IEC 60601-2-21:2009
(Main)Medical electrical equipment - Part 2-21: Particular requirements for the basic safety and essential performance of infant radiant warmers
Medical electrical equipment - Part 2-21: Particular requirements for the basic safety and essential performance of infant radiant warmers
IEC 60601-2-21:2009 specifies the safety requirements for infant radiant warmers. This particular standard amends and supplements IEC 60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance. This second edition cancels and replaces the first edition published in 1994 and its Amendment 1 (1996). This edition constitutes a technical revision. This edition of IEC 60601 2-21 was revised to structurally align with the 2005 edition of IEC 60601-1. The contents of the corrigendum of February 2013 have been included in this copy.
Appareils électromédicaux - Partie 2-21: Exigences particulières pour la sécurité de base et les performances essentielles des incubateurs radiants pour nouveau-nés
La CEI 60601-2-21:2009 spécifie les exigences de sécurité relatives aux incubateurs radiants pour nouveau-nés. La présente norme particulière modifie et complète la CEI 60601-1:2005, Appareils électromédicaux - Partie 1: Exigences générales pour la sécurité de base et les performances essentielles. Cette deuxième édition annule et remplace la première édition parue en 1994 et son Amendement 1 (1996). Cette édition constitue une révision technique. La présente édition de la CEI 60601-2-21 a été mise à jour de façon à correspondre structurellement à l'édition 2005 de la CEI 60601-1. Le contenu du corrigendum de février 2013 a été pris en considération dans cet exemplaire.
General Information
Relations
Standards Content (Sample)
IEC 60601-2-21 ®
Edition 2.0 2009-02
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Medical electrical equipment –
Part 2-21: Particular requirements for the basic safety and essential performance
of infant radiant warmers
Appareils électromédicaux –
Partie 2-21: Exigences particulières pour la sécurité de base et les
performances essentielles des incubateurs radiants pour nouveau-nés
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by
any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from either IEC or
IEC's member National Committee in the country of the requester.
If you have any questions about IEC copyright or have an enquiry about obtaining additional rights to this publication,
please contact the address below or your local IEC member National Committee for further information.
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite
ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie
et les microfilms, sans l'accord écrit de la CEI ou du Comité national de la CEI du pays du demandeur.
Si vous avez des questions sur le copyright de la CEI ou si vous désirez obtenir des droits supplémentaires sur cette
publication, utilisez les coordonnées ci-après ou contactez le Comité national de la CEI de votre pays de résidence.
IEC Central Office
3, rue de Varembé
CH-1211 Geneva 20
Switzerland
Email: inmail@iec.ch
Web: www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.
ƒ Catalogue of IEC publications: www.iec.ch/searchpub
The IEC on-line Catalogue enables you to search by a variety of criteria (reference number, text, technical committee,…).
It also gives information on projects, withdrawn and replaced publications.
ƒ IEC Just Published: www.iec.ch/online_news/justpub
Stay up to date on all new IEC publications. Just Published details twice a month all new publications released. Available
on-line and also by email.
ƒ Electropedia: www.electropedia.org
The world's leading online dictionary of electronic and electrical terms containing more than 20 000 terms and definitions
in English and French, with equivalent terms in additional languages. Also known as the International Electrotechnical
Vocabulary online.
ƒ Customer Service Centre: www.iec.ch/webstore/custserv
If you wish to give us your feedback on this publication or need further assistance, please visit the Customer Service
Centre FAQ or contact us:
Email: csc@iec.ch
Tel.: +41 22 919 02 11
Fax: +41 22 919 03 00
A propos de la CEI
La Commission Electrotechnique Internationale (CEI) est la première organisation mondiale qui élabore et publie des
normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications CEI
Le contenu technique des publications de la CEI est constamment revu. Veuillez vous assurer que vous possédez
l’édition la plus récente, un corrigendum ou amendement peut avoir été publié.
ƒ Catalogue des publications de la CEI: www.iec.ch/searchpub/cur_fut-f.htm
Le Catalogue en-ligne de la CEI vous permet d’effectuer des recherches en utilisant différents critères (numéro de référence,
texte, comité d’études,…). Il donne aussi des informations sur les projets et les publications retirées ou remplacées.
ƒ Just Published CEI: www.iec.ch/online_news/justpub
Restez informé sur les nouvelles publications de la CEI. Just Published détaille deux fois par mois les nouvelles
publications parues. Disponible en-ligne et aussi par email.
ƒ Electropedia: www.electropedia.org
Le premier dictionnaire en ligne au monde de termes électroniques et électriques. Il contient plus de 20 000 termes et
définitions en anglais et en français, ainsi que les termes équivalents dans les langues additionnelles. Egalement appelé
Vocabulaire Electrotechnique International en ligne.
ƒ Service Clients: www.iec.ch/webstore/custserv/custserv_entry-f.htm
Si vous désirez nous donner des commentaires sur cette publication ou si vous avez des questions, visitez le FAQ du
Service clients ou contactez-nous:
Email: csc@iec.ch
Tél.: +41 22 919 02 11
Fax: +41 22 919 03 00
IEC 60601-2-21 ®
Edition 2.0 2009-02
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Medical electrical equipment –
Part 2-21: Particular requirements for the basic safety and essential
performance of infant radiant warmers
Appareils électromédicaux –
Partie 2-21: Exigences particulières pour la sécurité de base et les
performances essentielles des incubateurs radiants pour nouveau-nés
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
W
CODE PRIX
ICS 11.040.10 ISBN 978-2-88910-219-8
– 2 – 60601-2-21 © IEC:2009
CONTENTS
FOREWORD.3
INTRODUCTION.5
201.1 Scope, object and related standards .6
201.2 Normative references .8
201.3 Terms and definitions.8
201.4 General requirements.10
201.5 General requirements for testing of ME EQUIPMENT.11
201.6 Classification of ME EQUIPMENT and ME SYSTEMS .11
201.7 ME EQUIPMENT identification, marking and documents.11
201.8 Protection against electrical HAZARDS from ME EQUIPMENT.13
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS.13
201.10 Protection against unwanted and excessive radiation HAZARDS.15
201.11 Protection against excessive temperatures and other HAZARDS.15
201.12 Accuracy of controls and instruments and protection against hazardous
outputs.16
201.13 HAZARDOUS SITUATIONS and fault conditions.20
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) .20
201.15 Construction of ME EQUIPMENT .20
201.16 ME SYSTEMS .22
*
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS .22
202 Electromagnetic compatibility – Requirements and tests .22
210 Requirements for the development of physiologic closed-loop controllers .22
Annexes .23
Annex AA (informative) Particular guidance and rationale.24
Bibliography.32
Index of defined terms used in this particular standard.34
Figure 201.101 – Layout of TEST DEVICES .9
Figure 201.102 – TEST DEVICE.10
Table 201.101 – Additional ESSENTIAL PERFORMANCE requirements.11
60601-2-21 © IEC:2009 – 3 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-21: Particular requirements for the basic safety and
essential performance of infant radiant warmers
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International standard IEC 60601-2-21 has been prepared by subcommittee 62D:
Electromedical equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
This second edition cancels and replaces the first edition published in 1994 and its
Amendment 1 (1996). This edition constitutes a technical revision. This edition of
IEC 60601-2-21 was revised to structurally align with the 2005 edition of IEC 60601-1.
The text of this particular standard is based on the following documents:
FDIS Report on voting
62D/735/FDIS 62D/762/RVD
Full information on the voting for the approval of this particular standard can be found in the
report on voting indicated in the above table.
– 4 – 60601-2-21 IEC:2009
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– Requirements and definitions: roman type.
– Test specifications: italic type.
– Informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
ERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
– T
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title Medical electrical
equipment, can be found on the IEC website.
The committee has decided that the contents of this publication will remain unchanged until
the maintenance result date indicated on the IEC web site under "http://webstore.iec.ch" in
the data related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
The contents of the corrigendum of February 2013 have been included in this copy.
60601-2-21 © IEC:2009 – 5 –
INTRODUCTION
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of INFANT RADIANT WARMER equipment.
This particular standard amends and supplements IEC 60601-1:2005, Medical electrical
equipment – Part 1: General requirements for basic safety and essential performance,
hereinafter referred to as the general standard.
The requirements are followed by specifications for the relevant tests.
A general guidance and rationale for the requirements of this particular standard are given in
Annex AA.
It is considered that knowledge of the reasons for these requirements will not only facilitate
the proper application of this particular standard but will, in due course, expedite any revision
necessitated by changes in clinical practice or as a result of developments in technology.
However, this annex does not form part of the requirements of this standard.
– 6 – 60601-2-21 © IEC:2009
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-21: Particular requirements for the basic safety and
essential performance of infant radiant warmers
201.1 Scope, object and related standards
1)
Clause 1 of the general standard applies, except as follows:
201.1.1 Scope
Replacement:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of INFANT
RADIANT WARMERS as defined in 201.3.204, also referred to as ME EQUIPMENT.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS
within the scope of this standard are not covered by specific requirements in this standard
except in 7.2.13 and 8.4.1 of the general standard.
NOTE See also 4.2 of the general standard.
This particular standard specifies the safety requirements for INFANT RADIANT WARMERS, but
alternate methods of compliance with a specific clause, by demonstrating equivalent safety,
will not be judged as non-compliant, if the MANUFACTURER has demonstrated in his RISK
MANAGEMENT FILE that the RISK presented by the HAZARD has been found to be of an
acceptable level when weighed against the benefit of treatment from the device.
This particular standard does not apply to:
– devices supplying heat via BLANKETS, PADS or MATTRESSES in medical use; for information
see IEC 80601-2-35;
– INFANT INCUBATORS; for information see IEC 60601-2-19;
– INFANT TRANSPORT INCUBATORS, for information see IEC 60601-2-20;
– INFANT PHOTOTHERAPY EQUIPMENT, for information see IEC 60601-2-50.
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements for INFANT RADIANT WARMERS as defined in 201.3.204, which
minimize HAZARDS to PATIENT and OPERATOR, and to specify tests by which compliance with
the requirements can be verified.
—————————
1)
The general standard is IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for
basic safety and essential performance.
60601-2-21 © IEC:2009 – 7 –
201.1.3 *Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 2 of this particular standard.
IEC 60601-1-2 and IEC 60601-1-10 apply as modified in Clauses 202 and 210 respectively.
IEC 60601-1-3 does not apply. All other published collateral standards in the IEC 60601-1
series apply as published.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard and collateral standards as appropriate for the particular ME
EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL PERFORMANCE
requirements.
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this standard addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the 60601-1-2 collateral standard,
203.4 in this particular standard addresses the content of Clause 4 of the 60601-1-3 collateral
standard, etc.). The changes to the text of the general standard are specified by the use of
the following words:
"Replacement" means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
"Addition" means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
"Amendment" means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However, due to the fact that definitions in the general
standard are numbered 3.1 through 3.139, additional definitions in this standard are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term "this standard" is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
– 8 – 60601-2-21 © IEC:2009
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
Clause 2 of the general standard applies, except as follows:
Amendment:
IEC 60601-1-2:2007, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral standard: Electromagnetic compatibility –
Requirements and tests
Addition:
IEC 60601-1-10:2007, Medical electrical equipment – Part 1-10: General requirements for
basic safety and essential performance – Collateral Standard: Requirements for the
development of physiologic closed-loop controllers
NOTE Informative references are listed in the bibliography beginning on page 32.
201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in the general standard
apply, except as follows:
NOTE An index of defined terms is found beginning on page 34.
Addition:
201.3.201
BABY CONTROLLED RADIANT WARMER
mode of operation in which the power output varies automatically in order to maintain the
SKIN TEMPERATURE SENSOR according to the CONTROL
temperature as measured by a
TEMPERATURE set by the OPERATOR
NOTE An INFANT RADIANT WARMER operating as a BABY CONTROLLED RADIANT WARMER is a PHYSIOLOGIC CLOSED-
LOOP CONTROLLER as defined in IEC 60601-1-10.
201.3.202
CONTROL TEMPERATURE
temperature selected at the temperature control
201.3.203
INFANT
PATIENT up to 3 months and with a weight of less than 10 kg
201.3.204
INFANT RADIANT WARMER
electrically powered device with a radiant heating source intended to maintain the thermal
balance of an INFANT by direct radiation of energy in the infrared region of the electromagnetic
spectrum
60601-2-21 © IEC:2009 – 9 –
201.3.205
MANUAL MODE
mode of operation in which the heater output is either at a fixed level or a proportion of its
maximum output set by the OPERATOR
201.3.206
MID-POINT AVERAGE TEMPERATURE
T
M
TEST DEVICE AVERAGE TEMPERATURE of the TEST DEVICE positioned at the mid-point of the
INFANT RADIANT WARMER mattress (See Figure 201.101)
C
D
M
A
B
IEC 255/09
Key
1 = Mattress
Figure 201.101 – Layout of TEST DEVICES
201.3.207
*
PREWARM MODE
mode of operation in which the heater output is maintained at a preset level (set by the
MANUFACTURER) for the purpose of pre-warming the INFANT RADIANT WARMER and maintaining
the level of warmth of the INFANT RADIANT WARMER prior to an INFANT being placed on the
device
201.3.208
SKIN TEMPERATURE SENSOR
sensing device intended to measure the INFANT’S SKIN TEMPERATURE
201.3.209
STEADY TEMPERATURE CONDITION
condition which is reached when the temperature, measured at the centre of the TEST DEVICE
positioned on the mid-point of the INFANT RADIANT WARMER mattress, does not vary by more
than 1 °C over a period of 1 h
201.3.210
TEST DEVICE
totally matt blackened disc used as a reproducible receiver of radiant energy during testing of
the INFANT RADIANT WARMER (see Figure 201.102)
– 10 – 60601-2-21 © IEC:2009
Dimensions in millimeters
50 ± 2
∅100 ± 2
IEC 372/09
Key
1 = Mattress
Surface finish: non-reflective black paint
Disc mass 500 g ± 10 g
3 3
Disc material: aluminum of density within the range 2,6 g/cm and 2,9 g/cm
Figure 201.102 – TEST DEVICE
201.3.211
TEST DEVICE AVERAGE TEMPERATURE
(T , T , T , T OR T )
1 2 3 4 M
average temperature reading taken during a STEADY TEMPERATURE CONDITION at regular
intervals at the centre of a TEST DEVICE
NOTE T is expressed in °C.
M
201.3.212
*
TEST LOAD
array of five TEST DEVICES used in a specified configuration (see Figure 201.101) for
performance tests of the INFANT RADIANT WARMER
201.4 General requirements
Clause 4 of the general standard applies, except as follows:
201.4.1 Conditions for application to ME EQUIPMENT OR ME SYSTEMS
Addition:
For ME EQUIPMENT or ME SYSTEMS, which combines alternative heat sources, for instance
incubators with integrated RADIANT WARMERS, devices supplying heat via BLANKETS, PADS or
MATTRESSES etc., safety requirements of other relevant particular standards shall be
considered. Further the safety requirements of this particular standard shall be fulfilled with
the combination of the other equipment, which is approved by the MANUFACTURER, as stated in
the ACCOMPANYING DOCUMENTS according to Clause 16 (ME SYSTEMS).
Compliance is checked by the test of Clause 201.11 and subclause 201.15.4.2.1 of the
relevant particular standards (e.g. IEC 60601-2-19 etc).
201.4.3 ESSENTIAL PERFORMANCE
Addition:
23 ± 1
∅5 max.
60601-2-21 © IEC:2009 – 11 –
*
201.4.3.101 Additional ESSENTIAL PERFORMANCE requirements
Additional ESSENTIAL PERFORMANCE requirements are found in the subclauses listed in Table
201.101.
Table 201.101 – Additional ESSENTIAL PERFORMANCE requirements
Requirement Subclause
ESSENTIAL PERFORMANCE requirement 1 201.12.1.103, and
generation of a visual
and audible alarm in
compliance with
201.15.4.2.1
201.5 General requirements for testing of ME EQUIPMENT
Clause 5 of the general standard applies, except as follows:
201.5.3 * Ambient temperature, humidity, atmospheric pressure
Addition to item a):
The ME EQUIPMENT shall comply with the requirements of this standard when operating
within the following conditions:
– an ambient temperature within the range 18 °C to 30 °C;
– an ambient air velocity is less than 0,3 m/s.
If not otherwise specified in this particular standard, all tests shall be carried out at an
ambient temperature within the range of 21 °C to 26 °C.
201.5.4 Other conditions
Addition:
If not otherwise specified, the CONTROL TEMPERATURE shall be 36 °C ± 1 °C and shall always
exceed the ambient temperature by at least 3 °C.
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies.
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
201.7.2 Marking on the outside of ME EQUIPMENT or ME EQUIPMENT parts
(see also Table C.1 of the general standard)
Additional subclauses:
*
201.7.2.101 Oxygen monitor
An INFANT RADIANT WARMER not equipped with an integral oxygen monitor and which provides
means for oxygen administration shall be marked in a prominent position with a text which
states: "Use an oxygen monitor when oxygen is administered".
– 12 – 60601-2-21 © IEC:2009
201.7.2.102 Distance markings
The INFANT RADIANT WARMER without integral bed areas shall be permanently and clearly
marked with an indication of the permissible distances between the INFANT RADIANT WARMER
heating systems and any mattress.
201.7.4.2 Control devices
Addition:
Means shall be provided for the clear selection and indication of CONTROL TEMPERATURE on or
adjacent to the controls. The means provided shall allow resolution at intervals not greater
than 0,2 °C.
201.7.9.2.2 Warning and safety notices
Addition:
The instructions for use shall additionally contain:
a) a statement that independent monitoring of the temperature of the INFANT by the OPERATOR
is essential and it is inadvisable to leave an INFANT unattended under the INFANT RADIANT
WARMER;
b) recommendations on the permissible distances between the INFANT RADIANT WARMER
heating system and any mattress used with it, and a statement on the effects which any
changes in this distance may have;
c) instructions on the recommended positions and methods of use and attachment of the
temperature sensors provided for use with the INFANT RADIANT WARMER;
d) for INFANT RADIANT WARMER with TYPE B APPLIED PART in which the INFANT might not be
isolated from earth, a warning that particular care shall be taken to ensure that additional
equipment connected to the INFANT is electrically safe;
e) if applicable, a recommendation to the OPERATOR to inspect regularly latches and closing
devices of barriers to prevent the INFANT falling out;
f) a statement of the maximum loads which can be applied to all supports and mounting
brackets for ACCESSORIES and ancillary equipment;
*g) information on the effects on the functioning of the INFANT RADIANT WARMER of detachment
of the SKIN TEMPERATURE SENSOR from the PATIENT skin;
h) if applicable, a statement that the tilting of the mattress from its horizontal position relative
to the INFANT RADIANT WARMER heater can affect the performance of the INFANT RADIANT
WARMER (see 201.12.1.102);
i) a statement that ACCESSORIES, e.g. for phototherapy or heated mattresses, or sunlight can
cause an increase in INFANT temperature to dangerous levels
j) a statement that the INFANT RADIANT WARMER is not suitable for use in the presence of
flammable anaesthetic gases or other flammable materials, such as some types of
cleaning fluids;
k) a statement that rectal temperatures are not appropriate for controlling the heater output
of the INFANT RADIANT WARMER;
*l) a statement that the INFANT RADIANT WARMER cannot differentiate between an increase in
core temperature with a cold skin (fever) and a low core and SKIN TEMPERATURE
(hypothermia), and a recommendation to monitor the temperature of the PATIENT;
m) a statement that environmental conditions (e.g. air movement) can affect the thermal
balance of the INFANT;
*n) a statement that an INFANT RADIANT WARMER shall be used only by appropriately trained
personnel and under the direction of qualified medical personnel who are familiar with
currently known RISKS and benefits of radiant warmer use;
60601-2-21 © IEC:2009 – 13 –
o) a statement that an INFANT RADIANT WARMER can increase the PATIENT’S insensible water
loss;
p) concentration of carbon dioxide (CO ): If the mattress of an INFANT RADIANT WARMER is
fitted with a COMPARTMENT which encloses the baby, the MANUFACTURER shall specify (see
201.12.4.2.101) in the ACCOMPANYING DOCUMENTS the maximum CO concentration which
will occur in the COMPARTMENT during NORMAL CONDITIONS;
q) a statement that the INFANT RADIANT WARMER does not adjust for PATIENT temperature in
PREWARM MODE and that the mode shall be changed to MANUAL MODE or BABY CONTROLLED
RADIANT WARMER (baby mode) immediately when the PATIENT is placed on the device. The
MANUFACTURER shall disclose the level of heat in mW/cm when operating in PREWARM
MODE.
201.7.9.2.9 Operating instructions
Addition:
The instructions for use shall also contain
a) for each mode of control, a detailed statement describing the method by which the amount
of radiation is controlled and the temperature of the baby is maintained;
*b) if BABY CONTROLLED RADIANT WARMER operation is available, a statement to explain why the
OPERATOR should use this mode whenever possible.
*
201.7.9.2.13 Maintenance
Addition:
If the source of radiation has a limited lifetime, the MANUFACTURER shall state, in the
ACCOMPANYING DOCUMENTS, the time after which the source of radiation shall be replaced
because of ageing.
201.7.9.2.14 ACCESSORIES, supplementary equipment, used material
Addition:
The instructions for use shall include details of any specified combinations of INFANT RADIANT
WARMER with other equipment (see 201.4.1).
201.7.9.3 Technical description
201.7.9.3.1 General
Additional item:
– the MANUFACTURER shall specify in the ACCOMPANYING DOCUMENTS the maximum CO
concentration (see 201.12.4.2.101)
201.8 Protection against electrical HAZARDS from ME EQUIPMENT
Clause 8 of the general standard applies.
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS
Clause 9 of the general standard applies, except as follows:
201.9.4.2.1 Instability in transport position
Addition:
– 14 – 60601-2-21 IEC:2009
The INFANT RADIANT WARMER and the mounting brackets and shelves are provided with the
most unfavourable combination of detachable parts and ACCESSORIES and are loaded with the
recommended maximum load.
201.9.6.2 Acoustic energy
201.9.6.2.1 Audible acoustic energy
Addition:
*
201.9.6.2.1.101 Audible alarms sound level
Auditory ALARM SIGNALS shall have a sound level of at least 65 dBA at a distance of 3 m from
the front of the INFANT RADIANT WARMER in a reflecting room. Other than the AUDIO PAUSED
specified in 201.12.3.103, the auditory ALARM SIGNAL may be adjusted by the OPERATOR to a
minimum lower level of 50 dBA.
The sound pressure level of the ALARM SIGNAL shall not exceed 80 dBA on the mattress.
If the frequency of the auditory ALARM SIGNAL is adjustable by the OPERATOR, these
requirements shall apply to all the individual selectable frequencies.
Compliance is checked with the microphone of a sound level meter complying with the
requirements of IEC 61672-1 placed 1,5 m above the floor and 3 m from the front of the
INFANT RADIANT WARMER.
Compliance of the maximum level is checked with each alarm sound means activated, the
sound level being measured at a point 5 cm above the centre of the mattress.
Ensure that the background sound pressure level is at least 10 dBA below the measured
levels.
201.9.8 HAZARDS associated with support systems
Additional subclause:
201.9.8.101 Supports and mounting brackets for ACCESSORIES
Supports and mounting brackets for ACCESSORIES shall be suitable and of adequate strength
for their purpose
Compliance is checked by inspection and by the following test:
A gradually increasing force is applied so as to act vertically through the centre of the
supports and mounting brackets, e.g. an ACCESSORY shelf in the extended position with a
MANUFACTURER'S recommended load. The force is increased from zero in a 5 s to 10 s
interval, until it equals three times the recommended load and is sustained for a period of
1 min. There shall be no evidence of damage to the items under test.
201.9.8.3 Strength of PATIENT or OPERATOR support or suspension systems
201.9.8.3.1 General
Addition:
The normal load for an INFANT is reduced to 10 kg
60601-2-21 © IEC:2009 – 15 –
Additional subclause:
201.9.8.3.101 Barriers
For INFANT RADIANT WARMERS with an integral bed area suitable barriers shall be provided to
prevent the PATIENT from falling off the mattress. Such barriers as intended to be opened or
removed to allow access to the PATIENT shall latch in their closed positions and shall remain
locked under the test conditions.
Compliance is checked by inspection and the following test: Apply to all the barriers (other
than those secured with the use of a TOOL) an outward horizontal force of 20 N to the centre
of each barrier for 5 s. The barriers shall remain closed.
201.10 Protection against unwanted and excessive radiation HAZARDS
Clause 10 of the general standard applies except as follows:
*
201.10.6 Infrared radiation
Addition:
in the
The maximum irradiance level at any point on the mattress shall not exceed 60 mW/cm
total infrared spectrum.
The maximum irradiance level shall not exceed 10 mW/cm in the near infrared spectrum
(760 nm to 1 400 nm).
Compliance is checked by measurements.
201.11 Protection against excessive temperatures and other HAZARDS
Clause 11 of the general standard applies, except as follows:
*
201.11.1.2.2 APPLIED PARTS not intended to supply heat to a PATIENT
Replacement:
The temperature of surfaces accessible to an INFANT on the mattress shall not exceed 40 °C
for metal surfaces and 43 °C for other materials when the INFANT RADIANT WARMER is operating
under STEADY TEMPERATURE CONDITION at its maximum CONTROL TEMPERATURE.
Under conditions of warm-up to STEADY TEMPERATURE CONDITION or that of a SINGLE FAULT
CONDITION these surfaces shall not exceed 42 °C for metal or 45 °C for other materials.
These requirements apply under NORNAL CONDITION and SINGLE FAULT CONDITIONS including:
– failure of the heater control circuit;
– failure of a SKIN TEMPERATURE SENSOR;
– disconnection of a SKIN TEMPERATURE SENSOR from the INFANT RADIANT WARMER.
201.11.1.4 GUARDS
Addition:
If the heater element surface temperature exceeds 85 °C in NORMAL USE, heater GUARDS which
cannot exceed 85 °C in NORMAL USE shall be fitted.
– 16 – 60601-2-21 © IEC:2009
Compliance is checked by measurement of the temperature and by performing the rigidity test
as described in 15.3.2 of the general standard. The heater GUARD shall not touch the heater
element.
*
201.11.2 Fire prevention
Subclause 11.2 of the general standard applies.
201.11.6.3 Spillage on ME EQUIPEMENT and ME SYSTEMS
Replacement:
The INFANT RADIANT WARMER or ME SYSTEM shall be so constructed that in the event of spillage
of water (accidental wetting) on the PATIENT support or SKIN TEMPERATURE SENSOR no HAZARD
shall result from the ingress of water.
Compliance is checked by the following tests: Position the INFANT RADIANT WARMER or
ME SYSTEM in the least favourable position of NORMAL USE. In the case of an INFANT RADIANT
WARMER or ME SYSTEM with BABY CONTROLLED RADIANT WARMER operation, the SKIN
TEMPERATURE SENSOR shall be placed at the centre of the upper surface of the mattress. Pour
200 ml of isotonic water (0,9 % saline) steadily on the centre of the mattress over a period of
15 s.
After this test, the INFANT RADIANT WARMER or ME SYSTEM shall meet the dielectric strength
requirements specified in 8.8.3 of the general standard and the INFANT RADIANT WARMER or
ME EQUIPMENT shall function normally.
201.11.8 Interruption of the power supply / SUPPLY MAINS to ME EQUIPMENT
Addition:
The INFANT RADIANT WARMER shall be so designed that an interruption and restoration of the
power supply up to 10 minutes does not change the CONTROL TEMPERATURE or other preset
values.
Compliance is checked by switching the SUPPLY MAINS off and then switching on, and
inspecting the INFANT RADIANT WARMER.
201.12 Accuracy of controls and instruments and protection against
hazardous outputs
Clause 12 of the general standard applies, except as follows:
201.12.1 Accuracy of controls and instruments
Additional subclauses:
*
Accuracy of SKIN TEMPERATURE SENSOR
201.12.1.101
The temperature measured by the SKIN TEMPERATURE SENSOR shall be continuously displayed
and clearly visible. The temperature displayed shall have an accuracy of ±0,3 °C. If the
display is used to present any other parameter, this shall only be obtained on demand, using
a momentary action switch. The range of displayed temperature shall be at least 30 °C to
40 °C.
Compliance is checked by inspection and the following test:
60601-2-21 © IEC:2009 – 17 –
Immerse the SKIN TEMPERATURE SENSOR in a water bath maintained at 36 °C ± 0,1 °C. Position
a calibrated thermometer, accurate to within ±0,05 °C, with its bulb adjacent to the SKIN
TEMPERATURE SENSOR. The reading of the standard thermometer shall be compared with the
displayed temperature and their difference shall not exceed 0,3 °C less the calibrated
thermometer error.
*
201.12.1.102 Accuracy of distribution of irradiation to the mattress
The difference between the MID-POINT AVERAGE TEMPERATURE and the TEST DEVICE AVERAGE
TEMPERATURE of any of the other devices comprising the TEST LOAD shall not exceed 2 °C.
Compliance is checked by the following test: Prepare five TEST DEVICES consisting of
aluminium discs each with a mass of 500 g ± 10 g and a diameter of 100 mm ± 2 mm. Drill
5 mm diameter holes 50 mm ± 2 mm deep as shown in Figure 201.102 and coat the entire
disc surface with non-reflective black paint.
NOTE The disc thickness will be approximately 23 mm.
Subject the INFANT RADIANT WARMER to the following test in a room in which the maximum air
velocity is 0,1 m/s and the ambient temperature is maintained at 23 °C ± 2 °C.
Place four individually identified TEST DEVICES, marked 1, 2, 3 and 4 on the horizontal
mattress at the centres of each of four rectangles formed by bisecting the length and width of
TEST DEVICE marked “M” on the mid-
the mattress as shown in Figure 201.101. Place a fifth
point of the mattress. Insert a temperature sensor in each of the five TEST DEVICE centres and,
in the case of an INFANT RADIANT WARMER with BABY CONTROLLED RADIANT WARMER operation,
attach the SKIN TEMPERATURE SENSOR to the centre of the upper surface of the TEST DEVICE
“M”, assuring a good thermal conductive contact (e.g. thermal paste). In the case of an INFANT
RADIANT WARMER with a BABY CONTROLLED RADIANT WARMER operation, set the temperature
control to a CONTROL TEMPERATURE of 36 °C ± 0,1 ° and operate the INFANT RADIANT WARMER
until a STEADY TEMPERATURE CONDITION is obtained. In the case of an INFANT RADIANT WARMER
with only a MANUAL MODE, set the heater output so that the TEST DEVICE will warm up to
STEADY TEMPERATURE CONDITION. Take at least 20 readings of
approximately 36 ° under
temperature of each TEST DEVICE at regular intervals over a 60 min. period.
Calculate the five values of the TEST DEVICE AVERAGE TEMPERATURE for each TEST DEVICE as
follows:
(t + t + t + t + .t )
11 12 13 14 1n
T =
n
where
T is the TEST DEVICE AVERAGE TEMPERATURE for TEST DEVICE No. 1;
t t are the individual temperature readings taken of TEST DEVICE No. 1 at regular
….
11 1n
intervals during the STEADY TEMPERATURE CONDITION;
n is the number of readings during the STEADY TEMPERATURE CONDITION.
Calculate the remaining TEST DEVICE AVERAGE TEMPERATURES T , T , T and T in the same
2 3 4 M
way.
Compare TEST DEVICE AVERAGE TEMPERATURES T T , T , T with T and verify the maximum
,
1 2 3 4 M
difference does not exceed 2,0 °C.
*
201.12.1.103 Accuracy of BABY CONTROLLED RADIANT WARMER operation
With the INFANT RADIANT WARMER working in the BABY CONTROLLED RADIANT WARMER operation
with horizontal mattress orientation in NORMAL CONDITION, the temperature as measured by the
...
IEC 60601-2-21 ®
Edition 2.1 2016-04
CONSOLIDATED VERSION
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-21: Particular requirements for the basic safety and essential performance
of infant radiant warmers
Appareils électromédicaux –
Partie 2-21: Exigences particulières pour la sécurité de base et les performances
essentielles des incubateurs radiants pour nouveau-nés
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite
ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie
et les microfilms, sans l'accord écrit de l'IEC ou du Comité national de l'IEC du pays du demandeur. Si vous avez des
questions sur le copyright de l'IEC ou si vous désirez obtenir des droits supplémentaires sur cette publication, utilisez
les coordonnées ci-après ou contactez le Comité national de l'IEC de votre pays de résidence.
IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé Fax: +41 22 919 03 00
CH-1211 Geneva 20 info@iec.ch
Switzerland www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.
IEC Catalogue - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
The stand-alone application for consulting the entire The world's leading online dictionary of electronic and
bibliographical information on IEC International Standards, electrical terms containing 20 000 terms and definitions in
Technical Specifications, Technical Reports and other English and French, with equivalent terms in 15 additional
documents. Available for PC, Mac OS, Android Tablets and languages. Also known as the International Electrotechnical
iPad. Vocabulary (IEV) online.
IEC publications search - www.iec.ch/searchpub IEC Glossary - std.iec.ch/glossary
The advanced search enables to find IEC publications by a 65 000 electrotechnical terminology entries in English and
variety of criteria (reference number, text, technical French extracted from the Terms and Definitions clause of
committee,…). It also gives information on projects, replaced IEC publications issued since 2002. Some entries have been
and withdrawn publications. collected from earlier publications of IEC TC 37, 77, 86 and
CISPR.
IEC Just Published - webstore.iec.ch/justpublished
Stay up to date on all new IEC publications. Just Published IEC Customer Service Centre - webstore.iec.ch/csc
details all new publications released. Available online and If you wish to give us your feedback on this publication or
also once a month by email. need further assistance, please contact the Customer Service
Centre: csc@iec.ch.
A propos de l'IEC
La Commission Electrotechnique Internationale (IEC) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications IEC
Le contenu technique des publications IEC est constamment revu. Veuillez vous assurer que vous possédez l’édition la
plus récente, un corrigendum ou amendement peut avoir été publié.
Catalogue IEC - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
Application autonome pour consulter tous les renseignements
Le premier dictionnaire en ligne de termes électroniques et
bibliographiques sur les Normes internationales,
électriques. Il contient 20 000 termes et définitions en anglais
Spécifications techniques, Rapports techniques et autres
et en français, ainsi que les termes équivalents dans 15
documents de l'IEC. Disponible pour PC, Mac OS, tablettes
langues additionnelles. Egalement appelé Vocabulaire
Android et iPad.
Electrotechnique International (IEV) en ligne.
Recherche de publications IEC - www.iec.ch/searchpub
Glossaire IEC - std.iec.ch/glossary
La recherche avancée permet de trouver des publications IEC
65 000 entrées terminologiques électrotechniques, en anglais
en utilisant différents critères (numéro de référence, texte,
et en français, extraites des articles Termes et Définitions des
comité d’études,…). Elle donne aussi des informations sur les
publications IEC parues depuis 2002. Plus certaines entrées
projets et les publications remplacées ou retirées.
antérieures extraites des publications des CE 37, 77, 86 et
CISPR de l'IEC.
IEC Just Published - webstore.iec.ch/justpublished
Restez informé sur les nouvelles publications IEC. Just Service Clients - webstore.iec.ch/csc
Published détaille les nouvelles publications parues. Si vous désirez nous donner des commentaires sur cette
Disponible en ligne et aussi une fois par mois par email. publication ou si vous avez des questions contactez-nous:
csc@iec.ch.
IEC 60601-2-21 ®
Edition 2.1 2016-04
CONSOLIDATED VERSION
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-21: Particular requirements for the basic safety and essential
performance of infant radiant warmers
Appareils électromédicaux –
Partie 2-21: Exigences particulières pour la sécurité de base et les
performances essentielles des incubateurs radiants pour nouveau-nés
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 11.040.10 ISBN 978-2-8322-3377-1
IEC 60601-2-21 ®
Edition 2.1 2016-04
CONSOLIDATED VERSION
REDLINE VERSION
VERSION REDLINE
colour
inside
Medical electrical equipment –
Part 2-21: Particular requirements for the basic safety and essential performance
of infant radiant warmers
Appareils électromédicaux –
Partie 2-21: Exigences particulières pour la sécurité de base et les performances
essentielles des incubateurs radiants pour nouveau-nés
– 2 – IEC 60601-2-21:2009+AMD1:2016 CSV
© IEC 2016
CONTENTS
FOREWORD. 3
INTRODUCTION . 6
201.1 Scope, object and related standards . 7
201.2 Normative references . 9
201.3 Terms and definitions . 9
201.4 General requirements . 11
201.5 General requirements for testing of ME EQUIPMENT . 12
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 12
201.7 ME EQUIPMENT identification, marking and documents . 12
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 14
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 14
201.10 Protection against unwanted and excessive radiation HAZARDS . 16
201.11 Protection against excessive temperatures and other HAZARDS . 16
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 17
201.13 HAZARDOUS SITUATIONS and fault conditions . 21
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 21
201.15 Construction of ME EQUIPMENT . 21
201.16 ME SYSTEMS . 23
*
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 23
202 Electromagnetic compatibility – Requirements and tests . 23
210 Requirements for the development of physiologic closed-loop controllers .
Annexes . 24
Annex AA (informative) Particular guidance and rationale . 25
Bibliography . 34
Index of defined terms used in this particular standard . 36
Figure 201.101 – Layout of TEST DEVICES . 10
Figure 201.102 – TEST DEVICE . 11
Figure AA.1 – Illustration of the main requirements of this standard . 25
Table 201.101 – Additional ESSENTIAL PERFORMANCE requirements . 12
© IEC 2016
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-21: Particular requirements for the basic safety and
essential performance of infant radiant warmers
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
This consolidated version of the official IEC Standard and its amendment has been prepared
for user convenience.
IEC 60601-2-21 edition 2.1 contains the second edition (2009-02) [documents 62D/735/FDIS and
62D/762/RVD], its corrigendum 1 (2013-02) and its amendment 1 (2016-04) [documents
62D/1326/FDIS and 62D/1347/RVD].
In this Redline version, a vertical line in the margin shows where the technical content is
modified by amendment 1. Additions are in green text, deletions are in strikethrough red text.
A separate Final version with all changes accepted is available in this publication.
– 4 – IEC 60601-2-21:2009+AMD1:2016 CSV
© IEC 2016
International standard IEC 60601-2-21 has been prepared by subcommittee 62D:
Electromedical equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
This second edition cancels and replaces the first edition published in 1994 and its
Amendment 1 (1996). This edition constitutes a technical revision. This edition of
IEC 60601-2-21 was revised to structurally align with the 2005 edition of IEC 60601-1.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– Requirements and definitions: roman type.
– Test specifications: italic type.
– Informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title Medical electrical
equipment, can be found on the IEC website.
© IEC 2016
The committee has decided that the contents of the base publication and its amendment will
remain unchanged until the stability date indicated on the IEC web site under
"http://webstore.iec.ch" in the data related to the specific publication. At this date, the
publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
– 6 – IEC 60601-2-21:2009+AMD1:2016 CSV
© IEC 2016
INTRODUCTION
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of INFANT RADIANT WARMER equipment.
This particular standard amends and supplements IEC 60601-1:2005, Medical electrical
equipment – Part 1: General requirements for basic safety and essential performance,
hereinafter referred to as the general standard.
The requirements are followed by specifications for the relevant tests.
A general guidance and rationale for the requirements of this particular standard are given in
Annex AA.
It is considered that knowledge of the reasons for these requirements will not only facilitate
the proper application of this particular standard but will, in due course, expedite any revision
necessitated by changes in clinical practice or as a result of developments in technology.
However, this annex does not form part of the requirements of this standard.
© IEC 2016
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-21: Particular requirements for the basic safety and
essential performance of infant radiant warmers
201.1 Scope, object and related standards
1)
Clause 1 of the general standard applies, except as follows:
201.1.1 Scope
Replacement:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of INFANT
RADIANT WARMERS as defined in 201.3.204, also referred to as ME EQUIPMENT.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS
within the scope of this standard are not covered by specific requirements in this standard
except in 7.2.13 and 8.4.1 of the general standard.
NOTE See also 4.2 of the general standard.
This particular standard specifies the safety requirements for INFANT RADIANT WARMERS, but
alternate methods of compliance with a specific clause, by demonstrating equivalent safety,
will not be judged as non-compliant, if the MANUFACTURER has demonstrated in his RISK
MANAGEMENT FILE that the RISK presented by the HAZARD has been found to be of an
acceptable level when weighed against the benefit of treatment from the device.
This particular standard does not apply to:
– devices supplying heat via BLANKETS, PADS or MATTRESSES in medical use; for information
see IEC 80601-2-35;
– INFANT INCUBATORS; for information see IEC 60601-2-19;
– INFANT TRANSPORT INCUBATORS, for information see IEC 60601-2-20;
– INFANT PHOTOTHERAPY EQUIPMENT, for information see IEC 60601-2-50.
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements for INFANT RADIANT WARMERS as defined in 201.3.204, which
minimize HAZARDS to PATIENT and OPERATOR, and to specify tests by which compliance with
the requirements can be verified.
—————————
1)
The general standard is IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for
basic safety and essential performance.
– 8 – IEC 60601-2-21:2009+AMD1:2016 CSV
© IEC 2016
201.1.3 * Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 2 of this particular standard.
IEC 60601-1-2 and IEC 60601-1-10 applies as modified in Clauses 202 and 210 respectively.
IEC 60601-1-3 and IEC 60601-1-10 do not apply. All other published collateral standards in
the IEC 60601-1 series apply as published.
201.1.4 * Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard and collateral standards as appropriate for the particular ME
EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL PERFORMANCE
requirements.
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this standard addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the 60601-1-2 collateral standard,
203.4 in this particular standard addresses the content of Clause 4 of the 60601-1-3 collateral
standard, etc.). The changes to the text of the general standard are specified by the use of
the following words:
"Replacement" means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
"Addition" means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
"Amendment" means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However, due to the fact that definitions in the general
standard are numbered 3.1 through 3.139, additional definitions in this standard are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term "this standard" is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
© IEC 2016
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
SKIN TEMPERATURE SENSORS which are applied to operate a BABY CONTROLLED RADIANT WARMER
including the displayed value are considered to be not a CLINICAL THERMOMETER in the sense
of the particular standard ISO 80601-2-56.
201.2 Normative references
Clause 2 of the general standard applies, except as follows:
Amendment:
IEC 60601-1-2:2007, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral standard: Electromagnetic compatibility –
Requirements and tests
Addition:
IEC 60601-1-10:2007, Medical electrical equipment – Part 1-10: General requirements for
basic safety and essential performance – Collateral Standard: Requirements for the
development of physiologic closed-loop controllers
NOTE Informative references are listed in the bibliography beginning on page 34.
201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in the general standard
apply, except as follows:
NOTE An index of defined terms is found beginning on page 36.
Addition:
201.3.201
BABY CONTROLLED RADIANT WARMER
mode of operation in which the power output varies automatically in order to maintain the
temperature as measured by a SKIN TEMPERATURE SENSOR according to the CONTROL
TEMPERATURE set by the OPERATOR
NOTE An INFANT RADIANT WARMER operating as a BABY CONTROLLED RADIANT WARMER is a PHYSIOLOGIC CLOSED-
LOOP CONTROLLER as defined in IEC 60601-1-10.
201.3.202
CONTROL TEMPERATURE
temperature selected at the temperature control
201.3.203
INFANT
PATIENT up to 3 months and with a weight of less than 10 kg
201.3.204
INFANT RADIANT WARMER
electrically powered device with a radiant heating source intended to maintain the thermal
balance of an INFANT by direct radiation of energy in the infrared region of the electromagnetic
spectrum
– 10 – IEC 60601-2-21:2009+AMD1:2016 CSV
© IEC 2016
201.3.205
MANUAL MODE
mode of operation in which the heater output is either at a fixed level or a proportion of its
maximum output set by the OPERATOR
201.3.206
MID-POINT AVERAGE TEMPERATURE
T
M
TEST DEVICE AVERAGE TEMPERATURE of the TEST DEVICE positioned at the mid-point of the
INFANT RADIANT WARMER mattress (See Figure 201.101)
C
D
M
A
B
IEC 255/09
Key
1 = Mattress
Figure 201.101 – Layout of TEST DEVICES
201.3.207
*
PREWARM MODE
mode of operation in which the heater output is maintained at a preset level (set by the
MANUFACTURER) for the purpose of pre-warming the INFANT RADIANT WARMER and maintaining
the level of warmth of the INFANT RADIANT WARMER prior to an INFANT being placed on the
device
201.3.208
SKIN TEMPERATURE SENSOR
sensing device intended to measure the INFANT’S SKIN TEMPERATURE
201.3.209
STEADY TEMPERATURE CONDITION
condition which is reached when the temperature, measured at the centre of the TEST DEVICE
positioned on the mid-point of the INFANT RADIANT WARMER mattress, does not vary by more
than 1 °C over a period of 1 h
201.3.210
TEST DEVICE
totally matt blackened disc used as a reproducible receiver of radiant energy during testing of
the INFANT RADIANT WARMER (see Figure 201.102)
© IEC 2016
Dimensions in millimeters
50 ± 2
∅100 ± 2
IEC 372/09
Key
1 = Mattress
Surface finish: non-reflective black paint
Disc mass 500 g ± 10 g
3 3
Disc material: aluminum of density within the range 2,6 g/cm and 2,9 g/cm
Figure 201.102 – TEST DEVICE
201.3.211
TEST DEVICE AVERAGE TEMPERATURE
(T , T , T , T OR T )
1 2 3 4 M
average temperature reading taken during a STEADY TEMPERATURE CONDITION at regular
intervals at the centre of a TEST DEVICE
NOTE T is expressed in °C.
M
201.3.212
*
TEST LOAD
array of five TEST DEVICES used in a specified configuration (see Figure 201.101) for
performance tests of the INFANT RADIANT WARMER
201.4 General requirements
Clause 4 of the general standard applies, except as follows:
201.4.1 Conditions for application to ME EQUIPMENT OR ME SYSTEMS
Addition:
For ME EQUIPMENT or ME SYSTEMS, which combines alternative heat sources, for instance
incubators with integrated RADIANT WARMERS, devices supplying heat via BLANKETS, PADS or
MATTRESSES etc., safety requirements of other relevant particular standards shall be
considered. Further the safety requirements of this particular standard shall be fulfilled with
the combination of the other equipment, which is approved by the MANUFACTURER, as stated in
the ACCOMPANYING DOCUMENTS according to Clause 16 (ME SYSTEMS).
Compliance is checked by the test of Clause 201.11 and subclause 201.15.4.2.1 of the
relevant particular standards (e.g. IEC 60601-2-19 etc).
201.4.3 ESSENTIAL PERFORMANCE
Addition:
23 ± 1
∅5 max.
– 12 – IEC 60601-2-21:2009+AMD1:2016 CSV
© IEC 2016
*
201.4.3.101 Additional ESSENTIAL PERFORMANCE requirements
Additional ESSENTIAL PERFORMANCE requirements are found in the subclauses listed in Table
201.101.
Table 201.101 – Additional ESSENTIAL PERFORMANCE requirements
Requirement Subclause
ESSENTIAL PERFORMANCE requirement 1 201.12.1.103, and
generation of a visual
and audible alarm in
compliance with
201.15.4.2.1
201.5 General requirements for testing of ME EQUIPMENT
Clause 5 of the general standard applies, except as follows:
201.5.3 * Ambient temperature, humidity, atmospheric pressure
Addition to item a):
The ME EQUIPMENT shall comply with the requirements of this standard when operating
within the following conditions:
– an ambient temperature within the range 18 °C to 30 °C;
– an ambient air velocity is less than 0,3 m/s.
If not otherwise specified in this particular standard, all tests shall be carried out at an
ambient temperature within the range of 21 °C to 26 °C.
201.5.4 Other conditions
Addition:
If not otherwise specified, the CONTROL TEMPERATURE shall be 36 °C ± 1 °C and shall always
exceed the ambient temperature by at least 3 °C.
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies.
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
201.7.2 Marking on the outside of ME EQUIPMENT or ME EQUIPMENT parts
(see also Table C.1 of the general standard)
Additional subclauses:
*
201.7.2.101 Oxygen monitor
An INFANT RADIANT WARMER not equipped with an integral oxygen monitor and which provides
means for oxygen administration shall be marked in a prominent position with a text which
states: "Use an oxygen monitor when oxygen is administered".
© IEC 2016
201.7.2.102 Distance markings
The INFANT RADIANT WARMER without integral bed areas shall be permanently and clearly
marked with an indication of the permissible distances between the INFANT RADIANT WARMER
heating systems and any mattress.
201.7.4.2 Control devices
Addition:
Means shall be provided for the clear selection and indication of CONTROL TEMPERATURE on or
adjacent to the controls. The means provided shall allow resolution at intervals not greater
than 0,2 °C.
201.7.9.2.2 Warning and safety notices
Addition:
The instructions for use shall additionally contain:
a) a statement that independent monitoring of the temperature of the INFANT by the OPERATOR
is essential and it is inadvisable to leave an INFANT unattended under the INFANT RADIANT
WARMER;
b) recommendations on the permissible distances between the INFANT RADIANT WARMER
heating system and any mattress used with it, and a statement on the effects which any
changes in this distance may have;
c) instructions on the recommended positions and methods of use and attachment of the
temperature sensors provided for use with the INFANT RADIANT WARMER;
d) for INFANT RADIANT WARMER with TYPE B APPLIED PART in which the INFANT might not be
isolated from earth, a warning that particular care shall be taken to ensure that additional
equipment connected to the INFANT is electrically safe;
e) if applicable, a recommendation to the OPERATOR to inspect regularly latches and closing
devices of barriers to prevent the INFANT falling out;
f) a statement of the maximum loads which can be applied to all supports and mounting
brackets for ACCESSORIES and ancillary equipment;
*g) information on the effects on the functioning of the INFANT RADIANT WARMER of detachment
of the SKIN TEMPERATURE SENSOR from the PATIENT skin;
h) if applicable, a statement that the tilting of the mattress from its horizontal position relative
to the INFANT RADIANT WARMER heater can affect the performance of the INFANT RADIANT
WARMER (see 201.12.1.102);
i) a statement that ACCESSORIES, e.g. for phototherapy or heated mattresses, or sunlight can
cause an increase in INFANT temperature to dangerous levels
j) a statement that the INFANT RADIANT WARMER is not suitable for use in the presence of
flammable anaesthetic gases or other flammable materials, such as some types of
cleaning fluids;
k) a statement that rectal temperatures are not appropriate for controlling the heater output
of the INFANT RADIANT WARMER;
*l) a statement that the INFANT RADIANT WARMER cannot differentiate between an increase in
core temperature with a cold skin (fever) and a low core and SKIN TEMPERATURE
(hypothermia), and a recommendation to monitor the temperature of the PATIENT;
m) a statement that environmental conditions (e.g. air movement) can affect the thermal
balance of the INFANT;
*n) a statement that an INFANT RADIANT WARMER shall be used only by appropriately trained
personnel and under the direction of qualified medical personnel who are familiar with
currently known RISKS and benefits of radiant warmer use;
– 14 – IEC 60601-2-21:2009+AMD1:2016 CSV
© IEC 2016
o) a statement that an INFANT RADIANT WARMER can increase the PATIENT’S insensible water
loss;
p) concentration of carbon dioxide (CO ): If the mattress of an INFANT RADIANT WARMER is
fitted with a COMPARTMENT which encloses the baby, the MANUFACTURER shall specify (see
201.12.4.2.101) in the ACCOMPANYING DOCUMENTS the maximum CO concentration which
will occur in the COMPARTMENT during NORMAL CONDITIONS;
q) a statement that the INFANT RADIANT WARMER does not adjust for PATIENT temperature in
PREWARM MODE and that the mode shall be changed to MANUAL MODE or BABY CONTROLLED
RADIANT WARMER (baby mode) immediately when the PATIENT is placed on the device. The
MANUFACTURER shall disclose the level of heat in mW/cm when operating in PREWARM
MODE.
201.7.9.2.9 Operating instructions
Addition:
The instructions for use shall also contain
a) for each mode of control, a detailed statement describing the method by which the amount
of radiation is controlled and the temperature of the baby is maintained;
*b) if BABY CONTROLLED RADIANT WARMER operation is available, a statement to explain why the
OPERATOR should use this mode whenever possible.
*
201.7.9.2.13 Maintenance
Addition:
If the source of radiation has a limited lifetime, the MANUFACTURER shall state, in the
ACCOMPANYING DOCUMENTS, the time after which the source of radiation shall be replaced
because of ageing.
201.7.9.2.14 ACCESSORIES, supplementary equipment, used material
Addition:
The instructions for use shall include details of any specified combinations of INFANT RADIANT
WARMER with other equipment (see 201.4.1).
201.7.9.3 Technical description
201.7.9.3.1 General
Additional item:
– the MANUFACTURER shall specify in the ACCOMPANYING DOCUMENTS the maximum CO
concentration (see 201.12.4.2.101)
201.8 Protection against electrical HAZARDS from ME EQUIPMENT
Clause 8 of the general standard applies.
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS
Clause 9 of the general standard applies, except as follows:
201.9.4.2.1 Instability in transport position
Addition:
© IEC 2016
The INFANT RADIANT WARMER and the mounting brackets and shelves are provided with the
most unfavourable combination of detachable parts and ACCESSORIES and are loaded with the
recommended maximum load.
201.9.6.2 Acoustic energy
201.9.6.2.1 Audible acoustic energy
Addition:
*
201.9.6.2.1.101 Audible alarms sound level
Auditory ALARM SIGNALS shall have a sound level of at least 65 dBA at a distance of 3 m from
the front of the INFANT RADIANT WARMER in a reflecting room. Other than the AUDIO PAUSED
specified in 201.12.3.103, the auditory ALARM SIGNAL may be adjusted by the OPERATOR to a
minimum lower level of 50 dBA.
The sound pressure level of the ALARM SIGNAL shall not exceed 80 dBA on the mattress.
If the frequency of the auditory ALARM SIGNAL is adjustable by the OPERATOR, these
requirements shall apply to all the individual selectable frequencies.
Compliance is checked with the microphone of a sound level meter complying with the
requirements of IEC 61672-1 placed 1,5 m above the floor and 3 m from the front of the
INFANT RADIANT WARMER.
Compliance of the maximum level is checked with each alarm sound means activated, the
sound level being measured at a point 5 cm above the centre of the mattress.
Ensure that the background sound pressure level is at least 10 dBA below the measured
levels.
201.9.8 HAZARDS associated with support systems
Additional subclause:
201.9.8.101 Supports and mounting brackets for ACCESSORIES
Supports and mounting brackets for ACCESSORIES shall be suitable and of adequate strength
for their purpose
Compliance is checked by inspection and by the following test:
A gradually increasing force is applied so as to act vertically through the centre of the
supports and mounting brackets, e.g. an ACCESSORY shelf in the extended position with a
MANUFACTURER'S recommended load. The force is increased from zero in a 5 s to 10 s
interval, until it equals three times the recommended load and is sustained for a period of
1 min. There shall be no evidence of damage to the items under test.
201.9.8.3 Strength of PATIENT or OPERATOR support or suspension systems
201.9.8.3.1 General
Addition:
The normal load for an INFANT is reduced to 10 kg
– 16 – IEC 60601-2-21:2009+AMD1:2016 CSV
© IEC 2016
Additional subclause:
201.9.8.3.101 Barriers
For INFANT RADIANT WARMERS with an integral bed area suitable barriers shall be provided to
prevent the PATIENT from falling off the mattress. Such barriers as intended to be opened or
removed to allow access to the PATIENT shall latch in their closed positions and shall remain
locked under the test conditions.
Compliance is checked by inspection and the following test: Apply to all the barriers (other
than those secured with the use of a TOOL) an outward horizontal force of 20 N to the centre
of each barrier for 5 s. The barriers shall remain closed.
201.10 Protection against unwanted and excessive radiation HAZARDS
Clause 10 of the general standard applies except as follows:
*
201.10.6 Infrared radiation
Addition:
The maximum irradiance level at any point on the mattress shall not exceed 60 mW/cm in the
total infrared spectrum.
in the near infrared spectrum
The maximum irradiance level shall not exceed 10 mW/cm
(760 nm to 1 400 nm).
Compliance is checked by measurements.
201.11 Protection against excessive temperatures and other HAZARDS
Clause 11 of the general standard applies, except as follows:
*
201.11.1.2.2 APPLIED PARTS not intended to supply heat to a PATIENT
Replacement:
The temperature of surfaces accessible to an INFANT on the mattress shall not exceed 40 °C
for metal surfaces and 43 °C for other materials when the INFANT RADIANT WARMER is operating
under STEADY TEMPERATURE CONDITION at its maximum CONTROL TEMPERATURE.
Under conditions of warm-up to STEADY TEMPERATURE CONDITION or that of a SINGLE FAULT
CONDITION these surfaces shall not exceed 42 °C for metal or 45 °C for other materials.
These requirements apply under NORNAL CONDITION and SINGLE FAULT CONDITIONS including:
– failure of the heater control circuit;
– failure of a SKIN TEMPERATURE SENSOR;
– disconnection of a SKIN TEMPERATURE SENSOR from the INFANT RADIANT WARMER.
201.11.1.4 GUARDS
Addition:
If the heater element surface temperature exceeds 85 °C in NORMAL USE, heater GUARDS which
cannot exceed 85 °C in NORMAL USE shall be fitted.
© IEC 2016
Compliance is checked by measurement of the temperature and by performing the rigidity test
as described in 15.3.2 of the general standard. The heater GUARD shall not touch the heater
element.
*
201.11.2 Fire prevention
Subclause 11.2 of the general standard applies.
201.11.6.3 Spillage on ME EQUIPEMENT and ME SYSTEMS
Replacement:
The INFANT RADIANT WARMER or ME SYSTEM shall be so constructed that in the event of spillage
of water (accidental wetting) on the PATIENT support or SKIN TEMPERATURE SENSOR no HAZARD
shall result from the ingress of water.
Compliance is checked by the following tests: Position the INFANT RADIANT WARMER or
ME SYSTEM in the least favourable position of NORMAL USE. In the case of an INFANT RADIANT
WARMER or ME SYSTEM with BABY CONTROLLED RADIANT WARMER operation, the SKIN
TEMPERATURE SENSOR shall be placed at the centre of the upper surface of the mattress. Pour
200 ml of isotonic water (0,9 % saline) steadily on the centre of the mattress over a period of
15 s.
After this test, the INFANT RADIANT WARMER or ME SYSTEM shall meet the dielectric strength
requirements specified in 8.8.3 of the general standard and the INFANT RADIANT WARMER or
ME EQUIPMENT shall function normally.
201.11.8 Interruption of the power supply / SUPPLY MAINS to ME EQUIPMENT
Addition:
The INFANT RADIANT WARMER shall be so designed that an interruption and restoration of the
power supply up to 10 minutes does not change the CONTROL TEMPERATURE or other preset
values.
Compliance is checked by switching the SUPPLY MAINS off and then switching on, and
inspecting the INFANT RADIANT WARMER.
201.12 Accuracy of controls and instruments and protection against
hazardous outputs
Clause 12 of the general standard applies, except as follows:
201.12.1 Accuracy of controls and instruments
Additional subclauses:
*
201.12.1.101 Accuracy of SKIN TEMPERATURE SENSOR
The temperature measured by the SKIN TEMPERATURE SENSOR shall be continuously displayed
and clearly visible. The temperature displayed shall have an accuracy of ±0,3 °C. If the
display is used to present any other parameter, this shall only be obtained on demand, using
a momentary action switch. The range of displayed temperature shall be at least 30 °C to
40 °C.
Compliance is checked by inspection and the following test:
– 18 – IEC 60601-2-2
...