FprEN 50527-2-3:2021
(Main)Procedure for the assessment of the exposure to electromagnetic fields of workers bearing active implantable medical devices - Part 2-3: Specific assessment for workers with implantable neurostimulators
Procedure for the assessment of the exposure to electromagnetic fields of workers bearing active implantable medical devices - Part 2-3: Specific assessment for workers with implantable neurostimulators
This document provides the procedure for the specific assessment required in EN 50527 1:2016, Annex A, for workers with implanted neurostimulator systems (NS), specifically of the type used for spinal cord stimulation (SCS). It is recognized that implantable neurostimulators have been developed for a wide variety of clinical applications, however the SCS devices within the scope of this document represent the largest segment of the implantable neurostimulator applications thus far. NOTE 1 If the worker has other Active Implantable Medical Devices (AIMDs) implanted additionally, they are assessed separately according to EN 50527 1 or other particular standards within the EN 50527 series. The purpose of the specific assessment is to determine the risk for workers with implanted SCS devices arising from exposure to electromagnetic fields (EMF) at the workplace. The assessment includes the likelihood of clinically significant effects and takes both transient and long-term exposure within specific areas of the workplace into account. NOTE 2 This document does not address risks from contact currents or the effects upon any associated external devices. The techniques described in the different approaches can also be used for the assessment of publicly accessible areas. The frequency range to be observed is from 0 Hz to 3 GHz. Above 3 GHz no interference with the devices within the scope of this document is expected to occur. NOTE 3 The rationale for limiting the observation range to 3 GHz can be found in ISO 14708 3. NOTE 4 Further information concerning the functions of neurostimulator systems can be found at https://www.aans.org/Patients/Neurosurgical-Conditions-and-Treatments/Spinal-Cord-Stimulation.
Verfahren zur Beurteilung der Exposition von Arbeitnehmern mit aktiven implantierbaren medizinischen Geräten gegenüber elektromagnetischen Feldern - Teil 2-3: Besondere Beurteilung für Arbeitnehmer mit implantierbaren Neurostimulatoren
Procédure pour l'évaluation de l'exposition des travailleurs porteurs de dispositifs médicaux implantables actifs aux champs électromagnétiques - Partie 2-3 : Evaluation spécifique aux travailleurs porteurs de neurostimulateurs implantés
Le présent document fournit la procédure d’évaluation spécifique exigée dans l’Annexe A de l’EN 50527-1:2016, pour les travailleurs porteurs de systèmes neurostimulateurs (SN) implantés, spécifiquement du type utilisé pour la stimulation médullaire (SM). Il est admis que des neurostimulateurs implantables ont été développés pour un grand nombre d’applications cliniques; toutefois, les dispositifs de SM relevant du domaine d’application du présent document représentent jusqu’à présent le plus grand segment des applications de neurostimulateurs implantables. NOTE 1 Si le travailleur porte en plus d’autres dispositifs médicaux implantables actifs (DMIA) implantés, ceux-ci sont évalués de manière individuelle conformément à l’EN 50527-1 ou à d’autres normes particulières de la série EN 50527. Le but de cette évaluation spécifique est de déterminer, pour des travailleurs porteurs de dispositifs de SM implantés, le risque provenant de l’exposition aux champs électromagnétiques (EMF - electromagnetic fields) sur le lieu de travail. L’évaluation comprend la probabilité d’effets cliniquement significatifs et tient compte à la fois de l’exposition transitoire et de l’exposition de longue durée dans des zones spécifiques du lieu de travail. NOTE 2 Le présent document ne traite pas des risques vis-à-vis des courants de contact ni des effets sur tout dispositif externe associé. Les techniques décrites dans les différentes approches peuvent aussi être utilisées pour l’évaluation des zones accessibles au public. La plage de fréquences à observer comprend le spectre entre 0 Hz et 3 GHz. Au-dessus de 3 GHz, aucune interférence n’est réputée se produire avec les dispositifs relevant du domaine d’application du présent document. NOTE 3 La justification de la limitation de la plage d’observation à 3 GHz peut être consultée dans l’ISO 14708-3. NOTE 4 Des informations complémentaires concernant les fonctions des systèmes neurostimulateurs peuvent être consultées à l’adresse https://www.aans.org/Patients/Neurosurgical-Conditions-and-Treatments/Spinal-Cord-Stimulation.
Postopek ocenjevanja izpostavljenosti delavcev z aktivnimi medicinskimi vsadki elektromagnetnim poljem - 2-3. del: Specifično ocenjevanje delavcev z vsadljivimi nevrostimulatorji
General Information
Standards Content (sample)
SLOVENSKI STANDARD
oSIST prEN 50527-2-3:2020
01-februar-2020
Postopek ocenjevanja izpostavljenosti delavcev z aktivnimi medicinskimi vsadki
elektromagnetnim poljem - 2-3. del: Specifično ocenjevanje delavcev z vsadljivimi
nevrostimulatorjiProcedure for the assessment of the exposure to electromagnetic fields of workers
bearing active implantable medical devices - Part 2-3: Specific assessment for workers
with implantable neurostimulatorsTa slovenski standard je istoveten z: prEN 50527-2-3
ICS:
11.040.40 Implantanti za kirurgijo, Implants for surgery,
protetiko in ortetiko prosthetics and orthotics
17.240 Merjenje sevanja Radiation measurements
oSIST prEN 50527-2-3:2020 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
---------------------- Page: 1 ----------------------oSIST prEN 50527-2-3:2020
---------------------- Page: 2 ----------------------
oSIST prEN 50527-2-3:2020
EUROPEAN STANDARD DRAFT
prEN 50527-2-3
NORME EUROPÉENNE
EUROPÄISCHE NORM
December 2019
ICS
English Version
Procedure for the assessment of the exposure to
electromagnetic fields of workers bearing active implantable
medical devices - Part 2-3: Specific assessment for workers with
implantable neurostimulators
To be completed To be completed
This draft European Standard is submitted to CENELEC members for enquiry.
Deadline for CENELEC: 2020-02-28.
It has been drawn up by CLC/TC 106X.
If this draft becomes a European Standard, CENELEC members are bound to comply with the CEN/CENELEC Internal Regulations which
stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
This draft European Standard was established by CENELEC in three official versions (English, French, German).
A version in any other language made by translation under the responsibility of a CENELEC member into its own language and notified to
the CEN-CENELEC Management Centre has the same status as the official versions.CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are aware and to
provide supporting documentation.Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without notice and
shall not be referred to as a European Standard.European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2019 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Project: 66708 Ref. No. prEN 50527-2-3 E---------------------- Page: 3 ----------------------
oSIST prEN 50527-2-3:2020
prEN 50527-2-3:2019 (E)
Contents Page
European foreword ........................................................................................................................................... 3
1 Scope ........................................................................................................................................................... 4
2 Normative references ................................................................................................................................. 4
3 Terms and definitions ................................................................................................................................ 4
4 Specific assessment .................................................................................................................................. 6
4.1 Description of the assessment process ............................................................................................ 6
4.1.1 General ................................................................................................................................ 6
4.1.2 Exclusion based on history and physician warnings ........................................................... 8
4.1.3 Specific assessment phase 1: Consideration of equipment and SCS therapy type ........... 9
4.1.4 Equipment consideration ................................................................................................... 10
4.1.5 Specific assessment phase 2: Ruling out extrinsic stimulation and tissue damage .......... 11
4.1.6 Specific assessment phase 3: Assessment of clinical effects using in vivo testing .......... 11
4.2 Assessment of extrinsic stimulation and tissue damage risk ......................................................... 12
4.2.1 General .............................................................................................................................. 12
4.2.2 Overview of the Assessments Method given in Annexes D and E .................................... 13
4.2.3 Selection of Parameters Affecting the Assessment ........................................................... 13
4.2.4 Examples of assessment for unipolar configuration .......................................................... 15
4.2.5 Overall Assessment ........................................................................................................... 17
4.2.6 Further Assessment ........................................................................................................... 19
4.2.7 Concluding phase 2 assessment ....................................................................................... 20
4.3 Specific Assessment Phase 3: Clinical Evaluation ......................................................................... 20
4.3.1 Considerations in choosing a clinical method .................................................................... 20
4.3.2 Conducting in vivo tests ..................................................................................................... 21
5 Documentation ......................................................................................................................................... 22
Annex A (normative) Device specific replacement of EN 50527-1:2016, Table 1 ................................... 23
Annex B (informative) Clinical investigation methods .............................................................................. 30
Annex C (informative) Rationale .................................................................................................................. 31
Annex D (informative) Nerve Stimulation for SCS ..................................................................................... 32
Annex E (informative) Modelling Nerve Stimulation for SCS.................................................................... 38
Bibliography .................................................................................................................................................... 70
---------------------- Page: 4 ----------------------oSIST prEN 50527-2-3:2020
prEN 50527-2-3:2019 (E)
European foreword
This document (prEN 50527-2-3:2019) has been prepared by CLC/TC 106X “Electromagnetic fields in the
human environment”.This document is currently submitted to the Enquiry.
The following dates are proposed:
• latest date by which the existence of this (doa) dor + 6 months
document has to be announced at national
level
• latest date by which this document has to be (dop) dor + 12 months
implemented at national level by publication of
an identical national standard or by
endorsement
• latest date by which the national standards (dow) dor + 36 months
conflicting with this document have to be (to be confirmed or
withdrawn modified when voting)
This document has been prepared under a mandate given to CENELEC by the European Commission and
the European Free Trade Association.---------------------- Page: 5 ----------------------
oSIST prEN 50527-2-3:2020
prEN 50527-2-3:2019 (E)
1 Scope
This document provides the procedure for the specific assessment required in EN 50527-1:2016, Annex A, for
workers with implanted neurostimulator systems (NS), specifically of the type used for spinal cord stimulation
(SCS).It is recognized that implantable neurostimulators have been developed for a wide variety of clinical
applications, however the SCS devices within the scope of this document represent the largest segment of the
implantable neurostimulator applications thus far.NOTE 1 If the worker has other Active Implantable Medical Devices (AIMDs) implanted additionally, they are assessed
separately according to EN 50527-1 or other particular standards within the EN 50527 series.
The purpose of the specific assessment is to determine the risk for workers with implanted SCS devices arising
from exposure to electromagnetic fields (EMF) at the workplace. The assessment includes the likelihood of
clinically significant effects and takes both transient and long-term exposure within specific areas of the
workplace into account.NOTE 2 This document does not address risks from contact currents or the effects upon any associated external
devices.The techniques described in the different approaches can also be used for the assessment of publicly
accessible areas.The frequency range to be observed is from 0 Hz to 3 GHz. Above 3 GHz no interference with the devices
within the scope of this document is expected to occur.NOTE 3 The rationale for limiting the observation range to 3 GHz can be found in ISO 14708-3.
NOTE 4 Further information concerning the functions of neurostimulator systems can be found at
https://www.aans.org/Patients/Neurosurgical-Conditions-and-Treatments/Spinal-Cord-Stimulation.
2 Normative referencesThe following documents are referred to in the text in such a way that some or all of their content constitutes
requirements of this document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
EN 50527-1:2016, Procedure for the assessment of the exposure to electromagnetic fields of workers bearing
active implantable medical devices - Part 1: GeneralEN ISO 14155, Clinical investigation of medical devices for human subjects — Good clinical practice
(ISO 14155)3 Terms and definitions
For the purposes of this document, the terms and definitions given in EN 50527-1:2016 and the following apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— IEC Electropedia: available at http://www.electropedia.org/— ISO Online browsing platform: available at http://www.iso.org/obp
3.1
implantable pulse generator
IPG
part of the active implantable medical device, including the power supply and electronic circuit that produces
an electrical output---------------------- Page: 6 ----------------------
oSIST prEN 50527-2-3:2020
prEN 50527-2-3:2019 (E)
3.2
neurostimulator system
active implantable medical device comprising an implantable pulse generator and includes therapy delivering
electrodes usually part of implanted electrical leads that are intended to deliver therapy to a patient by
electrically stimulating certain nerve structures, along with an associated external patient programming device
3.3electrode
electrically conducting part (usually the termination of a lead) which is designed to form an interface with body
tissue or body fluid3.4
bipolar lead
lead with at least two electrodes that are electrically isolated from each other
3.5
AIMD-Employee
worker with an active implantable medical device
Note 1 to entry: For the purposes of this document, the term AIMD-Employee refers to the patient whose implant consists
of a neurostimulator system of the type intended for spinal cord stimulation.3.6
assessment team
team consisting of:
— employer and if applicable, his occupational health and safety experts and/or occupational physician;
— AIMD-Employee and his responsible physician;— (technical and medical) experts as necessary, e.g. manufacturer of the device
3.7
unipolar stimulation
stimulation using a single electrode with reference to the outer shell of the implantable pulse generator
3.8single lead bipolar stimulation
stimulation using two or more adjacent electrodes of a single lead structure
Note 1 to entry: Leads for use with SCS devices typically have a minimum of 8 electrodes.
3.9two lead bipolar stimulation
stimulation using two electrodes located on two separate leads that are implanted in close proximity to one
another3.10
EM phantom
physical model containing tissue-equivalent material used to simulate the body in an experimental dose
measurementNote 1 to entry: EM phantoms are sometimes also referred to as torso simulator or phantom.
[SOURCE: World Health Organization]3.11
uninfluenced behaviour
conditions for uninfluenced behaviour are provided in EN 50527-1:2016, 4.1.3
---------------------- Page: 7 ----------------------
oSIST prEN 50527-2-3:2020
prEN 50527-2-3:2019 (E)
3.12
device
implanted spinal cord stimulator
3.13
electromagnetic field
EMF
alternating electric field, alternating magnetic field, or radio wave between
1 Hz and 3 GHz whether continuous, pulsed, or modulated in space or time3.14
static field
static magnetic field
SMF
static electric field
SEF
static (stationary) or quasi-static (below 1 Hz) electric field or magnetic field whether continuous, pulsed, or
modulated in space or time3.15
spinal cord stimulator
SCS
neurostimulator system designed specifically for stimulation of the human spinal cord to treat chronic pain by
electrically stimulating the spinal cord but not the Dorsal Root Ganglion (DRG)4 Specific assessment
4.1 Description of the assessment process
4.1.1 General
Spinal Cord Stimulation (SCS) is generally used for chronic pain reduction. The stimulation does not provide
therapy in direct relation to AIMD-Employee safety, so changes to therapy, or turning therapy off, does not
pose a specific risk to the AIMD-Employee. It is noted that an AIMD-Employee can also receive therapy
changes which may appear as a “shock” or “jolt” due to abrupt movements such as coughing and laughing as
well as from postural changes such as standing up from seating, or vice versa. These are not significant direct
risk situations to the AIMD-Employee, but there may be indirect implications due to the nature of the occupation
or positioning. A workplace risk assessment does not need to be based on an absence of risk or effect. It can
be made using a balance of the benefits and the identified possible effects, both direct and indirect, due to the
nature of the employment.The risks to an AIMD-Employee resulting from to EMF or SF exposure in the workplace includes the following
categories:— that the implanted SCS may itself be influenced in a way that leads to temporary or permanent loss of
therapy, or delivery of a corrupted form of therapy that might not meet the needs of the patient such that
they might be unable to carry out their employee functions. ISO 14708-3 is the product standard that
addresses the risks of malfunction and damage to the SCS NS when exposed to EMF or SF. The working
group has deemed the risks to the AIMD-Employee related to malfunction and device damage to be
acceptable, and they are therefore not addressed by this document. See C.2 for additional rationale.
— that the AIMD-Employee may experience additional effects upon their nervous system due to extrinsic
stimulation arising from induced currents in their implanted lead system. These effects can include
unpleasant sensations, up to and including “shocking” and “jolting”,— that the induced currents are of sufficient magnitude to cause nerve tissue damage. This risk would occur
at higher exposure levels and where the EMF frequencies are high enough that they would not be
perceived by the AIMD-Employee.---------------------- Page: 8 ----------------------
oSIST prEN 50527-2-3:2020
prEN 50527-2-3:2019 (E)
To address these risks, this document describes a multi-phase risk assessment as summarized in Figure 1.
— Exclusion from further assessment based upon a consideration of prior history— Exclusion from further assessment based upon a survey of the equipment in the workplace
— Exclusion from further assessment if the AIMD-Employee has a stimulator where outputs can be shut off
in high-impedance mode and is able to work with the stimulator in this mode receiving no therapy while at
the workplace.— Assessment of the likelihood that the extrinsically induced current arising from the EMF environment of
the workplace will exceed the stimulation threshold (i.e. limit of perceived stimulation) for the spinal cord
portion of the human nervous system. Further assessment is prescribed where extrinsic stimulation cannot
be ruled out.— Assessment of the likelihood that the extrinsically induced current arising from the EMF environment of
the workplace could cause damage to the tissues of the spinal cord. This risk is present when the
characteristics of the extrinsically induced currents are such that they are not perceived by the AIMD-
Employee, yet are of sufficient magnitude to result in a level of deposited power as to cause tissue damage
as a result of tissue heating. . The underlying approach is first to allow work without restriction if it can be
determined that no extrinsic stimulation and no tissue damage would occur,— Where the risk of tissue damage has been ruled out, but extrinsic stimulation has not been ruled out,
assessment of actual stimulation effects. Since the consequences of unintended stimulation vary widely
from one individual to another the assessment necessarily involves in vivo testing.
---------------------- Page: 9 ----------------------oSIST prEN 50527-2-3:2020
prEN 50527-2-3:2019 (E)
Figure 1 — Overview of the assessment process
4.1.2 Exclusion based on history and physician warnings
Further risk assessment is not necessary if a history of uninfluenced behaviour at the workplace exists and a
responsible physician has confirmed that this history is sufficient to exclude clinically significant interaction.
When a SCS NS is tested according to ISO 14708-3, the manufacturer is required to provide a warning to the
implanting physician in the accompanying technical information as to any settings available in the device that
if used, afford the device with a reduced immunity to certain types of EMI. A specific warning would only be
given to the patient receiving the implant if they were discharged with one of these settings in effect, or if at
follow-up, a change to one of these settings was made for clinical reasons.Figure 2 summarizes the situations where further risk assessment is unnecessary, and where a specific
assessment (beginning with 4.1.3) is required.---------------------- Page: 10 ----------------------
oSIST prEN 50527-2-3:2020
prEN 50527-2-3:2019 (E)
Figure 2 — Summary of assessment exclusion criteria
4.1.3 Specific assessment phase 1: Consideration of equipment and SCS therapy type
To minimize the burden on the employer and AIMD-employee, the specific assessment begins with a first
phase in which it is determined if work can be allowed based upon considerations of equipment in the
workplace.A specific risk assessment for the AIMD-Employee is required when there is history of influenced behaviour or
one of the following five conditions exists:a) there is equipment present in the workplace that is neither included in, nor used in accordance with
Table A.1;b) all equipment at the workplace is listed in Table A.1 (see Annex A) and is used accordingly, but the AIMD-
Employee has received warning(s) from the responsible physician that their device might be susceptible
to static fields (SF) or electromagnetic fields (EMF) associated with particular types of equipment, thereby
increasing the risk at the workplace.c) there is equipment present in the workplace that is neither included in, nor used in accordance with
Table A.1 and for which the AIMD-Employee does not have a history of device behaviour (uninfluenced
or otherwise) while in its presence, but the AIMD-Employee has received a specific warning as described
in 4.1.1 b).Figure 3 depicts phase 1 of the specific assessment. The steps to be taken are based upon a consideration of
equipment in the workplace and consultation with the responsible physician.When any of the conditions a) through c) exist, 4.1.4 applies. Otherwise, no further assessment is required,
and documentation of the assessment can proceed as required in Clause 5.---------------------- Page: 11 ----------------------
oSIST prEN 50527-2-3:2020
prEN 50527-2-3:2019 (E)
Figure 3 — Specific assessment phase 1
4.1.4 Equipment consideration
Information relevant to the equipment or other field generating sources under consideration shall be collected
to answer sufficiently the following two questions:— can it be determined that clinically significant interference with the device will not occur as a result of
expected exposure to the equipment under consideration? If so, no further assessment is required, and
documentation of the result can proceed, as required in Clause 5;---------------------- Page: 12 ----------------------
oSIST prEN 50527-2-3:2020
prEN 50527-2-3:2019 (E)
— can it be determined that the AIMD-Employee can return to the workplace only with restrictions placed on
the work tasks or areas of access? If so, no further assessment is required, and documentation of the
work restrictions can proceed as required in Clause 5.When neither of these questions can be answered positively, the assessment continues to phase 2, as
described in 4.1.5.The intent of this subclause is to find and utilize information that might already exist and that allows the
assessment to be completed without further, costlier and time-consuming effort. It is recommended that
experts who are likely to have such information be contacted. Examples of such experts are the device
manufacturer, equipment manufacturer, employer’s technical department, consultants, or others skilled in EMF
and SF effects with implanted devices.4.1.5 Specific assessment phase 2: Ruling out extrinsic stimulation and tissue damage
When steps of the specific assessment phase 1 shown in Figure 3 have been followed but fail to mitigate or
to dismiss risk to the AIMD-Employee from the effects of workplace EMF or SF, then additional investigation
shall be performed as shown in Figure 4. The goal of Phase 2 investigation is to rule out the risk of extrinsic
stimulation, and tissue damage.The first step of phase 2 is to determine the levels of SF or EMF associated with the equipment in question.
This is ordinarily done by performing an EMF site survey wherein the fields are measured directly in the
workplace at the separation distances expected in daily exposure for the AIMD-Employee. Such
measurements should account for seasonal variations (if any), as well as measurement at lesser distances to
better understand the boundaries where higher exposures might occur. The results are then assessed using
the methodology described in 4.2.Coincident with the workplace site survey, it is necessary to collect information concerning the specific implant
situation. As a minimum, it is necessary to know the type of stimulation in use (unipolar, bipolar or multipolar
single lead, or bipolar two-lead). Additional information that is useful includes an approximation of the
implanted loop area, proximity of lead electrodes to the spinal cord, the electrode area(s) for the specific leads
implanted, and whether the implantable pulse generator carries an MRI conditional approval. These items of
information might be available from the implanting physician. Otherwise, in the absence of such information, it
should be assumed that the nominal conditions of implant as described in 4.2 apply.
Device “off” exclusionIf it is learned that the AIMD-employee is receiving unipolar stimulation and SCS device re-programming to
bipolar stimulation is not possible, and the AIMD-Employee has a stimulator whose outputs can be shut off in
high-impedance mode, and the AIMD-Employee is able to work with the stimulator in this mode receiving no
therapy, the AIMD-employee is allowed to work with this restriction.4.1.6 Specific assessment phase 3: Assessment of clinical effects using in vivo testing
There can exist situations where the results from 4.2 indicate that the risk of tissue damage is negligible, yet
the EMF exposure levels are such that perception of them by the AIMD-Employee cannot be ruled out. In these
situations, the specific assessment continues with the consideration of applying in vivo testing. The
requirements for this phase of the assessment are described in 4.3.---------------------- Page: 13 ----------------------
oSIST prEN 50527-2-3:2020
prEN 50527-2-3:2019 (E)
Figure 4 — Specific assessment phase 2 — Ruling out malfunction and extrinsic stimulation
4.2 Assessment of extrinsic stimulation and tissue damage risk4.2.1 General
Assessment of the risk of extrinsic induced stimulation that might exceed the threshold of perception is a
complex topic. Annex E together with Annex D, has been developed to provide an analysis methodology for
determination of whether or not an external EMF will be of sufficient magnitude to result in stimulation of the
spinal cord by way of induced currents and internal fields.---------------------- Page: 14 ----------------------
oSIST prEN 50527-2-3:2020
prEN 50527-2-3:2019 (E)
The risk of extrinsic induced voltages that might result in damage to tissue adjacent to the electrodes is also a
developed in Annex E.4.2.2 Overview of the Assessments Method given in Annexes D and E
Annex D introduces the SCS devices and important parameters associated with their use.
The SENN model (Spatially Extended Nonlinear Node model), which is used for modelling stimulation of
nerves by internal electric fields in the body, is introduced in E.1 and E.2 and based on the work of J.P Reilly.
The values of the ICNIRP internal electric field Basic Restriction for nerve stimulation is used for assessing
nerve stimulation that may arise from external magnetic and electric fields. In E.3 the values of the ICNIRP
internal electric field Basic Restriction for local heating in the body, are used in the assessment of whether or
not tissue damage may occur.The method for assessing whether nerve stimulation will be caused by external fields is based on evaluating
the quantity which is referred to as the Nerve Stimulation Ratio (NSR). If the NSR is greater than 1 then nerve
stimulation may occur, and if it is less than 1 nerve stimulation is not expected.
The stages of the method for assessing nerve stimulation are set out in E.5.1, and the results of the
assessment are contained in the remainder of E.5.The method for assessing whether tissue damage will be caused by external fields is based on evaluating the
quantity which is referred to as the Tissue Damage Ratio (TDR). If the TDR is greater than one then nerve
stimulation may occur, and if it is less than 1 tissue damage is not expected.The stages of the method for assessing tissue damage are set out in E.6.1, and the results of the assessment
are contained in the remainder of E.6.The methods for assessing whether nerve stimulation and tissue damage occurs both depend on calculating
the voltage induced by the external fields between the electrodes by the external fields (electric and magnetic),
and the current flowing in the tissue between the electrodes. These are presented in E.4.
The assessments of E.5 and E.6 are brought together in 4.2.5 below and E.7.In some situations, further assessment is required. The approach for this is given in E.8 and reproduced here.
4.2.3 Selection of Parameters Affecting the AssessmentThe outcome of the assessment depends on many different parameters. To simplify this, assessments have
been carried out for a few particular sets of parameters representative of worst cases.
Lead configurationThe leads between the IPG and electrode site form a loop in which voltages may be induced by magnetic
fields. The lead configuration affects loop area which in turn affects the induced voltage. The areas used are
2 2 2as follows: 612 cm for unipolar, 140 cm for bipolar (using two leads) and 10 cm for bipolar (using one lead).
Electrode spacingThe spacing of implanted electrodes affects the voltage induced between them. The spacings used are 50 cm
for unipolar, 5,4 cm for percutaneous leads and 3,7 cm for paddle electrodes.Electric and magnetic fields
Values of electric and magnetic fields for this assessment are shown in Figure 5 and Figure 6, respectively.
Two sets of values are used for frequencies between 1 Hz and 10 GHz. The two lines used are:
— the Low Action Level from the EU EMF Directive for exposure of worker;— the Reference Level from the EU EMF Recommendation for exposures to the General Public.
These are representative of the higher fields that may be found...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.