Standard Test Methods for Nonvolatile Matter in Halogenated Organic<br> Solvents and Their Admixtures

SIGNIFICANCE AND USE
Nonvolatile matter in solvents can adversely affect their cleaning properties. These test methods can be used to control soil contamination in the boiling solvent, which if allowed to become too high, can decrease the stability of the solvent.
These test methods can be used to establish manufacturing and purchasing specifications.
SCOPE
1.1 These test methods cover the determination of nonvolatile matter in halogenated organic solvents and admixtures.
1.2 Five test methods are covered, as follows:
1.2.1 Test Method A—For halogenated organic solvents or admixtures having less than 50 ppm nonvolatile matter; or where precision better than ±10 ppm is required.
1.2.2 Test Method B—For halogenated organic solvents or admixtures having more than 50 ppm nonvolatile matter or where precision of ±0.001 % (10 ppm) is satisfactory.
1.2.3 Test Method C—For low-boiling halogenated organic solvents or their admixtures (for example, methylene chloride, trichlorotrifluoroethane) that may superheat and cause bumping while evaporating to dryness with steam. A precision of greater than± 10 ppm can be attained.
1.2.4 Test Method D—For rapid measurement of nonvolatile matter in halogenated organic solvents and their admixtures and where precision better than ±10 ppm is required.
1.2.5 Test Method E—For halogenated organic solvents or admixtures and where precision better than ±10 ppm is required.
1.3 The values stated in SI units are to be regarded as the standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
31-May-2006
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM D2109-01(2006)e1 - Standard Test Methods for Nonvolatile Matter in Halogenated Organic<br> Solvents and Their Admixtures
English language
4 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM D2109-01(2006)e1 - Standard Test Methods for Nonvolatile Matter in Halogenated Organic<br> Solvents and Their Admixtures
English language
4 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
´1
Designation:D2109–01(Reapproved2006)
Standard Test Methods for
Nonvolatile Matter in Halogenated Organic
Solvents and Their Admixtures
This standard is issued under the fixed designation D2109; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the Department of Defense.
´ NOTE—Removed research report references in Sections 7, 10, and 13 editorially in March 2008.
1. Scope 2.1.1 The term nonvolatile matter should not be construed
as equivalent to residue on ignition, ignition residue, or ash
1.1 These test methods cover the determination of nonvola-
content. Particulates, sediments, and suspended matter should
tile matter in halogenated organic solvents and admixtures.
not be considered part of nonvolatile matter. If these solids are
1.2 Five test methods are covered, as follows:
present in the sample, they should be removed by filtration or
1.2.1 Test Method A—For halogenated organic solvents or
decantation prior to beginning this test method. Nonvolatile
admixtures having less than 50 ppm nonvolatile matter; or
matter is considered to be “in solution” with the solvent and
where precision better than 610 ppm is required.
that which will become residual upon drying the solvent at a
1.2.2 Test Method B—For halogenated organic solvents or
specified temperature.
admixtures having more than 50 ppm nonvolatile matter or
2.1.2 Nonvolatile matter and nonvolatile residue are inter-
where precision of 60.001 % (10 ppm) is satisfactory.
changeable terms.
1.2.3 Test Method C—For low-boiling halogenated organic
solvents or their admixtures (for example, methylene chloride,
3. Significance and Use
trichlorotrifluoroethane) that may superheat and cause bump-
3.1 Nonvolatile matter in solvents can adversely affect their
ing while evaporating to dryness with steam. A precision of
cleaning properties. These test methods can be used to control
greater than6 10 ppm can be attained.
soil contamination in the boiling solvent, which if allowed to
1.2.4 Test Method D—For rapid measurement of nonvola-
become too high, can decrease the stability of the solvent.
tilematterinhalogenatedorganicsolventsandtheiradmixtures
3.2 These test methods can be used to establish manufac-
and where precision better than 610 ppm is required.
turing and purchasing specifications.
1.2.5 Test Method E—For halogenated organic solvents or
admixtures and where precision better than 610 ppm is
4. Apparatus
required.
4.1 Oven, thermostatically controlled at 105 6 5°C.
1.3 The values stated in SI units are to be regarded as the
4.2 Evaporating Dish, 125-mL capacity, platinum or high-
standard.
silica glass, Methods A, B, C.
1.4 This standard does not purport to address all of the
4.3 Evaporating Dish (80 3 45 or 115 3 50) (Method D).
safety concerns, if any, associated with its use. It is the
4.4 Steam Bath (or hot plate).
responsibility of the user of this standard to establish appro-
4.5 Hot Plate, (Method D).
priate safety and health practices and determine the applica-
4.6 Heat Lamp, 250 W, (Method E).
bility of regulatory limitations prior to use.
4.7 Analytical Balance, capable of measuring to 0.0001 g.
2. Terminology 4.8 Top Loading Balance, capable of weighing to 0.01 g.
4.9 Aluminum Weighing Dish,57 3 18 mm, (Method D).
2.1 Definitions of Terms Specific to This Standard:
4.10 Aluminum Weighing Dish, 200 ML capacity, (Method
E).
These test methods are under the jurisdiction of ASTM Committee D26 on
4.11 1000-mL Volumetric Flask (Test Method A).
Halogenated Organic Solvents and Fire Extinguishing Agents and are the direct
4.12 100-mL Volumetric Pipet (Test Method B and E).
responsibility of Subcommittee D26.04 on Test Methods.
4.13 1000-mL Graduated Cylinder (Test Method C).
Current edition approved June 1, 2006. Published June 2006. Originally
approved in 1962 as D2109 – 62 T. Last previous edition approved in 2001 as
4.14 1500-mL Erlenmeyer Flask (Test Method C).
D2109 – 01. DOI: 10.1520/D2109-01R06E01.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
´1
D2109–01 (2006)
TEST METHOD A 8.3 Place the dish and contents in an oven at 105 6 5°C for
approximately 1 h. Cool in a desiccator and weigh the dish and
5. Procedure
contents.
5.1 Dry a 125-mL capacity platinum (or high-silica glass)
9. Calculation
evaporating dish in an oven at 105 6 5°C and cool in a
desiccator. Repeat until the weight is constant or within 0.1 mg
9.1 Calculate the nonvolatile matter and report in weight
of the last weighing. Rinse a clean dry 1000-mL volumetric
percent as follows:
flask with the solvent and fill to the 1000-mL mark with the
~A! ~100! ~A!
solvent to be tested. Invert the evaporating dish, place it over Nonvolatile matter, ppm by weight 5 5 (2)
B 100 B
~ ! ~ ! ~ !
the mouth of the flask, hold it firmly in place, and invert the
flask. In this position place both dish and flask on a steam bath. where:
Adjust a ring support to hold the flask so the mouth of the flask A = grams of residue, and
B = density of sample.
is approximately 25 mm above the bottom of the evaporating
dish. Thus held, the flask automatically feeds the solvent to the
10. Precision and Bias
dish during the evaporation. (Warning—This test method
must be run in a ventilated, dust-free area.)
10.1 Precision—As the only significant difference between
5.2 Evaporate the 1000-mL sample to dryness. Remove the
Test Method B and Test Methods A and C is the sample size
dish from the steam bath with metal tongs and blot the outside
(100 mL versus 1000 mL), it is reasonable to assume that the
of the dish with lint-free paper tissue.
precision is roughly ten times that of Test MethodsAand C or
2 ppm for repeatability and 10 ppm for reproducibility.
NOTE 1—Hot plates develop high temperatures on the plate surface. If
10.2 The bias of this test method has not been determined.
a hot plate is used to evaporate the solvent, the evaporating dish should be
placed inside a water bath while on the hot plate to prevent the sample
TEST METHOD C
from reaching temperatures exceeding 105°C.
5.3 Place the dish and contents in an oven at 105 6 5°C for
11. Procedure
approximately 1 h. Cool in a desiccator and weigh the dish and
11.1 Dry a 125-mL capacity platinum (or high-silica glass)
contents.
evaporating dish in an oven at 105°C and cool in a desiccator.
6. Calculation
Repeat until the weight is constant or within 0.1 mg of the last
weighing. Rinse a clean dry 1000-mLgraduated cylinder and a
6.1 Calculate the nonvolatile matter and report in weight
1500-mL Erlenmeyer flask with the solvent to be tested.
percent or parts per million as follows:
11.2 Measure 1000 mL of the sample into the cleaned
~A! ~10 ! ~A! ~1000!
graduate and transfer 970 mL of it into the clean Erlenmeyer
Nonvolatile matter, ppm by weight 5 5 (1)
~B! ~1000! ~B!
flask. Evaporate to about 40 mL on a steam bath or warm hot
plate in a fume hood. Do not evaporate to dryness.
where:
11.3 Transfer the residue to the evaporating dish. Rinse the
A = grams of residue, and
flask twice with a 15-mL portion of the sample retained in the
B =
...


This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
e1
An American National Standard
Designation:D2109–01(Reapproved2006)
Designation:D2109–96 (Reapproved 2000)
Standard Test Methods for
Nonvolatile Matter in Halogenated Organic
Solvents and Their Admixtures
This standard is issued under the fixed designation D 2109; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the Department of Defense.
e NOTE—A research report footnote was added in November 2001.
—Removed research report references in Sections 7, 10, and 13 editorially in March 2008.
1. Scope
1.1 These test methods cover the determination of nonvolatile matter in halogenated organic solvents and admixtures.
1.2Three1.2 Five test methods are covered, as follows:
1.2.1 Test Method A— For halogenated organic solvents or admixtures having less than 50 ppm nonvolatile matter; or where
precision greaterbetter than 610 ppm is required.
1.2.2 Test Method B— For halogenated organic solvents or admixtures having more than 50 ppm nonvolatile matter or where
precision of 60.001 % (10 ppm) is satisfactory.
1.2.3 Test Method C—For low-boiling halogenated organic solvents or their admixtures (for example, methylene chloride,
trichlorotrifluoroethane) that may superheat and cause bumping while evaporating to dryness with steam. A precision of greater
than 610 ppm can be attained. — For low-boiling halogenated organic solvents or their admixtures (for example, methylene
chloride, trichlorotrifluoroethane) that may superheat and cause bumping while evaporating to dryness with steam.Aprecision of
greater than6 10 ppm can be attained.
1.2.4 Test Method D— For rapid measurement of nonvolatile matter in halogenated organic solvents and their admixtures and
where precision better than 610 ppm is required.
1.2.5 Test Method E— For halogenated organic solvents or admixtures and where precision better than 610 ppm is required.
1.3 The values stated in SI units are to be regarded as the standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Terminology
2.1 Definitions of Terms Specific to This Standard:
2.1.1 The term nonvolatile matter should not be construed as equivalent to residue on ignition, ignition residue, or ash content.
Particulates, sediments, and suspended matter should not be considered part of nonvolatile residue.matter. If these solids are
present in the sample, they should be removed by filtration or decantation prior to beginning this test method. Nonvolatile matter
is considered to be “in solution” with the solvent and that which will become residual upon drying the solvent at a specified
temperature.
2.1.2 Nonvolatile matter and nonvolatile residue are interchangeable terms.
3. Significance and Use
3.1 Nonvolatile matter in solvents can adversely affect their cleaning properties. These test methods can be used to control soil
contamination in the boiling solvent, which if allowed to become too high, can decrease the stability of the solvent.
3.2 These test methods can be used to establish manufacturing and purchasing specifications.
4. Apparatus
4.1 Oven, thermostatically controlled at 105 6 5°C.
These test methods are under the jurisdiction ofASTM Committee D26 on Halogenated Organic Solvents and Fire ExtinguishingAgents and are the direct responsibility
of Subcommittee D26.04 on Test Methods.
Current edition approved June 10, 1996. Published August 1996. Originally published as D2109 – 62 T. Last previous edition D2109–92.
Current edition approved June 1, 2006. Published June 2006. Originally approved in 1962 as D 2109 – 62 T. Last previous edition approved in 2001 as D 2109 – 01.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
e1
D2109–01 (2006)
D2109–96 (2000)
4.2 Evaporating Dish,125-mLcapacity,platinumorhigh-silicaglass.,125-mLcapacity,platinumorhigh-silicaglass,Methods
A, B, C.
4.3 Evaporating Dish (80 3 45 or 115 3 50) (Method D).
4.4 Steam Bath (or hot plate).
4.4Analytical Balance.
4.5 Hot Plate, (Method D).
4.6 Heat Lamp, 250 W, (Method E).
4.7 Analytical Balance, capable of measuring to 0.0001 g.
4.8 Top Loading Balance, capable of weighing to 0.01 g.
4.9 Aluminum Weighing Dish,57 3 18 mm, (Method D).
4.10 Aluminum Weighing Dish, 200 ML capacity, (Method E).
4.11 1000-mL Volumetric Flask (Test Method A).
4.6
4.12 100-mL Volumetric Pipet(Test Method B).
4.71000-mL Graduated Cylinder (Test Method B and E).
4.13 1000-mL Graduated Cylinder (Test Method C).
4.8
4.14 1500-mL Erlenmeyer Flask (Test Method C).
TEST METHOD A
5. Procedure
5.1 Dry a 125-mL capacity platinum (or high-silica glass) evaporating dish in an oven at 105 6 5°C and cool in a desiccator.
Repeat until the weight is constant or within 0.1 mg of the last weighing. Rinse a clean dry 1000-mL volumetric flask with the
solvent and fill to the 1000-mLmark with the solvent to be tested. Invert the evaporating dish, place it over the mouth of the flask,
hold it firmly in place, and invert the flask. In this position place both dish and flask on a steam bath.Adjust a ring support to hold
the flask so the mouth of the flask is approximately 25 mm above the bottom of the evaporating dish. Thus held, the flask
automatically feeds the solvent to the dish during the evaporation. ( Warning—This test method must be run in a ventilated,
dust-free area.)
5.2 Evaporate the 1000-mL sample to dryness. Remove the dish from the steam bath with metal tongs and blot the outside of
the dish with lint-free paper tissue.
NOTE1—Caution:This test method must be run in a ventilated, dust-free area.
5.2Evaporate the 1000-mL sample to dryness. Remove the dish from the steam bath with metal tongs and blot the outside of
the dish with lint-free paper tissue.
NOTE2—Hot plates develop high temperatures on the plate surface. If a hot plate is used to evaporate the solvent, the evaporating dish should be placed
inside a water bath while on the hot plate to prevent the sample from reaching temperatures exceeding 105°C. 1—Hot plates develop high temperatures
on the plate surface. If a hot plate is used to evaporate the solvent, the evaporating dish should be placed inside a water bath while on the hot plate to
prevent the sample from reaching temperatures exceeding 105°C.
5.3 Place the dish and contents in an oven at 105 6 5°C for approximately 1 h. Cool in a desiccator and weigh the dish and
contents.
6. Calculation
6.1 Calculate the nonvolatile matter and report in weight percent or parts per million as follows:
~A! ~10 ! ~A! ~1000!
Nonvolatile matter, ppm by weight 5 5 (1)
~B! ~1000! ~B!
where:
A = grams of residue, and
B = density of sample.
7. Precision and Bias
7.1 Repeatability (Single Analyst)—The standard deviation of results has been estimated to be 0.2 weight ppm. Two such
values should be considered suspect (95 % confidence level) if they differ by more than 0.7 ppm.
7.2 Reproducibility (Multilaboratory)—The standard deviation of results has been estimated to be 1.0 weight ppm. Two such
values should be considered suspect (95 % confidence level) if they differ by more than 4.6 weight ppm.
TEST METHOD B
8. Procedure
8.1 Dry a 125-mL capacity platinum (or high-silica glass) evaporating dish in an oven at 105 6 5°C and cool in a desiccator.
e1
D2109–01 (2006)
D2109–96 (2000)
Repeat until the weight is constant or with
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.