Abstract

Develop a product standard for Ultra Low Power Active Medical Implants ( ULP-AMI) operating in the frequency range of 401-402MHz & 405-406 MHz.

Status
Published
Publication Date
15-Jul-2008
Technical Committee
MOC - Mobile Communications
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
10-Jun-2008
Due Date
15-Aug-2008
Completion Date
16-Jul-2008
Mandate
M/284

Buy Documents

Standard

ETSI EN 302 537-2 V1.1.2 (2007-12) - Electromagnetic compatibility and Radio spectrum Matters (ERM); Short Range Devices (SRD); Ultra Low Power Medical Data Service Systems operating in the frequency range 401 MHz to 402 MHz and 405 MHz to 406 MHz; Part 2: Harmonized EN covering essential requirements of article 3.2 of the R&TTE Directive

English language (20 pages)
sale 15% off
Preview
sale 15% off
Preview
Standard

SIST EN 302 537-2 V1.1.2:2008

English language (20 pages)
Preview
Preview
e-Library read for
×1 day

SIST EN 302 537-2 V1.1.2:2008 is the SIST version of the ETSI Harmonized European Standard for Ultra Low Power Medical Data Service Systems in the 401 MHz to 402 MHz and 405 MHz to 406 MHz bands. It specifies the radio spectrum and electromagnetic compatibility requirements used to show conformity with article 3.2 of the R&TTE Directive. The document is aimed at manufacturers, test laboratories, and compliance teams working on MEDS equipment, including ULP-AMI devices and related peripherals.

What does SIST EN 302 537-2 V1.1.2:2008 specify?

SIST EN 302 537-2 V1.1.2:2008 specifies the essential radio requirements for MEDS radio equipment operating in the two medical bands. It is a harmonized product standard for ultra low power transmitters and related equipment, and it is organized to support presumption of conformity with article 3.2 of the R&TTE Directive.

The standard is structured around scope, references, definitions and abbreviations, technical requirements, testing, and three annexes.

AnnexWhat it covers
Annex AHS Requirements and conformance Test specifications Table (HS-RTT)
Annex BEN title in the official languages
Annex CBibliography

Clause 4 sets the technical requirements, clause 5 sets the compliance tests, and Annex A ties the requirements to the test specifications used for conformity work.

What are the key requirements of SIST EN 302 537-2 V1.1.2:2008?

SIST EN 302 537-2 V1.1.2:2008 focuses on radio performance, spectrum access, and testability. In practice, the standard tells you what the device must do, how it must behave under the declared operating conditions, and which test suite is used to check each requirement.

Environmental and mechanical design

Clause 4.1 requires compliance within the declared operational environmental profile from the user manual and the provider’s declaration. The provider also declares that interruption of the communications link will not compromise patient health and safety. Clause 4.2.1 requires sound engineering practice, an antenna arrangement suited to the service, restricted access to controls that could worsen interference, and a transmitter shut-off feature that can be disabled for testing if needed.

Radio performance limits

Clauses 4.2.2 to 4.2.7 cover frequency error, emission bandwidth, effective radiated power of the fundamental emission, spurious emissions, frequency stability under low voltage conditions, and spurious radiation of receivers. Each of these clauses points back to ETSI EN 302 537-1 for the detailed definition and limit, and clause 5.3 names the corresponding conformance test. In practice, this means the radio section must stay on frequency, fit within its allowed bandwidth, stay within its power limit, and avoid unwanted emissions that could disturb other users.

Spectrum access method

Clause 4.2.8 requires the provider to declare a spectrum access method. The standard allows either LBT/AFA or LP/LDC, and both may be used if the total emission bandwidth does not exceed 100 kHz. For LBT/AFA, the system follows the monitoring-based access requirements of EN 302 537-1 clause 10. For LP/LDC, the device follows the duty-cycle, repetitive transmission, and power limits referenced from EN 302 537-1.

Compliance testing

Clause 5 requires tests to be run within the declared environmental profile, using representative conditions where performance changes with environment. It also requires measurement uncertainty to be recorded in the test report and calculated in line with TR 100 028. For buyers and quality teams, this matters because the test report must show not only the measured values but also the uncertainty used to judge them.

What terms does SIST EN 302 537-2 V1.1.2:2008 define?

  • MEDS - Medical Data Service, a system used to transfer non-time-critical medical data for diagnosis, therapy, and patient control functions.
  • ULP-AMI - Ultra Low Power-Active Medical Implants, the radio part of medical devices regulated under the Active Implantable Medical Device Directive.
  • ULP-AMI-P - peripherals associated with ULP-AMI systems.
  • ULP-AMD - the radio part of medical devices regulated under the Medical Device Directive.
  • LBT/AFA - listen before talk / adaptive frequency agility, a spectrum access method based on monitoring before transmission.
  • LP/LDC - low power / low duty cycle, a spectrum access method controlled by duty-cycle, repetition, and power limits.
  • e.r.p. - effective radiated power, the radiated power measure used for fundamental emission limits.
  • HS-RTT - the Harmonized Standard Requirements and Conformance Test Specifications Table in Annex A.

Who uses SIST EN 302 537-2 V1.1.2:2008?

SIST EN 302 537-2 V1.1.2:2008 is used by manufacturers of ultra low power medical radio equipment, especially ULP-AMI and related MEDS devices. Test laboratories use it to select the right radio test suite and report measurement uncertainty. Regulatory and compliance staff use it to prepare conformity evidence for article 3.2, while quality managers and buyers use it to check that the intended device class, spectrum access method, and test approach fit the product.

Which standards are used with SIST EN 302 537-2 V1.1.2:2008?

The main companion standard is ETSI EN 302 537-1, which provides the technical characteristics, definitions, limits, and detailed test methods that this part references throughout. ETSI TR 100 028 contributes the method for estimating measurement uncertainty in the test report. CEPT/ERC/REC 70-03 is cited as the SRD usage reference for the band and application context.

What does the SIST EN 302 537-2 V1.1.2:2008 document contain?

The document contains a clause-by-clause set of essential radio requirements and the matching test specifications for MEDS equipment in the two medical bands. Annex A maps each essential requirement to its clause and test procedure and identifies which tests are essential for conformity. Clause 5 includes the test conditions, measurement uncertainty handling, and the essential radio test suites for frequency error, emission bandwidth, power, spurious emissions, low-voltage stability, receiver spurious radiation, and the two spectrum access methods. Annex B lists the EN title in multiple official languages, and Annex C gives a short bibliography.

Buy Documents

Standard

ETSI EN 302 537-2 V1.1.2 (2007-12) - Electromagnetic compatibility and Radio spectrum Matters (ERM); Short Range Devices (SRD); Ultra Low Power Medical Data Service Systems operating in the frequency range 401 MHz to 402 MHz and 405 MHz to 406 MHz; Part 2: Harmonized EN covering essential requirements of article 3.2 of the R&TTE Directive

English language (20 pages)
sale 15% off
Preview
sale 15% off
Preview
Standard

SIST EN 302 537-2 V1.1.2:2008

English language (20 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

ANCE

Mexican certification and testing association.

EMA Mexico Verified

Intertek Slovenia

Intertek testing, inspection, and certification services in Slovenia.

UKAS Slovenia Verified

LNE (Laboratoire National de Métrologie et d'Essais)

French national laboratory for metrology and testing.

COFRAC France Verified

Sponsored listings

Frequently Asked Questions

SIST EN 302 537-2 V1.1.2:2008 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Electromagnetic compatibility and Radio spectrum Matters (ERM) - Short Range Devices (SRD) - Ultra Low Power Medical Data Service Systems operating in the frequency range 401 MHz to 402 MHz and 405 MHz to 406 MHz - Part 2: Harmonized EN covering essential requirements of article 3.2 of the R&TTE Directive". This standard covers: Develop a product standard for Ultra Low Power Active Medical Implants ( ULP-AMI) operating in the frequency range of 401-402MHz & 405-406 MHz.

Develop a product standard for Ultra Low Power Active Medical Implants ( ULP-AMI) operating in the frequency range of 401-402MHz & 405-406 MHz.

SIST EN 302 537-2 V1.1.2:2008 is classified under the following ICS (International Classification for Standards) categories: 33.060.99 - Other equipment for radiocommunications; 33.100.01 - Electromagnetic compatibility in general. The ICS classification helps identify the subject area and facilitates finding related standards.

SIST EN 302 537-2 V1.1.2:2008 is associated with the following European legislation: EU Directives/Regulations: 1999/5/EC, TRRTP123; Standardization Mandates: M/284. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.

SIST EN 302 537-2 V1.1.2:2008 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


Harmonized European Standard (Telecommunications series)

Electromagnetic compatibility
and Radio spectrum Matters (ERM);
Short Range Devices (SRD);
Ultra Low Power Medical Data Service Systems
operating in the frequency range
401 MHz to 402 MHz and 405 MHz to 406 MHz;
Part 2: Harmonized EN covering essential requirements
of article 3.2 of the R&TTE Directive

2 ETSI EN 302 537-2 V1.1.2 (2007-12)

Reference
DEN/ERM-TG30-005-2
Keywords
health, regulation, SRD
ETSI
650 Route des Lucioles
F-06921 Sophia Antipolis Cedex - FRANCE

Tel.: +33 4 92 94 42 00  Fax: +33 4 93 65 47 16

Siret N° 348 623 562 00017 - NAF 742 C
Association à but non lucratif enregistrée à la
Sous-Préfecture de Grasse (06) N° 7803/88

Important notice
Individual copies of the present document can be downloaded from:
http://www.etsi.org
The present document may be made available in more than one electronic version or in print. In any case of existing or
perceived difference in contents between such versions, the reference version is the Portable Document Format (PDF).
In case of dispute, the reference shall be the printing on ETSI printers of the PDF version kept on a specific network drive
within ETSI Secretariat.
Users of the present document should be aware that the document may be subject to revision or change of status.
Information on the current status of this and other ETSI documents is available at
http://portal.etsi.org/tb/status/status.asp
If you find errors in the present document, please send your comment to one of the following services:
http://portal.etsi.org/chaircor/ETSI_support.asp
Copyright Notification
No part may be reproduced except as authorized by written permission.
The copyright and the foregoing restriction extend to reproduction in all media.

© European Telecommunications Standards Institute 2007.
All rights reserved.
TM TM TM TM
DECT , PLUGTESTS , UMTS , TIPHON , the TIPHON logo and the ETSI logo are Trade Marks of ETSI registered
for the benefit of its Members.
TM
3GPP is a Trade Mark of ETSI registered for the benefit of its Members and of the 3GPP Organizational Partners.
ETSI
3 ETSI EN 302 537-2 V1.1.2 (2007-12)
Contents
Intellectual Property Rights.5
Foreword.5
1 Scope.6
2 References.7
2.1 Normative references.7
2.2 Informative references.7
3 Definitions and abbreviations.8
3.1 Definitions.8
3.2 Abbreviations.8
4 Technical requirements and specifications.8
4.1 Environmental profile.8
4.2 Conformance requirements.8
4.2.1 Mechanical and electrical design .8
4.2.1.1 General.8
4.2.1.2 Antennas.8
4.2.1.3 Controls.8
4.2.1.4 Transmitter shut-off facility.8
4.2.2 Frequency error.8
4.2.2.1 Definition.8
4.2.2.2 Limits.9
4.2.2.3 Conformance.9
4.2.3 Emission bandwidth.9
4.2.3.1 Definition.9
4.2.3.2 Limits.9
4.2.3.3 Conformance.9
4.2.4 Effective radiated power of the fundamental emission.9
4.2.4.1 Definition.9
4.2.4.2 Limits.9
4.2.4.3 Conformance.9
4.2.5 Spurious emissions.9
4.2.5.1 Definition.9
4.2.5.2 Limits.9
4.2.5.3 Conformance.9
4.2.6 Frequency stability under low voltage conditions .10
4.2.6.1 Definition.10
4.2.6.2 Limits.10
4.2.6.3 Conformance.10
4.2.7 Spurious radiation of receivers .10
4.2.7.1 Definition.10
4.2.7.2 Limits.10
4.2.7.3 Conformance.10
4.2.8 Spectrum access.10
4.2.8.1 LBT/AFA spectrum access.10
4.2.8.1.1 Definition.10
4.2.8.1.2 Limits.10
4.2.8.1.3 Conformance .10
4.2.8.2 Low Power Low Duty Cycle spectrum access.11
4.2.8.2.1 Definition.11
4.2.8.2.2 Limits for low duty cycle low power spectrum access .11
4.2.8.2.3 Conformance .11
5 Testing for compliance with technical requirements.11
5.1 Environmental conditions for testing .11
5.2 Interpretation of the measurement results .11
ETSI
4 ETSI EN 302 537-2 V1.1.2 (2007-12)
5.3 Essential radio test suites.12
5.3.1 Frequency error.12
5.3.2 Emission bandwidth.12
5.3.3 Effective radiated power of the fundamental emission.12
5.3.4 Spurious emissions.12
5.3.5 Frequency stability under low voltage conditions .12
5.3.6 Spurious radiation of receivers .12
5.3.7 Spectrum access for systems using LBT/AFA.13
5.3.8 Spectrum access for devices using low power and low duty cycle.13
5.3.9 Normal and extreme test-conditions .13
5.3.10 Test power source .13
5.3.11 Choice of samples for test suites.13
Annex A (normative): HS Requirements and conformance Test specifications Table (HS-
RTT).14
Annex B (informative): The EN title in the official languages .17
Annex C (informative): Bibliography.19
History .20

ETSI
5 ETSI EN 302 537-2 V1.1.2 (2007-12)
Intellectual Property Rights
IPRs essential or potentially essential to the present document may have been declared to ETSI. The information
pertaining to these essential IPRs, if any, is publicly available for ETSI members and non-members, and can be found
in ETSI SR 000 314: "Intellectual Property Rights (IPRs); Essential, or potentially Essential, IPRs notified to ETSI in
respect of ETSI standards", which is available from the ETSI Secretariat. Latest updates are available on the ETSI Web
server (http://webapp.etsi.org/IPR/home.asp).
Pursuant to the ETSI IPR Policy, no investigation, including IPR searches, has been carried out by ETSI. No guarantee
can be given as to the existence of other IPRs not referenced in ETSI SR 000 314 (or the updates on the ETSI Web
server) which are, or may be, or may become, essential to the present document.
Foreword
This Harmonized European Standard (Telecommunications series) has been produced by ETSI Technical Committee
Electromagnetic compatibility and Radio spectrum Matters (ERM).
The present document has been produced by ETSI in response to a mandate from the European Commission issued
under Council Directive 98/34/EC [3] (as amended) laying down a procedure for the provision of information in the
field of technical standards and regulations.
The present document is intended to become a Harmonized Standard, the reference of which will be published in the
Official Journal of the European Communities referencing the Directive 1999/5/EC of the European Parliament and of
the Council of 9 March 1999 on radio equipment and telecommunications terminal equipment and the mutual
recognition of their conformity (R&TTE Directive) [6].
Technical specifications relevant to Directive 1999/5/EC are given in annex A.
The present document is part 2 of a multi-part deliverable covering radio equipment in the frequency range 401 MHz to
402 MHz and 405 MHz to 406 MHz for Ultra Low Power Active Medical Devices and implants. This includes Body
Worn, Hand-Held, Data systems, etc., the medical section of which is regulated under the Medical Device Directive [4]
and Active Medical Implants and Peripherals the medical section of which is regulated under the Active Implantable
Medical Device Directive [4], as identified below:
• Part 1: "Technical characteristics and test methods";
• Part 2: "Harmonized EN covering essential requirements of article 3.2 of the R&TTE Directive".

National transposition dates
Date of adoption of this EN: 14 December 2007
Date of latest announcement of this EN (doa): 31 March 2008
Date of latest publication of new National Standard
or endorsement of this EN (dop/e): 30 September 2008
Date of withdrawal of any conflicting National Standard (dow): 30 September 2009

ETSI
6 ETSI EN 302 537-2 V1.1.2 (2007-12)
1 Scope
The present document covers various individual devices which when operating together form a system operating as a
Medical Data Service (MEDS) that provides medical practitioners with therapeutic and/or diagnostic information used
to provide improved medical treatment of a patient and/or to provide an interactive system for patient control of
therapeutic devices. MEDS is intended only for transmission of non-time critical data, the loss of which will not
compromise the health and/or safety of the patient.
Devices covered by the present document are an evolving new technology to be made available worldwide by the
medical equipment industry that will provide high speed communications capability between devices associated with an
individual patient that are part of a complete MEDS system as defined in clause 3.1. Examples of MEDS devices falling
under the scope of the present document are portable body worn physiological sensors that allow ambulatory
monitoring, implanted devices and external system devices used to transfer data collected by a MEDS system to
medical practitioners that will use the data to diagnose and treat a patient.
These devices utilize ultra low power radio transmitters in combination with medical devices, the medical portion of
which is regulated by the Medical Device Directive [4] (MDD) or the Active Implantable Medical Device Directive
(AIMD [7]). The radio part of medical devices regulated by the MDD is hereafter referred to as ULP-AMD,
ULP-AMD-P for peripheral devices, and ULP-BWD for body worn devices. ULP-BWD are devices, such as a
physiological parameter sensor or handheld devices that are intended to operate in very close proximity to the human
body, including touching the body, whose radio antenna is external to the body and is used to communicate with a
device that is part of a MEDS system. The radio part of medical devices regulated under the AIMD is hereafter referred
to as Ultra Low Power-Active Medical Implants (ULP-AMI) and peripherals (ULP-AMI-P) used in a Medical Data
Service (MEDS).
The present document contains required characteristics considered necessary for the radio sections to meet in order to
efficiently use the available spectrum for the purpose of transferring data that is used in diagnosing and delivering
therapies to individuals with various illnesses. Of particular importance is the inclusion of spectrum monitoring and
access requirements (listen before talk protocol) designed to significantly reduce any interference potential between
MEDS systems operating in the band or between a MEDS system and the primary users of the band.
The present document is a specific product standard applicable to ultra low power transmitters that are part of a system
used in the MEDS operating in spectrum within the frequency bands 401 MHz to 402 MHz and 405 MHz to 406 MHz.
The present document contains the technical characteristics for ultra low power radio equipment and is addressed by
CEPT/ERC/REC 70-03 [5] and annex 12 to that document. It does not necessarily include all the characteristics, which
may be required by a user, nor does it necessarily represent the optimum performance achievable.
It applies to ultra low power systems and accessories operating in spectrum within the bands 401 MHz to 402 MHz and
405 MHz to 406 MHz that operate in the MEDS service:
• for telecommand and telemetry between any devices that are part of a MEDS (see definition of MEDS);
• with or without an integral antenna; and/or
• with an antenna connection provided only for the purpose of connecting an external dedicated antenna.
In addition to the present document, other ENs that specify technical requirements in respect of essential requirements
under other parts of article 3 of the R&TTE Directive [6] will apply to equipment within the scope of the present
document.
ETSI
7 ETSI EN 302 537-2 V1.1.2 (2007-12)
2 References
References are either specific (identified by date of publication and/or edition number or version number) or
non-specific.
• For a specific reference, subsequent revisions do not apply.
• Non-specific reference may be made only to a complete document or a part thereof and only in the following
cases:
- if it is accepted that it will be possible to use all future changes of the referenced document for the
purposes of the referring document;
- for informative references.
Referenced documents which are not found to be publicly available in the expected location might be found at
http://docbox.etsi.org/Reference.
For online referenced documents, information sufficient to identify and locate the source shall be provided. Preferably,
the primary source of the referenced document should be cited, in order to ensure traceability. Furthermore, the
reference should, as far as possible, remain valid for the expected life of the document. The reference shall include the
method of access to the referenced document and the full network address, with the same punctuation and use of upper
case and lower case letters.
NOTE: While any hyperlinks included in this clause were valid at the time of publication ETSI cannot guarantee
their long term validity.
2.1 Normative references
The following referenced documents are indispensable for the application of the present document. For dated
references, only the edition cited applies. For non-specific references, the latest edition of the referenced document
(including any amendments) applies.
[1] ETSI EN 302 537-1 (V1.2.1): "Electromagnetic compatibility and Radio spectrum Matters (ERM);
Short Range Devices (SRD); Ultra Low Power Medical Data Service Systems operating in the
frequency range 401 MHz to 402 MHz and 405 MHz to 406 MHz; Part 1: Technical
characteristics and test methods".
[2] ETSI TR 100 028 (V1.4.1): "ElectroMagnetic Compatibility and Radio Spectrum Matters (ERM);
Uncertainties in the measurement of mobile radio equipment characteristics".
2.2 Informative references
[3] Directive 98/34/EC of the European Parliament and of the Council of 22 June 1998 laying down a
procedure for the provision of information in the field of technical standards and regulations.
[4] Council Directive 93/42/EEC of 14 June 1993 concerning medical devices (MD Directive).
[5] CEPT/ERC/REC 70-03 (2006): "Relating to the use of Short Range Devices (SRD)".
[6] Directive 1999/5/EC of the European Parliament and of the Council of 9 March 1999 on radio
equipment and telecommunications terminal equipment and the mutual recognition of their
conformity (R&TTE Directive).
[7] Council Directive 90/385/EEC of 20 June 1990 on the approximation of the laws of the Member
States relating to active implantable medical devices.
ETSI
8 ETSI EN 302 537-2 V1.1.2 (2007-12)
3 Definitions and abbreviations
3.1 Definitions
For the purposes of the present document, the terms and definitions given in EN 302 537-1 [1], clause 3.1 apply.
3.2 Abbreviations
For the purposes of the present document, the abbreviations given in EN 302 537-1 [1], clause 3.3 apply.
4 Technical requirements and specifications
4.1 Environmental profile
The technical requirements of the present document apply under the environmental profile for operation of the
equipment as described in the user's manual and declared by the provider. The equipment shall comply with all the
technical requirements of the present document at all times when operating within the boundary limits of the operational
environmental profile as described above. The provider shall declare that interruption of the communications link for
his MEDS system shall not result in compromising the health and safety of the patient.
4.2 Conformance requirements
4.2.1 Mechanical and electrical design
4.2.1.1 General
The equipment shall be designed, constructed and manufactured in accordance with sound engineering practice and
with the aim of minimizing
...


SLOVENSKI STANDARD
01-september-2008
(OHNWURPDJQHWQD]GUXåOMLYRVWLQ]DGHYHY]YH]L]UDGLMVNLPVSHNWURP (50 
1DSUDYHNUDWNHJDGRVHJD 65' 6OXåEHQLVLVWHPLXOWUDPDMKQLKPRþL]D
PHGLFLQVNHSRGDWNHNLGHOXMHMRYIUHNYHQþQLKREPRþMLKRG0+]GR0+]LQ
RG0+]GR0+]GHO+DUPRQL]LUDQL(1NL]DMHPDELVWYHQH]DKWHYH
þOHQDGLUHNWLYH5 77(
Electromagnetic compatibility and Radio spectrum Matters (ERM) - Short Range Devices
(SRD) - Ultra Low Power Medical Data Service Systems operating in the frequency
range 401 MHz to 402 MHz and 405 MHz to 406 MHz - Part 2: Harmonized EN covering
essential requirements of article 3.2 of the R&TTE Directive
Ta slovenski standard je istoveten z: EN 302 537-2 Version 1.1.2
ICS:
33.060.99 Druga oprema za radijske Other equipment for
komunikacije radiocommunications
33.100.01 Elektromagnetna združljivost Electromagnetic compatibility
na splošno in general
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

Harmonized European Standard (Telecommunications series)

Electromagnetic compatibility
and Radio spectrum Matters (ERM);
Short Range Devices (SRD);
Ultra Low Power Medical Data Service Systems
operating in the frequency range
401 MHz to 402 MHz and 405 MHz to 406 MHz;
Part 2: Harmonized EN covering essential requirements
of article 3.2 of the R&TTE Directive

2 ETSI EN 302 537-2 V1.1.2 (2007-12)

Reference
DEN/ERM-TG30-005-2
Keywords
health, regulation, SRD
ETSI
650 Route des Lucioles
F-06921 Sophia Antipolis Cedex - FRANCE

Tel.: +33 4 92 94 42 00  Fax: +33 4 93 65 47 16

Siret N° 348 623 562 00017 - NAF 742 C
Association à but non lucratif enregistrée à la
Sous-Préfecture de Grasse (06) N° 7803/88

Important notice
Individual copies of the present document can be downloaded from:
http://www.etsi.org
The present document may be made available in more than one electronic version or in print. In any case of existing or
perceived difference in contents between such versions, the reference version is the Portable Document Format (PDF).
In case of dispute, the reference shall be the printing on ETSI printers of the PDF version kept on a specific network drive
within ETSI Secretariat.
Users of the present document should be aware that the document may be subject to revision or change of status.
Information on the current status of this and other ETSI documents is available at
http://portal.etsi.org/tb/status/status.asp
If you find errors in the present document, please send your comment to one of the following services:
http://portal.etsi.org/chaircor/ETSI_support.asp
Copyright Notification
No part may be reproduced except as authorized by written permission.
The copyright and the foregoing restriction extend to reproduction in all media.

© European Telecommunications Standards Institute 2007.
All rights reserved.
TM TM TM TM
DECT , PLUGTESTS , UMTS , TIPHON , the TIPHON logo and the ETSI logo are Trade Marks of ETSI registered
for the benefit of its Members.
TM
3GPP is a Trade Mark of ETSI registered for the benefit of its Members and of the 3GPP Organizational Partners.
ETSI
3 ETSI EN 302 537-2 V1.1.2 (2007-12)
Contents
Intellectual Property Rights.5
Foreword.5
1 Scope.6
2 References.7
2.1 Normative references.7
2.2 Informative references.7
3 Definitions and abbreviations.8
3.1 Definitions.8
3.2 Abbreviations.8
4 Technical requirements and specifications.8
4.1 Environmental profile.8
4.2 Conformance requirements.8
4.2.1 Mechanical and electrical design .8
4.2.1.1 General.8
4.2.1.2 Antennas.8
4.2.1.3 Controls.8
4.2.1.4 Transmitter shut-off facility.8
4.2.2 Frequency error.8
4.2.2.1 Definition.8
4.2.2.2 Limits.9
4.2.2.3 Conformance.9
4.2.3 Emission bandwidth.9
4.2.3.1 Definition.9
4.2.3.2 Limits.9
4.2.3.3 Conformance.9
4.2.4 Effective radiated power of the fundamental emission.9
4.2.4.1 Definition.9
4.2.4.2 Limits.9
4.2.4.3 Conformance.9
4.2.5 Spurious emissions.9
4.2.5.1 Definition.9
4.2.5.2 Limits.9
4.2.5.3 Conformance.9
4.2.6 Frequency stability under low voltage conditions .10
4.2.6.1 Definition.10
4.2.6.2 Limits.10
4.2.6.3 Conformance.10
4.2.7 Spurious radiation of receivers .10
4.2.7.1 Definition.10
4.2.7.2 Limits.10
4.2.7.3 Conformance.10
4.2.8 Spectrum access.10
4.2.8.1 LBT/AFA spectrum access.10
4.2.8.1.1 Definition.10
4.2.8.1.2 Limits.10
4.2.8.1.3 Conformance .10
4.2.8.2 Low Power Low Duty Cycle spectrum access.11
4.2.8.2.1 Definition.11
4.2.8.2.2 Limits for low duty cycle low power spectrum access .11
4.2.8.2.3 Conformance .11
5 Testing for compliance with technical requirements.11
5.1 Environmental conditions for testing .11
5.2 Interpretation of the measurement results .11
ETSI
4 ETSI EN 302 537-2 V1.1.2 (2007-12)
5.3 Essential radio test suites.12
5.3.1 Frequency error.12
5.3.2 Emission bandwidth.12
5.3.3 Effective radiated power of the fundamental emission.12
5.3.4 Spurious emissions.12
5.3.5 Frequency stability under low voltage conditions .12
5.3.6 Spurious radiation of receivers .12
5.3.7 Spectrum access for systems using LBT/AFA.13
5.3.8 Spectrum access for devices using low power and low duty cycle.13
5.3.9 Normal and extreme test-conditions .13
5.3.10 Test power source .13
5.3.11 Choice of samples for test suites.13
Annex A (normative): HS Requirements and conformance Test specifications Table (HS-
RTT).14
Annex B (informative): The EN title in the official languages .17
Annex C (informative): Bibliography.19
History .20

ETSI
5 ETSI EN 302 537-2 V1.1.2 (2007-12)
Intellectual Property Rights
IPRs essential or potentially essential to the present document may have been declared to ETSI. The information
pertaining to these essential IPRs, if any, is publicly available for ETSI members and non-members, and can be found
in ETSI SR 000 314: "Intellectual Property Rights (IPRs); Essential, or potentially Essential, IPRs notified to ETSI in
respect of ETSI standards", which is available from the ETSI Secretariat. Latest updates are available on the ETSI Web
server (http://webapp.etsi.org/IPR/home.asp).
Pursuant to the ETSI IPR Policy, no investigation, including IPR searches, has been carried out by ETSI. No guarantee
can be given as to the existence of other IPRs not referenced in ETSI SR 000 314 (or the updates on the ETSI Web
server) which are, or may be, or may become, essential to the present document.
Foreword
This Harmonized European Standard (Telecommunications series) has been produced by ETSI Technical Committee
Electromagnetic compatibility and Radio spectrum Matters (ERM).
The present document has been produced by ETSI in response to a mandate from the European Commission issued
under Council Directive 98/34/EC [3] (as amended) laying down a procedure for the provision of information in the
field of technical standards and regulations.
The present document is intended to become a Harmonized Standard, the reference of which will be published in the
Official Journal of the European Communities referencing the Directive 1999/5/EC of the European Parliament and of
the Council of 9 March 1999 on radio equipment and telecommunications terminal equipment and the mutual
recognition of their conformity (R&TTE Directive) [6].
Technical specifications relevant to Directive 1999/5/EC are given in annex A.
The present document is part 2 of a multi-part deliverable covering radio equipment in the frequency range 401 MHz to
402 MHz and 405 MHz to 406 MHz for Ultra Low Power Active Medical Devices and implants. This includes Body
Worn, Hand-Held, Data systems, etc., the medical section of which is regulated under the Medical Device Directive [4]
and Active Medical Implants and Peripherals the medical section of which is regulated under the Active Implantable
Medical Device Directive [4], as identified below:
• Part 1: "Technical characteristics and test methods";
• Part 2: "Harmonized EN covering essential requirements of article 3.2 of the R&TTE Directive".

National transposition dates
Date of adoption of this EN: 14 December 2007
Date of latest announcement of this EN (doa): 31 March 2008
Date of latest publication of new National Standard
or endorsement of this EN (dop/e): 30 September 2008
Date of withdrawal of any conflicting National Standard (dow): 30 September 2009

ETSI
6 ETSI EN 302 537-2 V1.1.2 (2007-12)
1 Scope
The present document covers various individual devices which when operating together form a system operating as a
Medical Data Service (MEDS) that provides medical practitioners with therapeutic and/or diagnostic information used
to provide improved medical treatment of a patient and/or to provide an interactive system for patient control of
therapeutic devices. MEDS is intended only for transmission of non-time critical data, the loss of which will not
compromise the health and/or safety of the patient.
Devices covered by the present document are an evolving new technology to be made available worldwide by the
medical equipment industry that will provide high speed communications capability between devices associated with an
individual patient that are part of a complete MEDS system as defined in clause 3.1. Examples of MEDS devices falling
under the scope of the present document are portable body worn physiological sensors that allow ambulatory
monitoring, implanted devices and external system devices used to transfer data collected by a MEDS system to
medical practitioners that will use the data to diagnose and treat a patient.
These devices utilize ultra low power radio transmitters in combination with medical devices, the medical portion of
which is regulated by the Medical Device Directive [4] (MDD) or the Active Implantable Medical Device Directive
(AIMD [7]). The radio part of medical devices regulated by the MDD is hereafter referred to as ULP-AMD,
ULP-AMD-P for peripheral devices, and ULP-BWD for body worn devices. ULP-BWD are devices, such as a
physiological parameter sensor or handheld devices that are intended to operate in very close proximity to the human
body, including touching the body, whose radio antenna is external to the body and is used to communicate with a
device that is part of a MEDS system. The radio part of medical devices regulated under the AIMD is hereafter referred
to as Ultra Low Power-Active Medical Implants (ULP-AMI) and peripherals (ULP-AMI-P) used in a Medical Data
Service (MEDS).
The present document contains required characteristics considered necessary for the radio sections to meet in order to
efficiently use the available spectrum for the purpose of transferring data that is used in diagnosing and delivering
therapies to individuals with various illnesses. Of particular importance is the inclusion of spectrum monitoring and
access requirements (listen before talk protocol) designed to significantly reduce any interference potential between
MEDS systems operating in the band or between a MEDS system and the primary users of the band.
The present document is a specific product standard applicable to ultra low power transmitters that are part of a system
used in the MEDS operating in spectrum within the frequency bands 401 MHz to 402 MHz and 405 MHz to 406 MHz.
The present document contains the technical characteristics for ultra low power radio equipment and is addressed by
CEPT/ERC/REC 70-03 [5] and annex 12 to that document. It does not necessarily include all the characteristics, which
may be required by a user, nor does it necessarily represent the optimum performance achievable.
It applies to ultra low power systems and accessories operating in spectrum within the bands 401 MHz to 402 MHz and
405 MHz to 406 MHz that operate in the MEDS service:
• for telecommand and telemetry between any devices that are part of a MEDS (see definition of MEDS);
• with or without an integral antenna; and/or
• with an antenna connection provided only for the purpose of connecting an external dedicated antenna.
In addition to the present document, other ENs that specify technical requirements in respect of essential requirements
under other parts of article 3 of the R&TTE Directive [6] will apply to equipment within the scope of the present
document.
ETSI
7 ETSI EN 302 537-2 V1.1.2 (2007-12)
2 References
References are either specific (identified by date of publication and/or edition number or version number) or
non-specific.
• For a specific reference, subsequent revisions do not apply.
• Non-specific reference may be made only to a complete document or a part thereof and only in the following
cases:
- if it is accepted that it will be possible to use all future changes of the referenced document for the
purposes of the referring document;
- for informative references.
Referenced documents which are not found to be publicly available in the expected location might be found at
http://docbox.etsi.org/Reference.
For online referenced documents, information sufficient to identify and locate the source shall be provided. Preferably,
the primary source of the referenced document should be cited, in order to ensure traceability. Furthermore, the
reference should, as far as possible, remain valid for the expected life of the document. The reference shall include the
method of access to the referenced document and the full network address, with the same punctuation and use of upper
case and lower case letters.
NOTE: While any hyperlinks included in this clause were valid at the time of publication ETSI cannot guarantee
their long term validity.
2.1 Normative references
The following referenced documents are indispensable for the application of the present document. For dated
references, only the edition cited applies. For non-specific references, the latest edition of the referenced document
(including any amendments) applies.
[1] ETSI EN 302 537-1 (V1.2.1): "Electromagnetic compatibility and Radio spectrum Matters (ERM);
Short Range Devices (SRD); Ultra Low Power Medical Data Service Systems operating in the
frequency range 401 MHz to 402 MHz and 405 MHz to 406 MHz; Part 1: Technical
characteristics and test methods".
[2] ETSI TR 100 028 (V1.4.1): "ElectroMagnetic Compatibility and Radio Spectrum Matters (ERM);
Uncertainties in the measurement of mobile radio equipment characteristics".
2.2 Informative references
[3] Directive 98/34/EC of the European Parliament and of the Council of 22 June 1998 laying down a
procedure for the provision of information in the field of technical standards and regulations.
[4] Council Directive 93/42/EEC of 14 June 1993 concerning medical devices (MD Directive).
[5] CEPT/ERC/REC 70-03 (2006): "Relating to the use of Short Range Devices (SRD)".
[6] Directive 1999/5/EC of the European Parliament and of the Council of 9 March 1999 on radio
equipment and telecommunications terminal equipment and the mutual recognition of their
conformity (R&TTE Directive).
[7] Council Directive 90/385/EEC of 20 June 1990 on the approximation of the laws of the Member
States relating to active implantable medical devices.
ETSI
8 ETSI EN 302 537-2 V1.1.2 (2007-12)
3 Definitions and abbreviations
3.1 Definitions
For the purposes of the present document, the terms and definitions given in EN 302 537-1 [1], clause 3.1 apply.
3.2 Abbreviations
For the purposes of the present document, the abbreviations given in EN 302 537-1 [1], clause 3.3 apply.
4 Technical requirements and specifications
4.1 Environmental profile
The technical requirements of the present document apply under the environmental profile for operation of the
equipment as described in the user's manual and declared by the provider. The equipment shall comply with all the
tec
...