Amendment 1 - Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis

Appareils électromédicaux - Partie 2-28: Exigences particulières pour la sécurité de base et les performances essentielles des gaines équipées pour diagnostic médical

Medicinska električna oprema - 2-28. del: Posebne zahteve za osnovno varnost in bistvene lastnosti rentgenskih sestavov za medicinsko diagnostiko - Dopolnilo A1

General Information

Status
Not Published
Publication Date
31-May-2027
Current Stage
4020 - Enquiry circulated - Enquiry
Start Date
07-Nov-2025
Due Date
31-Oct-2025
Completion Date
07-Nov-2025

Relations

Effective Date
23-Sep-2025

Overview

EN IEC 60601-2-28:2019/prA1:2025 (Amendment 1) updates the particular safety and performance requirements for X‑ray tube assemblies for medical diagnosis. This amendment aligns the Part 2-28 standard with later amendments to the general IEC 60601-1 series and related collateral/particular standards, corrects normative references, clarifies requirements and terms, and adds or refines test and marking obligations specific to X‑ray tube assemblies.

Key topics and technical requirements

  • Updated normative references: Incorporates newer editions/amendments such as IEC 60601‑1:2005/AMD2:2020, IEC 60601‑1‑3/AMD2:2021 and IEC 60522‑1:2020 for filtration methods.
  • Marking and traceability: X‑ray tubes must bear manufacturer name/trademark, model/type reference and unique identification, legible for the expected service life.
  • Temperature limits: Painted metal surfaces likely to be touched for up to 1 second must not exceed 85 °C; other temperature limits reference the general standard and IEC/ISO guidance (IEC Guide 117, ISO 13732).
  • Documentation (accompanying documents): Required technical data expanded to include:
    • High‑voltage generator and connector specifications
    • Electrical supply details for electron emitters, beam control and rotating anodes
    • Mass (with/without optional components)
    • Maximum symmetrical radiation field (if stated) per IEC 60806
  • Radiation protection & filtration:
    • Permanent filtration and half‑value layer (HVL) methods referenced to IEC 60522‑1:2020
    • X‑ray tube assemblies must be marked with permanent filtration or material/thickness data
  • Focal spot declaration: Nominal focal spot values must be stated in accordance with IEC 60336.
  • Design control: Limiting diaphragm requirement - aperture no larger than needed; aperture material located as close as practicable to the focal spot.
  • EMC and usability: Clarifies that stand‑alone X‑ray tube manufacturers are not required to demonstrate conformity with IEC 60601‑1‑2 (EMC) for the tube alone; usability assessment is covered within the risk management process (ISO 14971:2019) rather than IEC 60601‑1‑6.

Practical applications and users

This amendment is essential for:

  • Manufacturers designing or updating X‑ray tube assemblies for diagnostic equipment
  • Test laboratories performing type testing, marking checks and radiation/filtration verification
  • Regulatory and conformity assessment bodies assessing compliance with IEC/EN medical device safety requirements
  • Clinical engineers and purchasers specifying compatible high‑voltage generators, connectors and auxiliary equipment
  • Risk managers integrating usability and safety data into the risk management file

Related standards

  • IEC 60601‑1 and its amendments (general requirements)
  • IEC 60601‑1‑3 (radiation protection) and IEC 60601‑1‑2 (EMC)
  • IEC 60336 (focal spot), IEC 60522‑1:2020 (filtration), IEC 60806 (radiation field)
  • ISO 14971:2019 and ISO/TR 24971 (risk management/usability)

This amendment helps ensure X‑ray tube assemblies meet contemporary safety, labelling and radiation‑protection expectations and supports clear integration into diagnostic X‑ray systems.

Draft

EN IEC 60601-2-28:2020/oprA1:2025

English language
12 pages
Preview
Preview
e-Library read for
1 day

Frequently Asked Questions

EN IEC 60601-2-28:2019/prA1:2025 is a draft published by CLC. Its full title is "Amendment 1 - Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis". This standard covers: Amendment 1 - Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis

Amendment 1 - Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis

EN IEC 60601-2-28:2019/prA1:2025 is classified under the following ICS (International Classification for Standards) categories: 11.040.55 - Diagnostic equipment. The ICS classification helps identify the subject area and facilitates finding related standards.

EN IEC 60601-2-28:2019/prA1:2025 has the following relationships with other standards: It is inter standard links to EN IEC 60601-2-28:2019. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

EN IEC 60601-2-28:2019/prA1:2025 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


SLOVENSKI STANDARD
01-december-2025
Medicinska električna oprema - 2-28. del: Posebne zahteve za osnovno varnost in
bistvene lastnosti rentgenskih sestavov za medicinsko diagnostiko - Dopolnilo A1
Amendment 1 - Medical electrical equipment - Part 2-28: Particular requirements for the
basic safety and essential performance of X-ray tube assemblies for medical diagnosis
Appareils électromédicaux - Partie 2-28: Exigences particulières pour la sécurité de base
et les performances essentielles des gaines équipées pour diagnostic médical
Ta slovenski standard je istoveten z: EN IEC 60601-2-28:2019/prA1:2025
ICS:
11.040.50 Radiografska oprema Radiographic equipment
11.040.55 Diagnostična oprema Diagnostic equipment
SIST EN IEC 60601-2- en
28:2020/oprA1:2025
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

62B/1396/CDV
COMMITTEE DRAFT FOR VOTE (CDV)
PROJECT NUMBER:
IEC 60601-2-28/AMD1 ED3
DATE OF CIRCULATION: CLOSING DATE FOR VOTING:
2025-11-07 2026-01-30
SUPERSEDES DOCUMENTS:
62B/1392/RR
IEC SC 62B : MEDICAL IMAGING EQUIPMENT, SOFTWARE, AND SYSTEMS
SECRETARIAT: SECRETARY:
Germany Ms Regina Geierhofer
OF INTEREST TO THE FOLLOWING COMMITTEES: HORIZONTAL FUNCTION(S):
SC 62C
ASPECTS CONCERNED:
Safety
SUBMITTED FOR CENELEC PARALLEL VOTING NOT SUBMITTED FOR CENELEC PARALLEL VOTING

This document is still under study and subject to change. It should not be used for reference purposes.
Recipients of this document are invited to submit, with their comments, notification of any relevant patent rights of
which they are aware and to provide supporting documentation.
Recipients of this document are invited to submit, with their comments, notification of any relevant “In Some
Countries” clauses to be included should this proposal proceed. Recipients are reminded that the CDV stage is
the final stage for submitting ISC clauses. (SEE AC/22/2007 OR NEW GUIDANCE DOC).

TITLE:
Amendment 1 - Medical electrical equipment - Part 2-28: Particular requirements for the basic
safety and essential performance of X-ray tube assemblies for medical diagnosis

PROPOSED STABILITY DATE: 2028
NOTE FROM TC/SC OFFICERS:
electronic file, to make a copy and to print out the content for the sole purpose of preparing National Committee positions.
You may not copy or "mirror" the file or printed version of the document, or any part of it, for any other purpose without
permission in writing from IEC.

IEC CDV 60601-2-28/AMD1 © IEC 2025
1 INTERNATIONAL ELECTROTECHNICAL COMMISSION
2 ____________
4 MEDICAL ELECTRICAL EQUIPMENT
6 Part 2-28: Particular requirements for the basic safety
7 and essential performance of X-ray tube assemblies for medical diagnosis
9 AMENDMENT 1
11 FOREWORD
12 Amendment 1 to IEC 60601-2-28:2017 has been prepared by subcommittee 62B: Medical
13 imaging equipment, software, and systems, of IEC technical committee 62: Medical equipment,
14 software, and systems.
15 The text of this Amendment is based on the following documents:
Draft / FDIS Report on voting
62B/XXXX/FDIS 62B/XXXX/RVD
17 Full information on the voting for its approval can be found in the report on voting indicated in
18 the above table.
19 The language used for the development of this amendment is English.
20 This document was drafted in accordance with ISO/IEC Directives, Part 2, and developed in
21 accordance with ISO/IEC Directives, Part 1 and ISO/IEC Directives, IEC Supplement, available
22 at www.iec.ch/members_experts/refdocs. The main document types developed by IEC are
23 described in greater detail at www.iec.ch/publications.
24 The committee has decided that the contents of this document will remain unchanged until the
25 stability date indicated on the IEC website under webstore.iec.ch in the data related to the
26 specific document. At this date, the document will be
27 • reconfirmed,
28 • withdrawn,
29 • replaced by a revised edition, or
30 • amended.
31 NOTE The attention of National Committees is drawn to the fact that equipment MANUFACTURERS and testing
32 organizations may need a transitional period following publication of a new, amended or revised IEC publication in
33 which to make products in accordance with the new requirements and to equip themselves for conducting new or
34 revised tests. It is the recommendation of the committee that the content of this publication be adopted for
35 implementation nationally not earlier than 3 years from the date of publication.
37 ____________
IEC CDV 60601-2-28/AMD1 © IEC 2025
39 INTRODUCTION to Amendment 1
40 This first amendment to the third edition of IEC 60601-2-28 has been prepared to provide a
41 complete set of safety requirements for X-RAY TUBE ASSEMBLIES, based on the second
42 amendment (2020) to IEC 60601-1:2005 and associated collateral standards. It corrects
43 references to actual standards (e.g. IEC 60336, IEC 60522-1, and the IEC 60601 series) and
44 clarifies single requirements within the standard.
45 Users of this document should note that when constructing the dated references to specific
46 elements in a standard, such as definitions, amendments are only referenced if they modified
47 the text being cited. For example, if a reference is made to a definition that has not been
48 modified by an amendment, then the reference to the amendment is not included in the dated
49 reference.
IEC CDV 60601-2-28/AMD1 © IEC 2025
51 201.1.1 Scope, object and related standards
52 Replace, in the existing footnote 1), “IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012”,
53 with “IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020”.
55 201.1.3 Collateral standards
56 Replace, in the existing second paragraph of the addition, “IEC 60601-1-3:2008 and IEC 60601-
57 1-3:2008/AMD1:2013”, with “IEC 60601-1-3:2008 and IEC 60601-1-3:2008/AMD1:2013 and
58 IEC 60601-1-3:2008/AMD2:2021”.
60 201.1.4 Particular standards
61 Replace, in the existing third paragraph of the replacement, “IEC 60601-1:2005 and IEC 60601-
62 1:2005/AMD1:2012”, with “IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-
63 1:2005/AMD2:2020”.
65 Replace, in the existing eighth paragraph of the replacement, “3.147” with “3.154”.
67 201.2 Normative references
68 Replace the existing reference to “IEC 60601-1-3:2008” with:
69 IEC 60601-1-3:2008, Medical electrical equipment – Part 1-3: General requirements for basic
70 safety and essential performance – Collateral Standard: Radiation protection in diagnostic X-
71 ray equipment
72 IEC 60601-1-3:2008/AMD1:2013
73 IEC 60601-1-3:2008/AMD2:2021
75 Replace the existing reference to “IEC 60336” with:
76 IEC 60336, Medical electrical equipment – X-ray tube assemblies for medical diagnosis – Focal
77 spot dimensions and related characteristics
79 Replace the existing reference to “IEC 60522” with:
80 IEC 60522-1:2020, Medical electrical equipment – Diagnostic X-rays – Part 1: Determination of
81 quality equivalent filtration and permanent filtration
IEC CDV 60601-2-28/AMD1 © IEC 2025
84 201.3 Terms and definitions
85 Replace the existing reference to “IEC 60522” with “IEC 60522-1:2020”.
87 201.7.2.11 Mode of operation
88 Delete the existing NOTE.
90 201.7.2.101 Marking of X-RAY TUBES
91 Replace the existing subclause by:
92 X-RAY TUBES shall be provided with the following markings:
93 • name or trademark of the MANUFACTURER;
94 • MODEL OR TYPE REFERENCE;
95 • individual identification.
96 The markings shall enable individual products, series or types to be correlated with their
97 ACCOMPANYING DOCUMENTS.
98 The markings on the X-RAY TUBE shall remain CLEARLY LEGIBLE during the EXPECTED SERVICE
99 LIFE.
100 The above markings may be given in the form of a combined designation explained in the
101 ACCOMPANYING DOCUMENTS.
103 Add, after the existing subclause 201.7.3.2, the following new subclause:
104 201.7.3.7 Supply terminals
105 Replacement of the first paragraph by:
106 If terminals for supply conductors are non-interchangeable by design, or, if otherwise, there is
107 no unacceptable RISK if the terminals are interchanged, then marking of the supply terminals or
108 marking adjacent to the supply terminals is not required. Otherwise, supply terminals shall be
109 marked adjacent to the terminals.
110 NOTE For this modification of the general standard the term supply terminal according to IEV 845-28-064 has been
111 considered.
113 201.7.9 ACCOMPANYING DOCUMENTS
114 201.7.9.1 General
115 Replace the existing NOTE 104 with:
116 NOTE 104 Instead of the USABILITY ENGINEERING PROCESS (IEC 60601-1-6 is not applicable for an X-RAY TUBE
117 ASSEMBLY), the RISK MANAGEMENT PROCESS (based on ISO 14971:2019 and its accompanying TR – ISO/TR
118 24971:2020) is the carrier of the considerations. ISO 14971:2019, 5.2 (INTENDED USE and reasonable foreseeable
119 misuse) and ISO/TR 24971:2020 A.2.31 (USABILITY of the user interface) cover the USABILITY aspects of X-RAY TUBE
120 ASSEMBLIES sufficiently.
IEC CDV 60601-2-28/AMD1 © IEC 2025
121 201.7.9.3.101 Technical description of X-RAY TUBE ASSEMBLIES
122 Replace, in the existing list item e), the reference to “IEC 60522” with “IEC 60522-1:2020”.
124 Replace the existing list items h) to k) with:
125 h) type-designation of compatible electrical supply equipment required to power the X-RAY
126 TUBE ASSEMBLY, including HIGH-VOLTAGE GENERATOR(S), or specifications concerning:
127 1) the HIGH VOLTAGE required from the HIGH-VOLTAGE GENERATOR,
128 2) electrical supply of the ELECTRON emitter(s),
129 3) electrical supply for electron beam control, e.g., focusing, pulsing, deflecting, etc. (as
130 applicable)
131 4) electrical supply for rotating the ANODE (as applicable),
132 5) CATHODE EMISSION CHARACTERISTICS.
133 i) type-designation or specification of HIGH VOLTAGE connectors, if removable in NORMAL USE;
134 j) type-designation of compatible auxiliary equipment (such as for cooling purposes), or
135 requirements for such auxiliary functions;
136 k) any supply interface requirement(s) in addition to items h) through j) above which are
...

Questions, Comments and Discussion