ASTM D7783-21
(Practice)Standard Practice for Within-laboratory Quantitation Estimation (WQE)
Standard Practice for Within-laboratory Quantitation Estimation (WQE)
SIGNIFICANCE AND USE
5.1 Appropriate application of this practice should result in a WQE achievable by the laboratory in applying the tested method/matrix/analyte combination to routine sample analysis. That is, a laboratory should be capable of measuring concentrations greater than WQEZ %, with the associated RSD equal to Z % or less.
5.2 The WQE values may be used to compare the quantitation capability of different methods for analysis of the same analyte in the same matrix within the same laboratory.
5.3 The WQE procedure should be used to establish the within-laboratory quantitation capability for any application of a method in the laboratory where quantitation is important to data use. The intent of the WQE is not to impose reporting limits. The intent is to provide a reliable procedure for establishing the quantitative characteristics of the method (as implemented in the laboratory for the matrix and analyte) and thus to provide the laboratory with reliable information characterizing the uncertainty in any data produced. Then the laboratory can make informed decisions about censoring data and has the information necessary for providing reliable estimates of uncertainty with reported data.
SCOPE
1.1 This practice establishes a uniform standard for computing the within-laboratory quantitation estimate associated with Z % relative standard deviation (referred to herein as WQEZ %), and provides guidance concerning the appropriate use and application.
1.2 WQEZ % is computed to be the lowest concentration for which a single measurement from the laboratory will have an estimated Z % relative standard deviation (Z % RSD, based on within-laboratory standard deviation), where Z is typically an integer multiple of 10, such as 10, 20, or 30. Z can be less than 10 but not more than 30. The WQE10 % is consistent with the quantitation approaches of Currie (1)2 and Oppenheimer, et al. (2).
1.3 The fundamental assumption of the WQE is that the media tested, the concentrations tested, and the protocol followed in developing the study data provide a representative and fair evaluation of the scope and applicability of the test method, as written. Properly applied, the WQE procedure ensures that the WQE value has the following properties:
1.3.1 Routinely Achievable WQE Value—The laboratory should be able to attain the WQE in routine analyses, using the laboratory’s standard measurement system(s), at reasonable cost. This property is needed for a quantitation limit to be feasible in practical situations. Representative data must be used in the calculation of the WQE.
1.3.2 Accounting for Routine Sources of Error—The WQE should realistically include sources of bias and variation that are common to the measurement process and the measured materials. These sources include, but are not limited to intrinsic instrument noise, some typical amount of carryover error, bottling, preservation, sample handling and storage, analysts, sample preparation, instruments, and matrix.
1.3.3 Avoidable Sources of Error Excluded—The WQE should realistically exclude avoidable sources of bias and variation (that is, those sources that can reasonably be avoided in routine sample measurements). Avoidable sources include, but are not limited to, modifications to the sample, modifications to the measurement procedure, modifications to the measurement equipment of the validated method, and gross and easily discernible transcription errors (provided there is a way to detect and either correct or eliminate these errors in routine processing of samples).
1.4 The WQE applies to measurement methods for which instrument calibration error is minor relative to other sources, because this practice does not model or account for instrument calibration error, as is true of most quantitation estimates in general. Therefore, the WQE procedure is appropriate when the dominant source of variation is not instrument calibration, but is perhaps one or ...
General Information
- Status
- Published
- Publication Date
- 14-Nov-2021
- Technical Committee
- D19 - Water
- Drafting Committee
- D19.02 - Quality Systems, Specification, and Statistics
Relations
- Effective Date
- 01-May-2020
- Effective Date
- 01-Apr-2019
- Effective Date
- 01-Jun-2014
- Effective Date
- 01-Oct-2013
- Effective Date
- 01-Oct-2012
- Effective Date
- 15-Jun-2012
- Effective Date
- 15-Feb-2012
- Effective Date
- 01-Jan-2012
- Effective Date
- 01-Sep-2010
- Effective Date
- 15-May-2010
- Effective Date
- 01-Mar-2010
- Effective Date
- 15-Jan-2008
- Effective Date
- 01-Oct-2007
- Effective Date
- 01-Sep-2006
- Effective Date
- 01-Sep-2006
Overview
ASTM D7783-21: Standard Practice for Within-laboratory Quantitation Estimation (WQE) provides a uniform approach for laboratories to estimate their quantitation capability within their routine analytical processes. WQE is a statistical measure that defines the lowest concentration at which a laboratory can reliably quantify an analyte in a given matrix using a specific method, with a required relative standard deviation (RSD), usually 10%, 20%, or 30%. This practice is essential for ensuring data integrity, comparability, and transparency in laboratory reporting, particularly where quantitation is critical to data use.
By applying ASTM D7783-21, laboratories can objectively assess and communicate their within-laboratory quantitation estimate (WQE), supporting informed decisions regarding data censoring and uncertainty estimates.
Key Topics
- WQE Definition and Purpose
- The WQE represents the lowest concentration that can be measured with a specified level of precision (%RSD) in a routine laboratory setting.
- WQE is not intended as a reporting limit but as a reliable, objective measure of quantitation capability and data uncertainty.
- Statistical Approach
- The practice outlines requirements for representative data collection, minimum data points and concentrations, and exclusion of non-routine sources of error in WQE determination.
- Four models of standard deviation versus concentration are considered (constant, linear, hybrid, exponential) to ensure the best statistical fit.
- Application to Laboratory Methods
- WQE is applied to individual analyte-matrix-method combinations within a single laboratory, ensuring that quantitation limits reflect real-world routine conditions.
- Data used must include all sources of routine variability such as sample handling, preparation, and instrument performance.
- Quality Control and Data Integrity
- Standardizes how laboratories document and justify their quantitation limits.
- Promotes transparency by requiring that the method, matrix, analyte, and relevant procedural details be reported with each WQE.
Applications
- Routine Chemical Analysis
- Used in environmental, water, pharmaceutical, and food laboratories to validate that routine measurement processes meet necessary precision requirements.
- Method Validation and Comparison
- Provides a consistent statistical basis for comparing the quantitation performance of different analytical methods for the same analyte/matrix within a laboratory.
- Data Quality Management
- Assists laboratories in setting and justifying quantitation limits for regulatory compliance, accreditation, or internal quality control.
- Supports reporting of measurement uncertainty alongside analytical data, meeting international data quality objectives.
- Improving Laboratory Performance
- Identifies the achievable performance level for any new method or protocol under routine conditions.
- Illuminates impacts of routine sources of error, helping to optimize procedures and staff training.
Related Standards
- ASTM D1129: Terminology Relating to Water
- ASTM D2777: Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water
- ASTM D6512: Practice for Interlaboratory Quantitation Estimate
- ASTM D7510: Practice for Performing Detection and Quantitation Estimation and Data Assessment Utilizing DQCALC Software
- ASTM E1763: Guide for Interpretation and Use of Results from Interlaboratory Testing of Chemical Analysis Methods
- BIPM GUM: JCGM 100:2008, Evaluation of measurement data - Guide to the expression of uncertainty in measurement
ASTM D7783-21 is a critical international standard for laboratories seeking to establish and document their quantitation capability using best practices. Its implementation enhances data quality, supports regulatory compliance, and ensures that quantitation limits reflect practical, achievable routine laboratory performance. For laboratories analyzing complex samples and striving for accurate, comparable data, this standard is an essential component of their quality management system.
Buy Documents
ASTM D7783-21 - Standard Practice for Within-laboratory Quantitation Estimation (WQE)
REDLINE ASTM D7783-21 - Standard Practice for Within-laboratory Quantitation Estimation (WQE)
Get Certified
Connect with accredited certification bodies for this standard
BSMI (Bureau of Standards, Metrology and Inspection)
Taiwan's standards and inspection authority.
Sponsored listings
Frequently Asked Questions
ASTM D7783-21 is a standard published by ASTM International. Its full title is "Standard Practice for Within-laboratory Quantitation Estimation (WQE)". This standard covers: SIGNIFICANCE AND USE 5.1 Appropriate application of this practice should result in a WQE achievable by the laboratory in applying the tested method/matrix/analyte combination to routine sample analysis. That is, a laboratory should be capable of measuring concentrations greater than WQEZ %, with the associated RSD equal to Z % or less. 5.2 The WQE values may be used to compare the quantitation capability of different methods for analysis of the same analyte in the same matrix within the same laboratory. 5.3 The WQE procedure should be used to establish the within-laboratory quantitation capability for any application of a method in the laboratory where quantitation is important to data use. The intent of the WQE is not to impose reporting limits. The intent is to provide a reliable procedure for establishing the quantitative characteristics of the method (as implemented in the laboratory for the matrix and analyte) and thus to provide the laboratory with reliable information characterizing the uncertainty in any data produced. Then the laboratory can make informed decisions about censoring data and has the information necessary for providing reliable estimates of uncertainty with reported data. SCOPE 1.1 This practice establishes a uniform standard for computing the within-laboratory quantitation estimate associated with Z % relative standard deviation (referred to herein as WQEZ %), and provides guidance concerning the appropriate use and application. 1.2 WQEZ % is computed to be the lowest concentration for which a single measurement from the laboratory will have an estimated Z % relative standard deviation (Z % RSD, based on within-laboratory standard deviation), where Z is typically an integer multiple of 10, such as 10, 20, or 30. Z can be less than 10 but not more than 30. The WQE10 % is consistent with the quantitation approaches of Currie (1)2 and Oppenheimer, et al. (2). 1.3 The fundamental assumption of the WQE is that the media tested, the concentrations tested, and the protocol followed in developing the study data provide a representative and fair evaluation of the scope and applicability of the test method, as written. Properly applied, the WQE procedure ensures that the WQE value has the following properties: 1.3.1 Routinely Achievable WQE Value—The laboratory should be able to attain the WQE in routine analyses, using the laboratory’s standard measurement system(s), at reasonable cost. This property is needed for a quantitation limit to be feasible in practical situations. Representative data must be used in the calculation of the WQE. 1.3.2 Accounting for Routine Sources of Error—The WQE should realistically include sources of bias and variation that are common to the measurement process and the measured materials. These sources include, but are not limited to intrinsic instrument noise, some typical amount of carryover error, bottling, preservation, sample handling and storage, analysts, sample preparation, instruments, and matrix. 1.3.3 Avoidable Sources of Error Excluded—The WQE should realistically exclude avoidable sources of bias and variation (that is, those sources that can reasonably be avoided in routine sample measurements). Avoidable sources include, but are not limited to, modifications to the sample, modifications to the measurement procedure, modifications to the measurement equipment of the validated method, and gross and easily discernible transcription errors (provided there is a way to detect and either correct or eliminate these errors in routine processing of samples). 1.4 The WQE applies to measurement methods for which instrument calibration error is minor relative to other sources, because this practice does not model or account for instrument calibration error, as is true of most quantitation estimates in general. Therefore, the WQE procedure is appropriate when the dominant source of variation is not instrument calibration, but is perhaps one or ...
SIGNIFICANCE AND USE 5.1 Appropriate application of this practice should result in a WQE achievable by the laboratory in applying the tested method/matrix/analyte combination to routine sample analysis. That is, a laboratory should be capable of measuring concentrations greater than WQEZ %, with the associated RSD equal to Z % or less. 5.2 The WQE values may be used to compare the quantitation capability of different methods for analysis of the same analyte in the same matrix within the same laboratory. 5.3 The WQE procedure should be used to establish the within-laboratory quantitation capability for any application of a method in the laboratory where quantitation is important to data use. The intent of the WQE is not to impose reporting limits. The intent is to provide a reliable procedure for establishing the quantitative characteristics of the method (as implemented in the laboratory for the matrix and analyte) and thus to provide the laboratory with reliable information characterizing the uncertainty in any data produced. Then the laboratory can make informed decisions about censoring data and has the information necessary for providing reliable estimates of uncertainty with reported data. SCOPE 1.1 This practice establishes a uniform standard for computing the within-laboratory quantitation estimate associated with Z % relative standard deviation (referred to herein as WQEZ %), and provides guidance concerning the appropriate use and application. 1.2 WQEZ % is computed to be the lowest concentration for which a single measurement from the laboratory will have an estimated Z % relative standard deviation (Z % RSD, based on within-laboratory standard deviation), where Z is typically an integer multiple of 10, such as 10, 20, or 30. Z can be less than 10 but not more than 30. The WQE10 % is consistent with the quantitation approaches of Currie (1)2 and Oppenheimer, et al. (2). 1.3 The fundamental assumption of the WQE is that the media tested, the concentrations tested, and the protocol followed in developing the study data provide a representative and fair evaluation of the scope and applicability of the test method, as written. Properly applied, the WQE procedure ensures that the WQE value has the following properties: 1.3.1 Routinely Achievable WQE Value—The laboratory should be able to attain the WQE in routine analyses, using the laboratory’s standard measurement system(s), at reasonable cost. This property is needed for a quantitation limit to be feasible in practical situations. Representative data must be used in the calculation of the WQE. 1.3.2 Accounting for Routine Sources of Error—The WQE should realistically include sources of bias and variation that are common to the measurement process and the measured materials. These sources include, but are not limited to intrinsic instrument noise, some typical amount of carryover error, bottling, preservation, sample handling and storage, analysts, sample preparation, instruments, and matrix. 1.3.3 Avoidable Sources of Error Excluded—The WQE should realistically exclude avoidable sources of bias and variation (that is, those sources that can reasonably be avoided in routine sample measurements). Avoidable sources include, but are not limited to, modifications to the sample, modifications to the measurement procedure, modifications to the measurement equipment of the validated method, and gross and easily discernible transcription errors (provided there is a way to detect and either correct or eliminate these errors in routine processing of samples). 1.4 The WQE applies to measurement methods for which instrument calibration error is minor relative to other sources, because this practice does not model or account for instrument calibration error, as is true of most quantitation estimates in general. Therefore, the WQE procedure is appropriate when the dominant source of variation is not instrument calibration, but is perhaps one or ...
ASTM D7783-21 is classified under the following ICS (International Classification for Standards) categories: 17.020 - Metrology and measurement in general. The ICS classification helps identify the subject area and facilitates finding related standards.
ASTM D7783-21 has the following relationships with other standards: It is inter standard links to ASTM D1129-13(2020)e2, ASTM E2586-19e1, ASTM E2586-14, ASTM E2586-13, ASTM E2586-12b, ASTM D2777-12, ASTM E2586-12a, ASTM E2586-12, ASTM E2586-10a, ASTM E2586-10, ASTM D1129-10, ASTM D2777-08, ASTM E2586-07, ASTM D1129-06a, ASTM D1129-06ae1. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ASTM D7783-21 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D7783 − 21
Standard Practice for
Within-laboratory Quantitation Estimation (WQE)
This standard is issued under the fixed designation D7783; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope variation (that is, those sources that can reasonably be avoided
in routine sample measurements). Avoidable sources include,
1.1 This practice establishes a uniform standard for com-
but are not limited to, modifications to the sample, modifica-
puting the within-laboratory quantitation estimate associated
tions to the measurement procedure, modifications to the
with Z % relative standard deviation (referred to herein as
measurement equipment of the validated method, and gross
WQE ), and provides guidance concerning the appropriate
Z%
and easily discernible transcription errors (provided there is a
use and application.
way to detect and either correct or eliminate these errors in
1.2 WQE is computed to be the lowest concentration for
Z%
routine processing of samples).
which a single measurement from the laboratory will have an
estimated Z% relative standard deviation (Z% RSD, based on
1.4 The WQE applies to measurement methods for which
within-laboratory standard deviation), where Z is typically an
instrument calibration error is minor relative to other sources,
integer multiple of 10, such as 10, 20, or 30. Z can be less than
because this practice does not model or account for instrument
10 but not more than 30. The WQE is consistent with the
10 %
calibration error, as is true of most quantitation estimates in
quantitation approaches of Currie (1) and Oppenheimer, et al.
general. Therefore, the WQE procedure is appropriate when
(2).
the dominant source of variation is not instrument calibration,
but is perhaps one or more of the following:
1.3 The fundamental assumption of the WQE is that the
media tested, the concentrations tested, and the protocol
1.4.1 Sample Preparation, and especially when calibration
followed in developing the study data provide a representative
standards do not go through sample preparation.
and fair evaluation of the scope and applicability of the test
1.4.2 Differences in Analysts, and especially when analysts
method, as written. Properly applied, the WQE procedure
have little opportunity to affect instrument calibration results
ensures that the WQE value has the following properties:
(as is the case with automated calibration).
1.3.1 Routinely Achievable WQE Value—The laboratory
1.4.3 Differences in Instruments (measurement equipment),
shouldbeabletoattaintheWQEinroutineanalyses,usingthe
such as differences in manufacturer, model, hardware,
laboratory’s standard measurement system(s), at reasonable
electronics, sampling rate, chemical-processing rate, integra-
cost. This property is needed for a quantitation limit to be
tion time, software algorithms, internal signal processing and
feasible in practical situations. Representative data must be
thresholds, effective sample volume, and contamination level.
used in the calculation of the WQE.
1.3.2 Accounting for Routine Sources of Error—The WQE
1.5 Data Quality Objectives—For a given method, one
should realistically include sources of bias and variation that
typically would compute the WQE for the lowest RSD for
are common to the measurement process and the measured
which the data set produces a reliable estimate. Thus, if
materials.Thesesourcesinclude,butarenotlimitedtointrinsic
possible, WQE would be computed. If the data indicated
10%
instrument noise, some typical amount of carryover error,
that the method was too noisy, so that WQE could not be
10%
bottling, preservation, sample handling and storage, analysts,
computed reliably, one might have to compute instead
sample preparation, instruments, and matrix.
WQE , or possibly WQE . In any case, a WQE with a
20% 30%
1.3.3 Avoidable Sources of Error Excluded—The WQE
higherRSDlevel(suchasWQE )wouldnotbeconsidered,
50%
should realistically exclude avoidable sources of bias and
though a WQE with RSD < 10% (such as WQE ) could be
5%
acceptable. The appropriate level of RSD is based on the data
quality objective(s) for a particular use or uses. This practice
This practice is under the jurisdiction ofASTM Committee D19 on Water and
is the direct responsibility of Subcommittee D19.02 on Quality Systems,
allows for calculation of WQEs with user selected RSDs less
Specification, and Statistics.
than 30%.
Current edition approved Nov. 15, 2021. Published March 2022. Originally
approved in 2012. Last previous edition approved in 2013 as D7783–13. DOI:
1.6 This international standard was developed in accor-
10.1520/D7783-21.
2 dance with internationally recognized principles on standard-
Theboldfacenumbersinparenthesesrefertothelistofreferencesattheendof
this standard. ization established in the Decision on Principles for the
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D7783 − 21
Development of International Standards, Guides and Recom- (after optional outlier removal) from at least six independent
mendations issued by the World Trade Organization Technical measurements at a minimum of five concentrations.
Barriers to Trade (TBT) Committee.
4.2 Retaineddataareanalyzedtoidentifyandfitoneoffour
proposedstandard-deviationmodels: constant, linear(straight-
2. Referenced Documents
line), hybrid (proposed by Rocke and Lorenzato (3)), and
2.1 ASTM Standards:
exponential. These models describe the relationship between
D1129Terminology Relating to Water
the within-laboratory standard deviation of measurements and
D2777Practice for Determination of Precision and Bias of
the true concentration, T. The selection process involves
Applicable Test Methods of Committee D19 on Water
evaluating the models, starting with the linear model and
E2586Practice for Calculating and Using Basic Statistics
performing statistical tests to choose the simplest model that
adequately fits the data. Evaluation includes statistical signifi-
2.2 BIPM Documents:
cance testing and residual analysis, and requires the judgment
GUM: JCGM 100:2008Evaluation of measurement data—
of a qualified chemist.
Guide to the expression of uncertainty in measurement
4.3 Once the standard-deviation model has been selected, it
3. Terminology
determines the fitting technique for the model of measured
3.1 Definitions—For definitions of terms used in this
concentration versus true concentration, referred to in this
practice, refer to Terminology D1129, Practice E2586, and the practice as the mean-recovery model. If the standard deviation
GUM.
is constant, then ordinary least squares may be used. If the
standard deviation is not constant, the predicted standard
3.2 Definitions of Terms Specific to This Standard:
deviations are used to generate weights for use in weighted
3.2.1 censored measurement, n—a measurement that is not
least squares. Regardless of the fitting technique, the mean-
reported numerically, but is stated as a “nondetection” or a
recovery model fits a straight line to the data.
less-than (for example, “less than 0.1 ppb”).
4.4 The linear regression (measured versus true) is evalu-
3.2.2 quantitation limit (QL) or limit of quantitation (LQ),
ated for statistical significance, for lack of fit, and for residual
n—a numerical value, expressed in physical units or
patterns.
proportion, intended to represent the lowest level of
quantitation, based on a set of criteria for quantitation.
4.5 Thesetwomodels(standarddeviationandrecovery)are
3.2.2.1 Discussion—The WQE is an example of a QL.
then used to calculate the WQE values. Either a direct
3.2.3 Z % within-laboratory quantitation estimate calculation or iterative algorithm (depending on the model) is
(WQE ), n—(in accordance with Currie (1))—The lowest used to compute WQE , the lowest true concentration with
10%
Z%
estimated RSD = 10% (Z = 10); WQE (%RSD = 20 = Z);
concentration for which a single measurement from the exam-
20%
ined laboratory will have an estimated Z% relative standard and WQE (%RSD = 30 = Z). If needed for particular
30%
data-qualityobjectives(DQOs),WQE maybecomputedfor
deviation (Z% RSD, based on the within-laboratory standard
Z %
deviation). some Z<10.Theparticular Z%selectedforuseshoulddepend
upon the data-quality needs and the realized performance.
4. Summary of Practices
Typically, either 10% or 20% is used in environmental water
testing. The 30% RSD approaches the criterion for detection.
4.1 The WQE procedure provides an estimate of the true
Z values greater than 30 should not be used. An RSD of 5%
concentration at which a desired relative precision is achieved.
approximates a level at which at least one sure significant digit
Whether from analysis of routine quality samples or from
has been achieved.
studies undertaken from time to time (or both), the first step is
toacquiredatarepresentativeofthelaboratoryperformancefor
5. Significance and Use
use in the WQE calculations. Such data must include concen-
trations suitable for modeling the precision and bias over a
5.1 Appropriate application of this practice should result in
range of concentrations. Each datum for a method/matrix/
a WQE achievable by the laboratory in applying the tested
analyte should represent an independent sample where routine
method/matrix/analytecombinationtoroutinesampleanalysis.
sources of measurement variability occur at typical levels of
That is, a laboratory should be capable of measuring concen-
influence. Outlying individual measurements should be
trations greater than WQE , with the associated RSD equal
Z%
eliminated, using an accepted, scientifically-based procedure
to Z% or less.
for outlier identification and a documented, scientific basis for
5.2 The WQE values may be used to compare the quanti-
removal of data from the data set, such as found in Practice
tation capability of different methods for analysis of the same
D2777. WQE computations must be based on retained data
analyte in the same matrix within the same laboratory.
5.3 The WQE procedure should be used to establish the
within-laboratory quantitation capability for any application of
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
a method in the laboratory where quantitation is important to
Standards volume information, refer to the standard’s Document Summary page on
data use. The intent of the WQE is not to impose reporting
the ASTM website.
limits. The intent is to provide a reliable procedure for
Available from https://www.bipm.org/en/publications/guides/, accessed May
2021. establishing the quantitative characteristics of the method (as
D7783 − 21
implemented in the laboratory for the matrix and analyte) and of replicates at each concentration. The range chosen, exclud-
thus to provide the laboratory with reliable information char- ing the zero value for purposes of the discussion of range,
acterizing the uncertainty in any data produced. Then the should extend from below the estimated detection level to
laboratory can make informed decisions about censoring data abovetheWQEofinterest(forexample,10%,20%,or30%),
and has the information necessary for providing reliable to avoid the need for extrapolation.
estimates of uncertainty with reported data. 6.2.2.2 A single model (one of the four models in this
practice)shoulddescribethebehaviorofthestandarddeviation
6. Procedure in this range. The anticipated form of the relationship between
measurement standard deviation and true concentration, if
6.1 This procedure is described in stages as follows: Devel-
known, can help in choosing design spacing. Chemistry,
opment of Data, Data Screening, Modeling Standard
physics, empirical evidence, or informed judgment may make
Deviation, Fitting the Recovery Relationship, and Computing
one model more likely than others. Evaluation of interlabora-
the Quantitation Estimates.
tory method-validation studies may also provide information
6.2 Development of Data for Input to the Calculations—A
about these relationships. If a model of standard deviation is
single WQE calculation is performed per analyte, matrix/
likely to be one with curvature at lower concentrations (hybrid
medium and method. A minimum of five concentrations must
orexponential),thenasemi-geometricdesignisfavored.Ifthe
be used to allow for high-quality estimation of measured-
likely relationship is constant or linear, then equidistant spac-
versus-true concentration, and for modeling the relationship of
ing might be favored.
standard deviation to true concentration. At least six values at
6.2.2.3 Inclusion of additional concentrations, beyond the
each concentration are required to provide a high-quality
minimum of five concentrations, is strongly recommended
estimation of the standard-deviation and the recovery relation-
where knowledge of these relationships is unknown. Where
ships. Additional concentrations (especially additional
more than one order of magnitude is covered in the range
representative, independent samples at each concentration) are
selected (per range definition in 6.2.2.1), it is recommended
highly encouraged. Such inclusion will reduce the uncertainty
that four additional unique concentrations be added per addi-
in the estimate and better assure that after outlier removal, the
tional order of magnitude greater than one.
minimum requirements for concentrations and values will be
(1)Where ongoing quality-control (QC) information is
met. Data for each WQE calculation should come from only
available and it indicates that precision is good at the concen-
one laboratory and one method, and be for only one analyte in
tration of this quality control measure (for example, 5% RSD
one matrix/medium. Concentrations may be designed in
or less at higher concentrations), then establishing the maxi-
advance, or data already developed may be used.
mumconcentrationforthestudyatorbelowthatconcentration
6.2.1 Designing Concentrations—Where concentrations are
should be considered where the RSD criterion for the WQE is
being selected in advance of the collection of data, the
higher (for example, a WQE ).
10%
development of an optimized design should consider many
(2)Where ongoing QC demonstrates a high RSD (for
factors, including:
example,above30%),severalconcentrationsatandabovethe
6.2.1.1 Concentrations of available data, such as routine
concentration of the QC sample should be included.
quality-control samples.
NOTE 1—Where more than five concentrations are available, determi-
6.2.1.2 Potential use of the same data to calculate detection
nation of the WQE with and without the highest (and potentially the
limits and or other control limits.
lowest) concentration(s) included can provide insight into the effects of
6.2.1.3 The anticipated or previously determined WQE
thehighestconcentration(s)ontherecoveryrelationshipandthemodeling
(study range should exceed this value by at least a factor of 2).
of standard deviation. Calculation of the WQE values based on the most
appropriate and applicable concentrations, so long as minimums are met,
6.2.1.4 The potential need to eliminate the lowest concen-
is allowed.
tration(s) selected (see zero-concentration discussion above).
6.2.1.5 Where possible, select a WQE study design that has 6.2.2.4 The minimum of six independent values at each
concentration is required by this practice to provide a mini-
enoughdistinctconcentrationstoassessstatisticallackoffitof
the models (see Draper and Smith (4)). Recommended designs mally acceptable data set for calculation of standard deviation
are: (a) The semi-geometric design with five or more true at each concentration. Increasing the number of levels is
concentrations, T , T , and so forth, such as: 0, WQE /D , desirablewhereprojectconstraintsallow.Itisnotrequiredthat
1 2 0
WQE /D,WQE , D×WQE , D × WQE , where D is a the same number of replicates be used for each concentration;
0 0 0 0
number greater than or equal to 2 and WQE is an initial however, extreme differences (for example orders of magni-
estimate of the WQE, (b) equi-spaced design: 0, WQE /2, tude) should be avoided.
WQE , (3/2) × WQE,2×WQE , (5/2) × WQE . Other 6.2.2.5 Known, routine sources of measurement variability,
0 0 0 0
designs with at least five concentrations—provided the design
consistentwiththoseofroutineanalysisofsamples,musthave
includes blanks, one concentration that approximates 2 × been in action at the time of the generation of the data to be
WQE , and at least one nonzero concentration belowWQE —
used, if the WQE is to be used for characterizing routine
0 0
should be adequate. performance. That is, in order for the WQE to represent
6.2.2 Considerations for All Concentration Selections:
routinely achieved quantitation, the data used for WQE calcu-
6.2.2.1 The range of the data, the number of unique lation must be generated under routine analytical conditions.
concentrations, and the spacing of the concentrations are the Representative within-laboratory variation can only be seen if
primary decisions for study design, in addition to the number the number of qualified analysts and qualified measurement
D7783 − 21
systems in the laboratory are represented. The data used and However,formanymethods,itmaynotbepossibletoconduct
the more combinations included, the less effect any specific an unbiased sampling of the zero (blank) concentration
bias in these pairings should have on the WQE estimate. samples, since instruments and software systems routinely
Similarly,samplemanagement(forexample,holdingtime)and smoothelectronicinformation(rawdata)fromthedetectorand
allowed variations in routine sample-processing procedures
through software settings that censor reported data. Through
mustbeincluded.Thetimeperiodspannedmustallowroutine, these automated processes, many testing instruments return to
time-dependent sources of variation to affect the testing. This
the operator a result value of “zero,” when, if these processes
consideration should include factors such as the frequency of had been turned off, a non-zero numeric result (positive or
calibration of instruments, introduction of newly prepared or
negative) would have been produced. These “false-zero” val-
purchased standards, reagents and supplies, and sample-
ues adversely affect the use of the zero-concentration data in
holding times. Historically, the failure to utilize representative
statistics and should not be used for WQE studies. Most
data in determination of quantitation limits has been a primary
chromatography systems (and many other types of computer-
component in over-statements of quality through quantitation-
assisted instruments) have instrument set-points (such as (digi-
limit values and should be strictly avoided (that is, garbage in,
tal) bunch rate, slope sensitivity, and minimum area counts)
garbage out). Ideally, each measurement would be a double-
that are operator-controllable. For purposes of this study,
blind measurement made by a different analyst, using a
generating as much uncensored low-level data as practical is
different (qualified) measurement system on a different day.
important and the presence of these processes as well as the
Optimally,datatobeusedshouldbeeithercompletelyblind,or
setting of any operator-controllable setting should be evalu-
fromknownbutcompletelyroutine,integratedtesting(suchas
ated.
routine quality-control data). In any case, the goal is to
NOTE 2—Qualitative criteria used by the method to identify and
minimize special treatment of the WQE test samples.
discriminate among analytes are separate criteria, and must be satisfied
6.2.2.6 Where the WQE is meant to represent the best
according to the method.
possible performance, and not routine performance, then opti-
6.2.3.1 Once true-concentration-zero measurements have
mized conditions for data generation would be appropriate.
beengenerated,andpriortouse,itisimportanttoexamineand
Similarly, if the performance of only a single process, instru-
evaluate these data. A graph of measured concentration by
ment system, analyst, etc., is of interest, only the applicable
frequency of occurrence may be helpful. However, unless a
variables should be included. It is the responsibility of the user
fairly large sample size is represented (for example, n > 20),
of this practice to assure that the appropriate data are utilized
the distribution may be distorted by the random nature of
for the end use(s) ofWQE.Where the end use is unknown, the
samplingalone.Asageneralrule,iftherewerenobias,thenon
data generator who is using the WQE needs to disclose the
average and over a large sampling, a truly uncensored set of
specificattributesofthedatausedinthecalculation(aswellas
zero-concentration (blank) data would have a mean of zero
the RSD), and thus of the WQE.
with approximately half of the results being negative values
6.2.2.7 Where preexisting, routine-source data (for
andhalfpositive,andbeNormallydistributed.Ifsomepositive
example, quality-control data) are used, care must be taken to
or negative bias were present, the percentages would shift.
assure that: (1) each data point represents a true and indepen-
However, in general the frequency should be higher near the
dent sampling of the population (as well as of the sample
mean of the values and should decline as the concentrations
mediumbeingexamined,whereapplicable)and(2)allsample-
move away from the mean, with approximately half of the
processing steps and equipment (for example, bottles,
non-mean data above and half below the mean.
preservatives, holding, preparation, cleanup) are represented.
(1)Blank data are considered suspect if: (1) there is no
Also,“true”concentrationlevelsmusteitherbeknown(thatis,
variation in these data, (2) there are an inordinate number of
true“spiked”concentrationlevels),orknowable,afterthefact.
zerovalues(andnonegativevalues)relativetothefrequencies
Aconcentrationisconsidered knownifreferencestandardscan
ofpositivevalues(6.2.3above),(3)ifthereisahighfrequency
be purchased or constructed, and knowable if an accurate
of the lowest value in the data set (for example, where
determination can be made.
minimum-peak-area rejection has been used) relative to the
6.2.2.8 Transformation of other types of data (such as
frequency of higher concentration values, and few or no lower
laboratory replicates, which under-represent the variability as
values, or (4) a frequency graphic does not begin to approxi-
compared to independent samples and usually do not have
mate a bell curve (when there are 20 or more samples).
known true concentrations), using scientifically and statisti-
(2)If the distribution of the data is suspect, the literature,
cally sound approaches is not prohibited by this practice.
plus instrument-software and equipment manuals, should be
However, care must be taken and the validity of these trans-
consulted. These documents can provide an understanding of:
formations tested. It is also critical that any standards used to
(1) the theory of operation of the detection system, (2) the
prepare study samples be completely independent of the
signal processing, calibration, etc., and (3) other aspects of the
standards used to calibrate the instrument.
conversion of response to reported values. Judgment will be
6.2.2.9 Blank correction should not be performed, unless
needed to determine whether to use some or all of the
the method requires this correction to calculate result values.
true-concentration-zero (blank) data, or to exclude the data
6.2.3 True-Concentration Zero (Blank) Data Discussion— from the calculations. In general, if less than 10% of the
Where possible, it is preferable to include data from samples zero-concentration data are: (1) censored, (2) suspect, or (3)
with true concentration of zero (for example, blanks). false-zeros, then these “problem” data should be removed.
D7783 − 21
Only the remaining blank data are used in the WQE calcula- is required. See Carroll and Ruppert (6) for further discussion
tions; there must be at least six replicates. Where the zero of standard-deviation modeling. The models are as follows:
concentration is excluded or is not possible to obtain, it is
Model C Constant WLSD Model :□s 5 g1error (1)
~ !
importanttoincludeatrueconcentrationascloseaspossibleto
where:
zero in the study design.
s = standard deviation of measurement results, and
(3)Where 75% or less of the data are censored or
g = model parameter.
smoothed, and there are at least six remaining values, it is
reasonable to use statistical procedures to simulate the distri-
Under Model C, standard deviation does not change with
bution that is missing or smoothed. Software procedures are
concentration, resulting in a relative standard deviation that
commercially available. Additionally, procedures such as log-
declines with increasing concentration, T.
normal transformation may be used to accommodate data that
Model L ~Linear WLSD Model!:□s 5 g1h 3 T1error (2)
are not normally distributed. The decision about inclusion or
exclusion of zero-concentration data in aWQE data set should where:
weigh:(1)thenumberofotherconcentrationsavailable,(2)the
s = standard deviation of measurement results,
range of the other concentrations, and (3) the risk of extrapo-
T = true concentration, and
lation of the WQE outside the data-set concentration range
g and h = model parameters.
against the quality of the zero-concentration data.
Under Model L, standard deviation increases linearly with
6.2.3.2 True Concentrations Near Zero—As with concen-
concentration, resulting in an asymptotically constant relative
tration zero, true concentrations very near to zero may also
standard deviation as T increases.
have been censored, smoothed, and contain false-zeros. Ex-
2 2
Model H Hybrid WLSD Model :□s 5 =g 1 h 3 T 1error(3)
~ ! ~ !
amination of these very low concentrations, as above for zero
concentration, is important. The likelihood of occurrence and
where:
the percentage of data affected decreases with increasing
s = standard deviation of measurement results,
concentration.
T = true concentration, and
g and h = model parameters.
6.3 Data Screening, Outlier Identification, and Outlier Re-
moval: Under Model H, within-laboratory standard deviation in-
creases with concentration in such a way that the relative
6.3.1 Data that are to be the input to the WQE calculation
standard deviation declines as T increases, approaching an
should be screened for compliance with this practice’s
asymptote of h.
conditions, appropriateness for the intended use of the WQE,
obvious errors, and individual outliers. Graphing of the data
Model E Exponential WLSD Model :□s 5 g 3exp h 3 T 1error
~ ! ~ !
(true versus measured) is recommended as an assistive visual
(4)
tool.
where:
6.3.2 Outlying individual measurements must be evaluated;
s = standard deviation of measurement results,
if determined to be erroneous, they should be eliminated using
T = true concentration, and
scientifically-based reasoning. Identification of potential outli-
g and h = model parameters.
ers for data evaluation and validation may be accomplished
using statistical procedures or through visual examination of a Under Model E, within-laboratory standard deviation in-
graphical representation of the data. WQE computations must creasesexponentiallywithconcentration,resultinginarelative
be based on retained data from at least six independent standarddeviationthatmayinitiallydeclineas Tincreases,but
measurements at each of at least five concentration levels.The eventually increases with T, so that there is at most a bounded
data removed and the percentage of data removed must be quantitation range within which the RSD is less than Z ⁄100.
recorded and retained to document the WQE calculations. 6.4.1.1 The procedures for estimating the parameters of
each model and their uncertainties are given inAppendix X2 –
6.4 Modeling Standard Deviation Versus True
Appendix X5. Model L should be tried first, since its fitting
Concentration—Thepurposeistopredictthewithin-laboratory
procedure (in Appendix X2) provides criteria for choosing
measurement standard deviation (WLSD) as a function of true
among the various models.
concentration, σˆ(T).The relationship is used for two purposes:
6.4.1.2 In all cases, it is assumed that g > 0.Avalue of g <
(1)toprovideweights(ifneeded)forfittingthemean-recovery
0 has no practical interpretation and may indicate that a
model and (2) to provide the within-laboratory standard
different WLSD model should be used. Furthermore, it is
deviation estimates crucial to determining the WQEs.
assumed that g is not underestimated by censored data among
NOTE 3—See Caulcutt and Boddy (5) for more discussion of standard
measurementsofblanksorotherlow-concentrationsamples.If
deviation modeling and weighted least squares (WLS) in analytical
h < 0, it must not be statistically significant, and Model C
chemistry.
should be evaluated.
6.4.1 This practice uses four models as potential fits for the 6.4.2 If a model other than the best fit is chosen, the reason
Within-Laboratory Standard Deviation (WLSD) model. The for the choice should be scientific and should be recorded to
selection process considers a linear model first, performing document the WQE.
statistical tests to decide whether a simpler constant model can 6.4.2.1 It is recommended that the relationship of measure-
beusedorwhetheroneofthemorecomplicatedcurvedmodels ment standard deviation to true concentration be graphed and
D7783 − 21
used for visual verification of the appropriateness of each 7. Review, Documentation and Reporting
model and of the model selected for use.
7.1 The WQE analysis report should include: (1) the iden-
tification of laboratory and (2) identification of analytical
6.5 Fitting the Mean-Recovery Relationship (Measured ver-
method,analyte(s),matrix(ormatrices),sampleproperties(for
sus True Concentration)—Based on the standard-deviation
example,volumeormass)andspecificmethodoptions(ifany)
model selected (constant versus other models), the mean-
utilized. Where the laboratory uses standard operating proce-
recoveryconcentrationisfittedversustrueconcentration,using
dures(SOPs)toimplementmethodsormethodprotocols,these
ordinaryleastsquaresorweightedleastsquares,asappropriate.
SOPs should be referenced, including the identification of any
The mean recovery is evaluated for statistical significance and
revision/version.Documentationofeachdatumusedshouldbe
lack of fit.Agraph of mean recovery (along with the “calibra-
equivalent to that of reported data (for example, instrument,
tion” line) and a graph of the residuals should also be visually
analyst, date, etc.). There should be a description of all
examined. Many off-the-shelf statistical software packages
data-screening procedures employed, all results obtained, all
may be used for this purpose.
individualvaluesomittedfromfurtheranalysis(thatis,outliers
NOTE 4—Regression coefficients should not be used to assess goodness
that have been removed), all missing values, and the percent-
of fit.
ageofdatautilizedinthecalculationsrelativetotheinitialdata
set.Anyanomaliesencounteredshouldbelisted,includingand
6.5.1 The mean-recovery regression (measured versus true
anomalous calibration or quality control sample results (for
concentration) model is a straight line,
example, data validation qualifiers or flags). The data (statis-
Model R:□Y 5 a1bT1error (5)
tical) analysis should be included or referenced and the WQE
values determined recorded. The selected standard-deviation
The fitting procedure depends on the chosen standard-
model, plus the coefficient estimates for this model and for
deviationmodel.Iftheconstantmodel,ModelC,wasselected,
mean-recovery model, should also be recorded. Where a
then ordinary least squares (OLS) can be used to fit Model R
statistical model other than the mathematical best fit has been
for mean recovery. In all other cases, weighted least squares
chosen, the reasoning should be described.
(WLS) should be used. WLS approximately provides the
minimum-variance unbiased linear estimate of the parameters,
8. Report
a and b. The WLS procedure is described in Appendix X6.
8.1 The analysis report should at a minimum contain:
6.6 Compute the WQE for each Z (%RSD)—Using the
8.1.1 Identification of laboratory,
mean-recovery regression line determined above, the most
8.1.2 Analytical method,
appropriate model of the relationship of relative standard
8.1.3 Analyte(s),
deviation to true concentration (also determined above), and
8.1.4 Matrix (or matrices),
the Z value desired, the user obtains the WQE, which is an
8.1.5 Sample properties (for example, volume),
estimate of the lowest true concentration (corresponding to the
8.1.6 Study design,
measuredconcentration)atwhichthedesiredRSDisachieved.
Procedures for calculating theWQE and estimating its relative 8.1.7 Analyst, method, and date of testing for each study
sample,
standarduncertaintyaregiveninAppendixX2–AppendixX5.
8.1.8 Any anomalies in the study, including QA/QC sample
6.6.1 Given the standard-deviation model, its estimated
results,
parameters, and the mean-recovery regression line, there is a
8.1.9 Data-screening results, individual values and labora-
lower limit, Z , below which WQE cannot be calculated
lim Z%
tories omitted from further analysis, and missing values,
because there is no true concentration at which RSD equals
8.1.10 WLSD model selected, and
Z%.For Z> Z ,theWQEcanbecalculatedbutitstruevalue
lim
may nevertheless be extremely uncertain, especially when Z is 8.1.11 CoefficientestimatesfortheWLSDmodelandmean-
recovery model.
near Z . If the relative standard uncertainty, u (WQE),
lim rel
exceeds 25%, the calculated WQE should be considered an
8.2 The report should be given a second-party review to
unreliable estimate. The actual reliability of the WQE also
verify that:
depends on the adequacy of the standard-deviation model,
8.2.1 The data transcription and reporting have been per-
which is more difficult to quantify.
formed correctly,
8.2.2 The analysis of the data and the application of this
NOTE 5—Under Model C, Z is zero. For the other models described
lim
here, Z is always positive. standard have been performed correctly, and
lim
8.2.3 The results of the analysis have been used
6.6.2 The measured concentration (Y ) at which the desired
Q
appropriately, including assessment of assumptions necessary
RSDwasachievedmayalsobeofinterestforsomeuses.This
to compute a WQE.
value is the level at which the required RSD was obtained in
measured concentration units (that is, the value, paired with a
NOTE 6—Reviewer(s) should be qualified in one or more of the
following areas: (1) applied statistics, (2) metrology, and (3) analytical
WQE, that has not been corrected for bias through the
chemistry.
mean-recovery regression). Where the Y and the WQE are
Q
equal (following application of significant figures and
8.3 A statement of the review and the results of the review
rounding), there is no apparent bias at theWQE concentration. should accompany the report.
D7783 − 21
FIG. 1 Sample Standard Deviations Versus True Concentration, with Linear Fit, Hybrid Model Fit, and Residuals from Linear Fit (Lower
Plot), All in ppb
9. Rationale relative error of any measurement of a true concentration
greater than theWQE will not exceed 63 × Z%. For example,
9.1 The basic rationale for the WQE is contained in Currie
a measurement above the WQE (and assumed to have true
10%
(1). The WQE is a performance characteristic of an analytical
concentration above the WQE) could be reported as 6 ppb
measurement process, to paraphrase Currie. Like an estimated
(630%) = (6 6 2) ppb, with a high degree of certainty.
detection limit, theWQE is helpful for the planning and use of
chemical analyses. The WQE is another benchmark indicating
9.3 There are several real-world complications to this ide-
whether the method can adequately meet measurement needs.
alized situation. See Maddalone et al. (7), Gibbons (8), and
9.2 The idealized definition of WQE is that it is the Coleman et al. (9). Some of these complications are listed as
Z %
lowest concentration, L , that satisfies: L = (100 / Z) σ follows:
Q Q L
Q
(where σ istheactualstandarddeviationofwithin-laboratory
L 9.3.1 Analyte recovery is not perfect; the relationship be-
Q
measurements at concentration L ); this definition is equiva-
Q
tween measured values of concentrations and true concentra-
lent to satisfying RSD = σ ⁄ L = Z %. In other words,
L
Q
Q tions cannot be assumed to be trivial. There is bias between
WQE is the lowest concentration with Z% RSD (assuming
Z %
true and measured values. Recovery can and should be
suchaconcentrationexists).If,asiscommonlythecase, RSD
modeled. Usually, a straight line will suffice. In practice, when
declines with increasing true concentration, then the relative
both the standard deviation and recovery models are used,
uncertaintyofanymeasurementofatrueconcentrationgreater
WQEZ% is calculated to be the lowest concentration L that
Q
than the WQE will not exceed 6Z%. The range, 63σ ,isan
L
Q
satisfies L 5 100 ⁄ Z σˆ L ⁄ b.
~ ! ~ !
Q Q
approximate prediction or confidence interval very likely to
9.3.2 Variation is introduced by different laboratories,
contain the measurement, which is assumed to be normally
distributed. This assertion is based on critical values from the analysts, models, and pieces of equipment; environmental
factors; flexibility/ambiguity in a test method; contamination;
normal distribution (or from the student’s t-distribution if σ is
estimated rather than known). Then, with high confidence, the carryover; matrix influence; and other factors. It is intractable
D7783 − 21
to model these factors individually, but their collective contri- where the standard deviation used in the 10σ formula is
butions to measurement WLSD can be observed, if these estimated at a detection critical value, and then is taken to be
contributions are part of how a study is designed and con- a constant (over a trace-level range of concentrations) for the
ducted. 10σ computation. In contrast, WQE follows the “more
10%
statistically and conceptually rigorous” approach described by
9.3.3 The standard deviation of measurements is generally
unknown, and may change with true concentration, possibly Gibbons et al. (8), and contained in Currie (1). This greater
rigor comes at the risk of: (a) possibly being unattainable for
because of the physical principle of the test method.To ensure
that a particular RSD is attained at or above the WQE, there somemethods(forwhichonlyalessstrictlevelofRSDcanbe
ensured); (b) having uncertainty that is potentially complex,
must be a way to predict the WLSD at different true concen-
trations. Short of severely restricting the range of concentra- and depends both on the model used and on the data.
tions for a study, prediction is accomplished by an empirical
10. Keywords
WLSD model. In all of the respects discussed in 9.1 – 9.3,
WQE is similar to the AML developed by Gibbons et al. 10.1 critical limits; matrix effects; precision; quantitation;
10%
(10). However, the AML follows an approximate approach, quantitation limits
APPENDIXES
(Nonmandatory Information)
X1. GLOSSARY OF KEY SYMBOLS, ACRONYMS, AND LABELS
∆g—one iteration’s change in the estimate of parameter g, g—model parameter representing the intercept of the
the intercept parameter in the Hybrid model, using nonlinear prediction curve, equal to the predicted measurement standard
least squares deviation at zero concentration
∆h—one iteration’s change in the estimate of parameter h, g —initial estimate of g
the slope parameter in the Hybrid model, using nonlinear least g'—ln g, the natural logarithm of g
squares h—model parameter accounting for the dependence of the
∆m—bias-correction term for the natural logarithm of (the measurement standard deviation on concentration (Models L,
numerical value of) the sample standard deviation of m H, and E)
normally distributed results h —initial estimate of h
σ—true within-laboratory standard deviation j—index used to number iteration steps when using New-
σˆ—predicted value of the within-laboratory measurement ton’s Method, for example when estimating the parameters of
standard deviation the Hybrid Model (Model H)
σˆ(T)—predicted value of the within-laboratory measure- k—index used for different concentrations, T , and associ-
k
ment standard deviation at concentration T ated statistics
ζ —variance of the logarithm of (the numerical value of) LQ—another designation for theWQE, in accordance with
m
thesamplestandarddeviationof mnormallydistributedresults Currie’s notation
ζ—quotient of an estimated value and its standard uncer- Model C—Constant model for ILSD
tainty Model L—Linear model for ILSD; within-laboratory stan-
a—estimateoftheslopeinthemean-recoverycurve(linear dard deviation increases linearly with concentration
model) Model H—Hybrid model for ILSD; combines additive and
'
a —bias-correction factor for the sample standard devia- multiplicative error, with within-laboratory standard deviation
m
tion of a random sample of m normally distributed results that increases with increasing concentration, according to the
AML—Alternative Minimum Level, a quantitation limit model proposed by Rocke and Lorenzato
that is similar to the WQE (and compatible in approach) Model E—Exponential model for ILSD; within-laboratory
b—estimateoftheslopeinthemean-recoverycurve(linear standard deviation increases exponentially with concentration,
model) ensuring at most a finite quantitation range
c–intermediate variable used in estimating g and h for the Model R—the linear model for the mean-recovery curve
Hybrid model by nonlinear least squares; similar to d, p, q, u, n—number of true concentrations in the study (n ≥ 5)
and v NLLS—nonlinear least squares, where coefficients in a
d—similar to c nonlinearmodelarecomputedtominimizethe(weighted)sum
e—Euler’s number, 2.71828… of the squares of the residuals (that is, the differences between
e —the residual associated with T from a NLLS precision the observed and predicted values)
k k
model fit; defined as the difference in log sample standard OLS—ordinaryleastsquares,afittingtechniqueforalinear
deviation and log predicted standard deviation (that is, additive) model that minimizes the sum of the squares
f(T)—the natural log of the current estimate of the within- of the residuals (that is, the differences between observed and
laboratory standard deviation at concentration T predicted values)
D7783 − 21
p—similar to c u(x, y)—estimated measurement covariance of x and y
q—similar to c u (x)—relativestandarduncertaintyof x,equalto u(x)⁄|x|
rel
th 2
th
q —k residual for the formal regression of T versus T
w —statistical weighting factor used for the k data point
k
k
QL—quantitation limit (also called practical quantitation
when estimating model parameters by weighted linear or
limit, PQL); see L
Q nonlinear least squares
RSD—relative standard deviation, that is, the standard
WLS—weighted least squares, a modified form of ordinary
deviation divided by the concentration (both generally esti-
least squares that incorporates nonuniform variability in the
mated)
data
s —observed value of the within-laboratory standard de-
k
WLSD—within-laboratory standard deviation
viation at concentration T
k
WQE —within-laboratory quantitation estimate associ-
Z %
*
s —observed value of the standard deviation at concentra-
k
ated with approximately Z % RSD
tion T , corrected for bias
k
Y—random variable representing a measurement result
s —maximumsampleILSD:equaltomax(s ,s ,.,s )
max 1 2 n
Z—numerical value of RSD expressed as a percentage
s —pooled standard deviation estimate
p
Z —the estimated lowest limit of %RSD achievable,
lim
SS —residual sum of squares
e
based on study results, for a particular measurement system,
T—true concentration
th matrix, and analyte
T —k value of true concentra
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: D7783 − 13 D7783 − 21
Standard Practice for
Within-laboratory Quantitation Estimation (WQE)
This standard is issued under the fixed designation D7783; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
Note—Balloted information was included and the year date changed on March 28, 2013.
1. Scope
1.1 This practice establishes a uniform standard for computing the within-laboratory quantitation estimate associated with Z %
relative standard deviation (referred to herein as WQE ), and provides guidance concerning the appropriate use and
Z %Z %
application.
1.2 WQE is computed to be the lowest concentration for which a single measurement from the laboratory will have an
Z% %
estimated Z %Z % relative standard deviation (Z %(Z % RSD, based on within-laboratory standard deviation), where Z is typically
an integer multiple of 10, such as 10, 20, or 30. Z can be less than 10 but not more than 30. The WQE is consistent with the
10 %
quantitation approaches of Currie (1) and Oppenheimer, et a.lal. (2).
1.3 The fundamental assumption of the WQE is that the media tested, the concentrations tested, and the protocol followed in the
developing the study data provide a representative and fair evaluation of the scope and applicability of the test method, as written.
Properly applied, the WQE procedure ensures that the WQE value has the following properties:
1.3.1 Routinely Achievable WQE Value—The laboratory should be able to attain the WQE in routine analyses, using the
laboratory‘slaboratory’s standard measurement system(s), at reasonable cost. This property is needed for a quantitation limit to be
feasible in practical situations. Representative data must be used in the calculation of the WQE.
1.3.2 Accounting for Routine Sources of Error—The WQE should realistically include sources of bias and variation that are
common to the measurement process and the measured materials. These sources include, but are not limited to intrinsic instrument
noise, some typical amount of carryover error, bottling, preservation, sample handling and storage, analysts, sample preparation,
instruments, and matrix.
1.3.3 Avoidable Sources of Error Excluded—The WQE should realistically exclude avoidable sources of bias and variation (that
is, those sources that can reasonably be avoided in routine sample measurements). Avoidable sources would include, but are not
limited to, modifications to the sample, modifications to the measurement procedure, modifications to the measurement equipment
of the validated method, and gross and easily discernible transcription errors (provided there wasis a way to detect and either
correct or eliminate these errors in routine processing of samples).
1.4 The WQE applies to measurement methods for which instrument calibration error is minor relative to other sources, because
This practice is under the jurisdiction of ASTM Committee D19 on Water and is the direct responsibility of Subcommittee D19.02 on Quality Systems, Specification,
and Statistics.
Current edition approved March 28, 2013Nov. 15, 2021. Published April 2013March 2022. Originally approved in 2012. Last previous edition approved in 20122013 as
D7783 – 12.D7783 – 13. DOI: 10.1520/D7783-13.10.1520/D7783-21.
The boldface numbers in parentheses refer to athe list of references at the end of this standard.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D7783 − 21
this practice does not model or account for instrument calibration error, as is true of quantiation most quantitation estimates in
general. Therefore, the WQE procedure is appropriate when the dominant source of variation is not instrument calibration, but is
perhaps one or more of the following:
1.4.1 Sample Preparation, and especially when calibration standards do not go through sample preparation.
1.4.2 Differences in Analysts, and especially when analysts have little opportunity to affect instrument calibration results (as is the
case with automated calibration).
1.4.3 Differences in Instruments (measurement equipment), such as differences in manufacturer, model, hardware, electronics,
sampling rate, chemical-processing rate, integration time, software algorithms, internal signal processing and thresholds, effective
sample volume, and contamination level.
1.5 Data Quality Objectives—For a given method, one typically would compute the lowest % RSD possible for any given data
set. WQE for the lowest RSD for which the data set produces a reliable estimate. Thus, if possible, WQE would be computed.
10 %
If the data indicated that the method was too noisy, so that WQE could not be computed reliably, one might have to compute
10 %
instead WQE , or possibly WQE . In any case, a WQE with a higher % RSD higher RSD level (such as WQE ) would
20 % 30 % 50 %
not be considered, though a WQE with RSD <10 % < 10 % (such as WQE ) wouldcould be acceptable. The appropriate level
1 %5 %
of % RSD RSD is based on the data-quality data quality objective(s) for a particular use or uses. This practice allows for
calculation of WQEs with user selected % RSDs RSDs less than 30 %.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2.1 ASTM Standards:
D1129 Terminology Relating to Water
D2777 Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water
D6091E2586 Practice for 99 %/95 % Interlaboratory Detection Estimate (IDE) for Analytical Methods with Negligible
Calibration ErrorCalculating and Using Basic Statistics
D6512 Practice for Interlaboratory Quantitation Estimate
D7510 Practice for Performing Detection and Quantitation Estimation and Data Assessment Utilizing DQCALC Software, based
on ASTM Practices D6091 and D6512 of Committee D19 on Water
E1763 Guide for Interpretation and Use of Results from Interlaboratory Testing of Chemical Analysis Methods
2.2 BIPM Documents:
GUM: JCGM 100:2008 Evaluation of measurement data—Guide to the expression of uncertainty in measurement
3. Terminology
3.1 Definitions—For definitions of terms used in this practice, refer to Terminology D1129., Practice E2586, and the GUM.
3.2 Definitions of Terms Specific to This Standard:
3.2.1 censored measurement, n—a measurement that is not reported numerically, but is stated as a “nondetection” or a less-than
(for example, “less than 0.1 ppb”).
3.2.2 quantitation limit (QL) or limit of quantitation (LQ), n—a numerical value, expressed in physical units or proportion,
intended to represent the lowest level of quantitation, based on a set of criteria for quantitation.
3.2.2.1 Discussion—
The WQE is an example of a QLQL.
3.2.3 Z % within-laboratory quantitation estimate
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Available from https://www.bipm.org/en/publications/guides/, accessed May 2021.
D7783 − 21
(WQE ), n—(in accordance with Currie (1))—The lowest concentration for which a single measurement from the examined
Z %Z %
laboratory will have an estimated Z %Z % relative standard deviation (Z %(Z % RSD, based on the within-laboratory standard
deviation).
4. Summary of Practices
4.1 The WQE procedure provides an estimate of the true concentration at which a desired level of (relative) relative precision is
achieved. Whether from analysis of routine quality samples or from studies undertaken from time to time (or both), the first step
is to acquire data representative of the laboratory performance for use in the WQE calculations. Such data must include
concentrations suitable for modeling the precision and bias over a range of concentrations. Each datum for a method/matrix/analyte
should represent an independent sample where routine sources of measurement variability occur at typical levels of influence.
Outlying individual measurements should be eliminated, using an accepted, scientifically-based procedure for outlier identification
and a documented, scientific basis for removal of data from the data set, such as found in Practice D2777. WQE computations must
be based on retained data (after optional outlier removal) from at least six independent measurements at a minimum of five
concentrations.
4.2 Retained data are analyzed to identify and fit one of four proposed standard-deviation models.models: constant,linear
(straight-line), hybrid (proposed by Rocke and Lorenzato (3)), and exponential. These models describe the relationship between
the within-laboratory standard deviation of measurements and the true concentration, T. The identificationselection process
involves evaluating the models in order, from simplest to most complex: constant, straight-line, exponential, and hybrid (proposed
by Rocke and Lorenzatomodels, starting with the linear model and performing statistical tests to choose the (simplest3) and
Guidemodel that E1763. adequately fits the data. Evaluation includes statistical-significance statistical significance testing and
residual analysis, and is based on the best requires the judgment of a qualified chemist and the requirement to utilize the simplest
model that adequately fits the data.chemist.
4.3 Once the standard-deviation model has been determined, it is used to determine selected, it determines the fitting technique
for modeling the model of measured concentration (referred versus true concentration, referred to in this practice as the
mean-recovery model)model. to true concentration. If If the standard deviation is constant, then ordinary least squares is may be
used. If the standard deviation is not constant, the modeled standard-deviation predictionspredicted standard deviations are used
to generate weights for use in the weighted-least-squares fitting. With either fitting technique, a straight line is the model that is
fitted weighted least squares. Regardless of the fitting technique, the mean-recovery model fits a straight line to the data.
4.4 The linerlinear regression (true(measured versus measured)true) is evaluated for statistical significance, for lack of fit, and for
residual patterns.
4.5 These two models (standard-deviation (standard deviation and calibration)recovery) are then used to calculate the WQE
values. Either a direct calculation or interactiveiterative algorithm (depending on the model) is used to compute WQE , the
10 %
lowest true concentration with estimated RSD = 10 % (Z(Z = 10); WQE (% RSD=20 %=Z); (%RSD = 20 = Z); and WQE
20 % 30 %
30 % (% RSD=30 %=Z). (%RSD = 30 = Z). If needed for particular data-quality objectives (DQOs), WQE may be computed
Z %Z %
for some Z < 10. The particular Z %Z % selected for use should depend upon the data-quality needs and the realized performance.
Typically, either 10 % or 20 % is used in environmental-water environmental water testing. The 30 % RSD approaches the criterion
for detection. Z values greater than 30 should not be used. An RSD of 5 % approximates a level at which at least one sure
significant digit has been achieved.
5. Significance and Use
5.1 Appropriate application of this practice should result in a WQE achievable by the laboratory in applying the tested
method/matrix/analyte combination to routine sample analysis. That is, a laboratory should be capable of measuring concentrations
greater than WQE , with the associated RSD equal to Z %Z % or less.
Z %Z %
5.2 The WQE values may be used to compare the quantitation capability of different methods for analysis of the same analyte in
the same matrix within the same laboratory.
5.3 The WQE procedure should be used to establish the within-laboratory quantitation capability for any application of a method
in the laboratory where quantitation is important to data use. The intent of the WQE is not to impose reporting limits. The intent
is to provide a reliable procedure for establishing the quantitative characteristics of the method (as implemented in the laboratory
D7783 − 21
for the matrix and analyte) and thus to provide the laboratory with reliable information characterizing the uncertainty in any data
produced. Then the laboratory maycan make informed decisions about censoring data and has the information necessary for
providing reliable estimates of uncertainty with reported data.
6. Procedure
6.1 This procedure is described in stages as follows: Development of Data, Data Screening, Modeling Standard Deviation, Fitting
the Recovery Relationship, and Computing the Quantitation Estimates.
6.2 Development of Data for Input to the Calculations—A single WQE calculation is performed per analyte, matrix/medium and
method. A minimum of five concentrations must be used to allow for high-quality estimation of true-verses-measuredmeasured-
versus-true concentration, and for modeling the relationship of standard deviation to true concentration. A minimum of At least
six values at each concentration are required to provide a high-quality estimation of the standard-deviation and the recovery
relationships. Additional concentrations (especially additional,additional representative, independent samples at each concentra-
tion) are highly encouraged; suchencouraged. Such inclusion will reduce the uncertainty in the estimate and better assure that after
outlier removal, the minimum requirements for concentrations and values will be met. Data for each WQE calculation should come
from only one laboratory, laboratory and one method, and be for only one analyte in one matrix/medium. Concentrations may be
designed in advance, or data already developed may be used. For multi-laboratory determinations, see Practice D6091.
6.2.1 Designing Concentrations—Where concentrations are being selected in advance of the collection of data, the development
of an optimized design should consider many factors, including:
6.2.1.1 Concentrations of available data, such as routine quality-control samples.
6.2.1.2 Potential use of the same data to calculate detection limits and or other control limits.
6.2.1.3 The anticipated or previously determined WQE (study range should exceed this value by at least a factor of 2).
6.2.1.4 The potential need to eliminate the lowest concentration(s) selected (see zero-concentration discussion above).
6.2.1.5 Where possible, select a WQE study design that has enough distinct concentrations to assess statistical lack of fit of the
models (see Draper and Smith (4)). Recommended designs are: (a) The semi-geometric design with five or more true
2 2
concentrations, T , T , and so forth, such as: 0, WQE /D/D , WQE /D,/D, WQE , D × WQE , D × WQE , where D is a number
1 2 0 0 0 0 0
greater than or equal to 2 and WQE is an initial estimate of the WQE, (b) equi-spaced design: 0, WQE /2, WQE , (3/2) × WQE ,
0 0 0 0
2 × WQE , (5/2) × WQE . Other designs with at least five concentrations—provided the design includes blanks, one concentration
0 0
that approximates 2 × WQE , and at least one nonzero concentration below WQE —should be adequate.
0 0
D7783 − 21
6.2.2 Considerations for All Concentration Selections:
6.2.2.1 The range of the data, the number of unique concentrations, and the spacing of the concentrationconcentrations are the
primary decisions for study design, in addition to the number of replicates at each concentration. The range chosen, excluding the
zero value for purposes of the discussion of range, should beextend from below the estimated detection level to above the WQE
of interest (for example, 10 %, 20 %, or 30 %), so as to allow for performance of calculations without to avoid the need for
extrapolation.
6.2.2.2 A single model (one of the four models in this practice) should describe the behavior of the standard deviation in this range.
The anticipated form of the relationship between measurement standard deviation and true concentration, if known, can help in
choosing design spacing. Chemistry, physics, empirical evidence, or informed judgment may make one model more likely than
others. Evaluation of interlaboratory method-validation studies may also provide information about these relationships. If a model
of standard deviation is likely to be one with curvature at lower concentrations (hybrid or exponential)exponential), then a
semi-geometric design is favored. If the likely relationship is constant or straight-line,linear, then equidistant spacing might be
favored.
6.2.2.3 Additional Inclusion of additional concentrations, beyond the minimum of five concentrations, is strongly recommended
where knowledge of these relationships is unknown. Where more than one order of magnitude is covered in the range selected (per
range definition in 6.2.2.1), it is recommended that four additional unique concentrations be added per additional order of
magnitude greater than one.
(1) Where ongoing quality-control (QC) information is available and it indicates that precision is good at the concentration of
this quality control measure,measure (for example, 5 % RSD or less,less at higher concentrations), then establishing the maximum
concentration for the study at or below that concentration should be considered where the % RSD criterion for the WQE is higher
(for example, a WQE 10 %).).
10 %
(2) Where ongoing QC demonstrates a high % RSD (for example, above 30 %), several concentrations at and above the
concentration of the QC sample should be included.
NOTE 1—Where more than five concentrations are available, determination of the WQE with and without the highest (and potentially the lowest)
concentration(s) included can provide insight into the effects of the highest concentration(s) on the recovery relationship and the modeling of standard
deviation. Calculation of the WQE values based on the most appropriate and applicable concentrations, so long as minimums are met, is allowed.
6.2.2.4 The minimum of six independent values at each concentration is required by this practice to provide a minimally
acceptable data set for calculation of standard deviation at each concentration. Increasing the number of levels is desirable where
project constraints allow. It is not required that the same number of replicates be used for each concentration; however, extreme
differences (for example orders of magnitude) should be avoided.
6.2.2.5 Known, routine sources of measurement variability, consistent with those of routine analysis of samples, must have been
in action at the time of the generation of the data to be used, if the WQE is to be used for characterizing routine performance. That
is, in order for the WQE to represent routinely achieved quantitation, the data used for WQE calculation must be generated under
routine analytical conditions. Representative within-laboratory variation can only be seen if the number of qualified analysts and
qualified measurement systems in the laboratory are represented. The data used and the more combinations included, the less effect
any specific bias in these pairings should have on the WQE estimate. Similarly, sample management (for example, holding time)
and allowed variations in routine sample-processing procedures must be included. The time period spanned must allow routine,
time-dependent sources of variation to affect the testing. This consideration should include factors such as the frequency of
calibration of instruments, introduction of newly prepared or purchased standards, reagents and supplies, and sample-holding
times. Historically, the failure to utilize representative data in determination of quantitation limits has been a primary component
in over-statements of quality through quantitation-limit values and should be strictly avoided (that is, garbage in, garbage out).
Ideally, each measurement would be a double-blind measurement made by a different analyst, using a different (qualified)
measurement system on a different day. Optimally, data to be used should be either completely blind, or from known but
completely routine, integrated testing (such as routine quality-control data). In any case, the goal is to minimize special treatment
of the WQE test samples.
6.2.2.6 Where the WQE is meant to represent the best possible performance, and not routine performance, then optimized
conditions for data generation would be appropriate. Similarly, if the performance of only a single process, instrument system,
analyst, etc.etc., is of interest, only the applicable variables should be included. It is the responsibility of the user of this practice
D7783 − 21
to assure that the appropriate data are utilized for the end use(s) of WQE. Where the end use is unknown, the data generator who
is using the WQE needs to disclose the specific attributes of the data used in the calculation (as well as the % RSD), RSD), and
thus of the WQE.
6.2.2.7 Where preexisting, routine-source data (for example, quality-control data) are used, care must be taken to assure that: (1)
each data point represents a true and independent sampling of the population (as well as of the sample medium being examined,
where applicable) and (2) all sample-processing steps and equipment (for example, bottles, preservatives, holding, preparation,
cleanup) are represented. Also, “true” concentration levels must either be known (that is, true “spiked” concentration levels), or
knowable, after the fact. A concentration is considered known if reference standards can be purchased or constructed, and knowable
if an accurate determination can be made.
6.2.2.8 Transformation of other types of data (such as laboratory replicates, which under-represent the variability as compared to
independent samples and usually do not have known true concentrations), using scientifically and statistically sound approaches
is not prohibited by this practice. However, care must be taken and the validity of these transformations tested. It is also critical
that any standards used to prepare study samples be completely independent of the standards used to calibrate the instrument.
6.2.2.9 Blank correction should not be performed, unless the method requires this correction to calculate result values.
6.2.3 True-Concentration Zero (Blank) Data Discussion—Where possible, it is preferable to include data from samples with true
concentration of zero (for example, blanks). However, for many methods, it may not be possible to conduct an unbiased sampling
of the zero (blank) concentration samples, since instruments and software systems routinely smooth electronic information (raw
data) from the detector and through software settings that censor reported data. Through these automated processes, many testing
instruments return to the operator a result value of “zero,” when, if these processes had been turned off, a non-zero numeric result
(positive or negative) would have been produced. These “false-zero” values adversely affect the use of the zero-concentration data
in statistics and should not be used for WQE studies. Most chromatography systems (and many other types of computer-assisted
instruments) have instrument set-points (such as (digital) bunch rate, slope sensitivity, and minimum area counts) that are
operator-controllable. For purposes of this study, generating as much uncensored low-level data as practical is important and the
presence of these processes as well as the setting of any operator-controllable setting should be evaluated.
NOTE 2—Qualitative criteria used by the method to identify and discriminate among analytes are separate criteria, and must be satisfied according to the
method.
6.2.3.1 Once true-concentration-zero measurements have been generated, and prior to use, it is important to examine and evaluate
these data. A graph of measured concentration by frequency of occurrence may be helpful. However, unless a fairly large sample
size is represented (for example, n>20),n > 20), the distribution may be distorted by the random nature of sampling alone. As a
general rule, if there were no bias, then on average and over a large sampling, a truly uncensored set of zero-concentration (blank)
data would have a mean of zero with approximately half of the results being negative values and half positive, and be Normally
distributed. If some positive or negative bias were present, the percentages would shift. However, in general the frequency should
be higher near the mean of the values and should decline as the concentrations move away from the mean, with approximately
half of the non-mean data above and half below the mean.
(1) Blank data are considered suspect if: (1) there is no variation in these data, (2) there are an inordinate number of zero values
(and no negative values) relative to the frequencies of positive values (6.2.3 above), (3) if there is a high frequency of the lowest
value in the data set (for example, where minimum-peak-area rejection has been used) relative to the frequency of higher
concentration values, and few or no lower values, or (4) a frequency graphic does not begin to approximate a bell curve (when
there are 20 or more samples).
(2) If the distribution of the data is suspect, the literature, plus instrument-software and equipment manuals, should be
consulted. These documents can provide an understanding of: (1) the theory of operation of the detection system, (2) the signal
processing, calibration, etc., and (3) other aspects of the conversion of response to reported values. Judgment will be needed to
determine whether to use some or all of the true-concentration-zero (blank) data, or to exclude the data from the calculations. In
general, if less than 10 % of the zero-concentration data are: (1) censored, (2) suspect, or (3) false-zeros, then these “problem” data
should be removed. Only the remaining blank data are used in the WQE calculations; there must be at least six replicates. Where
the zero concentration is excluded or is not possible to obtain, it is important to include a true concentration as close as possible
to zero in the study design.
(3) Where 75 % or less of the data are censored or smoothed, and there are at least six remaining values, it is reasonable to
use statistical procedures to simulate the distribution that is missing or smoothed. Software procedures are commercially available.
Additionally, procedures such as log-normal transformation may be used to accommodate data that are not normally distributed.
The presence of zero-concentration in the study data and in the WQE is not as critical as inclusion of such data in the WDE
calculations. Therefore, the decision about inclusion or exclusion of zero-concentration data in a WQE data set should weigh: (1)
D7783 − 21
the number of other concentrations available, (2) the range of the other concentrations, and (3) the risk of extrapolation of the WQE
outside the data-set concentration range against the quality of the zero-concentration data.
6.2.3.2 True Concentrations Near Zero—As with concentration zero, true concentrations very near to zero may also have been
censored, smoothed, and contain false-zeros. Examination of these very low concentrations, as above for zero concentration, is
important. The likelihood of occurrence and the percentage of data affected decreases with increasing concentration.
6.3 Data Screening, Outlier Identification, and Outlier Removal:
6.3.1 Data that are to be the input to the WQE calculation should be screened for compliance with this practice’s conditions,
appropriateness for the intended use of the WDE,WQE, obvious errors, and individual outliers. Graphing of the data (true versus
measured) is recommended as an assistive visual tool. This graphic is available in the DQCALC software.
6.3.2 Outlying individual measurements must be evaluated; if determined to be erroneous, they should be eliminated using
scientifically-based reasoning. Identification of potential outliers for data evaluation and validation may be accomplished using
statistical procedures, such as the optional one provided in the DQCALC software, procedures or through visual examination of
a graphical representation of the data. WQE computations must be based on retained data from at least six independent
measurements at each of at least five concentration levels. The data removed and the percentage of data removed must be recorded
and retained to document the WQE calculations.
6.4 Modeling Standard Deviation versusVersus True Concentration—The purpose is to characterizepredict the
intralaboratorywithin-laboratory measurement standard deviation (ILSD)(WLSD) as a function of true concentration, σ = σˆ(T).
G (T). The relationship is used for two purposes: (1) to provide weights (if needed) for fitting the mean-recovery model and (2)
to provide the within-laboratory standard deviation estimates crucial to determining the WQEs.
NOTE 3—See Caulcutt and Boddy (5) for more discussion of standard deviation modeling and weighted least squares (WLS) in analytical chemistry.
6.4.1 This practice utilizesuses four models as potential fits for the IntraLaboratoryWithin-Laboratory Standard Deviation
(ILSD)(WLSD) model. The identificationselection process considers (that is, fits and evaluates) each model in turn, from simplest
to most complex, until a suitable model is found.a linear model first, performing statistical tests to decide whether a simpler
constant model can be used or whether one of the more complicated curved models is required. See Carroll and Ruppert (6) for
further discussion of standard-deviation modeling. The model order is models are as follows:
Model A Constant ILSD Model :□s 5 g1error (1)
~ !
Model C ~Constant WLSD Model!:□s 5 g1error (1)
where:
g = a fitted constant.
s = standard deviation of measurement results, and
g = model parameter.
Under Model A,C, standard deviation does not change with concentration, resulting in a relative standard deviation that declines
with increasing concentration, T.
Model B ~Straight 2 line ILSD Model!:□s 5 g1h 3T1error (2)
Model L ~Linear WLSD Model!:□s 5 g1h 3T1error (2)
where:
g and h = fitted constants.
s = standard deviation of measurement results,
T = true concentration, and
g and h = model parameters.
Under Model B,L, standard deviation increases linearly with concentration, resulting in an asymptotically constant relative
standard deviation as T increases.
2 2 1⁄2
Model C Hybrid ILSD Model :□s 5 g 1 h 3 T 1error (3)
~ ! $ ~ ! %
D7783 − 21
2 2
Model H Hybrid WLSD Model :□s 5=g 1 h 3 T 1error (3)
~ ! ~ !
where:
g and h = fitted constants.
s = standard deviation of measurement results,
T = true concentration, and
g and h = model parameters.
Under Model D,H, within-laboratory standard deviation increases with concentration in such a way that the relative standard
deviation declines as T increases, approaching an asymptote of h.
Model D Exponential ILSD Model :□s 5 g 3exp□ h 3T 1error (4)
~ ! ~ !
Model E ~Exponential WLSD Model!:□s 5 g 3exp~h 3 T!1error (4)
where:
g and h = fitted constants.
s = standard deviation of measurement results,
T = true concentration, and
g and h = model parameters.
Under Model D,E, within-laboratory standard deviation increases exponentially with concentration, resulting in a relative
standard deviation that may initially decline as T increases, but eventually increases aswith T, increases.so that there is at most a
bounded quantitation range within which the RSD is less than Z ⁄ 100.
6.4.1.1 The procedures for estimating the parameters of each model and their uncertainties are given in Appendix X2 – Appendix
X5. Model L should be tried first, since its fitting procedure (in Appendix X2) provides criteria for choosing among the various
models.
6.4.1.2 In all cases, it is assumed that g > 0. A value of g < 0 has no practical interpretation and may indicate that a different
ILSDWLSD model should be used. Furthermore, it is assumed that g is not underestimated by censored data among measurements
of blanks or other low-concentration samples. If h < 0, it must not be statistically significant, and Model AC should be evaluated.
6.4.2 The ASTM D19 Practice If D7510 describes the DQCALC software that can be used to perform the calculations for each
of the four models, as well as the fit of each (this product can be obtained by contacting ASTM and asking for the DQCALC
adjunct). The software identifies which model produced the best fit, and allows the user to select either this model or an alternative
model. The software provides various graphical representations of the data and residuals, and the user manual provides assistance
in using and interpreting the graphics and calculated values. Evaluation of the fit of each model to the data (as well as knowledge
of chemistry, the method, and the systems used to generate the data) and judgment are important when selecting the most
appropriate model. Where a model other than the best fit is chosen, the reason for the choice should be scientific and should be
recorded to document the WQE.
6.4.2.1 Users of this practice not using the ASTM D19 DQCALC software can consult Practice D6091, which contains a protocol
that provides the full procedural, consensus-balloted basis for these calculations. It is also recommended that those not using the
software graph the relationship of true concentration to measurement standard deviation, and visually verify It is recommended
that the relationship of measurement standard deviation to true concentration be graphed and used for visual verification of the
appropriateness of each model and of the model selected for use.
6.5 Fitting the Mean-Recovery Relationship (Measured versus True Concentration)—Based on the standard-deviation model
selected (constant versus other models), the mean-recovery concentration is fitted versus true concentration, using ordinary least
squares or weighted least squares, respectively. The mean-recovery as appropriate. The mean recovery is evaluated for statistical
significance and lack of fit. A graph of mean recovery (along with the “calibration” line) and a graph of the residuals should also
be visually examined. The ASTM D19 DQCALC software performs these activities automatically. Alternatively, many Many
off-the-shelf statistical software packages may also be used. be used for this purpose.
NOTE 4—Regression coefficients should not be used to assess goodness of fit.
6.5.1 The mean-recovery regression (true(measured versus measuredtrue concentration) model is a simple straight line,
D7783 − 21
Model R:□Y 5 a1b T1error (5)
The fitting procedure depends on the standard-deviation-model selection. chosen standard-deviation model. If the constant
model, Model A,C, was selected, then ordinary least squares (OLS) can be used to fit Model R for mean recovery (see the left
column of recovery. In all other cases, Table 1, or Caulcutt and Boddy (5)). If a non-constant standard-deviation model was
selected, then weighted least squares (WLS) should be used to fit mean recovery. The used. WLS approximately provides the
minimum-variance unbiased linear estimate of the coefficients,parameters, a and b. The WLS procedure is described in
theAppendix X6 IDE Practice D6091.
6.6 Compute the WQE for each Z (%RSD)—Using the mean-recovery regression line determined above, the most appropriate
model of the relationship of relative standard deviation to true concentration (also determined above), and the Z value desired, the
user obtains the WQE, which is the an estimate of the lowest true concentration (corresponding to the measured concentration)
at which the desired % RSD was achieved.is achieved. Procedures for calculating the WQE and estimating its relative standard
uncertainty are given in Appendix X2 – Appendix X5.
6.6.1 Given the standard-deviation model, its estimated parameters, and the mean-recovery regression line, there is a lower limit,
Z , below which WQE cannot be calculated because there is no true concentration at which RSD equals Z %. For Z > Z ,
lim Z % lim
the WQE can be calculated but its true value may nevertheless be extremely uncertain, especially when Z is near Z . If the relative
lim
standard uncertainty, u (WQE), exceeds 25 %, the calculated WQE should be considered an unreliable estimate. The actual
rel
reliability of the WQE also depends on the adequacy of the standard-deviation model, which is more difficult to quantify.
NOTE 5—Under Model C, Z is zero. For the other models described here, Z is always positive.
lim lim
6.6.2 The measured concentration (YQ)(Y ) at which the desired % RSD RSD was achieved may also be of interest for some
Q
uses. This value is the level at which the required % RSD RSD was obtained in measured concentration units (that is, the value,
paired with a WQE, that has not been corrected for bias through the mean-recovery regression). Where the YQY and the WQE
Q
are equal (following application of significant figures and rounding), there is no apparent bias present at the WQE concentration.
6.6.2 The WQE is the lowest true concentration at which (based on the modeling of standard deviation at that concentration and
including the required confidence for the sample size (90% tolerance interval)) the percent relative standard deviation is achieved
at the desired Z. The DQCALC adjunct software calculates the 10 %, 20 %, and 30 % WQE as the typical Z values.
6.6.2.1 Fig. 1 provides an example that demonstrates a case with positive bias (intercept greater than zero) and imperfect recovery
(slope of the calibration not equal to one), thereby highlighting the advantages of the WDE procedure. More simplistic quantitation
procedures often make inappropriate assumptions about slope (that is, assume it to be one) and y-intercept (that is, assume it to
be zero at a true concentration of zero), in addition to assuming that the standard deviation is constant. Additionally, where the
simplest model (constant) for standard deviation is rejected, the WDE procedure requires that weighted least squares be used for
fitting the recovery model, thus preventing higher concentrations from having an excessive effect on the resulting curve; most other
practices do not offer this protection.
7. Review, Documentation and Reporting
7.1 The WQE analysis report should include: (1) the identification of laboratory and (2) identification of analytical method,
analyte(s), matrix (or matrices), sample properties (for example, volume or mass) and specific method options (if any) utilized.
Where the laboratory uses standard operating procedures (SOPs) to implement methods or method protocols, these SOPs should
be referenced, including the identification of any revision/version. Documentation of each datum used should be equivalent to that
of reported data (for example, instrument, analyst, date, etc.). There should be a description of all data-screening procedures
employed, all results obtained, all individual values omitted from further analysis (that is, outliers that have been removed), all
missing values, and the percentage of data utilized in the calculations relative to the initial data set. Any anomalies encountered
should be listed, including and anomalous calibration or quality control sample results (for example, data validation qualifiers or
flags). The data (statistical) analysis should be included or referenced (for example, the output file from the DQCALC software)
and the WQE values determined recorded. The selected standard-deviation model, plus the coefficient estimates for this model and
for mean-recovery model, should also be recorded. Where a statistical model other than the mathematical best fit has been chosen,
the reasoning should be described.
8. Report
8.1 The analysis report should at a miminumminimum contain:
D7783 − 21
FIG. 1 Sample Standard Deviations (+) Versus True Concentration, with Straight-LineLinear Fit, Hybrid Model Fit, and Residuals from
Straight-LineLinear Fit (Lower Plot), All in ppb
8.1.1 Identification of laboratory,
8.1.2 Analytical method,
8.1.3 Analyte(s),
8.1.4 Matrix (or matrices),
8.1.5 Sample properties (for example, volume),
8.1.6 Study design,
8.1.7 Analyst, method, and date of testing for each study sample,
8.1.8 Any anomalies in the study, including QA/QC sample results,
8.1.9 Data-screening results, individual values and laboratories omitted from further analysis, and missing values,
8.1.10 ILSDWLSD model selected, and
8.1.11 Coefficient estimates for the ILSDWLSD model and mean-recovery model.
D7783 − 21
NOTE 5—The DQCALC input and output files provide much of this documentation.
8.2 The report should be given a second-party review to verify that:
8.2.1 The data transcription and reporting have been performed correctly,
8.2.2 The analysis of the data and the application of this standard have been performed correctly, and
8.2.3 The results of the analysis have been used appropriately, including assessment of assumptions necessary to compute a WQE.
D7783 − 21
NOTE 6—Reviewer(s) should be qualified in one or bothmore of the following areas: (1) applied statistics, and (2) metrology, and (3) analytical chemistry.
8.3 A statement of the review and the results of the review should accompany the report.
9. Rationale
9.1 The basic rationale for the WQE is contained in Currie (1). The WQE is a performance characteristic of an analytical method,
measurement process, to paraphrase Currie. As with the Within-Laboraotory Detection Estimate (WDE), the Like an estimated
detection limit, the WQE is helpful for the planning and use of chemical analyses. The WQE is another benchmark indicating
whether the method can adequately meet measurement needs.
9.2 The idealized definition of WQE is that it is the lowest concentration, LQ,L , that satisfies: TL = (100/Z)(100 / Z) σ
Z% % Q Q L
Q
ϛ (where ϛσ is the actual standard deviation of interlaboratorywithin-laboratory measurements at concentration TL ); this
T L T Q
Q
definition is equivalent to satisfying, % satisfying RSD = σ ϛ ⁄ L / = TZ = Z %. In other words, WQE is the lowest
L TQ Z% %
Q
concentration with Z %Z % RSD (assuming such a concentration exists). If, as is commonly the case, % RSD RSD declines with
increasing true concentration, then the relative uncertainty of any measurement of a true concentration greater than the WQE will
not exceed 6Z %.6Z %. The range, 63ϛ63 σ , is an approximate prediction or confidence interval very likely to contain the
LQ L
Q
measurement, which is assumed to be normally distributed. This assertion is based on critical values from the normal distribution
(or from the student’s t distribution -distribution if ϛσ is estimated rather than known). Then, with high confidence, the relative error
of any measurement of a true concentration greater than the WQE will not exceed 6363 × Z · Z %. %. For example, a
measurement above the WQE (and assumed to have true concentration above the WQE) could be reported as 6 ppb (630 %)
10 %
= 6 (62) (6 6 2) ppb, with a high degree of certainty.
9.3 There are several real-world complications to this idealized situation. See Maddalone et al. (7), Gibbons (8), and Coleman et
al. (9). Some of these complications are listed as follows:
9.3.1 Analyte recovery is not perfect; the relationship between measured values of concentrations and true concentrations cannot
be assumed to be trivial. There is bias between true and measured values. Recovery can and should be modeled. Usually, a straight
line will suffice. In practice, when both the standard deviation and recovery models are used, WQEZ % is calcul
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...