Amendment 1 - Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems

Amendement 1 - Appareils électromédicaux - Partie 1-8: Exigences générales pour la sécurité de base et les performances essentielles - Norme collatérale: Exigences générales, essais et guide pour les systèmes d'alarme des appareils et des systèmes électromédicaux

General Information

Status
Published
Publication Date
27-Nov-2012
Current Stage
PPUB - Publication issued
Start Date
15-Nov-2012
Completion Date
28-Nov-2012
Ref Project

Relations

Buy Standard

Standard
IEC 60601-1-8:2006/AMD1:2012 - Amendment 1 - Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
English and French language
55 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


IEC 60601-1-8
Edition 2.0 2012-11
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
AMENDMENT 1
AMENDEMENT 1
Medical electrical equipment –
Part 1-8: General requirements for basic safety and essential performance –
Collateral standard: General requirements, tests and guidance for alarm systems
in medical electrical equipment and medical electrical systems

Appareils électromédicaux –
Partie 1-8: Exigences générales pour la sécurité de base et les performances
essentielles – Norme collatérale: Exigences générales, essais et guide pour
les systèmes d’alarme des appareils et des systèmes électromédicaux

IEC 60601-1-8:2006/A1:2012
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester.
If you have any questions about IEC copyright or have an enquiry about obtaining additional rights to this publication,
please contact the address below or your local IEC member National Committee for further information.

Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni
utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et les
microfilms, sans l'accord écrit de la CEI ou du Comité national de la CEI du pays du demandeur.
Si vous avez des questions sur le copyright de la CEI ou si vous désirez obtenir des droits supplémentaires sur cette
publication, utilisez les coordonnées ci-après ou contactez le Comité national de la CEI de votre pays de résidence.

IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé Fax: +41 22 919 03 00
CH-1211 Geneva 20 info@iec.ch
Switzerland www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.

About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.

Useful links:
IEC publications search - www.iec.ch/searchpub Electropedia - www.electropedia.org
The advanced search enables you to find IEC publications The world's leading online dictionary of electronic and
by a variety of criteria (reference number, text, technical electrical terms containing more than 30 000 terms and
committee,…). definitions in English and French, with equivalent terms in
It also gives information on projects, replaced and additional languages. Also known as the International
withdrawn publications. Electrotechnical Vocabulary (IEV) on-line.

IEC Just Published - webstore.iec.ch/justpublished Customer Service Centre - webstore.iec.ch/csc
Stay up to date on all new IEC publications. Just Published If you wish to give us your feedback on this publication
details all new publications released. Available on-line and or need further assistance, please contact the
also once a month by email. Customer Service Centre: csc@iec.ch.

A propos de la CEI
La Commission Electrotechnique Internationale (CEI) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.

A propos des publications CEI
Le contenu technique des publications de la CEI est constamment revu. Veuillez vous assurer que vous possédez
l’édition la plus récente, un corrigendum ou amendement peut avoir été publié.

Liens utiles:
Recherche de publications CEI - www.iec.ch/searchpub Electropedia - www.electropedia.org
La recherche avancée vous permet de trouver des Le premier dictionnaire en ligne au monde de termes
publications CEI en utilisant différents critères (numéro de électroniques et électriques. Il contient plus de 30 000
référence, texte, comité d’études,…). termes et définitions en anglais et en français, ainsi que
Elle donne aussi des informations sur les projets et les les termes équivalents dans les langues additionnelles.
publications remplacées ou retirées. Egalement appelé Vocabulaire Electrotechnique
International (VEI) en ligne.
Just Published CEI - webstore.iec.ch/justpublished
Service Clients - webstore.iec.ch/csc
Restez informé sur les nouvelles publications de la CEI.
Just Published détaille les nouvelles publications parues. Si vous désirez nous donner des commentaires sur
Disponible en ligne et aussi une fois par mois par email. cette publication ou si vous avez des questions
contactez-nous: csc@iec.ch.
IEC 60601-1-8
Edition 2.0 2012-11
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
AMENDMENT 1
AMENDEMENT 1
Medical electrical equipment –

Part 1-8: General requirements for basic safety and essential performance –

Collateral standard: General requirements, tests and guidance for alarm systems

in medical electrical equipment and medical electrical systems

Appareils électromédicaux –
Partie 1-8: Exigences générales pour la sécurité de base et les performances

essentielles – Norme collatérale: Exigences générales, essais et guide pour

les systèmes d’alarme des appareils et des systèmes électromédicaux

INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
CODE PRIX U
ICS 11.040.01 ISBN 978-2-83220-442-9

– 2 – 60601-1-8 Amend. 1 © IEC:2012
FOREWORD
This amendment has been prepared by subcommittee 62A: Common aspects of electrical
equipment used in medical practice, of IEC technical committee 62: Electrical equipment in
medical practice, and ISO subcommittee SC 3: Lung ventilators and related devices of
ISO technical committee 121: Anaesthetic and respiratory equipment.
The text of this amendment is based on the following documents:
FDIS Report on voting
62A/824/FDIS 62A/837/RVD
Full information on the voting for the approval of this amendment can be found in the report
on voting indicated in the above table. In ISO, the standard has been approved by 19 P-
members out of 21 having cast a vote.
The committee has decided that the contents of this amendment and the base publication will
remain unchanged until the stability date indicated on the IEC web site under
"http://webstore.iec.ch" in the data related to the specific publication. At this date, the
publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
_____________
INTRODUCTION TO THE AMENDMENT
The second edition of IEC 60601-1-8 was published in 2006. Since its publication, an issue
has been identified with respect to pulse and burst testing. In addition, issues have been
raised by IEC/62D/MT 22, Electromedical diagnostic and patient monitoring equipment, during
implementation of alarm system requirements in particular standards within their scope of
work.
At the Brussels meeting, IEC/SC 62A accepted a proposal, based on ISO/TC 121/SC 3
st
Resolution Orebro 6, to develop the 1 amendment to IEC 60601-1-8:2006 to address the
issues identified above. IEC/SC 62A – ISO/TC 121/SC 3 Joint Working Group 2, Alarms, was
reactivated as a maintenance team to develop this amendment.
Foreword
Add the following note at the end of the Foreword:
NOTE The attention of National Committees is drawn to the fact that equipment manufacturers and testing
organizations may need a transitional period following publication of a new, amended or revised IEC or
ISO publication in which to make products in accordance with the new requirements and to equip themselves for
conducting new or revised tests. It is the recommendation of the committee that the content of this publication be
adopted for mandatory implementation nationally not earlier than 3 years from the date of publication.

60601-1-8 Amend. 1 © IEC:2012 – 3 –
1.2 Object
In the first paragraph, change the print type of the defined terms "basic safety" and "essential
performance" to small caps.
1.3.1 IEC 60601-1
Replace the existing first dash with:
– "the general standard" designates IEC 60601-1 alone (latest edition including any
amendments);
2 Normative references
Replace the introductory paragraph with:
The following documents, in whole or in part, are normatively referenced in this document and
are indispensable for its application. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any
amendments) applies.
Replace the normative references for IEC 60417 and IEC 60601-1 with the following:
IEC 60417, Graphical symbols for use on equipment. Available from: symbols.info/equipment>
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic
safety and essential performance
Amendment 1:2012
Delete the normative references for IEC 60601-1-2, and IEC 60601-1-6.
Replace the normative reference to IEC 60651:1979 and its Amendment 1 (1993) and
Amendment 2 (2000) with the following:
IEC 61672-1:2002, Electroacoustics – Sound level meters – Part 1: Specifications
Add the following normative reference:
IEC 62366:2007, Medical devices – Application of usability engineering to medical devices
Replace the normative references for ISO 3744:1994 and ISO 7000:1989 with the following:
ISO 3744:2010, Acoustics – Determination of sound power levels and sound energy levels of
noise sources using sound pressure – Engineering methods for an essentially free field over a
reflecting plane
ISO 7000, Graphical symbols for use on equipment. Available from: symbols.info/equipment>
– 4 – 60601-1-8 Amend. 1 © IEC:2012
3 Terms and definitions
Replace the existing first paragraph with the following:
For the purposes of this document, the terms and definitions given in
IEC 60601-1:2005+A1:2012 and IEC 62366:2007, and the following terms and definitions
apply.
3.1
* ALARM CONDITION
Replace the existing definition with the following:
state of the ALARM SYSTEM when it has determined that a potential or actual HAZARDOUS
SITUATION exists for which OPERATOR awareness or response is required
NOTE 1 An ALARM CONDITION can be invalid, i.e. a FALSE POSITIVE ALARM CONDITION.
NOTE 2 An ALARM CONDITION can be missed, i.e. a FALSE NEGATIVE ALARM CONDITION.
Add the following new definition:
3.37
* ACKNOWLEDGED
state of an ALARM SYSTEM initiated by OPERATOR action, where the auditory ALARM SIGNAL
associated with a currently active ALARM CONDITION is inactivated until the ALARM CONDITION no
longer exists
NOTE 1 ACKNOWLEDGED only affects ALARM SIGNALS that are active at the time of the OPERATOR action.
NOTE 2 ACKNOWLEDGED can terminate after a predetermined time interval has elapsed.
6 Alarm systems
6.1 ALARM CONDITION
6.1.2 * ALARM CONDITION priority
Replace the existing title and text of this subclause including Table 1 with the following:
6.1.2 * Determination of ALARM CONDITIONS and assignment of priority
For each HAZARDOUS SITUATION where the MANUFACTURER has chosen to use an ALARM SYSTEM
as a means of RISK CONTROL, the MANUFACTURER shall assign an ALARM CONDITION and its
priority using Table 1.
For HAZARDOUS SITUATIONS where the onset of potential HARM is delayed and the potential
result of a failure to respond is discomfort or minor reversible injury, the MANUFACTURER may
determine that no ALARM CONDITION is required. In such cases, the MANUFACTURER may
implement an INFORMATION SIGNAL.
NOTE Not all LOW PRIORITY ALARM CONDITIONS require prompt notification of the OPERATOR. On this basis an
auditory ALARM SIGNAL or repeating auditory ALARM SIGNAL can be omitted, when appropriate, since the OPERATOR is
expected to check the ME EQUIPMENT at intervals. In the event that the OPERATOR does not check the ME EQUIPMENT
in a timely fashion, the ALARM CONDITION should escalate from LOW PRIORITY to MEDIUM PRIORITY or HIGH PRIORITY,
and can additionally increase the sound pressure level of the related auditory ALARM SIGNALS, as appropriate.
The priority of each ALARM CONDITION shall be disclosed in the instructions for use. Priorities
may be identified in groups.
Compliance is checked by inspection of the instructions for use and RISK MANAGEMENT FILE.

60601-1-8 Amend. 1 © IEC:2012 – 5 –
Table 1 – Determination of ALARM CONDITIONS and assignment of priorities
a
Potential result of failure Onset of potential HARM
to respond to the cause
b c d
Immediate Prompt Delayed
of ALARM CONDITION
Death or irreversible injury HIGH PRIORITY ALARM HIGH PRIORITY ALARM MEDIUM PRIORITY ALARM
e
CONDITION  CONDITION CONDITION
Reversible injury HIGH PRIORITY ALARM MEDIUM PRIORITY ALARM LOW PRIORITY ALARM
CONDITION CONDITION CONDITION
Discomfort or reversible MEDIUM PRIORITY ALARM LOW PRIORITY ALARM LOW PRIORITY ALARM
minor injury CONDITION CONDITION CONDITION, no ALARM
CONDITION or
INFORMATION SIGNAL
a
Onset of potential HARM refers to when an injury occurs and not to when it is manifested.
b
Having the potential for the event to develop within a period of time not usually sufficient for manual corrective
action.
c
Having the potential for the event to develop within a period of time usually sufficient for manual corrective
action.
d
Having the potential for the event to develop within an unspecified time greater than that given under “prompt”.
e
Where practicable, ME EQUIPMENT with a therapeutic function incorporates automatic safety mechanisms to
prevent immediate death or irreversible injury caused by the ME EQUIPMENT. See also appropriate particular
standards.
6.3.2 * Visual ALARM SIGNALS
6.3.2.2 Characteristics of visual ALARM SIGNALS
Restructure the content of this subclause by creating two new subclauses as follows:
6.3.2.2.1 * 4 m (distant) visual ALARM SIGNALS
Under this subclause place the existing text of 6.3.2.2 through Table 2.
In the second paragraph delete the last sentence reading "A
...