prEN 1518
(Main)Non-destructive testing - Leak testing - Characterization of mass spectrometer leak detectors
Non-destructive testing - Leak testing - Characterization of mass spectrometer leak detectors
This document specifies terms and procedures for the characterization of mass spectrometer leak detectors (MSLD). It is not intended to give a complete set of specifications for an acceptance test but a description of procedures that can be used without particular calibration equipment.
The methods described in this document are applicable without restrictions to helium as the tracer gas. For other gases, additional precautions may be necessary.
These methods are applicable to commonly available MSLD, based on the present level of technology, which may be able to measure leakage rates down to 10−12 Pa⋅m3/s.
Zerstörungsfreie Prüfung - Dichtheitsprüfung - Charakterisierung von massenspektrometrischen Leckdetektoren
Dieses Dokument legt Begriffe und Verfahren für die Charakterisierung von massenspektrometrischen Leckdetektoren (MSLD) fest. Es soll kein kompletter Satz von Spezifikationen für einen Abnahmetest angegeben werden, sondern es werden Vorgehensweisen beschrieben, die ohne spezielle Kalibrierausrüstung angewendet werden können.
Die in diesem Dokument beschriebenen Methoden sind ohne Einschränkungen auf Helium als Prüfgas anwendbar. Für andere Gase können zusätzliche Vorsichtsmaßnahmen notwendig sein.
Die hier beschriebenen Methoden sind auf allgemein erhältliche MSLD des heutigen Standes der Technik anwendbar, die Leckageraten bis herab zu 10−12 Pa·m3/s messen können.
Essais non destructifs - Contrôle d’étanchéité - Caractérisation des détecteurs de fuite à spectrométrie de masse
Le présent document spécifie les méthodes et termes permettant de caractériser les détecteurs de fuites à spectrométrie de masse (DFSM). Il n’a pas pour objet de donner un ensemble complet de spécifications en vue d’un essai de réception, mais une description des procédures qui peuvent être utilisées sans équipement d’étalonnage particulier.
Les méthodes décrites dans le présent document sont applicables sans restriction à l’hélium utilisé comme gaz traceur. Pour d’autres gaz, des précautions supplémentaires peuvent être nécessaires.
Ces méthodes sont applicables aux DFSM couramment disponibles sur le marché, conçus suivant le niveau actuel de la technologie, et qui sont capables de mesurer des flux de fuite de 10−12 Pa⋅m3/s.
Neporušitveno preiskave - Preskus tesnosti – Karakterizacija masno-spektrometričnih detektorjev puščanja
General Information
Relations
Standards Content (Sample)
SLOVENSKI STANDARD
01-januar-2025
Neporušitveno preskušanje - Preskus tesnosti – Karakterizacija masno-
spektrometričnih detektorjev puščanja
Non-destructive testing - Leak testing - Characterization of mass spectrometer leak
detectors
Zerstörungsfreie Prüfung - Dichtheitsprüfung - Charakterisierung von
massenspektrometrischen Leckdetektoren
Essais non destructifs - Contrôle d’étanchéité - Caractérisation des détecteurs de fuite à
spectrométrie de masse
Ta slovenski standard je istoveten z: prEN 1518
ICS:
19.100 Neporušitveno preskušanje Non-destructive testing
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
DRAFT
EUROPEAN STANDARD
NORME EUROPÉENNE
EUROPÄISCHE NORM
November 2024
ICS 19.100 Will supersede EN 1518:1998
English Version
Non-destructive testing - Leak testing - Characterization of
mass spectrometer leak detectors
Essais non destructifs - Contrôle d'étanchéité - Zerstörungsfreie Prüfung - Dichtheitsprüfung -
Caractérisation des détecteurs de fuite à spectrométrie Charakterisierung von massenspektrometrischen
de masse Leckdetektoren
This draft European Standard is submitted to CEN members for enquiry. It has been drawn up by the Technical Committee
CEN/TC 138.
If this draft becomes a European Standard, CEN members are bound to comply with the CEN/CENELEC Internal Regulations
which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
This draft European Standard was established by CEN in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC
Management Centre has the same status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and
United Kingdom.
Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are
aware and to provide supporting documentation.
Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without
notice and shall not be referred to as a European Standard.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2024 CEN All rights of exploitation in any form and by any means reserved Ref. No. prEN 1518:2024 E
worldwide for CEN national Members.
Contents Page
European foreword . 4
1 Scope . 5
2 Normative references . 5
3 Terms and definitions . 5
Terms . 5
3.1.1 Terms related to the MSLD . 5
3.1.2 Terms related to the operation of the MSLD . 6
3.1.3 Terms related to the specification of the MSLD . 7
3.1.4 Terms relating to the tracer gas background signal . 7
3.1.5 Terms relating to the detection limit. 7
3.1.6 Terms relating to the display resolution . 7
4 Description of an MSLD . 8
General. 8
Main parts of an MSLD . 8
Operating principles of MSLD . 9
4.3.1 Direct flow leak detectors . 9
4.3.2 Counterflow leak detectors . 9
MLSD specification . 9
5 Apparatus for methods of characterization . 10
Leaks . 10
5.1.1 General. 10
5.1.2 Small tracer gas leaks . 10
5.1.3 Large air leak . 10
Leak isolation valve. 10
Test vessel . 11
Arrangement for test . 12
5.4.1 Auxiliary pumping system . 12
5.4.2 Materials . 13
Preparation for test . 13
5.5.1 Power supply . 13
5.5.2 Warming up . 13
5.5.3 Adjustment . 13
5.5.4 Constant pumping speed . 13
5.5.5 Constant compression ratio of the high vacuum pump . 13
6 Reference conditions for characterization . 13
Reference temperature . 13
Reference pressure . 13
Reference humidity . 13
7 Test procedures . 14
General. 14
Minimum detectable leakage rate . 14
7.2.1 Background drift and noise . 14
7.2.2 Auxiliary pumping system background . 14
7.2.3 Sensitivity . 14
7.2.4 Time constant . 15
Minimum detectable concentration . 15
7.3.1 General . 15
7.3.2 Background drift and noise . 17
7.3.3 Background signal . 17
7.3.4 Sensitivity . 17
Inlet pumping speed for tracer gas . 18
8 Results . 18
Minimum detectable leakage rate . 18
8.1.1 Evaluation. 18
8.1.2 Expression of minimum detectable leakage rate . 20
Minimum detectable concentration . 20
8.2.1 Evaluation. 20
8.2.2 Expression of minimum detectable concentration . 21
Inlet pumping speed . 21
8.3.1 Evaluation. 21
8.3.2 Expression of inlet pumping speed . 22
9 Test report . 22
Bibliography . 24
European foreword
This document (prEN 1518:2024) has been prepared by Technical Committee CEN/TC 138 “Non-
destructive testing”, the secretariat of which is held by AFNOR.
This document is currently submitted to the CEN Enquiry.
This document will supersede EN 1518:1998.
a) normative references have been updated;
b) clause “Terms and definitions” has been revised;
c) new Subclause 4.1, General, has been added;
d) Clause 5.2 has been updated;
e) in Clause 6.1, the ambient temperature has been adapted;
f) Clause 7 and 8 have been revised;
g) the test report has been revised;
h) the figures have been integrated in the text.
1 Scope
This document specifies terms and procedures for the characterization of mass spectrometer leak
detectors (MSLD). It is not intended to give a complete set of specifications for an acceptance test but a
description of procedures that can be used without particular calibration equipment.
The methods described in this document are applicable without restrictions to helium as the tracer gas.
For other gases, additional precautions may be necessary.
These methods are applicable to commonly available MSLD, based on the present level of technology,
−12 3
which may be able to measure leakage rates down to 10 Pa⋅m /s.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
EN ISO 20484:2017, Non-destructive testing — Leak testing — Vocabulary (ISO 20484:2017)
3 Terms and definitions
For the purposes of this document, the terms and definitions given in EN ISO 20484:2017 and the
following apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
• IEC Electropedia: available at https://www.electropedia.org/
• ISO Online browsing platform: available at https://www.iso.org/obp
Terms
3.1.1 Terms related to the MSLD
3.1.1.1
compression ratio
ratio of partial pressure of tracer gas with zero flow at the exhaust of the counterflow stages of the high
vacuum pump to the partial pressure at the inlet of the mass spectrometer (MS)
3.1.1.2
display
device which indicates visually the leakage rate measured
Note 1 to entry: The units in which the leakage rate is expressed can be selectable.
Note 2 to entry: The display can be analogue or quasi-analogue (continuous scale with definite divisions) or digital
(numbers with a definite number of digits) or a combination of both.
3.1.1.3
inlet pumping speed
volume rate of flow at the test port of an MSLD when the instrument is operating with the MS below its
maximum working pressure
Note 1 to entry: The inlet pumping speed can be different for different gases and different modes of operation. The
inlet pumping speed for the tracer gas determines the response time for the volume under test.
3.1.1.4
internal leak port
flange directly behind the inlet valve, used to connect a small leak for zero drift determination
3.1.1.5
intrinsic pumping speed
volume rate of flow of tracer gas at the inlet of the MS in a direct flow leak detector
3.1.1.6
ion collector
part of the MS where ions are collected and neutralized producing a current in the collector which is a
measure of the number of neutralized ions
3.1.1.7
sensitivity control
electrical hardware or software control which may be used to adjust the sensitivity of the instrument so
that a calibrated leak is indicated with its true leakage rate
3.1.1.8
sensitivity
ratio of the ion current at the output of the MS to the corresponding partial pressure of tracer gas inside
the MS
3.1.1.9
zero control
electrical hardware or software control which may be used to shift the output indication of the leak
detector, to a determined point of the scale range in use, usually zero
3.1.2 Terms related to the operation of the MSLD
3.1.2.1
peak
trace showing a maximum when the leak detector is scanned with respect to mass with gas
present, usually the tracer gas, to which the detector is tuned
3.1.2.2
peak
to set the scanning control (see 3.1.2.3) of a leak detector so that the output due to a given tracer
gas input is maximized
Note 1 to entry: It is a form of tuning (see 3.1.2.4).
3.1.2.3
scan
to adjust the accelerating voltage (or other equivalent operating parameter) of a leak detector,
particularly across that range of voltage which includes the voltage necessary to produce a tracer gas
peak
3.1.2.4
tune
in leak detection, to adjust one or more of the controls of a leak detector so that its response to a
tracer gas is maximized
Note 1 to entry: Tuning by means of the scanning control only is called “peaking” (see 3.1.2.2).
3.1.3 Terms related to the specification of the MSLD
3.1.3.1
optimum working pressure
pressure in the MS, at which the minimum detectable concentration can be measured
3.1.3.2
maximum working pressure
pressure in the MS above which operation is no longer possible
3.1.3.3
maximum inlet pressure
maximum pressure at the test port at which the MSLD is able to detect leaks in a given mode of operation
Note 1 to entry: For an MSLD to be connected directly to a system under test, the total pressure in the system shall
be less than the maximum inlet pressure of the MSLD.
3.1.3.4
maximum gas load
maximum pV-throughput of all gases emerging from the test specimen that the MSLD can pump during
leak detection in a given mode of operation
Note 1 to entry: For component testing, the MSLD is normally ready for measurement when the desorption of
water vapour from the inner surfaces is less than the maximum gas load.
3.1.4 Terms relating to the tracer gas background signal
3.1.4.1
background signal drift
relatively slow change in background signal, given by the maximum change in a given period of time
3.1.4.2
background signal noise
relatively rapid change in background signal given by an average measure of scatter in a specified period
of time
3.1.5 Terms relating to the detection limit
3.1.5.1
minimum detectable concentration ratio
smallest concentration of a given tracer gas in an air mixture that can be detected unambiguously when
the mixture is fed into the MSLD at such a rate as that is at its optimum working pressure
3.1.5.2
resolving power
ratio of a given mass number to the peak width measured at a specified (for example 10 %) height of the
peak (in units of mass numbers)
3.1.6 Terms relating to the display resolution
3.1.6.1
display resolution
quantitative expression of the ability of the leakage rate-display device to distinguish meaningfully
between closely adjacent values of the leakage rate indicated
3.1.6.2
linear display resolution
constant difference between adjacent scale intervals expressed in % full scale indication
3.1.6.3
logarithmic display resolution
constant ratio between two adjacent scale intervals expressed as a percentage of the indicated value
4 Description of an MSLD
General
An MSLD has an integral high vacuum system for maintaining the sensing element (mass spectrometer)
at low operating pressure and for establishing a partial pressure related to the incoming gas flow. This
pressure is measured quantitatively by the mass spectrometer. Such instruments are able to selectively
measure the flow of a tracer gas. In most cases the tracer gas will be helium, flowing in/out of an object
through a leak.
MSLD are commonly operated in either direct or counterflow configurations. These are both illustrated
in Figure 2.
Main parts of an MSLD
An MSLD (see definition given in EN ISO 20484) consists basically of a mass spectrometer (MS) and a
high vacuum pumping system for:
— maintaining the MS under appropriate vacuum conditions;
— producing a definite partial pressure of tracer gas when a specific throughput of tracer gas enters the
leak detector.
An MSLD includes also a number of valves and pressure gauges to ensure the appropriate vacuum
conditions within the system.
The leakage rate output can be displayed in a number of ways, for example an electrical meter, or digital
displays of different types, such as with a logarithmic or linear display resolution. In addition, a digital
and/or analogue output is usually available, which shall be used for the test procedures in this document.
The general structure of an MSLD is described by the following list:
a) mass spectrometer:
— ion source;
— separation system;
— ion collector.
b) pumping system:
— inlet system;
— inlet line;
— inlet valve;
— pump valve;
— test port;
— vent valve;
— internal leak port.
c) sensitivity zero control.
d) display.
Operating principles of MSLD
4.3.1 Direct flow leak detectors
The leakage rate is given by the formula:
q
x
(1)
q pS⋅
x xx
where
is the pV-throughput of tracer gas;
q
x
is the intrinsic pumping speed;
S
x
is the partial pressure of tracer
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...