Amendment 1 - Medical electrical equipment - Medical electron accelerators - Functional performance characteristics

Specifies requirements related to the use of multi-element beam limiting devices (BLDs).

Amendement 1 - Appareils électromédicaux - Accélérateurs médicaux d'électrons - Caractéristiques fonctionnelles

Spécifie des exigences relatives à l'utilisation des dispositifs de limitation du faisceau multi-élément (DLF multi-élément).

General Information

Status
Published
Publication Date
30-Jul-2000
Current Stage
DELPUB - Deleted Publication
Completion Date
16-Oct-2007
Ref Project

Relations

Buy Standard

Standard
IEC 60976:1989/AMD1:2000 - Amendment 1 - Medical electrical equipment - Medical electron accelerators - Functional performance characteristics Released:7/31/2000 Isbn:2831853761
English and French language
19 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NORME CEI
INTERNATIONALE IEC
INTERNATIONAL
STANDARD
AMENDEMENT 1
AMENDMENT 1
2000-07
Amendement 1
Appareils électromédicaux –
Accélérateurs médicaux d'électrons –
Caractéristiques fonctionnelles
Amendment 1
Medical electrical equipment –
Medical electron accelerators –
Functional performance characteristics

 IEC 2000 Droits de reproduction réservés  Copyright - all rights reserved
International Electrotechnical Commission 3, rue de Varembé Geneva, Switzerland
Telefax: +41 22 919 0300 e-mail: inmail@iec.ch IEC web site http://www.iec.ch
CODE PRIX
Commission Electrotechnique Internationale
K
PRICE CODE
International Electrotechnical Commission
Pour prix, voir catalogue en vigueur
For price, see current catalogue

– 2 – 60976 amend. 1 © CEI:2000

AVANT-PROPOS
Le présent amendement a été établi par le sous-comité 62C: Appareils de radiothérapie, de

médecine nucléaire et de dosimétrie du rayonnement, du comité d'études 62 de la CEI:

Equipements électriques dans la pratique médicale.

Le texte de cet amendement est issu des documents suivants:

FDIS Rapport de vote
62C/276/FDIS 62C/278/RVD
Le rapport de vote indiqué dans le tableau ci-dessus donne toute information sur le vote ayant
abouti à l'approbation de cet amendement.
___________
Introduction à cet amendement
Depuis la publication de la présente Norme internationale en 1989 les DISPOSITIFS DE
LIMITATION DU FAISCEAU multi-élément (DLF multi-élément) ont été introduits sur le marché et
sont maintenant largement utilisés pour délimiter les contours des CHAMPS DE RAYONNEMENT,
avec ou sans utilisation d'un DLF complémentaire. Un DLF multi-élément peut délimiter des
CHAMPS DE RAYONNEMENT irréguliers ou des CHAMPS DE RAYONNEMENT rectangulaires,
symétriques ou asymétriques. Dans certains cas, le DLF multi-élément remplace une paire de
mâchoires d'un DLF standard, dans d'autres cas, le DLF multi-élément est utilisé comme un
troisième dispositif additionnel au DLF standard. En plus de délimiter des contours des CHAMPS
DE RAYONNEMENT X, un DLF multi-élément peut dans certains cas être utilisé pour délimiter les
contours des CHAMPS DE RAYONNEMENT ÉLECTRONIQUE avec ou sans APPLICATEUR DU FAISCEAU
D'ÉLECTRONS.
–––––––––––––
Page 6
PRÉFACE
Ajouter à la préface existante, avant la liste des publications de la CEI, le nouveau texte
suivant:
Cette norme, ainsi que la CEI 60977, doit être lue conjointement avec la CEI 60601-2-1.
Ajouter la nouvelle norme suivante à la liste des publications:
CEI 61217:1996, Appareils utilisés en radiothérapie – Coordonnées, mouvements et échelles

60976 Amend. 1 © IEC:2000 – 3 –

FOREWORD
This amendment has been prepared by subcommittee 62C: Equipment for radiotherapy,

nuclear medicine and radiation dosimetry, of IEC technical committee 62: Electrical equipment

in medical practice.
The text of this amendment is based on the following documents:

FDIS Report on voting
62C/276/FDIS 62C/278/RVD
Full information on the voting for the approval of this amendment can be found in the report on
voting indicated in the above table.
___________
Introduction to this amendment
Since this International Standard was first published in 1989, multi-element BEAM LIMITING
DEVICES (BLDs) have been introduced and are now widely used for determining the shape of the
RADIATION FIELD, with or without the use of back-up BLDs. A multi-element BLD may produce
rectangular, symmetrical or asymmetrical and irregular RADIATION FIELDS. In some cases, the
multi-element BLD takes the place of one or more of the standard BLD pairs of jaws, in other
cases, the multi-element BLD may serve as a tertiary device, in addition to the standard BLDs. In
addition to the shaping of X-RAY FIELDS, multi-element BLDs may, in some cases, be used for
shaping ELECTRON FIELDS, with or without ELECTRON BEAM APPLICATORS.
––––––––––––
Page 7
PREFACE
Add, to the existing preface and before the list of IEC publications, the following new text:
This standard, together with IEC 60977, should be read in conjunction with IEC 60601-2-1.

Add to the list of publications, the following new standard:
IEC 61217:1996; Radiotherapy equipment – Coordinates, movements and scales

– 4 – 60976 amend. 1 © CEI:2000

Page 12
3.2 CHAMPS DE RAYONNEMENT disponibles

Ajouter, à la fin de ce paragraphe, le nouveau texte suivant:

Pour un DLF multi-élément, les DOCUMENTS D'ACCOMPAGNEMENT doivent indiquer:

• le nombre d'éléments;
• leurs dimensions projetées à la DISTANCE NORMALE DE TRAITEMENT;

• les dimensions des CHAMPS DE RAYONNEMENT maximal et minimal ainsi que les valeurs de
leurs coordonnées le long des axes Xb et Yb (voir CEI 61217); et
• la position de ces CHAMPS DE RAYONNEMENT par rapport à l'AXE DE RÉFÉRENCE.
Page 14
Ajouter, après 3.8, le nouveau paragraphe 3.9 suivant:
3.9 Dimensions, formes, dégagement des DISPOSITIFS DE LIMITATION DU FAISCEAU
par rapport à la TÊTE RADIOGÈNE et dans la zone située entre la TÊTE RADIOGÈNE et
l'ISOCENTRE
Pour les DLF multi-éléments, les DOCUMENTS D'ACCOMPAGNEMENT doivent contenir un schéma
donnant les dimensions en centimètres de l'APPAREIL et faisant apparaître de façon détaillée:
• la DISTANCE NORMALE DE TRAITEMENT;
• les distances de la SOURCE DE RAYONNEMENT X (surface frontale de la CIBLE de
RAYONNEMENT X), ou fenêtre de RAYONNEMENT ÉLECTRONIQUE si cela est applicable, aux
extrémités les plus proches ou les plus éloignées de tout DLF, y compris de type multi-
élément,
• les épaisseurs de chaque DLF; et
• les dimensions et la position par rapport à la DISTANCE NORMALE DE TRAITEMENT ou de la
SOURCE DE RAYONNEMENT X:
– du plan d'accrochage sur la TÊTE RADIOGÈNE au plus près de la DISTANCE NORMALE DE
TRAITEMENT, sur lequel se fixent les ACCESSOIRES amovibles; et
– de toute combinaison des dispositifs fixes ou amovibles pour les ACCESSOIRES de la
TÊTE RADIOGÈNE, et les dispositifs de conformation du FAISCEAU DE RAYONNEMENT tels
que les APPLICATEURS DU FAISCEAU D'ÉLECTRONS, les FILTRES EN COIN, les blocs de

conformation du CHAMP DE RAYONNEMENT ou les mâchoires y compris ceux qui sont
utilisés en combinaison avec les DLF multi-élément.
Se rapporter à la figure 10 qui donne un exemple d'un dessin de la TÊTE RADIOGÈNE, avec un DLF
multi-élément (type de modèle tertiaire) et de plusieurs ACCESSOIRES pour le RAYONNEMENT X.
Lorsqu'un DLF multi-élément est utilisé pour le RAYONNEMENT ÉLECTRONIQUE, un dessin similaire
avec le DLF multi-élément, et/ou les APPLICATEURS DU FAISCEAU D'ÉLECTRONS, y compris tout
ACCESSOIRE pour le FAISCEAU DE RAYONNEMENT ÉLECTRONIQUE, doit aussi être fourni.

60976 Amend. 1 © IEC:2000 – 5 –

Page 13
3.2 Available RADIATION FIELDS

Add, at the end of this subclause, the following new text:

For a multi-element BLD, the ACCOMPANYING DOCUMENTS shall state:

• the number of elements;
• their dimensions projected to the NORMAL TREATMENT DISTANCE;

• the dimensions of their minimum and maximum RADIATION FIELD sizes, with coordinates
stated along axes Xb and Yb (see IEC 61217); and
• the location of these RADIATION FIELDS in relation to the REFERENCE AXIS.
Page 15
Add, after 3.8, the following new subclause 3.9:
3.9 Dimensions, shapes, clearances, within the RADIATION HEAD, and in the region
RADIATION HEAD to ISOCENTRE, of BEAM LIMITING DEVICES
For multi-element BLDs, the ACCOMPANYING DOCUMENTS shall include an EQUIPMENT layout
drawing with all dimensions indicated in centimetres, in which is detailed the following
information:
• the NORMAL TREATMENT DISTANCE;
• the distances from the X-RADIATION SOURCE (front surface of the X-RADIATION TARGET), or
ELECTRON RADIATION window, if applicable, to the proximal or distal surfaces of all BLDs,
including any of the multi-element type;
• the thicknesses of all BLDs; and
• the dimensions, and location, relative to the NORMAL TREATMENT DISTANCE or to the X-RADIA-
TION source, of
– the fixed RADIATION HEAD surface, proximal to the NORMAL TREATMENT DISTANCE, to which
demountable ACCESSORIES may be attached, and
– any combinations of demountable or fixed RADIATION HEAD ACCESSORY structures, and
RADIATION FIELD ELECTRON BEAM APPLICATORS WEDGE FILTERS
shaping devices such as , ,
RADIATION FIELD shaping blocks or jaws, including those used in conjunction with multi-
element BLDs.
See figure 10 which provides an example of the layout of a RADIATION HEAD, with multi-element
BLD (in this case a tertiary type) and several X-RADIATION ACCESSORIES. Where a multi-element
BLD is used with ELECTRON RADIATION, a similar layout diagram, with multi-element BLD and/or
ELECTRON BEAM APPLICATOR, and any other ELECTRON RADIATION ACCESSORIES included, shall
also be provided.
– 6 – 60976 amend. 1 © CEI:2000

Page 16
4.6 CHAMPS DE RAYONNEMENT
Ajouter, à la fin de ce paragraphe, le nouveau texte suivant:

Sauf indication contraire, toutes les mesures doivent être effectuées avec des CHAMPS DE

RAYONNEMENT symétriques carrés ou rectangulaires.

Les DOCUMENTS D'ACCOMPAGNEMENT doivent indiquer que lorsque des CHAMPS DE RAYONNEMENT

asymétriques et irréguliers sont utilisés, la caractéristique fonctionnelle correspondante peut

être différente de celle obtenue avec des CHAMPS DE RAYONNEMENT symétriques carrés ou
rectangulaires, par exemple dans le cas des articles 5, 6 et 7.
Pour les CHAMPS DE RAYONNEMENT délimités par un DLF multi-élément, les DOCUMENTS
D'ACCOMPAGNEMENT doivent aussi indiquer quels sont les critères à suivre concernant
l'utilisation de tout DLF réglable complémentaire.
Page 54
7.3 PÉNOMBRE des CHAMPS DE RAYONNEMENT
7.3.1 Informations pour l'UTILISATEUR
Ajouter, à la fin de ce paragraphe, le nouveau texte suivant:
Pour les CHAMPS DE RAYONNEMENT délimités par un DLF multi-élément, les DOCUMENTS
D'ACCOMPAGNEMENT doivent indiquer la largeur maximale de la PÉNOMBRE le long des médianes
pour:
• un CHAMP DE RAYONNEMENT de 10 cm × 10 cm positionné n'importe où dans le FAISCEAU DE
RAYONNEMENT disponible; et
• le CHAMP DE RAYONNEMENT maximal (carré ou rectangulaire) délimité par le DLF multi-
élément.
7.3.2 Essai
Ajouter, à la fin de ce paragraphe, le nouveau texte suivant:

Pour les CHAMPS DE RAYONNEMENT délimités par un DLF multi-élément, les essais concernant la
largeur de la PÉNOMBRE le long de leurs médianes, doivent être effectués:
• pour deux CHAMPS DE RAYONNEMENT
de 10 cm × 10 cm positionnés le long de la diagonale
du CHAMP DE RAYONNEMENT maximal tel qu'indiqué à la figure 11; et
• pour le CHAMP DE RAYONNEMENT maximal (carré ou rectangulaire).

60976 Amend. 1 © IEC:2000 – 7 –

Page 17
4.6 RADIATION FIELDS
Add, at the end of this subclause, the following new text:

Except as otherwise indicated, all measurements shall be performed with symmetrical

rectangular or square RADIATION FIELDS.

The ACCOMPANYING DOCUMENTS shall state that when asymmetrical and irregular RADIATION

FIELDS are considered, the functional performance characteristic under investigation may differ

from that obtained from symmetrical rectangular or square RADIATION FIELDS, as in the case, for
example, of clauses 5, 6, and 7.
For RADIATION FIELDS shaped by a multi-element BLD, the ACCOMPANYING DOCUMENTS shall also
state the criteria governing the use of any back-up adjustable BLDs.
Page 55
7.3 PENUMBRA of RADIATION FIELDS
7.3.1 Information to the USER
Add, at the end of this subclause, the following new text:
For RADIATION FIELDS shaped by a multi-element BLD, the ACCOMPANYING DOCUMENTS shall state
the maximum width of the PENUMBRA along the major axes of:
• a 10 cm × 10 cm RADIATION FIELD situated anywhere within the available RADIATION BEAM;
and
• the maximum (rectangular or square) multi-element BLD RADIATION FIELD.
7.3.2 Test
Add, at the end of this subclause, the following new text:
For RADIATION FIELDS shaped by a multi-element BLD, tests of the width of the PENUMBRA along
their major axes shall be made

• for two 10 cm × 10 cm RADIATION FIELDS placed along the diagonal of the maximum
RADIATION FIELD in the manner indicated by figure 11, and
• for the maximum (rectangular or square) RADIATION FIELD.

– 8 – 60976 amend. 1 © CEI:2000

Page 56
8.1 RAYONNEMENT X
8.1.1 Indication numérique du champ

Ajouter, à la fin de ce paragraphe, le nouveau texte suivant:

Les DLF multi-éléments doivent comporter les indications numériques des coordonnées de la
position projetée à la DISTANCE NORMALE DE TRAITEMENT, du bord de chaque élément, comme

cela est défini par la CEI 61217. Une indication de la dimension du CHAMP DE RAYONNEMENT

délimité par chacune des paires d'éléments opposés peut également être fournie.
8.1.1.1 Informations pour l'UTILISATEUR
Ajouter, à la fin de ce paragraphe, le nouveau texte suivant:
Pour les DLF multi-éléments, les informations demandées dans ce paragraphe doivent être
données pour les CHAMPS DE RAYONNEMENT spécifiés en 7.3.1.
8.1.1.2 Essai
Ajouter, à la page 60, après le tableau 19, le nouveau texte suivant:
Pour les DLF multi-éléments, les mesures décrites dans ce paragraphe doivent être effectuées
pour les CHAMPS DE RAYONNEMENT spécifiés en 7.3.2, et pour l'ensemble des autres conditions
spécifiées dans les colonnes 2 et 3 du tableau 19.
Page 60
8.1.2 INDICATEUR LUMINEUX DU CHAMP
8.1.2.1 Informations pour l'UTILISATEUR
Ajouter, à la page 62, le nouveau texte suivant:
Pour les DLF multi-éléments, les informations demandées dans ce paragraphe doivent être
données pour les CHAMPS DE RAYONNEMENT spécifiés en 7.3.1.

Page 62
8.1.2.2 Essai
Ajouter, à la fin de ce paragraphe, le nouveau texte suivant:
Pour les DLF multi-éléments, les mesures décrites dans ce paragraphe doivent être effectuées
pour les CHAMPS DE RAYONNEMENT spécifiés en 8.1.1.2.

60976 Amend. 1 © IEC:2000 – 9 –

Page 57
8.1 X-RADIATION
8.1.1 Numerical field-indication

Add, at the end of this subclause, the following new text:

For a multi-element BLD there shall be numerical indications of the coordinates, as defined by
IEC 61217, of each element edge position projected to the NORMAL TREATMENT DISTANCE.

Additionally, the dimension of the RADIATION FIELD defined by each pair of opposed elements

may be indicated.
8.1.1.1 Information to the USER
Add, at the end of this subclause, the following new text:
For multi-element BLDs, the information required in this subclause shall be
...