEF) .20
Annexe B (informative) Exemple de méthodes d’extraction et de purification .22
Annexe C (informative) Exemples de dosages par CG/SMHR .29
Annexe D (informative) Données de répétabilité et de reproductibilité .33
Bibliographie .40
iv
Avant‑propos
L’ISO (Organisation internationale de normalisation) est une fédération mondiale d’organismes nationaux
de normalisation (comités membres de l’ISO). L’élaboration des Normes internationales est en général
confiée aux comités techniques de l’ISO. Chaque comité membre intéressé par une étude a le droit de faire
partie du comité technique créé à cet effet. Les organisations internationales, gouvernementales et non
gouvernementales, en liaison avec l’ISO participent également aux travaux. L’ISO collabore étroitement avec
la Commission électrotechnique internationale (IEC) en ce qui concerne la normalisation électrotechnique.
Les procédures utilisées pour élaborer le présent document et celles destinées à sa mise à jour sont
décrites dans les Directives ISO/IEC, Partie 1. Il convient, en particulier de prendre note des différents
critères d’approbation requis pour les différents types de documents ISO. Le présent document
a été rédigé conformément aux règles de rédaction données dans les Directives ISO/IEC, Partie 2
(voir www.iso.org/directives).
L’ISO attire l’attention sur le fait que la mise en application du présent document peut entraîner l’utilisation
d’un ou de plusieurs brevets. L’ISO ne prend pas position quant à la preuve, à la validité et à l’applicabilité
de tout droit de brevet revendiqué à cet égard. À la date de publication du présent document, l’ISO n’avait
pas reçu notification qu’un ou plusieurs brevets pouvaient être nécessaires à sa mise en application.
Toutefois, il y a lieu d’avertir les responsables de la mise en application du présent document que des
informations plus récentes sont susceptibles de figurer dans la base de données de brevets, disponible à
l’adresse www.iso.org/brevets. L’ISO ne saurait être tenue pour responsable de ne pas avoir identifié de tels
droits de brevet et averti de leur existence.
Les appellations commerciales éventuellement mentionnées dans le présent document sont données pour
information, par souci de commodité, à l’intention des utilisateurs et ne sauraient constituer un engagement.
Pour une explication de la nature volontaire des normes, la signification des termes et expressions
spécifiques de l’ISO liés à l’évaluation de la conformité, ou pour toute information au sujet de l’adhésion de
l’ISO aux principes de l’Organisation mondiale du commerce (OMC) concernant les obstacles techniques au
commerce (OTC), voir www.iso.org/avant-propos.
Le présent document a été élaboré par le comité technique ISO/TC 190, Qualité du sol, sous-comité SC 3,
Méthodes chimiques et caractéristiques physiques.
Cette troisième édition annule et remplace la deuxième édition (ISO 13914:2023), qui a fait l’objet d’une
révision technique.
Les principales modifications sont les suivantes:
— ajout de la spectrométrie de masse en tandem (SM/SM) comme mode d’analyse par chromatographie en
phase gazeuse pouvant remplacer la spectrométrie de masse haute résolution (SMHR);
— ajout de résultats de validation pour la SM/SM.
Il convient que l’utilisateur adresse tout retour d’information ou toute question concernant le présent
document à l’organisme national de normalisation de son pays. Une liste exhaustive desdits organismes se
trouve à l’adresse www.iso.org/fr/members.html.
v
Introduction
Il existe deux groupes d’éthers aromatiques chlorés apparentés, connus sous les noms de
polychlorodibenzo‑p‑dioxines (PCDD) et polychlorodibenzofuranes (PCDF). Ces groupes comprennent un
total de 210 substances individuelles (congénères): 75 PCDD et 135 PCDF dont 17 sont substitués par le
chlore en position 2,3,7,8.
Il existe un groupe de composés aromatiques chlorés similaires aux polychlorodibenzo-p-dioxines (PCDD)
et aux polychlorodibenzofuranes (PCDF), connu sous le nom de polychlorobiphényles (PCB); ce groupe
comprend 209 substances individuelles.
Les PCDD et les PCDF peuvent se former lors de la combustion de matériaux organiques; ils peuvent
être également générés en tant que sous-produits indésirables lors de la fabrication ou du traitement
ultérieur de produits chimiques organiques chlorés. Les PCDD/PCDF sont introduits dans l’environnement
principalement par ces voies d’émission et par l’utilisation de matériaux contaminés. En fait, ils sont
universellement présents à de très faibles concentrations. Les congénères chlorosubstitués en position 2,3,7,8
sont importants d’un point de vue toxicologique. Les 74 dibenzo-p-dioxines/dibenzofuranes monochlorés
à trichlorés sont beaucoup moins significatifs du point de vue toxicologique que les dibenzo‑p‑dioxines/
dibenzofuranes tétrachlorés à octachlorés.
Les PCB ont été produits pendant une cinquantaine d’années, jusqu’à la fin des années 1990, période durant
laquelle ils ont été utilisés à diverses fins dans des systèmes ouverts ou fermés, par exemple, comme
isolants électriques ou fluides diélectriques dans les condensateurs et les transformateurs, comme fluides
hydrauliques spéciaux, ou encore comme agents plastifiants pour matériaux d’étanchéité. La production
mondiale de PCB a dépassé le million de tonnes.
Les PCDD/PCDF et les PCB sont émis lors de processus thermiques, par exemple l’incinération des déchets.
En 1997, un groupe d’experts de l’Organisation mondiale de la santé (OMS) a établi des facteurs d’équivalence
toxique (TEF, pour toxicity equivalent factors") pour les PCDD et douze PCB connus sous le nom de PCB de
type dioxine (PCB-DL, voir l’Annexe A). Ces douze PCB de type dioxine comptent quatre PCB non-ortho et
huit PCB mono-ortho (aucun ou un seul atome de chlore aux positions 2, 2’, 6 et 6’) de structure planaire ou
principalement planaire. La contribution des PCB de type dioxine au TEQ-OMS total peut être très élevée.
Il convient que la méthode décrite dans le présent document ne soit appliquée que par des opérateurs
compétents ayant suivi une formation adéquate sur la manipulation des composés hautement toxiques.
vi
Norme internationale ISO 13914:2026(fr)
Sols, biodéchets traités et boues — Dosage des dioxines, des
furanes et des polychlorobiphényles de type dioxine par
chromatographie en phase gazeuse avec détection sélective
de masse (spectrométrie de masse à haute résolution ou
SMHR, et spectrométrie de masse en tandem ou SM/SM)
AVERTISSEMENT — Il convient que l’utilisateur du présent document maîtrise les pratiques
courantes de laboratoire. Le présent document ne prétend pas couvrir tous les problèmes de sécurité
potentiels associés à son utilisation. Il incombe à l’utilisateur de mettre en place des pratiques de
santé et de sécurité appropriées.
IMPORTANT — Les essais réalisés conformément au présent document doivent être exécutés par du
personnel ayant reçu une formation adéquate.
1 Domaine d’application
Le présent document spécifie une méthode de dosage quantitatif de 17 polychlorodibenzo‑p‑dioxines et
polychlorodibenzofuranes substitués par des atomes de chlore en 2,3,7,8 ainsi que de polychlorobiphényles
de type dioxine dans les boues, les biodéchets traités et les sols à l’aide de méthodes de purification
chromatographique sur colonne en phase liquide et d’analyse par chromatographie en phase gazeuse avec
spectrométrie de masse à haute résolution (CG/SMHR). La détection par spectrométrie de masse en tandem
(SM/SM) peut être utilisée de manière équivalente.
Les analytes qui peuvent être dosés avec la méthode spécifiée dans le présent document sont répertoriés
dans le Tableau 1.
Tableau 1 — Analytes et leurs abréviations
Substance Abréviation
Tétrachlorodibenzo-p-dioxine TCDD
Pentachlorodibenzo-p-dioxine PeCDD
Hexachlorodibenzo-p-dioxine HxCDD
Heptachlorodibenzo-p-dioxine HpCDD
Octachlorodibenzo-p-dioxine OCDD
Tétrachlorodibenzofurane TCDF
Pentachlorodibenzofurane PeCDF
Hexachlorodibenzofurane HxCDF
Heptachlorodibenzofurane HpCDF
Octachlorodibenzofurane OCDF
Polychlorobiphényle PCB
a
Trichlorobiphényle TCB
Tétrachlorobiphényle TeCB
Pentachlorobiphényle PeCB
a
Groupes ne contenant aucun PCB-DL, donnés uniquement à titre d’information.
TTabableleaauu 1 1 ((ssuuiitte)e)
Substance Abréviation
Hexachlorobiphényle HxCB
Heptachlorobiphényle HpCB
a
Décachlorobiphényle DecaCB
a
Groupes ne contenant aucun PCB-DL, donnés uniquement à titre d’information.
La limite de détection dépend du type d’échantillon, du congénère, de l’équipement utilisé et de la qualité
des produits chimiques utilisés pour l’extraction et la purification. Dans les conditions spécifiées dans le
présent document, des limites de détection supérieures à 1 ng/kg (exprimées en matière sèche) peuvent être
atteintes.
Cette méthode est "fondée sur les performances". Elle peut être modifiée si tous les critères de performance
donnés dans la méthode sont respectés.
Le présent document est applicable pour plusieurs types de matrices et validé pour les boues de stations
d’épuration des eaux usées urbaines (voir également l’Annexe D pour les résultats de la validation).
NOTE En principe, cette méthode peut être également appliquée pour les sédiments, les déchets minéraux et les
végétaux. Il appartient à l’utilisateur du présent document de valider l’application pour ces matrices. Pour les mesures
dans les matrices complexes telles que les cendres volantes adsorbées sur des végétaux, une amélioration de la
purification peut être nécessaire. Cela peut également s’appliquer aux sédiments et aux déchets minéraux.
2 Références normatives
Les documents suivants sont cités dans le texte de sorte qu’ils constituent, pour tout ou partie de leur
contenu, des exigences du présent document. Pour les références datées, seule l’édition citée s’applique. Pour
les références non datées, la dernière édition du document de référence s’applique (y compris les éventuels
amendements).
ISO 11465, Boues et matrices environnementales solides — Détermination de la teneur en résidu sec ou en eau et
calcul de la fraction massique de matière sèche
3 Termes et définitions
Pour les besoins du présent document, les termes et définitions suivants s’appliquent.
L’ISO et l’IEC tiennent à jour des bases de données terminologiques destinées à être utilisées en normalisation,
consultables aux adresses suivantes:
— ISO Online browsing platform: disponible à l’adresse https:// www .iso .org/ obp
— IEC Electropedia: disponible à l’adresse https:// www .electropedia .org/
3.1
étalon interne
homologue d’un analyte marqué au C , qui est ajouté aux échantillons avant l’extraction et par rapport
auquel les concentrations des analytes natifs sont calculées
[SOURCE: ISO 18073:2004, 3.1.5, modifié — "PCDD/PCDF substitués en positions 2,3,7,8" et "PCDD et PCDF"
ont été remplacés par "analytes".]
3.2
étalon interne de rendement
polychlorobiphényle de type dioxine (PCB-DL) et dibenzo-p-dioxine/dibenzofurane polychloré (PCDD/
PCDF) marqué au C , ajouté avant l’injection dans le chromatographe
[SOURCE: ISO 18073:2004, 3.1.12, modifié — "PCDD/PCDF chlorosubstitué en positions 2,3,7,8" a été
remplacé par "polychlorobiphényle de type dioxine (PCB‑DL) et dibenzo‑p‑dioxine/dibenzofurane
plolychloré (PCDD/PCDF)".]
4 Abréviations
PCB-DL polychlorobiphényle de type dioxine
MRM surveillance des réactions multiples
PCDD/PCDF ou PCDD/F polychlorodibenzo-p-dioxines/dibenzofuranes
I-TEQ équivalent toxique international
TEF‑OMS facteur d’équivalence toxique proposé par l’OMS en 2005
TEQ-OMS équivalent toxique obtenu en multipliant la masse déterminée par le TEF‑OMS
correspondant comprenant les PCDD, les PCDF et les PCB‑DL
NOTE 1 L’I‑TEQ est obtenu en multipliant la masse déterminée par l’I‑TEF correspondant comprenant les PCDD et
les PCDF (pour une description détaillée, voir l’Annexe A). Il convient de l’utiliser uniquement à des fins de comparaison
avec des données plus anciennes.
NOTE 2 Pour une description détaillée du TEF‑OMS, voir l’ Annexe A.
Pour une description détaillée du TEQ-OMS, voir l’ Annexe A. Il convient d’utiliser les termes TEQPCB-OMS
et TEQPCDD/PCDF‑OMS pour faire la distinction entre diverses classes de composés.
5 Principe
Le présent document repose sur l’utilisation de la chromatographie en phase gazeuse couplée à la
spectrométrie de masse combinée à la technique par dilution isotopique pour permettre la séparation,
la détection et la quantification de PCDD/PCDF et de PCB de type dioxine dans les boues, les biodéchets et
les sols. Pour la mise en œuvre de la méthode par dilution isotopique, au moins 16 des 17 étalons internes
de PCDD/PCDF marqués et 12 étalons internes de PCB marqués sont utilisés. Les extraits destinés aux
analyses par chromatographie en phase gazeuse couplée à la spectrométrie de masse (CG-SM) contiennent
au moins un étalon interne de rendement par groupe, c’est-à-dire au moins un C ‑PCDD/PCDF et un
C -PCB. L’utilisation d’un plus grand nombre d’étalons internes de rendement est recommandée. Les
paramètres de chromatographie en phase gazeuse fournissent des informations permettant l’identification
de congénères (position des atomes de chlore de substitution), tandis que les paramètres de spectrométrie
de masse permettent la différenciation entre des isomères ayant des nombres différents d’atomes de chlore
de substitution et entre des polychlorodibenzo-p-dioxines, des furanes et des PCB.
Les congénères de PCDD/PCDF et de PCB marqués au C sont ajoutés à l’échantillon avant l’extraction et
l’analyse par CG‑SM (SMHR ou SM/SM). Les pertes subies au cours de l’extraction et de la purification sont
détectées et compensées en utilisant ces congénères ajoutés comme étalons internes pour la quantification
conjointement avec les étalons internes de rendement qui sont ajoutés juste avant l’analyse. En ce qui
concerne le dosage de ces substances, il est nécessaire de séparer les PCB des PCDD/PCDF et inversement.
L’objectif principal du mode opératoire de purification de l’extrait d’échantillon brut est d’éliminer les
composants de la matrice d’échantillon susceptibles de surcharger la méthode de séparation, de perturber
la quantification ou d’avoir une incidence sérieuse sur la performance de la méthode d’identification et de
quantification ainsi que sur la séparation des PCDD/PCDF et des PCB de type dioxine. De plus, ce mode
opératoire assure la concentration des analytes dans l’extrait d’échantillon final. Les modes opératoires
d’extraction s’appuient généralement sur la méthode d’extraction Soxhlet ou sur des méthodes équivalentes
d’extraction d’échantillons séchés, de préférence lyophilisés. Généralement la purification de l’échantillon
est effectuée à l’aide de techniques chromatographiques en phase liquide à plusieurs colonnes utilisant
différents adsorbants. Le dosage des PCDD/PCDF et des PCB est basé sur la quantification par la méthode
par dilution isotopique en utilisant l’analyse par CG-SM (SMHR ou SM/SM).
6 Réactifs
6.1 Produits chimiques
Les solvants utilisés pour l’extraction et la purification doivent être de qualité "pour analyse de résidus" ou
équivalente; ils doivent être contrôlés par des blancs. Les adsorbants tels que l’oxyde d’aluminium, le gel de
silice, la terre de diatomées et autres adsorbants utilisés pour la purification doivent être au minimum de
qualité analytique et être préalablement purifiés et activés si nécessaire.
NOTE Voir l’Annexe B pour une liste spécifique de solvants et de produits chimiques.
6.2 Étalons
— Solution de dopage au C pour les PCDD/PCDF (étalon interne);
— solution de dopage au C pour les PCB (étalon interne);
— solutions d’étalonnage pour les PCDD/PCDF;
— solutions d’étalonnage pour les PCB;
— solution de dopage au C comme étalon interne de rendement pour les PCDD/PCDF
(remarque: les congénères types marqués au C utilisés comme étalons internes de rendement sont le
13 13 13
C -123789-Hexa-CDD, le C -1234-Tétra-CDD, le C ‑1234‑Tétra‑CDD ou ‑DF);
12 12 12
— solution de dopage au C comme étalon interne de rendement pour les PCB-DL.
NOTE Voir l’Annexe B pour des exemples de concentrations de solutions étalons.
7 Appareillage
7.1 Généralités
L’appareillage et le matériel énumérés ci-dessous sont censés satisfaire aux exigences minimales pour un
traitement "classique" d’échantillons par extraction Soxhlet et purification chromatographique sur colonne.
Des appareils et du matériel supplémentaires peuvent être nécessaires en raison des diverses méthodes
d’extraction et de purification d’échantillons.
7.2 Matériel pour la préparation des échantillons
7.2.1 Sorbonne de laboratoire, de taille suffisante pour contenir le matériel de préparation des
échantillons décrit ci-dessous.
7.2.2 Dessiccateur.
7.2.3 Balances, comprenant une balance analytique permettant de peser à 0,1 mg près et une balance à
plateau supérieur permettant de peser à 10 mg près.
7.2.4 Colonne de Snyder.
7.3 Extracteur Soxhlet
D’autres dispositifs automatisés de type Soxhlet peuvent également être utilisés, le cas échéant. Il appartient
à l’utilisateur du présent document de valider l’application pour ces autres dispositifs.
7.3.1 Soxhlet, d’un diamètre interne de 50 mm, d’une capacité de 150 ml ou 250 ml avec un ballon à fond
rond de 500 ml.
7.3.2 Cartouche, 43 mm × 123 mm, adaptée au Soxhlet.
7.3.3 Chauffe‑ballon hémisphérique, adapté au ballon à fond rond de 500 ml.
7.4 Matériel de purification
7.4.1 Pipettes jetables, soit des pipettes Pasteur jetables, soit des pipettes sérologiques jetables.
7.4.2 Colonnes en verre pour chromatographie ayant les dimensions suivantes:
— 150 mm de longueur × 8 mm de diamètre interne, avec un fritté en verre ou un tampon de laine de verre,
un réservoir de 250 ml et un robinet d’arrêt en verre ou en polytétrafluoroéthylène (PTFE);
— 200 mm de longueur × 15 mm de diamètre interne, avec un fritté en verre ou un tampon de laine de
verre, un réservoir de 250 ml et un robinet d’arrêt en verre ou en PTFE;
— 300 mm de longueur × 25 mm de diamètre interne, avec un fritté en verre ou un tampon de laine de
verre, un réservoir de 300 ml et un robinet d’arrêt en verre ou en PTFE.
7.4.3 Four, permettant de maintenir une température constante (±5 °C) dans la plage de 105 °C à 450 °C
pour l’étuvage et la conservation des adsorbants.
7.5 Matériel de concentration
D’autres dispositifs d’évaporation peuvent également être utilisés, le cas échéant. Il appartient à l’utilisateur
du présent document de valider l’application pour ces autres dispositifs.
7.5.1 Évaporateur rotatif, avec bain-marie à température variable et les éléments suivants:
— source de vide pour l’évaporateur rotatif, munie d’un robinet d’arrêt au niveau de l’évaporateur et d’un
manomètre à dépression;
— pompe et refroidisseur d’eau à recirculation, fournissant de l’eau de refroidissement à (9 ± 4) °C
(l’utilisation d’eau du robinet pour refroidir l’évaporateur nécessite d’en consommer de grandes quantités
et peut entraîner une irrégularité des performances due aux variations de température et de pression de
l’eau);
— ballons à fond rond, d’une capacité de 100 ml et de 500 ml ou plus, avec raccord en verre rodé compatible
avec l’évaporateur rotatif.
7.5.2 Dispositif de balayage à l’azote, avec un bain-marie contrôlé dans la plage de 30 °C à 60 °C ou un
flux chauffé d’azote ou d’un autre gaz inerte approprié, installé dans une sorbonne.
1)
7.5.3 Concentrateur de type Kuderna‑Danish .
1) Kuderna Danish est un exemple de produit adapté disponible sur le marché. Cette information est donnée à l’intention
des utilisateurs du présent document et ne signifie nullement que l’ISO approuve l’emploi du produit ainsi désigné.
7.5.4 Flacons à échantillons, répondant aux critères suivants:
— en verre ambré, capacité nominale de 2 ml à 5 ml, avec bouchon à vis revêtu de PTFE;
— en verre, capacité de 0,3 ml, coniques, avec bouchon à vis revêtu de PTFE ou bouchon serti.
7.6 Autres équipements
7.6.1 Chromatographe en phase gazeuse, équipé d’un injecteur "splitless" (sans division de flux) ou
"on column" (dans la colonne) ou à programmation de températures, pour utilisation avec des colonnes
capillaires, et avec programmation de la température du four permettant le maintien d’une température
constante.
7.6.2 Colonne CG pour PCDD/PCDF et pour la spécificité isomérique du 2,3,7,8‑TCDD (par exemple, 60 m
de longueur × 0,32 mm de diamètre interne; épaisseur de film de 0,25 µm; colonne capillaire en silice fondue
à phase liée composée de 5 % de phényle, 94 % de méthyle, 1 % de vinyle silicone).
7.6.3 Spectromètre de masse, avec ionisation par impact électronique de 28 eV à 80 eV, permettant un
balayage sélectif et répétitif d’au moins douze masses exactes pendant une période d’environ 1 s à haute
résolution (≥ 10 000 à 10 % du ratio pic‑vallée ou équivalent), ou ionisation par impact électronique de 70 eV
après analyse avec un système SM/SM triple quadripôle sur la surveillance spécifique de réactions multiples
(MRM) (voir le Tableau 5).
7.6.4 Système de traitement de données, permettant de collecter, d’enregistrer et de mémoriser des
données relatives à la spectrométrie de masse.
8 Conservation et prétraitement des échantillons
8.1 Conservation des échantillons
Il convient de conserver les échantillons dans des récipients adaptés munis d’un dispositif de fermeture en
matériau approprié, tel que le polytétrafluoroéthylène (PTFE). Les échantillons destinés à être congelés
peuvent être conservés dans des récipients en aluminium nettoyés au préalable par chauffage à 450 °C
pendant au moins 4 h ou par rinçage à l’aide d’un solvant non chloré.
Il convient de conserver les échantillons au froid et à l’abri de la lumière si nécessaire afin d’éviter toute
modification de la constitution de l’échantillon, par exemple en ce qui concerne la détermination de la teneur
en eau.
8.2 Prétraitement des échantillons
Sauf spécification contraire, il convient que le séchage et l’homogénéisation soient effectués conformément
à l’EN 16179. Conserver le matériau broyé dans un dessiccateur ou un récipient en verre muni d’un dispositif
de fermeture étanche.
La détermination de la teneur en eau doit être effectuée conformément à l’ISO 11465 ou à toute méthode
équivalente normalisée à l’échelle internationale.
9 Extraction et purification
9.1 Généralités
Le présent document décrit les exigences minimales à respecter pour l’extraction et la purification et fournit
des exemples d’utilisation. L’opérateur chargé de l’analyse peut utiliser l’un des modes opératoires décrits
ci-dessous et à l’Annexe B ou tout autre mode opératoire approprié.
Le dosage des PCDD/PCDF et des PCB‑DL est basé sur la quantification par la méthode par dilution isotopique
en utilisant l’analyse par CG-SM (SMHR ou SM/SM). Des congénères de PCB-DL marqués au C et de PCDD/
PCDF chlorosubstitués en position 2,3,7,8 sont ajoutés à différentes étapes du mode opératoire complet. Les
pertes subies au cours de l’extraction et de la purification peuvent être détectées et compensées en utilisant
ces congénères ajoutés comme étalons internes pour la quantification ainsi que comme étalons internes de
rendement ajoutés juste avant l’analyse par CG/SM. Cependant, en raison des différences possibles au niveau
des caractéristiques de liaison et d’adsorption entre les analytes natifs et les congénères marqués au C
ajoutés au cours de l’analyse, il n’est pas sûr que l’efficacité de l’extraction et la compensation des pertes
lors de la purification puissent être garanties. Par conséquent, les méthodes appliquées doivent en plus être
rigoureusement validées. Des exemples de méthodes d’extraction et de purification éprouvées sont donnés à
l’Annexe B. L’Annexe C donne des exemples de dosages par CG/SMHR.
L’objectif principal du mode opératoire de purification de l’extrait d’échantillon brut est d’éliminer les
composants de la matrice d’échantillon susceptibles de surcharger la méthode de séparation, de perturber
la quantification ou sinon d’avoir une incidence notable sur la performance de la méthode d’identification
et de quantification, et de séparer les PCB de type dioxine des PCDD/PCDF. De plus, ce mode opératoire
assure une concentration des analytes dans l’extrait d’échantillon final. Les modes opératoires d’extraction
reposent normalement sur l’extraction Soxhlet de la fraction inférieure à 2 mm (<2 mm) de l’échantillon
solide sec et broyé ou tamisé. Généralement la purification de l’échantillon est effectuée à l’aide de techniques
chromatographiques en phase liquide à plusieurs colonnes utilisant différents adsorbants.
En principe, il est possible d’utiliser toute méthode de purification permettant de récupérer les analytes
en quantités suffisantes. De plus, l’extrait d’échantillon final ne doit pas avoir d’incidence néfaste sur la
performance du système d’analyse ou sur l’étape de quantification. Toutefois, toutes les méthodes appliquées
doivent être rigoureusement vérifiées et doivent satisfaire à un ensemble d’exigences de validation de
méthode avant de pouvoir être utilisées (voir l’Annexe D). En outre, la vérification de la performance de la
méthode pour chaque échantillon individuel doit faire partie du protocole d’assurance qualité appliqué.
9.2 Extraction
La quantité d’échantillon utilisée pour l’extraction peut varier de 5 g à 50 g selon le niveau de contamination
prévu.
L’étalon interne, qui est l’un des congénères marqués au C répertoriés dans le Tableau 2, doit être
directement ajouté à l’échantillon avant extraction.
Le mode opératoire d’extraction est mis en œuvre par extraction Soxhlet au toluène. Il convient d’ajuster
la durée d’extraction en fonction du type et de la quantité d’échantillon utilisé. L’exigence minimale est de
50 cycles d’extraction (environ 12 h).
D’autres solvants ou d’autres méthodes, telles que l’extraction par liquide pressurisé, peuvent être également
utilisés mais il doit être démontré que leur performance est équivalente à celle de la méthode préconisée.
Tableau 2 — Congénères marqués au C utilisés comme étalons internes
Solution de dopage au C – étalon interne
Congénères de PCDD/PCDF Congénères de PCB
13 13
2,3,7,8- C -TCDD C -PCB 77
12 12
13 13
1,2,3,7,8- C -PeCDD C -PCB 81
12 12
13 13
1,2,3,4,7,8- C -HxCDD C -PCB 126
12 12
13 13
1,2,3,6,7,8- C -HxCDD C -PCB 169
12 12
13 a
1,2,3,7,8,9- C -HxCDD
13 13
1,2,3,4,6,7,8- C -HpCDD C -PCB 105
12 12
13 13
C -OCDD C -PCB 114
12 12
C -PCB 118
a 1,2,3,7,8,9- C -HxCDD est traditionnellement très utilisé comme étalon interne de rendement. Son utilisation en tant qu’étalon
interne n’est donc pas obligatoire.
TTabableleaauu 2 2 ((ssuuiitte)e)
Solution de dopage au C – étalon interne
Congénères de PCDD/PCDF Congénères de PCB
13 13
2,3,7,8- C ‑TCDF C -PCB 123
12 12
13 13
1,2,3,7,8- C ‑PeCDF C -PCB 156
12 12
13 13
2,3,4,7,8- C ‑PeCDF C -PCB 157
12 12
13 13
1,2,3,4,7,8- C ‑HxCDF C -PCB 167
12 12
13 13
1,2,3,6,7,8- C ‑HxCDF C -PCB 189
12 12
2,3,4,6,7,8- C ‑HxCDF
1,2,3,7,8,9- C ‑HxCDF
1,2,3,4,6,7,8- C ‑HpCDF
1,2,3,4,7,8,9- C ‑HpCDF
C ‑OCDF
a 1,2,3,7,8,9- C -HxCDD est traditionnellement très utilisé comme étalon interne de rendement. Son utilisation en tant qu’étalon
interne n’est donc pas obligatoire.
9.3 Purification
9.3.1 Généralités
Les méthodes de purification doivent permettre de préparer l’extrait d’échantillon de manière appropriée
pour le dosage quantitatif ultérieur. Les méthodes de purification doivent concentrer les PCDD/PCDF et
les PCB de type dioxine dans les extraits et éliminer les composants matriciels interférents présents dans
l’extrait brut.
Des modes opératoires de purification éprouvés doivent être utilisés, ceux‑ci comprenant habituellement
au moins deux des techniques suivantes qui peuvent être combinées selon un ordre quelconque.
L’Annexe B fournit une description détaillée de certains modes opératoires.
D’autres méthodes peuvent être également utilisées, mais il doit être prouvé que leur performance est
équivalente à celle des techniques décrites ci-dessous.
9.3.2 Chromatographie par perméation de gel
La plage de poids moléculaires étudiée pour les PCDD/PCDF et les PCB de type dioxine, située entre 200 g/
mol et 500 g/mol, peut être isolée des molécules et polymères de poids moléculaires plus élevés qui sont
susceptibles de surcharger d’autres méthodes de purification. Cette méthode peut être également utilisée
pour éliminer le soufre.
9.3.3 Colonne multicouche
La chromatographie liquide sur colonne multicouche, avec de la silice présentant différents niveaux d’activité
et modifications de surface, est utilisée. Les composés ayant des propriétés chimiques différentes de celles
des PCDD/PCDF et des PCB de type dioxine (PCB‑DL) peuvent être éliminés.
9.3.4 Traitement à l’acide sulfurique
Un traitement direct de l’extrait d’échantillon à l’acide sulfurique est possible, mais non recommandé en
raison du risque d’accident. De plus, il doit être effectué avec le plus grand soin pour éviter toute perte de
PCDD/PCDF et de PCB‑DL sur les surfaces carbonifères formées.
9.3.5 Colonne de charbon actif
La chromatographie d’adsorption sur colonne de charbon actif peut être utilisée pour séparer les molécules
de PCDD/PCDF planaires et de PCB coplanaires des molécules de PCB mono‑ortho et d’autres molécules non
planaires interférentes.
9.3.6 Colonne d’oxyde d’aluminium
La chromatographie liquide sur colonne d’oxyde d’aluminium de divers niveaux d’activité et d’acidité ou de
basicité est utilisée. Les composés interférents présentant de petites différences de polarité ou de structure
par rapport aux PCDD/PCDF et aux PCB de type dioxine peuvent être éliminés.
Par ailleurs, les colonnes d’oxyde d’aluminium peuvent être utilisées pour séparer les PCDD/PCDF des PCB
de type dioxine.
9.3.7 Élimination du soufre
L’élimination du soufre peut être effectuée en chauffant l’extrait à reflux avec de la poudre de cuivre ou par
chromatographie par perméation de gel.
9.4 Concentration finale de l’extrait d’échantillon purifié
Pour atteindre des limites de détection suffisantes, l’extrait d’échantillon purifié doit être concentré à un
volume de l’ordre de 25 µl à 100 µl avant quantification. Le solvant final doit être le nonane, le toluène ou un
autre solvant à point d’ébullition élevé.
Bien que les PCDD/PCDF et les PCB‑DL aient des points d’ébullition plutôt élevés (>320 °C), les mécanismes
de transfert en phase vapeur et la formation d’aérosols lors de l’évaporation du solvant peuvent entraîner
des pertes importantes lors de la concentration de volumes inférieurs à 10 ml.
...