EN ISO 8835-3:2007
(Main)Inhalational anaesthesia systems - Part 3: Transfer and receiving systems of active anaesthetic gas scavenging systems (ISO 8835-3:2007)
Inhalational anaesthesia systems - Part 3: Transfer and receiving systems of active anaesthetic gas scavenging systems (ISO 8835-3:2007)
ISO 8835-3:2007 specifies requirements for transfer and receiving systems of active anaesthetic gas scavenging systems (active AGSSs) intended to reduce exposure of healthcare personnel to anaesthetic gases and vapours while providing patient protection (e.g. against excessive flow and pressure). ISO 8835-3:2007 also specifies requirements for transfer and receiving systems of active anaesthetic gas scavenging systems in which the power device is integral with the transfer and receiving system.
Systeme für die Inhalationsanästhesie - Teil 3: Weiterleitungs- und Aufnahmesysteme von aktiven Anästhesiegas-Fortleitungssystemen (ISO 8835-3:2007)
Dieser Teil von ISO 8835 legt Anforderungen für Weiterleitungs und Aufnahmesysteme von aktiven Anästhesiegas-Fortleitungssystemen (aktive AGFS) fest, deren Ziel es ist, die Exposition des medizinischen Personals gegenüber Anästhesiegasen und dämpfen zu verringern und gleichzeitig den Patienten zu schützen (z. B. gegen übermäßigen Gasdurchfluss und Druck). Dieser Teil von ISO 8835 legt auch Anforderungen für Weiterleitungs- und Aufnahmesysteme von aktiven Anästhesiegas-Fortleitungssystemen fest, in denen die Antriebseinrichtung in das Weiterleitungs und Aufnahmesystem integriert ist.
Dieser Teil von ISO 8835 legt keine Anforderungen fest für
Entsorgungssysteme, die in ISO 7396-2 behandelt werden,
nichtaktive AGFS (passive AGFS),
in der Nähe untergebrachte Gasfortleitungssysteme (d. h. Systeme, die nicht direkt an das Atemsystem angeschlossen sind, oder zugehörige Geräte),
Weiterleitungs und Aufnahmesysteme, die für die Verwendung mit brennbaren Anästhesiemitteln nach IEC 60601-2-13:2003, Anhang DD, bestimmt sind.
Systèmes d'anesthésie par inhalation - Partie 3: Systèmes de transfert et de réception des systèmes d'évacuation des gaz d'anesthésie (ISO 8835-3:2007)
L'ISO 8835-3:2007 spécifie les exigences relatives aux systèmes de transfert et aux systèmes de réception des systèmes d'évacuation des gaz d'anesthésie actifs (SEGA actifs), conçus pour réduire l'exposition du personnel de santé aux gaz et aux vapeurs d'anesthésie, tout en protégeant le patient (par exemple contre une pression ou un débit de gaz excessifs). L'ISO 8835-3:2007 spécifie également les exigences applicables aux systèmes de transfert et de réception des systèmes d'évacuation des gaz d'anesthésie actifs dans lesquels la source d'alimentation fait partie intégrante du système de transfert et de réception.
Inhalacijski anestezijski sistemi - 3. del: Sistemi za prenos in sprejem sistemov za odstranjevanje anestezijskih plinov (ISO 8835-3:2007)
General Information
Relations
Standards Content (Sample)
SLOVENSKI STANDARD
01-april-2008
1DGRPHãþD
SIST EN 740:2000
SIST EN 740:2000/A1:2005
SIST EN 740:2000/AC:2000
Inhalacijski anestezijski sistemi - 3. del: Sistemi za prenos in sprejem sistemov za
odstranjevanje anestezijskih plinov (ISO 8835-3:2007)
Inhalational anaesthesia systems - Part 3: Transfer and receiving systems of active
anaesthetic gas scavenging systems (ISO 8835-3:2007)
Systeme für die Inhalationsanästhesie - Anästhesiegas-Fortleitungssysteme - Teil 3:
Weiterleitungs- und Aufnahmesysteme von aktiven Anästhesiegas-Fortleitungssystemen
(ISO 8835-3:2007)
Systemes d'anesthésie par inhalation - Partie 3: Systemes de transfert et de réception
des systemes d'évacuation des gaz d'anesthésie (ISO 8835-3:2007)
Ta slovenski standard je istoveten z: EN ISO 8835-3:2007
ICS:
11.040.10
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD
EN ISO 8835-3
NORME EUROPÉENNE
EUROPÄISCHE NORM
August 2007
ICS 11.040.10 Supersedes EN 740:1998
English Version
Inhalational anaesthesia systems - Part 3: Transfer and
receiving systems of active anaesthetic gas scavenging systems
(ISO 8835-3:2007)
Systèmes d'anesthésie par inhalation - Partie 3: Systèmes Systeme für die Inhalationsanästhesie - Teil 3:
de transfert et de réception des systèmes d'évacuation des Weiterleitungs- und Aufnahmesysteme von aktiven
gaz d'anesthésie (ISO 8835-3:2007) Anästhesiegas-Fortleitungssystemen (ISO 8835-3:2007)
This European Standard was approved by CEN on 9 June 2007.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European
Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national
standards may be obtained on application to the CEN Management Centre or to any CEN member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CEN member into its own language and notified to the CEN Management Centre has the same status as the
official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Estonia, Finland,
France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal,
Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre: rue de Stassart, 36 B-1050 Brussels
© 2007 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN ISO 8835-3:2007: E
worldwide for CEN national Members.
Foreword
This document (EN ISO 8835-3:2007) has been prepared by Technical Committee ISO/TC
121 "Anaesthetic and respiratory equipment" in collaboration with Technical Committee
CEN/TC 215 "Respiratory and anaesthetic equipment", the secretariat of which is held by
BSI.
This European Standard shall be given the status of a national standard, either by publication
of an identical text or by endorsement, at the latest by February 2008, and conflicting national
standards shall be withdrawn at the latest by May 2009.
This document has been prepared under a mandate given to CEN by the European
Commission and the European Free Trade Association, and supports essential requirements
of EU Directive(s).
For relationship with EU Directive(s), see informative Annex ZA, which is an integral part of
this document.
Other European Standards relating to anaesthetic workstations and their components
prepared by CEN/TC 215 which, together with EN 60601-2-13:2006, replace EN 740:1998 in
total, are:
– EN ISO 8835-2:2007, Inhalational anaesthesia systems – Part 2: Anaesthetic breathing
systems (ISO 8835-2:2007)
– EN ISO 8835-3:2007, Inhalational anaesthesia systems – Part 3: Transfer and receiving
systems of active anaesthetic gas scavenging systems (ISO 8835-3:2007)
– EN ISO 8835-4:2004, Inhalational anaesthesia systems – Part 4: Anaesthetic vapour
delivery devices (ISO 8835-4:2004)
– EN ISO 8835-5:2004, Inhalational anaesthesia systems – Part 5: Anaesthetic ventilators
(ISO 8835-5:2004)
Attention is also drawn to ISO/TS 18835:2004, Inhalational anaesthesia systems — Draw-
over vaporizers and associated equipment.
Attention is drawn to the possibility that some of the elements of this document may be the
subject of patent rights. CEN shall not be held responsible for identifying any or all such
patent rights.
According to the CEN/CENELEC Internal Regulations, the national standards organizations of
the following countries are bound to implement this European Standard: Austria, Belgium,
Bulgaria, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece,
Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United
Kingdom.
Endorsement notice
The text of ISO 8835-3:2007 has been approved by CEN as EN ISO 8835-3:2007 without any
modifications.
ANNEX ZA
(informative)
Relationship between this standard and the Essential
Requirements of EU Directive 93/42/EEC
This European Standard has been prepared under a mandate given to CEN by the European
Commission and the European Free Trade Association to provide a means to conforming to
Essential Requirements of the New Approach Directive 93/42/EEC, Council Directive of 14
June 1993 on the approximation of the laws of the Member States concerning medical
devices (Medical Device Directive).
Once this European Standard is cited in the Official Journal of the European Communities
under that Directive and has been implemented as a national standard in at least one
Member State, compliance with the clauses of this European Standard given in Table ZA.1
confers, within the limits of the scope of this European Standard, a presumption of conformity
with the corresponding Essential Requirements of that Directive and associated EFTA
regulations.
Table ZA.1 — Correspondence between this International Standard and EU Directive
93/42/EEC
Clause(s)/sub-clause(s) of Essential requirements (ERs) of EU Qualifying remarks/Notes
this International Standard Directive 93/42/EEC
All 2, 12.7.4
Warning – Other requirements and other EU Directives may be applicable to the
product(s) falling within the scope of this standard.
INTERNATIONAL ISO
STANDARD 8835-3
Second edition
2007-08-15
Inhalational anaesthesia systems —
Part 3:
Transfer and receiving systems of active
anaesthetic gas scavenging systems
Systèmes d'anesthésie par inhalation —
Partie 3: Systèmes de transfert et de réception des systèmes
d'évacuation des gaz d'anesthésie
Reference number
ISO 8835-3:2007(E)
©
ISO 2007
ISO 8835-3:2007(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.
© ISO 2007
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2007 – All rights reserved
ISO 8835-3:2007(E)
Contents Page
Foreword. iv
Introduction . v
* 1 Scope . 1
2 Normative references . 1
3 Terms and definitions. 2
4 General requirements and alternative test methods. 4
4.1 Materials . 4
4.2 Means of pressure relief . 4
4.3 Alternative test methods. 4
* 5 Patient and environmental protection . 4
5.1 Normal operating conditions. 4
5.2 Single fault conditions . 5
* 6 Connectors . 5
* 7 Transfer systems . 5
7.1 Inlet of transfer systems . 5
7.2 Outlet of transfer systems . 6
* 8 Receiving systems. 6
8.1 General. 6
8.2 Inlet of receiving systems. 6
8.3 Outlet of receiving systems. 7
8.4 Visual indicator . 7
8.5 Particle filter . 7
8.6 Receiving hoses. 7
9 Extract flow resistance. 7
9.1 Low-flow transfer and receiving systems . 7
9.2 High-flow transfer and receiving systems . 7
10 Electrical requirements . 8
* 11 Information to be supplied by the manufacturer. 8
12 Marking . 8
* 13 Colour coding. 8
Annex A (informative) Typical test arrangement and method for pressure measurement at inlet to
AGSS. 9
Annex B (informative) Typical test arrangement and method for sub-atmospheric pressure
limitation . 13
Annex C (informative) Typical test arrangement and method for testing of induced flow. 14
Annex D (informative) Typical test arrangement and method for resistance to extract flow. 15
Annex E (informative) Typical test arrangement and method for spillage. 17
Annex F (informative) Rationale . 20
Annex G (informative) Environmental aspects . 22
Bibliography . 23
----------------------
...