Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood - Amendment 1 (ISO 10993-4:2017/DAmd 1:2024)

Biologische Beurteilung von Medizinprodukten - Teil 4: Auswahl von Prüfungen zur Wechselwirkung mit Blut - Änderung 1 (ISO 10993 4:2017/DAM 1:2024)

Évaluation biologique des dispositifs médicaux - Partie 4: Choix des essais pour les interactions avec le sang - Amendement 1 (ISO 10993-4:2017/DAmd 1:2024)

Biološko ovrednotenje medicinskih pripomočkov - 4. del: Izbira preskusov za ugotavljanje interakcij s krvjo - Dopolnilo A1 (ISO 10993 4:2017/DAmd 1:2024)

General Information

Status
Not Published
Publication Date
09-Jun-2025
Current Stage
4599 - Dispatch of FV draft to CMC - Finalization for Vote
Start Date
07-Oct-2024
Completion Date
07-Oct-2024

Relations

Buy Standard

Draft
EN ISO 10993-4:2017/oprA1:2024
English language
11 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI STANDARD
01-maj-2024
Biološko ovrednotenje medicinskih pripomočkov - 4. del: Izbira preskusov za
ugotavljanje interakcij s krvjo - Dopolnilo A1 (ISO 10993 4:2017/DAmd 1:2024)
Biological evaluation of medical devices - Part 4: Selection of tests for interactions with
blood - Amendment 1 (ISO 10993-4:2017/DAmd 1:2024)
Biologische Beurteilung von Medizinprodukten - Teil 4: Auswahl von Prüfungen zur
Wechselwirkung mit Blut - Änderung 1 (ISO 10993 4:2017/DAM 1:2024)
Évaluation biologique des dispositifs médicaux - Partie 4: Choix des essais pour les
interactions avec le sang - Amendement 1 (ISO 10993-4:2017/DAmd 1:2024)
Ta slovenski standard je istoveten z: EN ISO 10993-4:2017/prA1
ICS:
11.100.20 Biološko ovrednotenje Biological evaluation of
medicinskih pripomočkov medical devices
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

DRAFT
Amendment
ISO 10993-4:2017/
DAM 1
ISO/TC 194
Biological evaluation of medical
Secretariat: DIN
devices —
Voting begins on:
Part 4:
2024-03-07
Selection of tests for interactions
Voting terminates on:
with blood 2024-05-30
AMENDMENT 1
Évaluation biologique des dispositifs médicaux —
Partie 4: Choix des essais pour les interactions avec le sang
AMENDEMENT 1
ICS: 11.100.20
THIS DOCUMENT IS A DRAFT CIRCULATED
FOR COMMENTS AND APPROVAL. IT
IS THEREFORE SUBJECT TO CHANGE
AND MAY NOT BE REFERRED TO AS AN
INTERNATIONAL STANDARD UNTIL
PUBLISHED AS SUCH.
This document is circulated as received from the committee secretariat.
IN ADDITION TO THEIR EVALUATION AS
BEING ACCEPTABLE FOR INDUSTRIAL,
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
STANDARDS MAY ON OCCASION HAVE TO
ISO/CEN PARALLEL PROCESSING
BE CONSIDERED IN THE LIGHT OF THEIR
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
NATIONAL REGULATIONS.
RECIPIENTS OF THIS DRAFT ARE INVITED
TO SUBMIT, WITH THEIR COMMENTS,
NOTIFICATION OF ANY RELEVANT PATENT
RIGHTS OF WHICH THEY ARE AWARE AND TO
PROVIDE SUPPORTING DOCUMENTATION.
Reference number
ISO 10993-4:2017/DAM 1:2024(en)
DRAFT
ISO 10993-4:2017/DAM 1:2024(en)
Amendment
ISO 10993-4:2017/
DAM 1
ISO/TC 194
Biological evaluation of medical
Secretariat: DIN
devices —
Voting begins on:
Part 4:
Selection of tests for interactions
Voting terminates on:
with blood
AMENDMENT 1
Évaluation biologique des dispositifs médicaux —
Partie 4: Choix des essais pour les interactions avec le sang
AMENDEMENT 1
ICS: 11.100.20
THIS DOCUMENT IS A DRAFT CIRCULATED
FOR COMMENTS AND APPROVAL. IT
IS THEREFORE SUBJECT TO CHANGE
AND MAY NOT BE REFERRED TO AS AN
INTERNATIONAL STANDARD UNTIL
PUBLISHED AS SUCH.
This document is circulated as received from the committee secretariat.
IN ADDITION TO THEIR EVALUATION AS
BEING ACCEPTABLE FOR INDUSTRIAL,
© ISO 2024
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
STANDARDS MAY ON OCCASION HAVE TO
ISO/CEN PARALLEL PROCESSING
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
BE CONSIDERED IN THE LIGHT OF THEIR
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
or ISO’s member body in the country of the requester.
NATIONAL REGULATIONS.
ISO copyright office
RECIPIENTS OF THIS DRAFT ARE INVITED
CP 401 • Ch. de Blandonnet 8
TO SUBMIT, WITH THEIR COMMENTS,
CH-1214 Vernier, Geneva
NOTIFICATION OF ANY RELEVANT PATENT
Phone: +41 22 749 01 11
RIGHTS OF WHICH THEY ARE AWARE AND TO
PROVIDE SUPPORTING DOCUMENTATION.
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland Reference number
ISO 10993-4:2017/DAM 1:2024(en)
ii
ISO 10993-4:2017/DAM 1:2024(en)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO documents should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent
rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a)
patent(s) which may be required to implement this document. However, implementers are cautioned that
this may not represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO's adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 194, Biological and clinical evaluation of
medical devices.
This amendment to ISO 10993-4:2017 refines certain language, clarifies examples of devices with indirect
blood contact and direct blood contact, and adds additional current references.
A list of all parts in the ISO 10993 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.

iii
ISO 10993-4:2017/DAM 1:2024(en)
Biological evaluation of medical devices —
Part 4:
Selection of tests for interactions with blood
AMENDMENT 1
5.2.2
Replace the heading with the following:
5.2.2  External communicating devices that have indirect blood contact

Delete the following bullet points “cannulae”, “cell savers” and “intravascular catheters”.

Replace “blood and blood product” with “saline and/or therapeutics” to read:
— devices for the storage and administration of saline and/or therapeutics (e.g. tubing and bags);

Add the bullet point “- blood monitors with indirect blood contact.” at the end of the list.

5.2.3
Add “blood collection devices” as a second bullet point, “cannulae” as a fourth bullet point, “cell savers” as a
sixth bullet point and “devices for the storage and administration of blood and blood products (e.g. tubing
and bags)” as a eighth bullet point.

Delete “or indirect” in the third bullet point to read:
— blood monitoring devices with direct blood contact;

6.1.2
Add to the second paragraph after the sentence “Only direct or indirect blood-contacting parts should be
tested.” the following sentences:
For direct contact haemocompatibility testing (e.g., direct haemolysis, complement activation,
coagulation, platelet activation, haematology, in vitro/ex vivo thrombosis), testing should be conducted
using only the direct blood contacting components of the device to minimize interference of non-
direct blood contacting components on the results. For extract-based haemocompatibility testing (e.g.,
indirect haemolysis), testing should be conducted using only the direct and indirect blood-contacting

ISO 10993-4:2017/DAM 1:2024(en)
components of the device. The test article shall be described, and a justification shall be provided if the
test article includes device components that include different tissue contact than described above.

Remove the fourth paragraph and replace with:
As many tests for haemocompatibility are recognized to be generally surface-contact dependent (e.g.,
direct contact haemolysis, complement activation, coagulation, platelet activation, haematology, and
in vitro/ex vivo thrombosis) such tests will not apply to indirect contact applications. For externally
communicating medical devices or components that have indirect blood contact, generally only an
indirect contact hemolysis test is recommended.

6.1.4
Change NOTE 1 to read:
NOTE 1 Changes in manufacturing process e.g., change in manufacturer, use of different manufacturing

aids that could affect the surface properties, or chemistry of the complete sterilized device, could also impact
haemocompatibility.
6.1.6
Change Table 1 to read:
ISO 10993-4:2017/DAM 1:2024(en)
Table 1 — Devices or device components and categories of appropriate testing for consideration
Test category
c
Haemolysis Thrombosis
in vitro
In
Device examples
Me-
Plate-
vivo/
Materi- chani- Coag-
let Comple- Haema-
Ex
al- cally- ula-
d
activa- ment tology
a
vivo
induced in- tion
tion
duced
e
External communicating devices indirectly contacting blood
Blood monitors with indirect blood contact X
Devices for storage and administration of saline
and/or therapeutics (e.g. tubing and bags), exten- X
sion sets
b
External communicating devices directly contacting circulating blood
Blood collection devices X X X X
Blood administration sets and extension sets X X X X X
Catheters in place for less than 24 h (e.g. atherecto-
my devices,
intravascular ultrasound catheters, antegrade/ X X X X X
retrograde coronary
perfusion catheters, guide wires); cannulae
Catheters in place for more than 24 h (e.g. paren-
teral nutrition catheters, central venous cathe- X X X X X
ters); cannulae
Cell savers X X X
Devices for adsorption of specific substances from
X X X X X
blood
Donor and therapeutic aphaeresis equipment and
X X X X X
cell separation systems
Cardiopulmonary bypass system X X X X X X X
Haemodialysis/haemofiltration equipment X X X X X X X
Leukocyte removal filter X X X X X X
Percutaneous circulatory support devices X X X X X X X
Implant devices
Annuloplasty rings, mechanical heart valves X X X
Embolization devices X X
Endovascular grafts X X
Implantable defibrillator and cardioverter leads X X
a
Thrombosis is an in vivo or ex vivo phenomenon but can be simulated with in vitro conditions. In vivo or ex vivo testing might
not be necessary if clinically relevant in vitro thrombosis testing is performed.
b
Some examples here may contain other components with indirect blood contact. For device components that only have
indirect blood contact, direct contact material-induced haemolysis, mechanical haemolysis, thrombosis, and complement
activation may not be necessary. For externally communicating medical devices with indirect blood contact, generally only an
indirect contact hemolysis test is recommended.
c
It is recognized that coagulation, platelet, and leucocyte responses are primarily involved in the process of thrombosis.
Therefore, in vitro thrombogenicity methods can be acceptable in place of in vivo testing if scientifically justified. The
manufacturer should justify which specific testing in the coagulation, platelet and haematology test categories is appropriate for
their devices.
d
Complement activation testing is also requested by certain regulatory authorities to address other end points such as
anaphylaxis for all devices with direct blood contact.
e
Except for devices composed of novel materials, thrombogenicity testing is generally not necessary for indirect blood contact
devices composed of commonly used materials.

ISO 10993-4:2017/DAM 1:2024(en)
TTabablele 1 1 ((ccoonnttiinnueuedd))
Test category
c
Haemolysis Thrombosis
in vitro
In
Device examples
Me-
Plate-
vivo/
Materi- chani- Coag-
let Comple- Haema-
Ex
al- cally- ula-
d
activa- ment tology
a
vivo
induced in- tion
tion
duced
Intra-aortic balloon pumps X X X
Pacemaker leads X X
Prosthetic (synthetic) vascular grafts and patches,
including X X
arteriovenous shunts
Stents (vascular) X X
Tissue heart valves, vascular grafts and patches
X X
and AV shunts
Total artificial hearts X X X
Vena cava filters X X
Ventricular-assist devices X X X
a
Thrombosis is an in vivo or ex vivo phenomenon but can be simulated with in vitro conditions. In vivo or ex vivo testing might
not be necessary if clinically relevant in vitro thrombosis testing is performed.
b
Some examples here may contain other components with indirect blood contact. For device components that only have
indirect blood contact, direct contact material-induced haemolysis, mechanical haemolysis, thrombosis, and complement
activation may not be necessary. For externally communicating medical devices with indirect blood contact, generally only an
indirect contact hemolysis test is recommended.
c
It is recognized that coagulation, platelet, and leucocyte responses are primarily involved in the process of thrombosis.
Therefore, in vitro thrombogenicity methods can be acceptable in place of in vivo testing if scientifically justified. The
manufacturer should justify which specific testing in the coagulation, platelet and haematolo
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.