EN ISO 10360-12:2016
(Main)Geometrical product specifications (GPS) - Acceptance and reverification tests for coordinate measuring systems (CMS) - Part 12: Articulated arm coordinate measurement machines (CMM) (ISO 10360-12:2016)
- Abstract
ISO 10360-12:2016 specifies the acceptance tests for verifying the performance of an articulated arm CMM by measuring calibrated test lengths as stated by the manufacturer. It also specifies the reverification tests that enable the user to periodically reverify the performance of the articulated arm CMM. It applies to articulated arm CMMs using tactile probes and optionally optical distance sensors (also referred to as laser line scanners or laser line probes). Details on tests for scanner accessories are given in Annex E.
ISO 10360-12:2016 does not specify how often or when testing is performed, if at all, nor does it specify which party should bear the cost of testing.
This part of ISO 10360 specifies
- performance requirements that can be assigned by the manufacturer or the user of the articulated arm CMM,
- the manner of execution of the acceptance and reverification tests to demonstrate the stated requirements,
- rules for proving conformance, and
- applications for which the acceptance and reverification tests can be used.
- Status
- Published
- Publication Date
- 01-Nov-2016
- Withdrawal Date
- 30-May-2017
- Current Stage
- 6060 - Definitive text made available (DAV) - Publishing
- Start Date
- 02-Nov-2016
- Due Date
- 02-Nov-2016
- Completion Date
- 02-Nov-2016
Overview
EN ISO 10360-12:2016 (ISO 10360-12:2016) defines acceptance and reverification tests for articulated arm coordinate measurement machines (CMMs) - a subset of coordinate measuring systems (CMS). The standard specifies how to verify performance by measuring calibrated test lengths (traceable to the metre) and covers articulated arms using tactile probes and, optionally, optical distance sensors (e.g., laser line scanners/laser line probes). It enables manufacturers to declare maximum permissible errors (MPEs) and allows users to reverify that articulated arm CMMs meet those stated requirements. The standard does not set test intervals or allocate testing costs.
Key Topics and Requirements
- Performance requirements and MPEs: Defines performance parameters that can be assigned by the manufacturer or user.
- Acceptance and reverification procedures: Prescribes the manner of execution for tests that demonstrate conformance to declared performance.
- Test types covered:
- Probing size and form errors - assessing probe-related contributions.
- Articulated location errors - evaluating joint/pose accuracy across the arm workspace.
- Length measurement errors - measuring point-to-point and concatenated lengths using calibrated artefacts.
- Measured evidence and rules for conformance: Procedures for data collection, rejection, repeated measurements and criteria for proving compliance.
- Operating conditions: Rated environmental and operating conditions under which tests are valid.
- Scanner accessory testing: Detailed test methods for scanning probing systems and scanner accessories (see Annex E).
- Documentation: Requirements for indicating test results and specifications in product literature and data sheets.
Applications and Who Uses It
- Manufacturers of articulated arm CMMs - to specify and declare performance (MPEs) and to support product claims.
- Calibration laboratories and metrology/service providers - to perform acceptance testing, reverification and calibration using traceable artefacts.
- Quality and inspection teams in industry - aerospace, automotive, tooling, heavy equipment, and reverse-engineering operations that rely on articulated arm measurements for part inspection, assembly verification, and portable inspection tasks.
- R&D and procurement - to compare articulated arm CMM performance when selecting systems or verifying supplier claims.
Practical benefits include traceable length verification, repeatable acceptance procedures, and standardized reporting that aids procurement, compliance, and process control.
Related Standards
- ISO 10360 series (other parts for different CMS types)
- ISO 14638 (GPS masterplan)
- ISO 8015 (fundamental GPS rules)
- ISO 14253-1 (decision rules for conformity assessment)
Keywords: ISO 10360-12, EN ISO 10360-12:2016, articulated arm CMM, coordinate measuring system, acceptance tests, reverification tests, tactile probe, laser line scanner, MPE, calibrated test length, traceability.
Relations
- Effective Date
- 09-Feb-2026
- Effective Date
- 09-Feb-2026
- Effective Date
- 12-Feb-2026
Get Certified
Connect with accredited certification bodies for this standard
BSMI (Bureau of Standards, Metrology and Inspection)
Taiwan's standards and inspection authority.
Sponsored listings
Frequently Asked Questions
EN ISO 10360-12:2016 is a standard published by the European Committee for Standardization (CEN). Its full title is "Geometrical product specifications (GPS) - Acceptance and reverification tests for coordinate measuring systems (CMS) - Part 12: Articulated arm coordinate measurement machines (CMM) (ISO 10360-12:2016)". This standard covers: ISO 10360-12:2016 specifies the acceptance tests for verifying the performance of an articulated arm CMM by measuring calibrated test lengths as stated by the manufacturer. It also specifies the reverification tests that enable the user to periodically reverify the performance of the articulated arm CMM. It applies to articulated arm CMMs using tactile probes and optionally optical distance sensors (also referred to as laser line scanners or laser line probes). Details on tests for scanner accessories are given in Annex E. ISO 10360-12:2016 does not specify how often or when testing is performed, if at all, nor does it specify which party should bear the cost of testing. This part of ISO 10360 specifies - performance requirements that can be assigned by the manufacturer or the user of the articulated arm CMM, - the manner of execution of the acceptance and reverification tests to demonstrate the stated requirements, - rules for proving conformance, and - applications for which the acceptance and reverification tests can be used.
ISO 10360-12:2016 specifies the acceptance tests for verifying the performance of an articulated arm CMM by measuring calibrated test lengths as stated by the manufacturer. It also specifies the reverification tests that enable the user to periodically reverify the performance of the articulated arm CMM. It applies to articulated arm CMMs using tactile probes and optionally optical distance sensors (also referred to as laser line scanners or laser line probes). Details on tests for scanner accessories are given in Annex E. ISO 10360-12:2016 does not specify how often or when testing is performed, if at all, nor does it specify which party should bear the cost of testing. This part of ISO 10360 specifies - performance requirements that can be assigned by the manufacturer or the user of the articulated arm CMM, - the manner of execution of the acceptance and reverification tests to demonstrate the stated requirements, - rules for proving conformance, and - applications for which the acceptance and reverification tests can be used.
EN ISO 10360-12:2016 is classified under the following ICS (International Classification for Standards) categories: 17.040.30 - Measuring instruments. The ICS classification helps identify the subject area and facilitates finding related standards.
EN ISO 10360-12:2016 has the following relationships with other standards: It is inter standard links to ISO 10360-8:2013, ISO 10360-9:2013, ISO 10360-12:2016. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
EN ISO 10360-12:2016 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
SLOVENSKI STANDARD
01-januar-2017
6SHFLILNDFLMDJHRPHWULMVNLKYHOLþLQL]GHOND*363UHVNXVL]DVSUHMHPOMLYRVWLQ
SRQRYQRRYHUMDQMHNRRUGLQDWQLKPHULOQLKVWURMHY.06GHO*LEOMLYDURND
NRRUGLQDWQLKPHULOQLKVWURMHY,62
Geometrical Product Specifications (GPS) - Acceptance and reverification tests for
coordinate measuring systems (CMS) - Part 12: Articulated arm coordinate
measurement machines (CMM) (ISO 10360-12:2016)
Geometrische Produktspezifikation (GPS) - Annahme- und Bestätigungsprüfung für
Koordinatenmesssysteme (KMS) - Teil 12: Koordinatenmessgeräte (KMG) mit
Gelenkausleger (ISO 10360-12:2016)
Spécification géométrique des produits (GPS) - Essais de réception et de vérification
périodique des machines à mesurer tridimensionnelles (MMT) - Partie 12: MMT à bras
articulés (ISO 10360-12:2016)
Ta slovenski standard je istoveten z: EN ISO 10360-12:2016
ICS:
17.040.30 Merila Measuring instruments
17.040.40 6SHFLILNDFLMDJHRPHWULMVNLK Geometrical Product
YHOLþLQL]GHOND*36 Specification (GPS)
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EN ISO 10360-12
EUROPEAN STANDARD
NORME EUROPÉENNE
November 2016
EUROPÄISCHE NORM
ICS 17.040.30
English Version
Geometrical product specifications (GPS) - Acceptance and
reverification tests for coordinate measuring systems
(CMS) - Part 12: Articulated arm coordinate measurement
machines (CMM) (ISO 10360-12:2016)
Spécification géométrique des produits (GPS) - Essais Geometrische Produktspezifikation (GPS) - Annahme-
de réception et de vérification périodique des systèmes und Bestätigungsprüfung für Koordinatenmesssysteme
de mesure tridimensionnels (SMT) - Partie 12: (KMS) - Teil 12: Koordinatenmessgeräte (KMG) mit
Machines à mesurer tridimensionnelles à bras articulés Gelenkausleger (ISO 10360-12:2016)
(MMT) (ISO 10360-12:2016)
This European Standard was approved by CEN on 27 August 2016.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this
European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references
concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN
member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by
translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management
Centre has the same status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania,
Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and
United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2016 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN ISO 10360-12:2016 E
worldwide for CEN national Members.
Contents Page
European foreword . 3
European foreword
This document (EN ISO 10360-12:2016) has been prepared by Technical Committee
ISO/TC 213 “Dimensional and geometrical product specifications and verification” in collaboration with
Technical Committee CEN/TC 290 “Dimensional and geometrical product specification and verification”
the secretariat of which is held by AFNOR.
This European Standard shall be given the status of a national standard, either by publication of an
identical text or by endorsement, at the latest by May 2017, and conflicting national standards shall be
withdrawn at the latest by May 2017.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CEN [and/or CENELEC] shall not be held responsible for identifying any or all such patent
rights.
According to the CEN-CENELEC Internal Regulations, the national standards organizations of the
following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria,
Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia,
France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta,
Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.
Endorsement notice
The text of ISO 10360-12:2016 has been approved by CEN as EN ISO 10360-12:2016 without any
modification.
INTERNATIONAL ISO
STANDARD 10360-12
First edition
2016-10-01
Geometrical product specifications
(GPS) — Acceptance and reverification
tests for coordinate measuring
systems (CMS) —
Part 12:
Articulated arm coordinate
measurement machines (CMM)
Spécification géométrique des produits (GPS) — Essais de
réception et de vérification périodique des systèmes de mesure
tridimensionnels (SMT) —
Partie 12: Machines à mesurer tridimensionnelles à bras articulés
(MMT)
Reference number
ISO 10360-12:2016(E)
©
ISO 2016
ISO 10360-12:2016(E)
© ISO 2016, Published in Switzerland
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Ch. de Blandonnet 8 • CP 401
CH-1214 Vernier, Geneva, Switzerland
Tel. +41 22 749 01 11
Fax +41 22 749 09 47
copyright@iso.org
www.iso.org
ii © ISO 2016 – All rights reserved
ISO 10360-12:2016(E)
Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Symbols . 6
5 Rated operating conditions . 7
5.1 Environmental conditions . 7
5.2 Operating conditions . 8
6 Acceptance tests and reverification tests . 8
6.1 General . 8
6.2 Probing size and form errors . 8
6.2.1 Principle . 8
6.2.2 Measuring equipment . 8
6.2.3 Procedure . 8
6.2.4 Derivation of test results .10
6.3 Articulated location errors .10
6.3.1 Principle .10
6.3.2 Measuring equipment .10
6.3.3 Procedure .10
6.3.4 Derivation of test results .11
6.4 Length measurement errors .11
6.4.1 Principle .11
6.4.2 Measuring equipment .12
6.4.3 Procedure .12
7 Compliance with specification .16
7.1 Acceptance tests .16
7.1.1 Acceptance criteria .16
7.1.2 Data rejection and repeated measurements .16
7.2 Reverification tests .17
8 Applications .17
8.1 Acceptance test .17
8.2 Reverification test .17
8.3 Interim check .18
9 Indication in product documentation and data sheets .19
Annex A (informative) Forms .20
Annex B (normative) Artefacts that represent a calibrated test length.22
Annex C (informative) Alignment of artefacts .28
Annex D (informative) Interim testing .29
Annex E (normative) Testing a scanning probing system of an articulated arm CMM .31
Annex F (normative) Length error measurement by concatenating test lengths .32
Annex G (informative) Optional probing articulated size and forms errors .37
Annex H (informative) Optional repeatability range of the length measurement error .38
Annex I (informative) Relation to the GPS matrix model .39
Bibliography .40
ISO 10360-12:2016(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation on the meaning of ISO specific terms and expressions related to conformity assessment,
as well as information about ISO’s adherence to the World Trade Organization (WTO) principles in the
Technical Barriers to Trade (TBT) see the following URL: www.iso.org/iso/foreword.html.
The committee responsible for this document is ISO/TC 213, Dimensional and geometrical product
specifications and verification.
ISO 10360 consists of the following parts, under the general title Geometrical Product specifications
(GPS) — Acceptance and reverification tests for coordinate measuring systems (CMS):
— Part 1: Vocabulary
— Part 2: CMMs used for measuring linear dimensions
— Part 3: CMMs with the axis of a rotary table as the fourth axis
— Part 4: CMMs used in scanning measuring mode
— Part 5: CMMs using single and multiple stylus contacting probing system
— Part 6: Estimation of errors in computing of Gaussian associated features
— Part 7: CMMs equipped with imaging probing systems
— Part 8: CMMs with optical distance sensors
— Part 9: CMMs with multiple probing systems
— Part 10: Laser trackers for measuring point-to-point distances
— Part 12: Articulated arm coordinate measuring machines (CMM)
iv © ISO 2016 – All rights reserved
ISO 10360-12:2016(E)
Introduction
This part of ISO 10360 is a general GPS standard (see ISO 14638). For more detailed information about
the relation of this part of ISO 10360 to other standards and the GPS matrix model, see Annex I.
This part of ISO 10360 is included in the ISO/GPS Masterplan given in ISO 14638, which gives an
overview of the ISO/GPS system. The fundamental rules of ISO/GPS given in ISO 8015 apply to this
part of ISO 10360 and the default decision rules given in ISO 14253-1 apply to specifications made in
accordance with this part of ISO 10360, unless otherwise indicated.
The objective of this part of ISO 10360 is to provide a well-defined testing procedure to
— enable manufacturers of articulated arm CMMs to provide specification MPEs, and
— enable users to test articulated arm CMMs to manufacturer specifications using calibrated traceable
reference artefacts.
The benefits of these tests are that the measured result has a direct traceability to the unit length, the
metre, and that they give information on how the articulated arm CMM will perform on similar length
measurements.
INTERNATIONAL STANDARD ISO 10360-12:2016(E)
Geometrical product specifications (GPS) — Acceptance
and reverification tests for coordinate measuring
systems (CMS) —
Part 12:
Articulated arm coordinate measurement machines (CMM)
1 Scope
This part of ISO 10360 specifies the acceptance tests for verifying the performance of an articulated
arm CMM by measuring calibrated test lengths as stated by the manufacturer. It also specifies the
reverification tests that enable the user to periodically reverify the performance of the articulated arm
CMM. It applies to articulated arm CMMs using tactile probes and optionally optical distance sensors
(also referred to as laser line scanners or laser line probes). Details on tests for scanner accessories are
given in Annex E.
This part of ISO 10360 does not specify how often or when testing is performed, if at all, nor does it
specify which party should bear the cost of testing.
This part of ISO 10360 specifies
— performance requirements that can be assigned by the manufacturer or the user of the articulated
arm CMM,
— the manner of execution of the acceptance and reverification tests to demonstrate the stated
requirements,
— rules for proving conformance, and
— applications for which the acceptance and reverification tests can be used.
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.
ISO 10360-8:2013, Geometrical product specifications (GPS) — Acceptance and reverification tests for
coordinate measuring systems (CMS) — Part 8: CMMs with optical distance sensors
ISO 10360-9:2013, Geometrical product specifications (GPS) — Acceptance and reverification tests for
coordinate measuring systems (CMS) — Part 9: CMMs with multiple probing systems
3 Terms and definitions
For the purposes of this document, the following terms and definitions given in ISO 10360-1 and the
following apply.
NOTE The definitions in this section are intended to concisely state the meaning of terms. For metrological
characteristics that have numerical values, the complete description of the procedure and derivation of test
results in Clause 6 and Annex E are to be followed in determining values.
ISO 10360-12:2016(E)
3.1
articulated arm coordinate measuring machine
system that measures spatial coordinates and comprises
— an open chain of fixed-length segments,
— joint assemblies interconnecting the segments and the probing system and attaching them to the
stationary environment, and
— a probing system at the free end of the chain
Note 1 to entry: The probing system may comprise a rigid probe or a sensing system such as a scanner.
Note 2 to entry: Rotary joint assemblies connected to the fixed-length segments are equipped with angular
encoders. Cartesian coordinates of each measuring point are calculated from the measured angles and segment
lengths.
3.2
joint
connection between adjacent elements of an articulated arm CMM that allows a single rotational degree
of freedom between these elements
Note 1 to entry: There are two types of joints: hinge joints, which cause a hinging movement between adjacent
arm segments, and swivel joints, which cause a rotary movement around the axis of the connected arm segment.
Note 2 to entry: Each joint ordinarily includes an angle measuring device (rotary encoder).
3.3
joint assembly
assembly of two or more joints between two adjacent elements of an articulated arm CMM
Note 1 to entry: Usually, a joint assembly includes at least a hinge joint and a swivel joint.
Note 2 to entry: In analogy to the human arm, the three main joint assemblies are designated the shoulder, elbow,
and wrist.
Note 3 to entry: Current machines have 2 or 3 degrees of freedom each for shoulder (a, b), elbow (c, d), and wrist
(e, f, g), as shown in Figure 1. Consequently, articulated arm CMMs are referred to as either six or seven axis
machines.
a) With six rotary axes
2 © ISO 2016 – All rights reserved
ISO 10360-12:2016(E)
b) With seven rotary axes
Figure 1 — Articulated arm CMM
3.4
measuring range
diameter of the spherical volume within which an articulated arm CMM is capable of measuring
Note 1 to entry: The measuring range is specified by the manufacturer.
Note 2 to entry: The measuring range is twice the reach of the articulated arm. However, some of the regions that
can be reached by the articulated arm may not be within the measuring volume.
3.5
measuring volume
region in space over which the manufacturer specifies the performance of the articulated arm CMM
Note 1 to entry: The measuring volume is restricted by inaccessible zones specified by the manufacturer. For
example, there may be an inaccessible zone close to the vertical main axis.
Note 2 to entry: Manufacturers may specify more than one measuring volume for a machine, each measuring
volume having a separate performance specification.
Note 3 to entry: Because of the possibility of binding up a joint when adjacent arm segments are brought close
together, the size of the measuring volume may depend on the direction of the probe stylus in relation to the
outside of the measuring volume or inaccessible zones within the measuring volume. The manufacturer may
specify one or more measuring volumes according to the direction of the probe stylus.
3.6
useful arm length
half the measuring range
3.7
coefficient of thermal expansion
CTE
α
linear thermal expansion coefficient of a material at 20 °C
Note 1 to entry: The above definition for CTE does not imply that a user is required to make measurements at 20 °C.
ISO 10360-12:2016(E)
3.8
normal CTE material
−6 −6
material with a CTE between 8 × 10 /°C and 13 × 10 /°C
Note 1 to entry: Some documents may express CTE in units 1/K, which is equivalent to 1/°C.
[SOURCE: ISO 10360-2]
3.9
kinematic seat
mechanical seat (nest) that repeatably holds the centre of a spherical surface in a fixed position in space
Note 1 to entry: An example of a kinematic seat is a trihedral seat that includes three hardened spheres, each
sphere placed on a circle and separated from the other spheres by nominally 120°. Each of the three spheres
contacts the surface of a larger sphere (or spherical surface) so as to permit repeatable positioning of the centre
of the larger sphere in space.
Note 2 to entry: As used in this part of ISO 10360, a kinematic seat provides constraint for 3 degrees of freedom
rather than 6 degrees of freedom.
3.10
single-point articulation test
test in which articulated arm CMM probe is held within a kinematic seat while the elbow location is
rotated by 180°
Note 1 to entry: The single-point articulation test is an interim test described in Annex D.
3.11
articulated location error, tactile
L
Dia.5x5:Art:Tact.AArm
diameter of the minimum circumscribed sphere encompassing the points that are the centres of the
five spheres obtained from performing the articulated location test when using a tactile probe
Note 1 to entry: In the context of this part of ISO 10360, the local abbreviation L is used.
Dia.5x5:Art
3.12
length measurement error, bidirectional
E
Bi:0:Tact.AArm
error of indication when performing a bidirectional point-to-point distance measurement
Note 1 to entry: In the context of this part of ISO 10360, the local abbreviation E is used.
Bi
Note 2 to entry: The subscript 0 indicates that there is no tip offset. There may be an offset in some other parts of
ISO 10360.
3.13
length measurement error, unidirectional
E
Uni:0:Tact.AArm
error of indication when performing a unidirectional point-to-point distance measurement
Note 1 to entry: Annex B discusses unidirectional and bidirectional measurements.
Note 2 to entry: In the context of this part of ISO 10360, the local abbreviation E is used.
Uni
3.14
probing form error, tactile
P
Form.Sph.1x25::Tact.AArm
error of indication within which the range of Gaussian radial distances can be determined by a Gaussian
(least-squares) fit of 25 points measured by a tactile probe on a test sphere
Note 1 to entry: In the context of this part of ISO 10360, the local abbreviation P is used.
Form.Sph.1x25
4 © ISO 2016 – All rights reserved
ISO 10360-12:2016(E)
3.15
probing size error, tactile
P
Size.Sph.1x25::Tact.AArm
error of indication of the diameter of a spherical material standard of size as determined by a Gaussian
(least-squares) fit of 25 points measured by a tactile probe
Note 1 to entry: In the context of this part of ISO 10360, the local abbreviation P is used.
Size.Sph.1x25
3.16
maximum permissible error of articulated location error, tactile
L
Dia.5x5:Art:Tact.AArm,MPE
extreme value of the articulated location error, tactile, L , permitted by specifications
Dia.5x5:Art:Tact.AArm
Note 1 to entry: In the context of this part of ISO 10360, the local abbreviation L is used.
Dia.5x5:Art,MPE
3.17
maximum permissible error of bidirectional length measurement
E
Bi:0:Tact.AArm,MPE
extreme value of the bidirectional length measurement error, E , permitted by specifications
Bi:0:Tact.AArm
Note 1 to entry: In the context of this part of ISO 10360, the local abbreviation E is used.
Bi,MPE
3.18
maximum permissible error of unidirectional length measurement
E
Uni:0:Tact.AArm,MPE
extreme value of the unidirectional length measurement error, E , permitted by
Uni:0:Tact.AArm
specifications
Note 1 to entry: In the context of this part of ISO 10360, the local abbreviation E is used.
Uni,MPE
3.19
maximum permissible error of probing form, tactile
P
Form.Sph.1x25::Tact.AArm,MPE
extreme value of the probing form error for tactile probe, P , permitted by
Form.Sph.1x25::Tact.AArm
specifications
Note 1 to entry: In the context of this part of ISO 10360, the local abbreviation P is used.
Form.Sph.1x25,MPE
3.20
maximum permissible error of probing size, tactile
P
Size.Sph.1x25::Tact.AArm,MPE
extreme value of the probing size error for a tactile probe, P , permitted by
Size.Sph.1x25::Tact.AArm
specifications
Note 1 to entry: In the context of this part of ISO 10360, the local abbreviation P is used.
Size.Sph.1x25,MPE
3.21
rated operating condition
operating condition that must be fulfilled during measurement in order that a measuring instrument or
measuring system performs as designed
Note 1 to entry: Rated operating conditions generally specify intervals of values for a quantity being measured
and for any influence quantity.
[SOURCE: ISO/IEC Guide 99:2007, 4.9]
Note 2 to entry: Within the ISO 10360 series, the term “as designed” means as specified by MPEs.
Note 3 to entry: If an MPE specification is thought of as a function (where different MPE values could be given for
different conditions), then the rated operating conditions define the domain of that function.
ISO 10360-12:2016(E)
4 Symbols
For the purpose of this part of ISO 10360, the symbols of Table 1 apply.
Table 1 — Symbols
Global symbols Local abbreviations Term
L L Articulated location error, tactile
Dia.5x5:Art:Tact.AArm Dia.5x5:Art
E E Length measurement error, unidirectional
Uni:0:Tact.AArm Uni
E E Length measurement error, bidirectional
Bi:0:Tact.AArm Bi
P P Probing form error, tactile
Form.Sph.1x25::Tact.AArm Form.Sph.1x25
P P Probing size error, tactile
Size.Sph.1x25::Tact.AArm Size.Sph.1x25
E E Maximum permissible error of unidirectional length
Uni:0:Tact.AArm,MPE Uni,MPE
measurement
E E Maximum permissible error of bidirectional length
Bi:0:Tact.AArm,MPE Bi,MPE
measurement
P P Maximum permissible error for probing form, tactile
Form.Sph.1x25::Tact. Form.Sph.1x25,MPE
AArm,MPE
P P Maximum permissible error of probing size, tactile
Size.Sph.1x25::Tact. Size.Sph.1x25,MPE
AArm,MPE
L L Maximum permissible error of articulated location
Dia.5x5:Art:Tact.AArm,MPE Dia.5x5:Art,MPE
error, tactile
P P Probing form error, ODS
Form.Sph.1x25::ODS.AArm Form.Sph.1x25::ODS
a
(based on ISO 10360-8)
P P Probing dispersion error
Form.Sph.D95%::ODS.AArm Form.Sph.D95%::ODS
a
(based on ISO 10360-8)
P P Probing size error, ODS
Size.Sph.1x25::ODS.AArm Size.Sph.1x25::ODS
a
(based on ISO 10360-8)
P P Probing size error All
Size.Sph.All::ODS.AArm Size.Sph.All::ODS
a
(based on ISO 10360-8)
P P Maximum permissible error for probing form, ODS
Form.Sph.1x25::ODS. Form.Sph.1x25::ODS,MPE
a
(based on ISO 10360-8)
AArm,MPE
P P Maximum permissible error for probing dispersion
Form.Sph.D95%::ODS. Form.Sph.D95%::ODS,MPE
a
(based on ISO 10360-8)
AArm,MPE
P P Maximum permissible error for probing size, ODS
Size.Sph.1x25::ODS. Size.Sph.1x25::ODS,MPE
a
(based on ISO 10360-8)
AArm,MPE
P P Maximum permissible error for probing size All
Size.Sph.All::ODS.AArm,MPE Size.Sph.All::ODS,MPE
a
(based on ISO 10360-8)
P P Probing flat form error
Form.Pla.D95%::ODS.AArm Form.Pla.D95%::ODS
a
(based on ISO 10360-8)
P P Maximum permissible error for probing flat form error
Form.Pla.D95%::ODS. Form.Pla.D95%::ODS,MPE
a
(based on ISO 10360-8)
AArm,MPE
P P Multiple system probing form error
Form.Sph.1x25::MPS.AArm Form.Sph.1x25::MPS
a
(based on ISO 10360-9)
P P Multiple system probing size error
Size.Sph.1x25::MPS.AArm Size.Sph.1x25::MPS
a
(based on ISO 10360-9)
L L Multiple system probing location error
Dia.1x25::MPS.AArm Dia.1x25::MPS
a
(based on ISO 10360-9)
P P Maximum permissible error for multiple system prob-
Form.Sph.1x25::MPS. Form.Sph.1x25::MPS,MPE
a
ing form error (based on ISO 10360-9)
AArm,MPE
a
The trailing qualifier, “.AArm”, indicates that this metrological characteristic as defined in the relevant ISO 10360 part
is tested with an articulating arm CMM.
b
These symbols relate to metrological characteristics for which specification and testing are optional (not normative).
6 © ISO 2016 – All rights reserved
ISO 10360-12:2016(E)
Table 1 (continued)
Global symbols Local abbreviations Term
P P Maximum permissible error for multiple system prob-
Size.Sph.1x25::MPS. Size.Sph.1x25::MPS,MPE
a
ing size error (based on ISO 10360-9)
AArm,MPE
L L Maximum permissible error for multiple system prob-
Dia.1x25::MPS.AArm,MPE Dia.1x25::MPS,MPE
a
ing location error (based on ISO 10360-9)
R R Repeatability range of the bidirectional length meas-
Uni.0::Tact.AArm Uni.0::Tact
a, b
urement errors
R R Repeatability range of the unidirectional length meas-
Bi.0::Tact.AArm Bi.0::Tact
a, b
urement errors
a, b
P P Probing articulated size error
Size.5x5:Art:Tact.AArm Size.5x5:Art:Tact
a, b
P P Probing articulated form error
Form.5x5:Art:Tact.AArm Form.5x5:Art:Tact
R R Maximum permissible error for the repeatability range
Uni.0::Tact.AArm,MPE Uni.0::Tact,MPE
a, b
of unidirectional length measurement errors
R R Maximum permissible error for the repeatability range
Bi.0::Tact.AArm,MPE Bi.0::Tact,MPE
a, b
of bidirectional length measurement errors
P P Maximum permissible error for probing articulated size
Size.5x5:Art:Tact.AArm,MPE Size.5x5:Art:Tact,MPE
a, b
error
P P Maximum permissible error for probing articulated
Form.5x5:Art:Tact.AArm,MPE Form.5x5:Art:Tact,MPE
a, b
form error
a
The trailing qualifier, “.AArm”, indicates that this metrological characteristic as defined in the relevant ISO 10360 part
is tested with an articulating arm CMM.
b
These symbols relate to metrological characteristics for which specification and testing are optional (not normative).
Local abbreviations are used in this part of ISO 10360 for the sake of simplicity; however, local
abbreviations used in different part(s) of ISO 10360 may not be the same, i.e. the same abbreviation
may refer to different complete symbols. Use of abbreviated symbols is not recommended outside the
exclusive context of this part of ISO 10360.
5 Rated operating conditions
5.1 Environmental conditions
Limits for permissible environmental conditions such as temperature conditions, air pressure, humidity
and vibration at the site of installation that influence the measurements shall be stated by
— the manufacturer, in the case of acceptance tests, and
— the user, in the case of reverification tests.
In both cases, the user is free to choose the environmental conditions under which the testing will be
performed within the stated limits. (Form 1 in Annex A shows an example method for stating these
conditions.) The manufacturer shall provide at a single location in published literature MPE values and
operating conditions under which the MPE values are valid.
If the user wishes to have testing performed under environmental conditions other than the ambient
conditions of the test site (e.g. at an elevated or lowered temperature), agreement between parties
regarding who bears the cost of environmental conditioning should be obtained.
ISO 10360-12:2016(E)
5.2 Operating conditions
The articulated arm CMM shall be operated by an appropriately trained and skilled operator using the
procedures given in the manufacturer’s operating manual when conducting the tests given in Clause 6.
Specific areas in the manufacturer’s manual to be adhered to are, for example,
a) machine start-up/warm-up cycles,
b) machine compensation procedures,
c) location, type and number of environmental sensors,
d) location, type and number of thermal workpiece sensors, and
e) mounting constraints.
6 Acceptance tests and reverification tests
6.1 General
Acceptance tests are executed according to the manufacturer’s specifications and procedures.
Reverification tests are executed according to the user’s specifications and the manufacturer’s
procedures.
Each error obtained in an acceptance test of this part of ISO 10360 shall be compared to the MPE value
or values stated by the manufacturer for the stylus configuration used in the test.
Optionally, a test report may include a plot of the measurement errors and the corresponding MPE values.
NOTE For an articulated arm CMM, it is often the case that a single MPE value is valid throughout the entire
measuring volume.
6.2 Probing size and form errors
6.2.1 Principle
The principle of this test procedure is to measure the size and form of a test sphere by probing 25
points on the surface of the sphere. A Gaussian (least-squares) sphere fit of the 25 points is examined
for the errors of indication for form and size. This analysis yields the form error, P , and the
Form.Sph.1x25
size error, P .
Size.Sph.1x25
6.2.2 Measuring equipment
The material standard of size, i.e. the test sphere, shall have a diameter not less than 10 mm and not
greater than 51 mm.
Note 1 The uncertainty in the calibrated size value of the test sphere and the form error of the test sphere
contribute to the test value uncertainty.
Note 2 A test sphere smaller than 20 mm may be difficult to probe without encountering interference
problems from a sphere support.
6.2.3 Procedure
Set up and qualify the probing system of the articulated arm CMM in accordance with the manufacturer’s
normal procedures. Choose two locations for the test sphere anywhere in the measuring volume
taking into consideration the direction of the stylus during the test (see NOTE 3 of 3.5). By default, one
location is near the vertical main axis of the articulated arm CMM, and the other location is near the
outer surface of the measuring volume.
8 © ISO 2016 – All rights reserved
ISO 10360-12:2016(E)
The test sphere and articulated arm CMM shall be mounted rigidly, both individually and with respect
to one another, to minimize errors due to bending.
Measure and record 25 points without rotating the stylus or changing the direction of the stylus. The
points shall be approximately evenly distributed over at least a hemisphere of the test sphere. Their
positions shall be at the discretion of the user and, if not specified, the following probing pattern is
recommended (see Figure 2):
— one point on the pole (defined by the direction of the stylus shaft) of the test sphere;
— four points (equally spaced) 22,5° below the pole;
— eight points (equally spaced) 45° below the pole and rotated 22,5° relative to the previous group;
— four points (equally spaced) 67,5° below the pole and rotated 22,5° relative to the previous group;
— eight points (equally spaced) 90° below the pole (i.e. on the equator) and rotated 22,5° relative to
the previous group.
Key
A pole
Figure 2 — Location of probing points
ISO 10360-12:2016(E)
6.2.4 Derivation of test results
For each of the first and second locations, using all 25 measured points, compute the Gaussian (least-
squares) associated sphere. Subtract the calibrated diameter, D , from the calculated diameter, D ,
Ref Meas
of the Gaussian (least-squares) associated sphere to get P = D – D . The absolute value
Size.Sph.1x25 Meas Ref
of the probing size error at each location shall be compared to the MPE value or values specified by the
manufacturer.
For each of the first and second locations, using all 25 measured points, compute the Gaussian (least-
squares) associated sphere. For each of the 25 measured points, calculate R, the distance from the
measured point to the Gaussian (least-squares) sphere centre (i.e. the Gaussian radial distance).
Calculate the probing form error by taking the range of Gaussian radial distances: P = R
Form.Sph.1x25 max
– R . The probing form error at each location shall be compared to the MPE value or values specified
min
by the manufacturer.
NOTE The term Gaussian (least-squares) radial distance is defined in ISO 10360-1.
6.3 Articulated location errors
6.3.1 Principle
The principle of 6.3 and 6.4 is to include all of the articulation joints of the articulated arm CMM in
a performance evaluation of this part of ISO 10360. The evaluation described in 6.4 includes all
articulation joints except the wrist joint assembly, which includes the fifth swivel joint, the sixth hinge
joint, and the seventh (if available) swivel joint; the evaluation in 6.3 focuses on the evaluation of this
joint assembly. The test procedure is to probe the surface of a test sphere in a manner that extensively
articulates the wrist by measuring with five orthogonal stylus orientations articulated at the wrist. For
each of the five orientations, the test sphere is probed with five points and a Gaussian (least-squares)
sphere centre location is calculated. The set of the five sphere centres, which ideally should have the
same coordinates, is used to evaluate the articulated location error, L .
Dia.5x5:Art
If the articulated arm CMM is equipped with a swivel joint on a seventh axis, the stylus may be rotated
by an arbitrary angle between each of the five sphere centre measurements.
If the articulated arm CMM is equipped with multiple styli, e.g. “cluster styli”, then the multiple styli
test in ISO 10360-5 should also be performed.
6.3.2 Measuring equipment
The test sphere, calibrated for size and form, shall have a diameter not less than 10 mm and not greater
than 51 mm.
NOTE 1 The uncertainty in the size of the test sphere and form error of the test sphere contribute to the test
value uncertainty.
NOTE 2 A test sphere smaller than 20 mm may be difficult to probe without encountering interference
problems from a sphere support.
6.3.3 Procedure
Set up and qualify the probing system on the articulated arm CMM in accordance with the manufacturer’s
normal procedures. Choose two locations for the test sphere anywhere in the measuring volume taking
into consideration the direction of the stylus during the test (see NOTE 3 of 3.5).
The test sphere and articulated arm CMM shall be mounted rigidly, both individually and with respect
to one another, to minimize errors due to bending.
Measure on the test sphere each of the five pole points as shown in Figure 3 a). In this figure, four of the
five pole points are placed on four quadrants of an equatorial plane of the test sphere, and the fifth pole
is placed on the test sphere so as to be symmetrical with respect to the other four. For each pole point,
10 © ISO 2016 – All rights reserved
ISO 10360-12:2016(E)
m
...