Medical electrical equipment - Part 2-34: Particular requirements for the basic safety and essential performance of invasive blood pressure monitoring equipment

IEC 60601-2-34:2011 concerns the basic safety and essential performance of invasive blood pressure monitoring equipment. It amends and supplements IEC 60601-1 (third edition, 2005): Medical electrical equipment - Part 1: General requirements for basic safety and essential performance. The third edition of IEC 60601-2-34 cancels and replaces the second edition published in 2001 and constitutes a technical revision in order to align structurally with the 2005 edition of IEC 60601-1. IEC 60601-2-34 applies to basic safety and essential performance of invasive blood pressure monitoring equipment. It does not apply to catheter tubing, catheter needles, Luer locks, taps and tap tables connected. It does not apply to non-invasive blood pressure monitoring equipment.

Medizinische elektrische Geräte - Teil 2-34: Besondere Festlegungen für die Sicherheit einschließlich der wesentlichen Leistungsmerkmale von invasiven Blutdruck-Überwachungsgeräten

Appareils électromédicaux - Partie 2-34: Exigences particulières pour la sécurité de base et les performances essentielles des appareils de surveillance de la pression sanguine prélevée directement

La CEI 60601-2-34:2011 concerne la sécurité de base et les performances essentielles pour les appareils de surveillance de la pression sanguine prélevée directement. Elle modifie et complète la CEI 60601-1 (troisième édition, 2005): Appareils électromédicaux - Partie 1: Exigences générales pour la sécurité de base et les performances essentielles. La troisième édition de CEI 60601-2-34 annule et remplace la deuxième édition publiée en 2001et constitue une révision technique pour l'alignement structurel à l'édition 2005 de la CEI 60601-1. La CEI 60601-2-34 s'applique à la sécurité de base et aux performances essentielles des appareils de surveillance de la pression sanguine prélevée directement. Cette norme particulière ne s'applique pas aux tubes et aiguilles de cathéter, aux connecteurs Luer, et aux prises et panneaux de prises connectés. Cette norme ne s'applique pas non plus aux appareils de surveillance de la pression sanguine prélevée indirectement.

Medicinska električna oprema - 2-34. del: Posebne zahteve za osnovno varnost in bistvene lastnosti opreme za invazivno nadzorovanje krvnega tlaka (IEC 60601-2-34:2011)

Ta mednarodni standard se uporablja za OSNOVNO VARNOST in BISTVENE LASTNOSTI OPREME ZA INVAZIVNO NADZOROVANJE KRVNEGA TLAKA, kot je določeno v 201.3.63, v nadaljevanju: ELEKTROMEDICINSKA OPREMA. Ta standard ne velja za katetrske cevke, katetrske igle, nastavke »Luer-lock«, pipe in mize s pipami, ki so povezane z DOME. Ta standard ne zajema OPREME ZA NEINVAZIVNO NADZOROVANJE KRVNEGA TLAKA.

General Information

Status
Published
Publication Date
26-Jun-2014
Withdrawal Date
26-Jun-2017
Current Stage
6060 - Document made available - Publishing
Start Date
27-Jun-2014
Completion Date
27-Jun-2014

Relations

Buy Standard

Standard
EN 60601-2-34:2014
English language
69 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI STANDARD
01-september-2014
0HGLFLQVNDHOHNWULþQDRSUHPDGHO3RVHEQH]DKWHYH]DRVQRYQRYDUQRVWLQ
ELVWYHQHODVWQRVWLRSUHPH]DLQYD]LYQRQDG]RURYDQMHNUYQHJDWODND ,(&

Medical electrical equipment - Part 2-34: Particular requirements for the basic safety and
essential performance of invasive blood pressure monitoring equipment
Medizinische elektrische Geräte - Teil 2-34: Besondere Festlegungen für die Sicherheit
einschließlich wesentlicher Leistungsmerkmale, von invasiven Blutdruck-
Überwachungsgeräten
Appareils électromédicaux - Partie 2-34: Règles particulières de sécurité de base et de
performances essentielles pour les appareils de surveillance de la pression sanguine
prélevée directement
Ta slovenski standard je istoveten z: EN 60601-2-34:2014
ICS:
11.040.55 'LDJQRVWLþQDRSUHPD Diagnostic equipment
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EUROPEAN STANDARD EN 60601-2-34

NORME EUROPÉENNE
EUROPÄISCHE NORM
June 2014
ICS 11.040.55 Supersedes EN 60601-2-34:2000
English Version
Medical electrical equipment - Part 2-34: Particular requirements
for the basic safety and essential performance of invasive blood
pressure monitoring equipment
(IEC 60601-2-34:2011)
Appareils électromédicaux - Partie 2-34: Exigences Medizinische elektrische Geräte - Teil 2-34: Besondere
particulières pour la sécurité de base et les performances Festlegungen für die Sicherheit einschließlich der
essentielles des appareils de surveillance de la pression wesentlichen Leistungsmerkmale von invasiven Blutdruck-
sanguine prélevée directement Überwachungsgeräten
(CEI 60601-2-34:2011) (IEC 60601-2-34:2011)
This European Standard was approved by CENELEC on 2011-05-16. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.

European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2014 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN 60601-2-34 E
Foreword
The text of document 62D/906/FDIS, future edition 3 of IEC 60601-2-34, prepared by SC 62D,
"Electromedical equipment", of IEC TC 62, "Electrical equipment in medical practice", was submitted to
the IEC-CENELEC parallel vote and approved by CENELEC as EN 60601-2-34:2014.
The following dates are fixed:
(dop) 2014-12-27
• latest date by which the document has
to be implemented at national level by
publication of an identical national
standard or by endorsement
• latest date by which the national (dow) 2017-06-27
standards conflicting with the
document have to be withdrawn
This document supersedes EN 60601-2-34:2000.

In this standard, the following print types are used:
– Requirements and definitions: roman type.
– Test specifications: italic type.
– Informative material appearing outside of tables, such as notes, examples and references: in smaller type. Normative text of
tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS NOTED:
SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents, inclusive of all
subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all subclauses of
Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the clause
number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any combination
of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC Directives,
Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance with this
standard;
– “should” means that compliance with a requirement or a test is recommended but is not mandatory for
compliance with this standard;

- 3 - EN 60601-2-34:2014
– “may” is used to describe a permissible way to achieve compliance with a requirement or test.

This document has been prepared under a mandate given to CENELEC by the European Commission
and the European Free Trade Association, and supports essential requirements of EU Directive(s).

For the relationship with EU Directive(s) see informative Annex ZZ, which is an integral part of this
document.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC [and/or CEN] shall not be held responsible for identifying any or all such patent
rights.
Endorsement notice
The text of the International Standard IEC 60601-2-34:2011 was approved by CENELEC as a European
Standard without any modification.
In the official version, for Bibliography, the following note has to be added for the standard indicated:
IEC 80601-2-30 NOTE  Harmonized as EN 80601-2-30.

Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications

The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.

NOTE  When an international publication has been modified by common modifications, indicated by (mod), the relevant EN/HD
applies.
Annex ZA of the general standard applies, except as follows:

Publication Year Title EN/HD Year

Replacement:
IEC 60601-1-2 2007 Medical electrical equipment - EN 60601-1-2 2007
(mod) Part 1-2: General requirements for basic + corr. March 2010
safety and essential performance - Collateral
standard: Electromagnetic compatibility -
Requirements and tests
IEC 60601-1-8 2006 Medical electrical equipment - EN 60601-1-8 2007
Part 1-8: General requirements for basic + corr. March 2010
safety and essential performance - Collateral
Standard: General requirements, tests and
guidance for alarm systems in medical
electrical equipment and medical electrical
systems
Addition:
IEC 60601-2-2 2009 Medical electrical equipment -  EN 60601-2-2 2009
Part 2-2: Particular requirements for basic + A11 2011
safety and essential performance of high
frequency surgical equipment and high
frequency surgical accessories

1)
IEC 60601-2-27 - Medical electrical equipment - EN 60601-2-27 -
Part 2-27: Particular requirements for the
basic safety and essential performance of
electrocardiographic monitoring equipment

1)
IEC 60601-2-49 - Medical electrical equipment - EN 60601-2-49 -
Part 2-49: Particular requirements for the
basic safety and essential performance of
multifunction patient monitoring equipment

ISO 15223-1 2007 Medical devices - Symbols to be used with EN ISO 15223-1 2012
medical device labels, labelling and
information to be supplied -
Part 1: General requirements
1)
At draft stage.
- 5 - EN 60601-2-34:2014
Annex ZZ
(informative)
Coverage of Essential Requirements of EC Directives
This European Standard has been prepared under a mandate given to CENELEC by the European
Commission and the European Free Trade Association and within its scope the standard covers all
relevant essential requirements as given in Annex I of the EC Directive 93/42/EEC except as follows:
– Essential Requirement 6a
– Essential Requirement 7.4
– Essential Requirement 7.5 paragraph 2 & 3
– Essential Requirement 13.6 (q)
Compliance with this standard provides one means of conformity with the specified essential
requirements of the Directive concerned.
WARNING - Other requirements and other EC Directives may be applicable to the products falling within
the scope of this standard.
IEC 60601-2-34 ®
Edition 3.0 2011-05
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Medical electrical equipment –
Part 2-34: Particular requirements for the basic safety and essential performance
of invasive blood pressure monitoring equipment

Appareils électromédicaux –
Partie 2-34: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils de surveillance de la pression sanguine prélevée
directement
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
CODE PRIX XB
ICS 11.040.55 ISBN 978-2-88912-495-4

– 2 – 60601-2-34  IEC:2011
CONTENTS
FOREWORD . 4
INTRODUCTION . 6
201.1 Scope, object and related standards . 7
201.2 Normative references . 8
201.3 Terms and definitions . 9
201.4 General requirements . 10
201.5 General requirements for testing of ME EQUIPMENT . 10
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 11
201.7 ME EQUIPMENT identification, marking and documents. 11
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 13
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 20
201.10 Protection against unwanted and excessive radiation HAZARDS . 21
201.11 Protection against excessive temperatures and other HAZARDS . 21
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 23
201.13 HAZARDOUS SITUATIONS and fault conditions . 29
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 29
201.15 Construction of ME EQUIPMENT . 29
201.16 ME SYSTEMS . 30
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 30
202 Electromagnetic compatibility – Requirements and tests . 30
208 General requirements, tests and guidance for alarm systems in medical
electrical equipment and medical electrical systems . 36
Annexes . 45
Annex AA (informative) Particular guidance and rationale . 46
Annex BB (informative) Alarm diagrams 208/IEC 60601-1-8:2006 . 59
Bibliography . 61
Index of defined terms used in this particular standard. 62

Figure 201.101 – Dynamic test for limitation of energy from different parts – Recovery
test . 15
Figure 201.102 – Diaphragm leak test . 16
Figure 201.103 – Measuring circuit for PATIENT LEAKAGE CURRENT via the PATIENT
CONNECTION(S) of an F-TYPE APPLIED PART to earth caused by an external voltage on the
PATIENT CONNECTION(S) . 17
Figure 201.104 – Measuring circuit for th
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.