ISO 80601-2-61:2017
(Main)Medical electrical equipment — Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
Medical electrical equipment — Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
ISO 80601-2-61:2017 applies to the basic safety and essential performance of pulse oximeter equipment intended for use on humans, hereafter referred to as me equipment. This includes any part necessary for normal use, including the pulse oximeter monitor, pulse oximeter probe, and probe cable extender. These requirements also apply to pulse oximeter equipment, including pulse oximeter monitors, pulse oximeter probes and probe cable extenders, which have been reprocessed. The intended use of pulse oximeter equipment includes, but is not limited to, the estimation of arterial oxygen haemoglobin saturation and pulse rate of patients in professional healthcare institutions as well as patients in the home healthcare environment and the emergency medical services environment. ISO 80601-2-61:2017 is not applicable to pulse oximeter equipment intended for use in laboratory research applications nor to oximeters that require a blood sample from the patient. If a clause or subclause is specifically intended to be applicable to me equipment only, or to me systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to me equipment and to me systems, as relevant. Hazards inherent in the intended physiological function of me equipment or me systems within the scope of this document are not covered by specific requirements in this document except in 201.11 and in 7.2.13 and 8.4.1 of the general standard. NOTE 1 See also 4.2 of the general standard. "The general standard" is IEC 60601-1:2005+AMD1:2012, Medical electrical equipment ? Part 1: General requirements for basic safety and essential performance. ISO 80601-2-61:2017 can also be applied to me equipment and their accessories used for compensation or alleviation of disease, injury or disability. ISO 80601-2-61:2017 is not applicable to pulse oximeter equipment intended solely for foetal use. ISO 80601-2-61:2017 is not applicable to remote or slave (secondary) equipment that displays SpO2 values that are located outside of the patient environment. NOTE 2 Me equipment that provides selection between diagnostic and monitoring functions is expected to meet the requirements of the appropriate document when configured for that function. ISO 80601-2-61:2017 is applicable to pulse oximeter equipment intended for use under extreme or uncontrolled environmental conditions outside the hospital environment or physician's office, such as in ambulances and air transport. Additional standards can apply pulse oximeter equipment for those environments of use. ISO 80601-2-61:2017 is a particular standard in the IEC 60601-1 and ISO/IEC 80601 series of standards.
Appareils électromédicaux — Partie 2-61: Exigences particulières pour la sécurité de base et les performances essentielles pour les oxymètres de pouls
ISO 80601-2-61:2017 s'applique à la sécurité de base et aux performances essentielles des oxymètres de pouls conçus pour une utilisation chez les êtres humains, ci-après dénommés appareils em. Sont inclus tous les éléments de l'appareil nécessaires pour une utilisation normale, y compris le moniteur de l'oxymètre de pouls, le capteur de l'oxymètre de pouls et le câble de raccordement du capteur. Ces exigences s'appliquent de la même manière aux oxymètres de pouls, y compris aux moniteurs d'oxymètre de pouls, aux capteurs d'oxymètre de pouls et aux câbles de raccordement du capteur, qui ont été remis en état. L'utilisation prévue des oxymètres de pouls inclut notamment, sans toutefois s'y limiter, l'estimation de la saturation en oxygène de l'hémoglobine artérielle et de la fréquence du pouls chez des patients en établissement de soins professionnel ainsi que chez les patients dans l'environnement des soins à domicile et dans l'environnement des services médicaux d'urgence. ISO 80601-2-61:2017 ne s'applique pas aux oxymètres de pouls destinés à être utilisés dans le cadre de recherches en laboratoire, de même qu'aux oxymètres nécessitant un échantillon de sang du patient. Si un article ou un paragraphe est spécifiquement destiné à être applicable uniquement aux appareils em ou uniquement aux systèmes em, le titre et le contenu de cet article ou de ce paragraphe l'indiquent. Si cela n'est pas le cas, l'article ou le paragraphe s'applique à la fois aux appareils em et aux systèmes em, selon le cas. Les dangers inhérents à la fonction physiologique prévue de l'appareil em ou des systèmes em dans le cadre du domaine d'application du présent document ne sont pas couverts par des exigences spécifiques contenues dans le présent document, à l'exception de 201.11 et de 7.2.13 et 8.4.1 de la norme générale. NOTE 1 Voir également 4.2 de la norme générale. La norme générale est l'IEC 60601-1:2005+AMD1:2012, Appareils électromédicaux ? Partie 1: Exigences générales pour la sécurité de base et les performances essentielles. ISO 80601-2-61:2017 peut également être appliqué aux appareils em et à leurs accessoires utilisés pour l'atténuation d'une maladie, la compensation ou l'atténuation d'une blessure ou d'une incapacité. ISO 80601-2-61:2017 ne s'applique pas aux oxymètres de pouls destinés uniquement à la surveillance de f?tus. ISO 80601-2-61:2017 ne s'applique pas aux appareils distants ou esclaves (secondaires) qui affichent les valeurs de la SpO2 et qui sont situés à l'extérieur de l'environnement du patient. NOTE 2 Il est attendu que les appareils em qui permettent une sélection entre des fonctions de diagnostic et de surveillance satisfassent aux exigences du document approprié lorsqu'ils sont configurés pour ladite fonction. ISO 80601-2-61:2017 est applicable aux oxymètres de pouls destinés à être utilisés dans des conditions environnementales extrêmes ou non maîtrisées à l'extérieur de l'environnement hospitalier ou d'un cabinet de médecin, telles que dans des ambulances ou dans les transports aériens. Des normes supplémentaires peuvent s'appliquer aux oxymètres de pouls pour ces environnements d'utilisation. ISO 80601-2-61:2017 est une norme particulière des séries de normes de l'IEC 60601-1 et de l'ISO/IEC 80601.
General Information
Relations
Buy Standard
Standards Content (Sample)
INTERNATIONAL ISO
STANDARD 80601-2-61
Second edition
2017-12
Corrected version
2018-02
Medical electrical equipment —
Part 2-61:
Particular requirements for basic
safety and essential performance of
pulse oximeter equipment
Appareils électromédicaux —
Partie 2-61: Exigences particulières pour la sécurité de base et les
performances essentielles pour les oxymètres de pouls
Reference number
ISO 80601-2-61:2017(E)
©
 ISO 2017
---------------------- Page: 1 ----------------------
ISO 80601-2-61:2017(E)
COPYRIGHT PROTECTED DOCUMENT
© ISO 2017
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting
on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address
below or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva, Switzerland
Tel. +41 22 749 01 11
Fax +41 22 749 09 47
copyright@iso.org
www.iso.org
Published in Switzerland
© ISO 2017 – All rights reserved
ii
---------------------- Page: 2 ----------------------
ISO 80601-2-61:2017(E)
Contents Page
Foreword . v
Introduction . vi
201. 1	 Scope, object and related standards . 1
201. 2	 Normative references . 3
201. 3	 Terms and definitions . 4
201. 4	 General requirements . 9
201. 5	 General requirements for testing of ME EQUIPMENT . 10
201. 6	 Classification of ME EQUIPMENT and ME SYSTEMS . 10
201. 7	 ME EQUIPMENT identification, marking and documents . 11
201. 8	 Protection against electrical HAZARDS from ME EQUIPMENT . 15
201. 9	 Protection against mechanical HAZARDS of ME EQUIPMENT and ME SYSTEMS . 15
201. 10	Protection against unwanted and excessive radiation HAZARDS . 15
201. 11	Protection against excessive temperatures and other HAZARDS . 15
201. 12	Accuracy of controls and instruments and protection against hazardous outputs . 17
201. 13	HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT . 21
201. 14	PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 21
201. 15	Construction of ME EQUIPMENT . 21
201. 16	ME SYSTEMS . 23
201. 17	Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 23
201. 101	 * PULSE OXIMETER PROBES and PROBE CABLE EXTENDERS . 23
201. 102	 Saturation pulse INFORMATION SIGNAL . 23
201. 103	 FUNCTIONAL CONNECTION . 24
202	 Electromagnetic disturbances – Requirements and tests . 24
206	 Usability . 25
208	 General requirements, tests and guidance for alarm systems in medical electrical equipment
and medical electrical systems . 26
211	 Requirements for medical electrical equipment and medical electrical systems used in the
home healthcare environment . 26
212	 Requirements for medical electrical equipment and medical electrical systems used in the
emergency medical services environment . 27
Annex C (informative) Guide to marking and labelling requirements for ME EQUIPMENT and ME SYSTEMS . 28
Annex D (informative) Symbols on marking . 31
Annex AA (informative) Particular guidance and rationale . 32
Annex BB (informative) Skin temperature at the PULSE OXIMETER PROBE . 40
ANNEX CC (informative) Determination of ACCURACY . 44
Annex DD (informative) Calibration standards. 53
Annex EE (informative) Guideline for evaluating and documenting SpO ACCURACY in human subjects . 54
2
Annex FF (informative) Simulators, calibrators and FUNCTIONAL TESTERS for PULSE OXIMETER EQUIPMENT . 61
©	ISO	2017	–	All	rights	reserved iii
---------------------- Page: 3 ----------------------
ISO 80601-2-61:2017(E)
Annex GG (informative) Concepts of ME EQUIPMENT response time . 70
Annex HH (normative) Data interface requirements . 74
Annex II (informative) Reference to the ESSENTIAL PRINCIPLES . 78
Annex JJ (informative) Terminology — alphabetized index of defined terms . 82
Bibliography . 87
iv	 ©	ISO	2017	–	All	rights	reserved
---------------------- Page: 4 ----------------------
ISO 80601-2-61:2017(E)
Foreword
ISO	(the	International	Organization	for	Standardization)	is	a	worldwide	federation	of	national	standards	bodies
(ISO	member	bodies).	The	work	of	preparing	International	Standards	is	normally	carried	out	through	ISO
technical	committees.	Each	member	body	interested	in	a	subject	for	which	a	technical	committee	has	been
established	has	the	right	to	be	represented	on	that	committee.	International	organizations,	governmental	and
non‐governmental,	in	liaison	with	ISO,	also	take	part	in	the	work.	ISO	collaborates	closely	with	the	International
Electrotechnical	Commission	(IEC)	on	all	matters	of	electrotechnical	standardization.
The	procedures	used	to	develop	this	document	and	those	intended	for	its	further	maintenance	are	described	in
the	ISO/IEC	Directives,	Part	1.	In	particular	the	different	approval	criteria	needed	for	the	different	types	of	ISO
documents	 should	 be	 noted.	 This	 document	 was	 drafted	 in	 accordance	with	the	editorial	rules	of	the
ISO/IEC	Directives,	Part	2	(see	www.iso.org/directives).
Attention	is	drawn	to	the	possibility	that	some	of	the	elements	of	this	document	may	be	the	subject	of	patent
rights.	ISO	shall	not	be	held	responsible	for	identifying	any	or	all	such	patent	rights.	Details	of	any	patent	rights
identified	during	the	development	of	the	document	will	be	in	the	Introduction	and/or	on	the	ISO	list	of	patent
declarations	received	(see	www.iso.org/patents).
trade	name	used	in	this	document	is	information	given	for	the	convenience	of	users	and	does	not	constitute
Any
an	endorsement.
For	an	explanation	on	the	voluntary	nature	of	standards,	the	meaning	of	ISO	specific	terms	and	expressions
related	to	conformity	assessment,	as	well	as	information	about	ISO's	adherence	to	the	World	Trade	Organization
(WTO)	principles	in	the	Technical	Barriers	to	Trade	(TBT)	see	the	following	URL:	www.iso.org/iso/foreword.html.
This	document	was	prepared	jointly	by	ISO/TC	121,	 Anaesthetic and respiratory equipment,	Subcommittee	SC	3,
Lung ventilators and related equipment,	and	Technical	Committee	IEC/TC	62,	 Electrical equipment in medical
practice,	Subcommittee	SC	D,	Electrical equipment.	The	draft	was	circulated	for	voting	to	the	national	bodies	of
both	ISO	and	IEC.
This	second	edition	of	ISO	80601‐2‐61	cancels	and	replaces	the	first	edition(ISO	80601‐2‐61:2011),	which	has
been	technically	revised.	It	includes	an	alignment	with	Amendment	1	of	both	the	third	edition	of	IEC	60601‐1	and
the	second	edition	of	IEC	60601‐1‐8,	as	well	as	the	fourth	edition	of	IEC	60601‐1‐2,	the	third	edition	of	IEC	60601‐
1‐6,	the	second	edition	of	IEC	60601‐1‐11	and	IEC	60601‐1‐12.
The	most	significant	changes	are	the	following	modifications:
– updated	rationale	(Annex	AA)	and	references	related	to	advances	in	the	understanding	of	hypoxaemia,
electronic	health	records	and	ALARM	SYSTEMS;
– ingress	protection	changed	from	IPX1	to	IPX2;
and	the	following	additions:
– Clause	211,	requirements	for	use	in	the	HOME	HEALTHCARE	ENVIRONMENT;
– Clause	212,	requirements	for	use	in	the	emergency	medical	services	(EMS)	environment;
– Annex	HH,	Data	interface	requirements.
This	corrected	version	of	ISO	80601‐2‐61:2017	incorporates	the	following	correction:
– headers	have	been	corrected.
A	list	of	all	the	parts	of	the	ISO/IEC	80601	series	is	available	on	the	ISO	website.
©	ISO	2017	–	All	rights	reserved v
---------------------- Page: 5 ----------------------
ISO 80601-2-61:2017(E)
Introduction
The	approximation	of	arterial	haemoglobin	saturation	and	pulse	rate	using	pulse	oximetry	is	common	practice	in
many	areas	of	medicine.	This	document	covers	BASIC	SAFETY	and	ESSENTIAL	PERFORMANCE	requirements	achievable
within	the	limits	of	existing	technology.
The	committees	recognized	the	need	to	revise	the	first	edition	of	this	document	because	of	the	publication	of	the
first	edition	of	IEC	60601‐1‐12,	as	well	as	the	fourth	edition	of	IEC	60601‐1‐2,	the	second	edition	of	IEC	60601‐1‐
11	and	the	first	Amendments	to	both	the	third	edition	of	IEC	60601‐1,	the	third	edition	of	IEC	60601‐1‐6	and	the
second	edition	of	IEC	60601‐1‐8.
Annex	AA	contains	a	rationale	for	some	of	the	requirements.	It	is	included	to	provide	additional	insight	into	the
reasoning	of	the	committees	that	led	to	a	requirement	and	identifying	the	HAZARDS	that	the	requirement	addresses.
Annex	BB	is	a	literature	survey	relevant	to	the	determination	of	the	maximum	safe	temperature	of	the	interface
between	a	PULSE	OXIMETER	PROBE	and	a	PATIENT'S	tissue.
Annex	CC	 discusses	 both	 the	 formulae	 used	 to	 evaluate	 the	 SpO 	 ACCURACY	of	 PULSE	 OXIMETER	 EQUIPMENT
2
measurements,	and	the	names	that	are	assigned	to	those	formulae.
Annex	DD	presents	guidance	on	when	in	vitro	blood	calibration	of	PULSE	OXIMETER	EQUIPMENT	is	needed.
Annex	EE	presents	a	guideline	for	a	CONTROLLED	DESATURATION	STUDY	for	the	calibration	of	PULSE	OXIMETER	EQUIPMENT.
Annex	FF	is	a	tutorial	introduction	to	several	kinds	of	testers	used	in	pulse	oximetry.
Annex	GG	describes	concepts	of	PULSE	OXIMETER	EQUIPMENT	response	time.
Annex	HH	describes	data	interface	requirements.
Annex	II	contains	Reference	to	the	ESSENTIAL	PRINCIPLES	formerly	found	in	Annex	HH.
This	publication	has	been	drafted	in	accordance	with	the	ISO/IEC	Directives,	Part	2.
In	this	document,	the	following	print	types	are	used:
– requirements	and	definitions:	roman	type;
– test specifications: italic type;
– informative	material	appearing	outside	of	tables,	such	as	notes,	examples	and	references:	in	smaller	type;
normative	text	of	tables	is	also	in	a	smaller	type;
1
– TERMS	DEFINED	IN	CLAUSE	3	OF	THE	GENERAL	STANDARD 	IN	THIS	DOCUMENT	OR	AS	NOTED:	SMALL	CAPITALS.
In	referring	to	the	structure	of	this	document,	the	term
– “clause”	means	one	of	the	six	numbered	divisions	within	the	table	of	contents,	inclusive	of	all	subdivisions	(e.g.
Clause	201	includes	subclauses	201.7.1,	201.7.2)	and
– “subclause”	means	a	numbered	subdivision	of	a	clause	(e.g.	201.7.1,	7.2	and	201.7.2.1	are	all	subclauses	of
Clause	201.7).
1
 The	general	standard	is	IEC	60601‐1:2005	and	IEC	60601‐1:2005/AMD1:2012,	Medical	electrical	equipment	–	Part	1:
General	requirements	for	basic	safety	and	essential	performance.
vi	 ©	ISO	2017	–	All	rights	reserved
---------------------- Page: 6 ----------------------
ISO 80601-2-61:2017(E)
References	to	clauses	within	this	document	are	preceded	by	the	term	“Clause”	followed	by	the	clause	number.
References	to	subclauses	within	this	document	are	by	number	only.
In	this	document,	the	conjunctive	“or”	is	used	as	an	“inclusive	or”	so	a	statement	is	true	if	any	combination	of	the
conditions	is	true.
The	verbal	forms	used	in	this	document	conform	to	usage	described	in	Annex	H	of	the	ISO/IEC	Directives,	Part	2.
For	the	purposes	of	this	document,	the	auxiliary	verb:
– “shall”	means	that	compliance	with	a	requirement	or	a	test	is	mandatory	for	compliance	with	this	document;
– “should”	means	that	compliance	with	a	requirement	or	a	test	is	recommended	but	is	not	mandatory	for
compliance	with	this	document;	and
– “may”	is	used	to	describe	a	permissible	way	to	achieve	compliance	with	a	requirement	or	test.
An	asterisk	(*)	as	the	first	character	of	a	title	or	at	the	beginning	of	a	paragraph	or	table	title	indicates	that	there	is
guidance	or	rationale	related	to	that	item	in	Annex	AA.
©	ISO	2017	–	All	rights	reserved vii
---------------------- Page: 7 ----------------------
INTERNATIONAL STANDARD ISO 80601-2-61:2017(E)
Medical electrical equipment —
Part	2‐61:
Particular requirements for basic safety and essential
performance of pulse oximeter equipment
201.1 Scope, object and related standards
IEC	60601‐1:2005+AMD1:2012,	Clause	1	applies,	except	as	follows:
201.1.1 * Scope
Replacement:
This	document	applies	to	the	BASIC	SAFETY	and	ESSENTIAL	PERFORMANCE	of	PULSE	OXIMETER	EQUIPMENT	intended	for
use	on	humans,	hereafter	referred	to	as	ME	EQUIPMENT.	This	includes	any	part	necessary	for	NORMAL	USE,
including	the	PULSE	OXIMETER	MONITOR,	PULSE	OXIMETER	PROBE,	and	PROBE	CABLE	EXTENDER.
These	requirements	also	apply	to	PULSE	OXIMETER	EQUIPMENT,	including	PULSE	OXIMETER	MONITORS,	PULSE	OXIMETER
PROBES	and	PROBE	CABLE	EXTENDERS,	which	have	been	REPROCESSED.
The	intended	use	of	PULSE	OXIMETER	EQUIPMENT	includes,	but	is	not	limited	to,	the	estimation	of	arterial	oxygen
haemoglobin	saturation	and	pulse	rate	of	PATIENTS	in	professional	healthcare	institutions	as	well	as	PATIENTS
in	the	HOME	HEALTHCARE	ENVIRONMENT	and	the	EMERGENCY	MEDICAL	SERVICES	ENVIRONMENT.
This	document	is	not	applicable	to	 PULSE	OXIMETER	EQUIPMENT	intended	for	use	in	laboratory	research
applications	nor	to	oximeters	that	require	a	blood	sample	from	the	PATIENT.
If	a	clause	or	subclause	is	specifically	intended	to	be	applicable	to	ME	EQUIPMENT	only,	or	to	ME	SYSTEMS	only,
the	title	and	content	of	that	clause	or	subclause	will	say	so.	If	that	is	not	the	case,	the	clause	or	subclause
applies	both	to	ME	EQUIPMENT	and	to	ME	SYSTEMS,	as	relevant.
HAZARDS	inherent	in	the	intended	physiological	function	of	ME	EQUIPMENT	or	ME	SYSTEMS	within	the	scope	of
this	document	are	not	covered	by	specific	requirements	in	this	document	except	in	201.11	and	in	7.2.13	and
8.4.1	of	the	general	standard.
NOTE	1	 See	also	4.2	of	the	general	standard.	“The	general	standard”	is	IEC	60601‐1:2005+AMD1:2012,	 Medical
electrical equipment – Part 1: General requirements for basic safety and essential performance.
This	document	can	also	be	applied	to	ME	EQUIPMENT	and	their	ACCESSORIES	used	for	compensation	or	alleviation
of	disease,	injury	or	disability.
This	document	is	not	applicable	to	PULSE	OXIMETER	EQUIPMENT	intended	solely	for	foetal	use.
This	document	is	not	applicable	to	remote	or	slave	(secondary)	equipment	that	displays	SpO 	values	that	are
2
located	outside	of	the	PATIENT	ENVIRONMENT.
NOTE	2	 ME	EQUIPMENT	that	provides	selection	between	diagnostic	and	monitoring	functions	is	expected	to	meet	the
requirements	of	the	appropriate	document	when	configured	for	that	function.
©	ISO	2017	–	All	rights	reserved 1
---------------------- Page: 8 ----------------------
ISO 80601-2-61:2017(E)
This	document	is	applicable	to	PULSE	OXIMETER	EQUIPMENT	intended	for	use	under	extreme	or	uncontrolled
environmental	conditions	outside	the	hospital	environment	or	physician’s	office,	such	as	in	ambulances	and
air	transport.	Additional	standards	can	apply	PULSE	OXIMETER	EQUIPMENT	for	those	environments	of	use.
This	document	is	a	particular	standard	in	the	IEC	60601‐1	and	ISO/IEC	80601	series	of	standards.
201.1.2 Object
Subclause	1.2	of	the	general	standard	is	replaced	by:
The	object	of	this	document	is	to	establish	particular	BASIC	SAFETY	and	ESSENTIAL	PERFORMANCE	requirements	for
PULSE	OXIMETER	EQUIPMENT	[as	defined	in	201.3.217]	and	its	ACCESSORIES.
NOTE	 ACCESSORIES	are	included	because	the	combination	of	the	PULSE	OXIMETER	MONITOR	and	the	ACCESSORIES	needs	to
be	adequately	safe.	ACCESSORIES	can	have	a	significant	impact	on	the	BASIC	SAFETY	or	ESSENTIAL	PERFORMANCE	of	PULSE	OXIMETER
EQUIPMENT.
201.1.3 Collateral standards
IEC	60601‐1:2005+AMD1:2012,	subclause	1.3	applies	with	the	following	addition:
This	document	refers	to	those	applicable	collateral	standards	that	are	listed	in	Clause	2	of	the	general
standard	and	Clause	201.2	of	this	document.
IEC	60601‐1‐2,	IEC	60601‐1‐6,	IEC	60601‐1‐8,	IEC	60601‐1‐11	and	IEC	60601‐1‐12	apply	as	modified	in
[2]
Clauses	202,	206,	208,	211	and	212	respectively.	IEC	60601‐1‐3 	does	not	apply.	All	other	published
collateral	standards	in	the	IEC	60601‐1	series	apply	as	published.
201.1.4 Particular standards
IEC	60601‐1:2005+AMD1:2012,	subclause	1.4	is	replaced	by:
In	the	IEC	60601	series,	particular	standards	may	modify,	replace	or	delete	requirements	contained	in	the
general	standard,	including	the	collateral	standards,	as	appropriate	for	the	particular	ME	EQUIPMENT	under
consideration,	and	may	add	other	BASIC	SAFETY	or	ESSENTIAL	PERFORMANCE	requirements.
A	requirement	of	a	particular	standard	takes	priority	over	the	general	standard	or	the	collateral	standards.
For	brevity,	IEC	60601‐1:2005+AMD1:2012	is	referred	to	in	this	document	as	the	general	standard.	Collateral
standards	are	referred	to	by	their	document	number.
The	numbering	of	clauses	and	subclauses	of	this	document	corresponds	to	those	of	the	general	standard	with
the	prefix	“201”	(e.g.	201.1	in	this	document	addresses	the	content	of	Clause	1	of	the	general	standard)	or
applicable	collateral	standard	with	the	prefix	“2xx”	where	xx	is	the	final	digits	of	the	collateral	standard
document	number	(e.g.	202.4	in	this	document	addresses	the	content	of	Clause	4	of	the	IEC	60601‐1‐2
collateral	standard,	208.4	in	this	document	addresses	the	content	of	Clause	4	of	the	IEC	60601‐1‐8	collateral
standard,	etc.).	The	changes	to	the	text	of	the	general	standard	are	specified	by	the	use	of	the	following
words:
"Replacement"	means	that	the	clause	or	subclause	of	the	general	standard	or	applicable	collateral	standard	is
replaced	completely	by	the	text	of	this	document.
"Addition"	means	that	the	text	of	this	document	is	additional	to	the	requirements	of	the	general	standard	or
applicable	collateral	standard.
"Amendment"	means	that	the	clause	or	subclause	of	the	general	standard	or	applicable	collateral	standard	is
amended	as	indicated	by	the	text	of	this	document.
2 ©	ISO	2017	–	All	rights	reserved
---------------------- Page: 9 ----------------------
ISO 80601-2-61:2017(E)
Subclauses	or	figures	that	are	additional	to	those	of	the	general	standard	are	numbered	starting	from
201.101,	additional	annexes	are	lettered	AA,	BB,	etc.,	and	additional	items	aa), bb),	etc.
Subclauses	or	figures	that	are	additional	to	those	of	a	collateral	standard	are	numbered	starting	from	2xx,
where	“x”	is	the	number	of	the	collateral	standard,	e.g.	202	for	IEC	60601‐1‐2,	203	for	IEC	60601‐1‐3,	etc.
The	term	"this	document"	is	used	to	make	reference	to	the	general	standard,	any	applicable	collateral
standards	and	this	particular	document	taken	together.
Where	there	is	no	corresponding	clause	or	subclause	in	this	document,	the	section,	clause	or	subclause	of	the
general	 standard	 or	 applicable	 collateral	 standard,	 although	 possibly	not	relevant,	applies	without
modification;	where	it	is	intended	that	any	part	of	the	general	standard	or	applicable	collateral	standard,
although	possibly	relevant,	is	not	to	be	applied,	a	statement	to	that	effect	is	given	in	this	particular	document.
201.2 Normative references
The	following	documents	are	referred	to	in	the	text	in	such	a	way	that	some	or	all	of	their	content	constitutes
requirements	of	this	document.	For	dated	references,	only	the	edition	cited	applies.	For	undated	references,
the	latest	edition	of	the	referenced	document	(including	any	amendments)	applies.
NOTE	1	 The	way	in	which	these	referenced	documents	are	cited	in	normative	requirements	determines	the	extent	(in
whole	or	in	part)	to	which	they	apply.
NOTE	2	 Informative	references	are	listed	in	the	Bibliography.
IEC	60601‐1:2005+AMD1:2012,	Clause	2	applies,	except	as	follows:
Replacement:
IEC	60417,	Graphical symbols for use on equipment
IEC	60529:2013,	Degrees of protection provided by enclosures (IP code)
IEC	60601‐1‐2:2014,	 Medical electrical equipment — Part 1-2: General requirements for basic safety and
essential performance — Collateral Standard: Electromagnetic disturbances — Requirements and tests
IEC	60601‐1‐6:2010,	 Medical electrical equipment — Part 1-6: General requirements for basic safety and
essential performance — Collateral standard: Usability
+	Amendment	1:2013
IEC	60601‐1‐8:2006,	 Medical electrical equipment — Part 1-8: General requirements for basic safety and
essential performance — Collateral Standard: General requirements, tests and guidance for alarm systems in
medical electrical equipment and medical electrical systems
+	Amendment	1:2012
IEC	60825‐1:2014,	Safety of laser products - Part 1: Equipment classification and requirements
Addition:
ISO	3744:2010,	Acoustics — Determination of sound power levels and sound energy levels of noise sources using
sound pressure — Engineering methods for an essentially free field over a reflecting plane
ISO	4135:2001,	Anaesthetic and respiratory equipment — Vocabulary
ISO	7000,	Graphical symbols for use on equipment — Registered symbols
ISO	7010,	Graphical symbols — Safety colours and safety signs — Registered safety signs
©	ISO	2017	–	All	rights	reserved 3
---------------------- Page: 10 ----------------------
ISO 80601-2-61:2017(E)
ISO	14155:2011,	Clinical investigation of medical devices for human subjects — Good clinical practice
ISO	14937:2009,	 Sterilization of health care products — General requirements for characterization of a
sterilizing agent and the development, validation and routine control of a sterilization process for medical
devices
ISO	15223‐1:2016,	Medical devices — Symbols to be used with medical device labels, labelling and information
to be supplied — Part 1: General requirements
IEC	60068‐2‐27:2008,	 Environmental testing — Part 2-27: Tests — Test Ea and guidance: Shock
+	Amendment	1:2013
IEC	60068‐2‐31:2008,	 Environmental testing — Part 2-31: Tests — Test Ec: Rough handling shocks, primarily
for equipment-type specimens
IEC	60068‐2‐64:2008,	Environmental testing — Part 2-64: Tests — Test Fh: Vibration, broadband random and
guidance
IEC	60601‐1‐11:2015,	 Medical electrical equipment — Part 1-11: General requirements for basic safety and
essential performance — Collateral Standard: Requirements for medical electrical equipment and medical
electrical systems used in the home healthcare environment
IEC	60601‐1‐12:2014,	 Medical electrical equipment — Part 1-12: General requirements for basic safety and
essential performance — Collateral Standard: Requirements for medical electrical equipment and medical
electrical systems used in the emergency medical services environment
IEC	60825‐2:2004,	 Safety of laser products — Part 2: Safety of optical fibre communication systems (OFCS) +
Amendment	1:2006+	Amendment	2:2012
IEC	62366‐1:2015,	Medical devices — Part 1: Application of usability engineering to medical devices
201.3 Terms and definitions
For	the	purposes	of	this	document,	the	terms	and	definitions	given	in	ISO	3744:2010,	ISO	4135:2001,
IEC	60601‐1:2005+AMD	1:2012,	 IEC	60601‐1‐2:2014,	 IEC	60601‐1‐6:2010+AMD	1:2013,
IEC	60601‐1‐8:2006+AMD	1:2012,	IEC	60601‐1‐11:2015,	IEC	60601‐1‐12:2014,	IEC	62366‐1:2015	and	the
following	apply.
ISO	and	IEC	maintain	terminological	databases	for	use	in	standardization	at	the	following	addresses:
 ISO	Online	browsing	platform:	available	at	http://www.iso.org/obp
 IEC	Electropedia:	available	at	http://www.electropedia.org/
NOTE		 An	alphabetized	index	of	defined	terms	is	found	in	Annex	JJ.
201.3.201
accuracy
closeness	of	agreement	between	a	test	result	and	an	accepted	reference	value
Note	1	to	entry:	Subclause	201.12.1.101.2	provides	the	method	of	calculating	the	 SpO 	 ACCURACY	of	 PULSE	OXIMETER
2
EQUIPMENT.
Note	2	to	entry:	Additional	information	is	found	in	Annex	CC.
[3]
[SOURCE:	 ISO	3534‐2:2006 ,	3.3.1,	modified	—	Replaced	note	1	to	note	3	with	new	notes	and	 ‘or
measurement	result	and	the	true’	with	‘and	an	accepted	reference	’.]
4 ©	ISO	2017	–	All	rights	reserved
---------------------- Page: 11 ----------------------
ISO 80601-2-61:2017(E)
201.3.202
controlled desaturation study
hypoxaemia	induced	in	a	human	subject	performed	under	laboratory	conditions
Note	1	to	entry:	This	can	also	be	referred	to	as	a	controlled	hypoxaemia	(breathedown)	study.	Additional	information	is
found	in	Annex	EE.
201.3.203
co-oximeter
multiwavelength,	 optical	 blood	 analyser	 that	 measures	 TOTAL	 HAEMOGLOBIN	 CONCENTRATION	and	the
concentrations	of	various	haemoglobin	derivatives
Note	1	to	entry:	The	relevant	CO‐oximetry	value	is	functional	saturation	of	arterial	blood,	 SaO ,	which	PULSE	OXIMETER
2
EQUIPMENT	estimates	and	reports	as	SpO .
2
201.3.204
data update period
interval	in	which	the	PULSE	OXIMETER	EQUIPMENT	algorithm	provides	new	valid	data	to	the	display	or	the
FUNCTIONAL	CONNECTION
Note	1	to	entry:	This	definition	does	not	refer	to	the	regular	refresh	period	of	the	display,	which	is	typically	on	the	order
of	1	s,	but	rather	to	the	(typically	longer)	interval	defined	above.
201.3.205
declared range
portion	of	the	DISPLAYED	RANGE	of	SpO 	and	pulse	rate	values	over	which	there	is	specified	ACCURACY
2
201.3.206
displayed range
range	of	SpO 	or	pulse	rate	values	that	can	be	displayed	by	the	PULSE	OXIMETER	EQUIPMENT
2
Note	1	to	entry:	The	DISPLAYED	RANGE	can	extend	beyond	the	DECLARED	RANGE.
201.3.207
fractional oxyhaemoglobin
FO Hb fractional saturation (deprecated)
2
oxyhaemoglobin	concentration	cO Hb	divided	by	the	TOTAL	HAEMOGLOBIN	CONCENTRATION,	ctHb,	in	the	blood
2
cOHb
2
FOHb
2
ctHb
Note	1	to	entry:	cO Hb	is	the	concentration	of	oxyhaemoglobin;	ctHb	is	the	concentration	of	total	haemoglobin.
2
Note	2	to	entry:	This	is	sometimes	reported	as	a	percentage	(multiplying	the	fraction	by	100).
Note	3	to	entry:	FRACTIONAL	OXYHAEMOGLOBIN	is	the	term	used	by	the	Clinical	and	Laboratory	Standards	Institute	(CLSI),
(formerly	NCCLS	or	National	Committee	for	Clinical	Laboratory	Sciences)	for	this	ratio.
Note	4	to	entry:	CLSI	denotes	“concentration”	by	a	prefixed	letter	c,	while	in	the	past	the	convention	of	square	brackets
(e.g.	[O Hb]	)	was	used.
2
[4]
Note	5	to	entry:	CLSI 	uses	the	following	notations:
 oxyhaemoglobin	(O Hb);
2
 deoxyhaemoglobin	(HHb);
©	ISO	2017	–	All	rights	reserved 5
---------------------- Page: 12 ----------------------
ISO 80601-2-61:2017(E)
 carboxyhaemoglobin	(COHb);methaemoglobin	(MetHb);
 sulfhaemoglobin	(SulfHb);	and
[5]
 total	haemoglo
 ...
DRAFT INTERNATIONAL STANDARD
ISO/DIS 80601-2-61
ISO/TC 121/SC 3 Secretariat: ANSI
Voting begins on: Voting terminates on:
2016-10-24 2017-01-15
Medical electrical equipment —
Part 2-61:
Particular requirements for basic safety and essential
performance of pulse oximeter equipment
Appareils électromédicaux —
Partie 2-61: Exigences particulières pour la sécurité de base et les performances essentielles pour les
oxymètres de pouls
ICS: 11.040.10
This document is circulated as received from the committee secretariat.
THIS DOCUMENT IS A DRAFT CIRCULATED
This draft is submitted to a parallel vote in ISO and in IEC.
FOR COMMENT AND APPROVAL. IT IS
THEREFORE SUBJECT TO CHANGE AND MAY
NOT BE REFERRED TO AS AN INTERNATIONAL
STANDARD UNTIL PUBLISHED AS SUCH.
IN ADDITION TO THEIR EVALUATION AS
ISO/CEN PARALLEL PROCESSING
BEING ACCEPTABLE FOR INDUSTRIAL,
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
STANDARDS MAY ON OCCASION HAVE TO
BE CONSIDERED IN THE LIGHT OF THEIR
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
Reference number
NATIONAL REGULATIONS.
ISO/DIS 80601-2-61:2016(E)
RECIPIENTS OF THIS DRAFT ARE INVITED
TO SUBMIT, WITH THEIR COMMENTS,
NOTIFICATION OF ANY RELEVANT PATENT
RIGHTS OF WHICH THEY ARE AWARE AND TO
©
PROVIDE SUPPORTING DOCUMENTATION. ISO 2016
---------------------- Page: 1 ----------------------
ISO/DIS 80601-2-61:2016(E)
COPYRIGHT PROTECTED DOCUMENT
© ISO 2016, Published in Switzerland
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Ch. de Blandonnet 8 • CP 401
CH-1214 Vernier, Geneva, Switzerland
Tel. +41 22 749 01 11
Fax +41 22 749 09 47
copyright@iso.org
www.iso.org
ii © ISO 2016 – All rights reserved
---------------------- Page: 2 ----------------------
ISO/DIS 80601-2-61:2016(E) – 3 –
1 CONTENTS
2 CONTENTS .3
3 FOREWORD .5
4 INTRODUCTION .9
5 201.1 Scope, object and related standards . 11
6 201 2 Normative references. 13
7 201.3 Terminology and definitions. 15
8 201.5 General requirements for testing of ME EQUIPMENT . 21
9 201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 22
10 201.7 ME EQUIPMENT identification, marking and documents . 22
11 201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 26
12 201.9 Protection against mechanical HAZARDS of ME EQUIPMENT and ME SYSTEMS . 27
13 201.10 Protection against unwanted and excessive radiation HAZARDS . 27
14 201.11 Protection against excessive temperatures and other HAZARDS. 27
15 201.12 Accuracy of controls and instruments and protection against hazardous outputs. 29
16 201.13 HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT . 33
17 201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 34
18 201.15 Construction of ME EQUIPMENT. 34
19 201.16 ME SYSTEMS . 35
20 201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 35
21 201.101 PULSE OXIMETER PROBES and PROBE CABLE EXTENDER S . 35
22 201.102 Saturation pulse INFORMATION SIGNAL. 36
23 201.103 FUNCTIONAL CONNECTION. 36
24 202 Electromagnetic disturbances – Requirements and tests . 37
25 202.4.3.1 Configurations . 37
26 202.5.2.2.1 Requirements applicable to all ME EQUIPMENT and ME SYSTEMS . 37
27 202.8.1.101 Additional general requirements. 37
28 202.8.2 PATIENT physiological simulation . 37
29 206 Usability . 38
30 208 General requirements, tests and guidance for alarm systems
31 in medical electrical equipment and medical electrical systems . 38
32 208.6.1.2.101 Additional requirements for ALARM CONDITION priority. 38
33 208.6.5.4.101 Additional requirements for DEFAULT ALARM PRESET . 39
34 208.6.8.5.101 Additional requirements for ALARM SIGNAL inactivation states,
35 indication and access. 39
36 211 Requirements for medical electrical equipment and medical electrical systems used in
37 the home healthcare environment . 39
© ISO 2016 – All rights reserved
---------------------- Page: 3 ----------------------
– 4 – ISO/DIS 80601-2-61:2016(E)
38 212 General requirements, tests and guidance for EMS systems in medical electrical
39 equipment and medical electrical systems. 39
40 Annex C (informative) Guide to marking and labelling requirements for ME EQUIPMENT and
41 ME SYSTEMS . 40
42 Annex D (informative) Symbols on marking . 43
43 Annex AA (informative) Particular guidance and rationale . 44
44 Annex BB (informative) Skin temperature at the PULSE OXIMETER PROB E. 54
45 Annex CC (informative) Determination of ACCURACY. 59
46 Annex DD (informative) Calibration standards. 69
47 Annex EE (informative) Guideline for evaluating and documenting SpO ACCURACY in human
2
48 subjects . 71
49 Annex FF (informative) Simulators, calibrators and FUNCTIONAL TESTERS for PULSE OXIMETER
50 EQUIPMENT . 79
51 Annex GG (informative) Concepts of ME EQUIPMENT response time. 89
52 Annex HH (normative) Data interface requirements . 93
53 Annex II (informative) Reference to the Essential Principles. 97
54 Annex JJ (informative) Terminology — alphabetized index of defined terms . 101
55 Bibliography. 105
56 Annex ZA (informative) Relationship between this Document and the Essential Requirements of
57 EU Directive 93/42/EEC . 111
58
59
© ISO 2016 – All rights reserved
---------------------- Page: 4 ----------------------
ISO/DIS 80601-2-61:2016(E) – 5 –
60 INTERNATIONAL ORGANIZATION for STANDARDISATION
61 ____________
62
63 MEDICAL ELECTRICAL EQUIPMENT –
64 Part 2-61: Particular requirements for the basic safety and
65 essential performance of pulse oximetry equipment
66
67
68 FOREWORD
69 ISO (the International Organization for Standardization) is a worldwide federation of national standards
70 bodies (ISO member bodies). The work of preparing International Standards is normally carr ied out
71 through ISO technical committees. Each member body interested in a subject for which a technical
72 committee has been established has the right to be represented on that committee. International
73 organizations, governmental and non-governmental, in liaison with ISO, also take part in the work. ISO
74 collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
75 electrotechnical standardization.
76 The procedures used to develop this document and those intended for its furthe r maintenance are
77 described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
78 different types of ISO documents should be noted. This document was drafted in accordance with the
79 editorial rules of the ISO/IEC Directives, Part 2. www.iso.org/directives
80 Attention is drawn to the possibility that some of the elements of this document may be the subject of
81 patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of any
82 patent rights identified during the development of the document will be in the Introduction and/or on
83 the ISO list of patent declarations received. www.iso.org/patents
84 Any trade name used in this document is information given for the convenience of users and does not
85 constitute an endorsement.
86 ISO 80601-2-61 was prepared by Technical Committee ISO/TC 121, Anaesthetic and respiratory
87 equipment, Subcommittee SC 3, Lung ventilators and related equipment, and Technical Committee
88 IEC/TC 62, Electrical equipment in medical practice, Subcommittee SC D, Electrical equipment. The draft
89 was circulated for voting to the national bodies of both ISO and IEC.
1
[1]
90 This second edition of ISO 80601-2-61 cancels and replaces the first edition of ISO 80601-2-61 (2011).
91 This edition of ISO 80601-2-61 constitutes a technical revision and includes an alignment with
92 Amendment 1 of both the third edition of IEC 60601-1 and the second edition of IEC 60601-1-8, as well
93 as the fourth edition of IEC 60601-1-2, the third edition of IEC 60601-1-6, the second edition of
94 IEC 60601-1-11 and IEC 60601-1-12.
1
 Numbers in square brackets refer to the Bibliography.
© ISO 2016 – All rights reserved
---------------------- Page: 5 ----------------------
– 6 – ISO/DIS 80601-2-61:2016(E)
95 The most significant changes are the following modifications:
96 – Updated Rationale (Annex AA) and references related to advances in our understanding of hypoxia,
97 electronic health records and ALARM SYSTEMS;
98 and the following additions:
99 – Clause 211, requirements for use in the HOME HEALTHCARE ENVIRONMENT
100 – Clause 212, requirements for use in the emergency medical services (EMS) environment
101 – Annex HH, Data interface requirements
102
© ISO 2016 – All rights reserved
---------------------- Page: 6 ----------------------
ISO/DIS 80601-2-61:2016(E) – 7 –
103 European Foreword
104 The following referenced documents are indispensable for the application of this document. For undated
105 references, the latest edition of the referenced document (including any amendments) applies. For dated
106 references, only the edition cited applies. However, for any use of this standard “within the meaning of
107 Annex ZA", the user should always check that any referenced document has not been superseded and that
108 its relevant contents can still be considered the generally acknowledged state -of-art.
109 When the ISO or IEC standard is referred to in the ISO text standard, this must be understood as a
110 normative reference to the parallel EN standard or dated ISO standard, as outlined below, including the
111 foreword and the Annexes ZZ.
112 NOTE The way in which these references documents are cited in normative requirements determines the extent
113 (in whole or in part) to which they apply.
114 Table – Correlations between normative references and dated EN and ISO/IEC standards
Equivalent dated standard
Normative references as listed in
Clause 2
EN ISO/IEC
ISO 7000 (database) — ISO 7000 (database)
ISO 14155:2011 EN ISO 14155:2011+AC:2011 ISO 14155:2011
ISO 14937:2009 — ISO 14937:2009
2
prEN ISO 15223-1:— ISO 15223-1:—
ISO 15223-1:—
IEC 60068-2-27:2008+A1:2013 EN 60068-2-27:2009+A1:— IEC 60068-2-27:2008+A1:2013
IEC 60068-2-31:2008 EN 60068-2-31:2008 IEC 60068-2-31:2008
IEC 60068-2-64:2008 EN 60068-2-64:2008 IEC 60068-2-64:2008
IEC 60601-1:2005+A1:2012 EN 60601-1:2006+A1:2013 IEC 60601-1:2005+A1:2012
+AC:2014+A12:2014
IEC 60601-1-2:2014 EN 60601-1-2:—1 IEC 60601-1-2:2014
IEC 60601-1-6:2010 +A1:2013 EN 60601-1-6:2010+A1:—11 IEC 60601-1-6:2010 +A1:20131
IEC 60601-1-8:2006+A1:2012 EN 60601-1-8:2007 +A1:2013 IEC 60601-1-8:2006 +A1:2012
+AC:2014
IEC 60601-1-11:2015 EN 60601-1-11:2015 IEC 60601-1-11:2015
IEC 60601-1-12:2014 EN 60601-1-12:2015 IEC 60601-1-12:2014
IEC 60417 (database) — IEC 60417 (database)
IEC 60529:2013 EN 60529:— IEC 60529:2013
IEC 60825-1:2014 EN 60825-1:2014 IEC 60825-1:2014
2
 Under preparation. Stage at the time of publication: ISO FDIS 15223-1:2016.
© ISO 2016 – All rights reserved
---------------------- Page: 7 ----------------------
– 8 – ISO/DIS 80601-2-61:2016(E)
Equivalent dated standard
Normative references as listed in
Clause 2
EN ISO/IEC
IEC 60825-2:2004+A1:2006+A2:2010 EN 60825-2:2004+A1:2007 IEC 60825-
+A2:2010 2:2004+A1:2006+A2:2010
115
© ISO 2016 – All rights reserved
---------------------- Page: 8 ----------------------
ISO/DIS 80601-2-61:2016(E) – 9 –
116 INTRODUCTION
117 The approximation of arterial haemoglobin saturation and pulse rate using pulse oximetry is common
118 practice in many areas of medicine. This document covers BASIC SAFETY and ESSENTIAL PERFORMANCE
119 requirements achievable within the limits of existing technology.
120 The committees recognized the need to revise the first edition of this document due to the publication of
121 the first edition of IEC 60601-1-12, as well as the fourth edition of IEC 60601-1-2, the second edition of
122 IEC 60601-1-11 and the first Amendments to both the third edition of IEC 60601-1, the third edition of
123 IEC 60601-1-6 and the second edition of IEC 60601-1-8.
124 Annex AA contains a rationale for some of the requirements. It is included to provide additional insight
125 into the reasoning of the committee that led to a requirement and identifying the HAZARDS that the
126 requirement addresses.
127 Annex BB is a literature survey relevant to the determination of the maximum safe temperature of the
128 interface between a PULSE OXIMETER PROBE and a PATIENT'S tissue.
129 Annex CC discusses both the formulae used to evaluate the SpO ACCURACY of PULSE OXIMETER EQUIPMENT
2
130 measurements, and the names that are assigned to those formulae.
131 Annex DD presents guidance on when in vitro blood calibration of PULSE OXIMETER EQUIPMENT is needed.
132 Annex EE presents a guideline for a CONTROLLED DESATURATION STUDY for the calibration of PULSE OXIMETER
133 EQUIPMENT.
134 Annex FF is a tutorial introduction to several kinds of testers used in pulse oximetry.
135 Annex GG describes concepts of PULSE OXIMETER EQUIPMENT response time.
136 Annex HH describes data interface requirements.
137 Annex II contains Reference to the Essential Principles formerly found in Annex HH.
138 This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
139 In this document, the following print types are used:
140 – Requirements and definitions: roman type
141 – Test specifications: italic type
142 – Informative material appearing outside of tables, such as notes, examples and references: in smaller type.
143 Normative text of tables is also in a smaller type
3
144 – TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD IN THIS DOCUMENT OR AS NOTED: SMALL CAPITALS
145 In referring to the structure of this document, the term
3
 The general standard is IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, Medical electrical equipment – Part 1:
General requirements for basic safety and essential performance
© ISO 2016 – All rights reserved
---------------------- Page: 9 ----------------------
– 10 – ISO/DIS 80601-2-61:2016(E)
146 – “clause” means one of the six numbered divisions within the table of contents, inclusive of all
147 subdivisions (e.g. Clause 201includes subclauses 201.7.1, 201.7.2, etc.) and
148 – “subclause” means a numbered subdivision of a clause (e.g. 201.7.1, 7.2 and 201.7.2.1 are all
149 subclauses of Clause 201.7).
150 References to clauses within this document are preceded by the term “Clause” followed by the clause
151 number. References to subclauses within this document are by number only.
152 In this document, the conjunctive “or” is used as an “inclusive or” so a statement is true if any combination
153 of the conditions is true.
154 The verbal forms used in this document conform to usage described in Annex H of the ISO/IEC Directives,
155 Part 2. For the purposes of this document, the auxiliary verb:
156 – “shall” means that compliance with a requirement or a test is mandatory for compliance with this
157 document;
158 – “should” means that compliance with a requirement or a test is recommended but is not ma ndatory
159 for compliance with this document; and
160 – “may” is used to describe a permissible way to achieve compliance with a requirement or test.
161 An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title indicates
162 that there is guidance or rationale related to that item in Annex AA.
163 The attention of Member Bodies and National Committees is drawn to the fact that equipment
164 manufacturers and testing organizations may need a transitional period following publication of a new,
165 amended or revised ISO or IEC publication in which to make products in accordance with the new
166 requirements and to equip themselves for conducting new or revised tests. It is the recommendation of
167 the committees that the content of this document not be adopted for mandatory implementation
168 nationally earlier than 3 years from the date of publication for equipment newly designed, and not earlier
169 than 5 years from the date of publication for equipment already in production.
170
© ISO 2016 – All rights reserved
---------------------- Page: 10 ----------------------
ISO/DIS 80601-2-61:2016(E) – 11 –
171 Medical Electrical Equipment — Part 2-61: Particular
172 requirements for basic safety and essential performance of
173 pulse oximeter equipment
174 201.1 Scope, object and related standards
175 IEC 60601-1:2005+AMD1:2012, Clause 1 applies, except as follows:
176 201.1.1 * Scope
177 Replacement:
178 This document applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of PULSE OXIMETER EQUIPMENT
179 intended for use on humans, hereafter referred to as ME EQUIPMENT. This includes any part
180 necessary for NORMAL USE, including the PULSE OXIMETER MONITOR, PULSE OXIMETER PROBE, and PROBE
181 CABLE EXTENDER.
182 These requirements also apply to PULSE OXIMETER EQUIPMENT, including PULSE OXIMETER MONITORS,
183 PULSE OXIMETER PROBES and PROBE CABLE EXTENDERS, which have been REPROCESSED.
184 The intended use of PULSE OXIMETER EQUIPMENT includes, but is not limited to, the estimation of
185 arterial oxygen haemoglobin saturation and pulse rate of PATIENTS in professional healthcare
186 institutions as well as PATIENTS in the HOME HEALTHCARE ENVIRONMENT and the EMERGENCY SERVICES
187 ENVIRONMENT.
188 This document is not applicable to PULSE OXIMETER EQUIPMENT intended for use in laboratory
189 research applications nor to oximeters that require a blood sample from the PATIENT.
190 If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME
191 SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the
192 clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
193 HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the
194 scope of this document are not covered by specific requirements in this document except in
195 201.11 and in 7.2.13 and 8.4.1 of the general standard.
196 NOTE 1 See also 4.2 of the general standard.
197 This document can also be applied to ME EQUIPMENT and their ACCESSORIES used for compensation
198 or alleviation of disease, injury or disability.
199 This document is not applicable to PULSE OXIMETER EQUIPMENT intended solely for foetal use.
200 This document is not applicable to remote or slave (secondary) equipment that displays SpO
2
201 values that are located outside of the PATIENT ENVIRONMENT.
202 This document is not applicable to pulse haemoglobin monitors.
© ISO 2016 – All rights reserved
---------------------- Page: 11 ----------------------
– 12 – ISO/DIS 80601-2-61:2016(E)
203 NOTE 2 ME EQUIPMENT that provides selection between diagnostic and monitoring functions is expected to
204 meet the requirements of the appropriate document when configured for that function.
205 This document is applicable to PULSE OXIMETER EQUIPMENT intended for use under extreme or
206 uncontrolled environmental conditions outside the hospital environment or physician’s office,
207 such as in ambulances and air transport. Additional standards can apply PULSE OXIMETER EQUIPMENT
208 for those environments of use.
209 This document is a particular standard in the IEC 60601-1 and ISO/IEC 80601 series of standards.
210 201.1.2 Object
211 Subclause 1.2 of the general standard is replaced by:
212 The object of this document is to establish particular BASIC SAFETY and ESSENTIAL PERFORMANCE
213 requirements for PULSE OXIMETER EQUIPMENT [as defined in 201.3.203] and its ACCESSORIES.
214 NOTE ACCESSORIES are included because the combination of the PULSE OXIMETER MONITOR and the
215 ACCESSORIES needs to be adequately safe. ACCESSORIES can have a significant impact on the BASIC SAFETY or
216 ESSENTIAL PERFORMANCE of PULSE OXIMETER EQUIPMENT.
217 201.1.3 Collateral standards
218 IEC 60601-1:2005+AMD1:2012, subclause 1.3 applies with the following addition:
219 This document refers to those applicable collateral standards that are listed in Clause 2 of the
4
220 general standard and Clause 201.2 of this document.
221 IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-1-11 and IEC 60601-1-12 apply as
[2]
222 modified in Clauses 202, 206, 208, 211 and 212 respectively. IEC 60601-1-3 does not apply. All
223 other published collateral standards in the IEC 60601-1 series apply as published.
224 201.1.4 Particular standards
225 IEC 60601-1:2005+AMD1:2012, subclause 1.4 is replaced by:
226 In the IEC 60601 series, particular standards may modify, replace or delete requirements
227 contained in the general standard, including the collateral standards, as appropriate for the
228 particular ME EQUIPMENT under consideration, and may add other BASIC SAFETY or ESSENTIAL
229 PERFORMANCE requirements.
230 A requirement of a particular standard takes priority over the general standard or the collateral
231 standards.
232 For brevity, IEC 60601-1:2005+AMD1:2012 is referred to in this document as the general
233 standard. Collateral standards are referred to by their document number.
4
 The general standard is IEC 60601-1:2005+AMD1:2012, Medical electrical equipment – Part 1: General
requirements for basic safety and essential performance
© ISO 2016 – All rights reserved
---------------------- Page: 12 ----------------------
ISO/DIS 80601-2-61:2016(E) – 13 –
234 The numbering of clauses and subclauses of this document corresponds to those of the general
235 standard with the prefix “201” (e.g. 201.1 in this document addresses the content of Clause 1 of
236 the general standard) or applicable collateral standard with the prefix “2xx” where xx is the final
237 digits of the collateral standard document number (e.g. 202.4 in this document addresses the
238 content of Clause 4 of the IEC 60601-1-2 collateral standard, 208.4 in this document addresses the
239 content of Clause 4 of the IEC 60601-1-8 collateral standard, etc.). The changes to the text of the
240 general standard are specified by the use of the following words:
241 "Replacement" means that the clause or subclause of the general standard or applicable collateral
242 standard is replaced completely by the text of this document.
243 "Addition" means that the text of this document is additional to the requirements of the general
244 standard or applicable collateral standard.
245 "Amendment" means that the clause or subclause of the general standard or applicable collateral
246 standard is amended as indicated by the text of this document.
247 Subclauses or figures that are additional to those of the general standard are numb ered starting
248 from 201.101, additional annexes are lettered AA, BB, etc., and additional items aa), bb), etc.
249 Subclauses or figures that are additional to those of a collateral standard are numbered starting
250 from 2xx, where “x” is the number of the collateral standard, e.g. 202 for IEC 60601-1-2, 203 for
251 IEC 60601-1-3, etc.
252 The term "this document" is used to make reference to the general standard, any applicable
253 collateral standards and this particular document taken together.
254 Where there is no corresponding clause or subclause in this document, the section, clause or
255 subclause of the general standard or applicable collateral standard, although possibly not
256 relevant, applies without modification; where it is intended that any part of the general standard
257 or applicable collateral standard, although possibly relevant, is not to be applied, a statement to
258 that effect is given in this particular document.
259 201.2 Normative references
260 The following documents are referred to in the text in such a way that some or all of their content
261 constitutes requirements of this document. For dated references, only the edition cited applies.
262 For undated references, the latest edition of the referenced document (including any
263 amendments) applies.
264 NOTE 1 The way in which these referenced documents are cited in normative requirements determines
265 the extent (in whole or in part) to which they apply.
266 NOTE 2 Informative references are listed in the Bibliography.
267 IEC 60601-1:2005+AMD1:2012, Clause 2 applies, except as follows:
268 Replacement:
© ISO 2016 – All rights reserved
---------------------- Page: 13 ----------------------
– 14 – ISO/DIS 80601-2-61:2016(E)
5
269 IEC 60417 , Graphical symbols for use on equipment
270 IEC 60529:2013, Degrees of protection provided by enclosures (IP code)
271 IEC 60601-1-2:2014, Medical electrical equipment — Part 1-2: General requirements for basic
272 safety and essential performance — Collateral Standard: Electromagnetic compatibility —
273 Requirements and tests
274 IEC 60601-1-6:2010, Medical electrical equipment — Part 1-6: General requirements for basic
275 safety and essential performance — Collateral standard: Usability
276 + Amendment 1:2013
277 IEC 60601-1-8:2006
 ...
NORME ISO
INTERNATIONALE 80601-2-61
Deuxième édition
2017-12
Version corrigée
2018-02
Appareils électromédicaux —
Partie 2-61:
Exigences particulières pour la
sécurité de base et les performances
essentielles pour les oxymètres de
pouls
Medical electrical equipment —
Part 2-61: Particular requirements for basic safety and essential
performance of pulse oximeter equipment
Numéro de référence
ISO 80601-2-61:2017(F)
©
 ISO 2017
---------------------- Page: 1 ----------------------
ISO 80601-2-61:2017(F)
DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2017
Tous droits réservés. Sauf prescription différente ou nécessité dans le contexte de sa mise en oeuvre, aucune partie de cette
publication ne peut être reproduite ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique,
y compris la photocopie, ou la diffusion sur l’internet ou sur un intranet, sans autorisation écrite préalable. Une autorisation peut
être demandée à l’ISO à l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva, Switzerland
Tel. +41 22 749 01 11
Fax +41 22 749 09 47
copyright@iso.org
www.iso.org
Publié en Suisse
© ISO 2017 – Tous droits réservés
ii
---------------------- Page: 2 ----------------------
ISO 80601-2-61:2017(F)
Sommaire Page
Avant-propos . v
Introduction . vii
201. 1 Domaine d’application, objet et normes apparentées . 1
201. 2 Références normatives. 3
201. 3 Termes et définitions . 5
201. 4 Exigences générales . 11
201. 5 Exigences générales relatives aux essais des APPAREILS EM . 12
201. 6 Classification des APPAREILS EM et SYSTEMES EM. 12
201. 7 Identification, marquage et documentation des APPAREILS EM . 12
201. 8 Protection contre les DANGERS d’origine électrique provenant des APPAREILS EM . 17
201. 9 Protection contre les DANGERS mécaniques des APPAREILS EM et SYSTEMES EM . 17
201. 10 Protection contre les DANGERS dus aux rayonnements involontaires ou excessifs . 17
201. 11 Protection contre les températures excessives et les autres DANGERS . 17
201. 12 Exactitude des commandes et des appareils de mesure et protection contre les
caractéristiques de sortie dangereuses . 19
201. 13 SITUATIONS DANGEREUSES et conditions de défaut pour les APPAREILS EM . 24
201. 14 SYSTEMES ELECTROMEDICAUX PROGRAMMABLES (SEMP) . 24
201. 15 Construction des APPAREILS EM . 24
YSTEMES EM . 26
201. 16 S
201. 17 Compatibilité électromagnétique des APPAREILS EM et des SYSTEMES EM . 26
201. 101 * CAPTEURS D’OXYMETRE DE POULS ET CABLES DE RACCORDEMENT DE CAPTEURS . 26
201. 102 SIGNAL D’INFORMATION relatif au pouls de saturation . 26
201. 103 CONNEXION FONCTIONNELLE . 27
202 Perturbations électromagnétiques — Exigences et e ssais . 27
206 Aptitude à l’utilisation . 28
208 Exigences générales, essais et guide pour les systèmes d’alarme des appareils et
des systèmes électromédicaux . 29
211 Exigences pour les appareils électromédicaux et les systèmes électromédicaux
utilisés dans l’environnement des soins à domicile . 30
212 Exigences pour les appareils électromédicaux et les systèmes électromédicaux
destinés à être utilisés dans l’environnement des services médicaux d’urgence . 30
Annexe C (informative) Guide pour l’application des exigences relatives au marquage et à
l’étiquetage des APPAREILS EM et SYSTEMES EM . 31
Annexe D (informative) Symboles utilisés pour le marquage . 35
Annexe AA	(informative) Recommandations particulières et justifications . 36
©	ISO	2017	–	Tous	droits	réservés
iii
---------------------- Page: 3 ----------------------
ISO 80601-2-61:2017(F)
Annexe BB	(informative) Température de la peau au contact du capteur
de l’OXYMETRE DE
POULS . 46
ANNEXE CC	(informative) Détermination de l’EXACTITUDE. 51
Annexe DD	(informative) Normes d’étalonnage . 61
Annexe EE	(informative) Ligne directrice pour l’évaluation et la documentation des sujets
humains en matière d’EXACTITUDE de la SpO . 62
2
Annexe FF	(informative) Simulateurs, dispositifs d’étalonnage et DISPOSITIFS D’ESSAI
FONCTIONNELS pour OXYMETRES DE POULS . 70
Annexe GG	(informative) Concepts relatifs au temps de réponse des APPAREILS EM . 80
Annexe HH	(normative) Exigences relatives aux interfaces de données . 84
Annexe II	(informative) Référence aux PRINCIPES ESSENTIELS . 89
Annexe JJ	(informative) Terminologie — Index alphabétique des termes définis . 93
Bibliographie . 96
©	ISO	2017	–	Tous	droits	réservés
iv
---------------------- Page: 4 ----------------------
ISO 80601-2-61:2017(F)
Avant-propos
L’ISO	 (Organisation	 internationale	 de	 normalisation)	 est	 une	 fédération	 mondiale	 d’organismes
nationaux	de	normalisation	(comités	membres	de	l’ISO).	L’élaboration	des	Normes	internationales	est
en	général	confiée	aux	comités	techniques	de	l’ISO.	Chaque	comité	membre	intéressé	par	une	étude	a	le
droit	 de	 faire	 partie	 du	 comité	 technique	 créé	 à	 cet	 effet.	 Les	 organisations	 internationales,
gouvernementales	et	non	gouvernementales,	en	liaison	avec	l’ISO	participent	également	aux	travaux.
L’ISO	 collabore	 étroitement	 avec	 la	 Commission	 électrotechnique	 internationale	 (IEC)	 en	 ce	 qui
concerne	la	normalisation	électrotechnique.
Les	procédures	utilisées	pour	élaborer	le	présent	document	et	celles	destinées	à	sa	mise	à	jour	sont
décrites	dans	les	Directives	ISO/IEC,	Partie	1.	Il	convient,	en	particulier	de	prendre	note	des	différents
critères	d’approbation	requis	pour	les	différents	types	de	documents	ISO.	Le	présent	document	a	été
rédigé	 conformément	 aux	 règles	 de	 rédaction	 données	 dans	 les	 Directives	 ISO/IEC,	 Partie	2
(voir	www.iso.org/directives).
L’attention	est	attirée	sur	le	fait	que	certains	des	éléments	du	présent	document	peuvent	faire	l’objet	de
droits	de	propriété	intellectuelle	ou	de	droits	analogues.	L’ISO	ne	saurait	être	tenue	pour	responsable
de	ne	pas	avoir	identifié	de	tels	droits	de	propriété	et	averti	de	leur	existence.	Les	détails	concernant	les
références	 aux	 droits	 de	 propriété	 intellectuelle	 ou	 autres	 droits	 analogues	 identifiés	 lors	 de
l’élaboration	du	document	sont	indiqués	dans	l’Introduction	et/ou	dans	la	liste	des	déclarations	de
brevets	reçues	par	l’ISO	(voir	www.iso.org/brevets).
Les	appellations	commerciales	éventuellement	mentionnées	dans	le	présent	document	sont	données
pour	information,	par	souci	de	commodité,	à	l’intention	des	utilisateurs	et	ne	sauraient	constituer	un
engagement.
Pour	une	explication	de	la	nature	volontaire	des	normes,	la	signification	des	termes	et	expressions
spécifiques	de	l’ISO	liés	à	l’évaluation	de	la	conformité,	ou	pour	toute	information	au	sujet	de	l’adhésion
de	l’ISO	aux	principes	de	l’Organisation	mondiale	du	commerce	(OMC)	concernant	les	obstacles
techniques	au	commerce	(OTC),	voir	le	lien	suivant:	www.iso.org/iso/fr/avant‐propos.html.
Le	présent	document	a	été	élaboré	conjointement	par	le	comité	technique	ISO/TC	121,	 Matériel
d’anesthésie et de réanimation respiratoire,	sous‐comité	SC	3,	Ventilateurs pulmonaires et équipements
connexes	et	le	comité	technique	IEC/TC	62,	 Équipements électriques dans la pratique médicale,
sous‐comité	SC	D,	Appareils électromédicaux.	Le	présent	projet	de	document	a	été	mis	en	circulation
auprès	des	organismes	nationaux	de	l’ISO	et	de	l’IEC	en	vue	d’un	vote.
Cette	deuxième	édition	annule	et	remplace	la	première	édition	(ISO	80601‐2‐61:2011),	qui	a	fait	l’objet
d’une	révision	technique.	Il	comprend	un	alignement	avec	l’Amendement	1	de	la	troisième	édition	de
l’IEC	60601‐1	et	de	la	deuxième	édition	de	l’IEC	60601‐1‐8,	ainsi	que	de	la	quatrième	édition	de
l’IEC	60601‐1‐2,	la	troisième	édition	de	l’IEC	60601‐1‐6,	la	deuxième	édition	de	l’IEC	60601‐1‐11	et
l’IEC	60601‐1‐12.
©	ISO	2017	–	Tous	droits	réservés
v
---------------------- Page: 5 ----------------------
ISO 80601-2-61:2017(F)
Les	principales	modifications	apportées	sont	les	suivantes:
 mise	à	jour	de	la	justification	(Annexe	AA)	et	des	références	relatives	aux	progrès	réalisés	dans	la
compréhension	de	l’hypoxémie,	des	dossiers	médicaux	informatisés	et	des	SYSTEMES	D’ALARME;
 modification	de	la	protection	contre	la	pénétration	de	IPX1	à	IPX2;
et	les	ajouts	suivants:
ENVIRONNEMENT	DES	SOINS	A	DOMICILE;
 article	211,	exigences	d’utilisation	dans	l’
 article	212,	exigences	d’utilisation	dans	l’environnement	des	services	médicaux	d’urgence	(SMU);
 annexe	HH,	exigences	relatives	aux	interfaces	de	données.
La	présente	version	corrigée	de	l’ISO	80601‐2‐61:2017	inclut	les	corrections	suivantes:
 les	en‐têtes	ont	été	corrigés.
Une	liste	de	toutes	les	parties	de	la	série	ISO/IEC	80601	se	trouve	sur	le	site	Web	de	l’ISO.
©	ISO	2017	–	Tous	droits	réservés
vi
---------------------- Page: 6 ----------------------
ISO 80601-2-61:2017(F)
Introduction
La	mesure	approchée	de	la	saturation	de	l’hémoglobine	artérielle	et	de	la	fréquence	du	pouls	par	le
biais	de	l’oxymétrie	de	pouls	est	une	pratique	courante	dans	de	nombreuses	spécialités	de	la	médecine.
Le	présent	document	couvre	les	exigences	de	SECURITE	DE	BASE	et	de	PERFORMANCES	ESSENTIELLES	pouvant
être	atteintes	dans	les	limites	de	la	technologie	existante.
Les	comités	reconnaissent	la	nécessité	de	réviser	la	première	édition	du	présent	document	en	raison	de
la	publication	de	la	première	édition	de	l’IEC	60601‐1‐12,	ainsi	que	de	la	quatrième	édition	de
l’IEC	60601‐1‐2,	de	la	deuxième	édition	de	l’IEC	60601‐1‐11	et	des	premiers	Amendements	de	la
troisième	édition	de	l’IEC	60601‐1,	de	la	troisième	édition	de	l’IEC	60601‐1‐6	et	de	la	deuxième	édition
de	l’IEC	60601‐1‐8.
L’Annexe	AA	contient	une	justification	de	certaines	exigences.	Il	est	inclus	de	façon	à	fournir	un	aperçu
supplémentaire	du	raisonnement	suivi	au	sein	des	comités	et	ayant	abouti	à	une	exigence,	ainsi	que	de
l’identification	des	DANGERS	traités	par	l’exigence.
L’Annexe	BB	propose	une	enquête	bibliographique	relative	à	la	détermination	de	la	température	de
sécurité	maximale	de	l’interface	entre	le	CAPTEUR	DE	L’OXYMETRE	DE	POULS	et	le	tissu	du	PATIENT.
L’Annexe	CC	s’intéresse	à	la	fois	aux	formules	mathématiques	utilisées	pour	évaluer	l’EXACTITUDE	de	la
SpO 	des	mesurages	de	l’OXYMETRE	DE	POULS	et	aux	noms	attribués	à	ces	formules.
2
L’Annexe	DD	présente	des	recommandations	relatives	aux	situations	dans	lesquelles	un	étalonnage	de
l’OXYMETRE	DE	POULS	par	une	analyse	de	sang	in	vitro	est	nécessaire.
L’Annexe	EE	 propose	 des	 lignes	 directrices	 pour	 l’ETUDE	 DE	 DESATURATION	 CONTROLEE	 permettant
l’étalonnage	de	l’OXYMETRE	DE	POULS.
L’Annexe	FF	est	une	introduction	didactique	à	différents	types	de	dispositifs	d’essai	utilisés	dans
l’oxymétrie	de	pouls.
L’Annexe	GG	définit	les	concepts	relatifs	au	temps	de	réponse	de	l’OXYMETRE	DE	POULS.
L’Annexe	HH	décrit	les	exigences	relatives	aux	interfaces	de	données.
L’Annexe	II	 comprend	 une	 Référence	 aux	 PRINCIPES	 ESSENTIELS	qui	se	trouvait	auparavant	dans
l’Annexe	HH.
Cette	publication	a	été	rédigée	selon	les	Directives	ISO/IEC,	Partie	2.
Dans	le	présent	document,	les	types	de	polices	suivants	sont	utilisés:
 exigences	et	définitions:	caractères	cambria;
 spécifications d’essai: caractères italiques;
 les	indications	de	nature	informative	apparaissant	hors	des	tableaux,	comme	les	notes,	les
exemples	et	les	références:	en	petits	caractères;	le	texte	normatif	des	tableaux	est	également	en
petits	caractères;
©	ISO	2017	–	Tous	droits	réservés
vii
---------------------- Page: 7 ----------------------
ISO 80601-2-61:2017(F)
1)
 les	TERMES	DEFINIS	DANS	L’ARTICLE	3	DE	LA	NORME	GENERALE ,	DANS	LE	PRESENT	DOCUMENT	OU	EN	NOTE:	EN	PETITES
MAJUSCULES.
Concernant	la	structure	du	présent	document,	le	terme:
 «article»	désigne	l’une	des	six	sections	numérotées	dans	la	table	des	matières,	avec	toutes	ses
subdivisions	(par	exemple	l’Article	201	inclut	les	paragraphes	201.7.1	et	201.7.2);	et
 «paragraphe»	 désigne	 une	 subdivision	 numérotée	 d’un	 article	 (par	 exemple	 201.7.1,	 7,2	 et
201.7.2.1	sont	tous	des	paragraphes	de	l’Article	201.7).
Les	références	à	des	articles	du	présent	document	sont	précédées	du	terme	«Article»	suivi	du	numéro
de	l’article	concerné.	Dans	le	présent	document,	les	références	aux	paragraphes	utilisent	uniquement	le
numéro	du	paragraphe	concerné.
Dans	le	présent	document,	la	conjonction	«ou»	est	utilisée	avec	la	valeur	d’un	«ou	inclusif»;	ainsi	un
énoncé	est	vrai	si	une	combinaison	des	conditions	quelle	qu’elle	soit	est	vraie.
Les	formes	verbales	utilisées	dans	le	présent	document	sont	conformes	à	l’usage	décrit	dans	l’Annexe	H
des	directives	ISO/IEC,	Partie	2.	Pour	les	besoins	du	présent	document,	la	forme	verbale:
 «	devoir	»	signifie	que	la	satisfaction	à	une	exigence	ou	à	un	essai	est	obligatoire	pour	la	conformité
au	présent	document;
 «il	convient/il	est	recommandé»	signifie	que	la	satisfaction	à	une	exigence	ou	à	un	essai	est
recommandée,	mais	n’est	pas	obligatoire	pour	la	conformité	au	présent	document;	et
 «pouvoir»	mis	au	présent	de	l’indicatif	est	utilisé	pour	décrire	un	moyen	admissible	pour	satisfaire
à	une	exigence	ou	à	un	essai.
Lorsqu’un	astérisque	(*)	est	utilisé	comme	premier	caractère	devant	un	titre,	au	début	d’un	alinéa	ou
devant	un	titre	de	tableau,	il	indique	l’existence	d’une	recommandation	ou	d’une	justification	à
consulter	dans	l’Annexe	AA.
1)
	 La	norme	générale	est	l’IEC	60601‐1:2005	et	l’IEC	60601‐1:2005/AMD1:2012,	Appareils	électromédicaux	—
Partie	1:	Exigences	générales	pour	la	sécurité	de	base	et	les	performances	essentielles.
©	ISO	2017	–	Tous	droits	réservés
viii
---------------------- Page: 8 ----------------------
NORME INTERNATIONALE ISO 80601-2-61:2017(F)
Appareils électromédicaux — Partie 2-61: Exigences
particulières pour la sécurité de base et les performances
essentielles pour les oxymètres de pouls
201.1 Domaine d’application, objet et normes apparentées
L’Article	1	de	l’IEC	60601‐1:2005+AMD1:2012	s’applique,	avec	l’exception	suivante:
201.1.1 * Domaine d’application
Remplacement:
Le	présent	document	s’applique	à	la	SECURITE	DE	BASE	et	aux	PERFORMANCES	ESSENTIELLES	des	OXYMETRES
DE	POULS	conçus	pour	une	utilisation	chez	les	êtres	humains,	ci‐après	dénommés	APPAREILS	EM.	Sont
inclus	tous	les	éléments	de	l’appareil	nécessaires	pour	une	UTILISATION	NORMALE,	y	compris	le	MONITEUR
DE	L’OXYMETRE	DE	POULS,	le	CAPTEUR	DE	L’OXYMETRE	DE	POULS	et	le	CABLE	DE	RACCORDEMENT	DU	CAPTEUR.
Ces	exigences	s’appliquent	de	la	même	manière	aux	OXYMETRES	DE	POULS,	y	compris	aux	MONITEURS
D’OXYMETRE	DE	POULS,	aux	CAPTEURS	D’OXYMETRE	DE	POULS	et	aux	CABLES	DE	RACCORDEMENT	DU	CAPTEUR,	qui
ont	été	REMIS	EN	ETAT.
L’utilisation	prévue	des	OXYMETRES	DE	POULS	inclut	notamment,	sans	toutefois	s’y	limiter,	l’estimation	de
la	saturation	en	oxygène	de	l’hémoglobine	artérielle	et	de	la	fréquence	du	pouls	chez	des	PATIENTS	en
établissement	de	soins	professionnel	ainsi	que	chez	les	PATIENTS	dans	l’ENVIRONNEMENT	DES	SOINS	A
DOMICILE	et	dans	l’ENVIRONNEMENT	DES	SERVICES	MEDICAUX	D’URGENCE.
Le	présent	document	ne	s’applique	pas	aux	OXYMETRES	DE	POULS	destinés	à	être	utilisés	dans	le	cadre	de
recherches	en	laboratoire,	de	même	qu’aux	oxymètres	nécessitant	un	échantillon	de	sang	du	PATIENT.
Si	 un	 article	 ou	 un	 paragraphe	 est	 spécifiquement	 destiné	 à	 être	 applicable	 uniquement	 aux
APPAREILS	EM	ou	uniquement	aux	SYSTEMES	EM,	le	titre	et	le	contenu	de	cet	article	ou	de	ce	paragraphe
l’indiquent.	Si	cela	n’est	pas	le	cas,	l’article	ou	le	paragraphe	s’applique	à	la	fois	aux	APPAREILS	EM	et	aux
SYSTEMES	EM,	selon	le	cas.
Les	DANGERS	inhérents	à	la	fonction	physiologique	prévue	de	l’APPAREIL	EM	ou	des	SYSTEMES	EM	dans	le
cadre	 du	 domaine	 d’application	 du	 présent	 document	 ne	 sont	 pas	couverts	par	des	exigences
spécifiques	contenues	dans	le	présent	document,	à	l’exception	de	201.11	et	de	7.2.13	et	8.4.1	de	la
norme	générale.
NOTE	1	 Voir	également	4.2	de	la	norme	générale.	La	norme	générale	est	l’IEC	60601‐1:2005+AMD1:2012,
Appareils électromédicaux — Partie 1: Exigences générales pour la sécurité de base et les performances essentielles.
Le	présent	document	peut	également	être	appliqué	aux	APPAREILS	EM	et	à	leurs	ACCESSOIRES	utilisés	pour
l’atténuation	d’une	maladie,	la	compensation	ou	l’atténuation	d’une	blessure	ou	d’une	incapacité.
Le	présent	document	ne	s’applique	pas	aux	OXYMETRES	DE	POULS	destinés	uniquement	à	la	surveillance
de	fœtus.
Le	présent	document	ne	s’applique	pas	aux	appareils	distants	ou	esclaves	(secondaires)	qui	affichent	les
valeurs	de	la	SpO 	et	qui	sont	situés	à	l’extérieur	de	l’ENVIRONNEMENT	DU	PATIENT.
2
NOTE	2	 Il	est	attendu	que	les	APPAREILS	EM	qui	permettent	une	sélection	entre	des	fonctions	de	diagnostic	et	de
surveillance	satisfassent	aux	exigences	du	document	approprié	lorsqu’ils	sont	configurés	pour	ladite	fonction.
©	ISO	2017	–	Tous	droits	réservés
1
---------------------- Page: 9 ----------------------
ISO 80601-2-61:2017(F)
Le	présent	document	est	applicable	aux	OXYMETRES	DE	POULS	destinés	à	être	utilisés	dans	des	conditions
environnementales	extrêmes	ou	non	maîtrisées	à	l’extérieur	de	l’environnement	hospitalier	ou	d’un
cabinet	de	médecin,	telles	que	dans	des	ambulances	ou	dans	les	transports	aériens.	Des	normes
supplémentaires	peuvent	s’appliquer	aux	OXYMETRES	DE	POULS	pour	ces	environnements	d’utilisation.
Le	 présent	 document	 est	 une	 norme	 particulière	 des	 séries	 de	 normes	 de	 l’IEC	60601‐1	 et	 de
l’ISO/IEC	80601.
201.1.2 Objet
Le	paragraphe	1.2	de	la	norme	générale	est	remplacé	par:
L’objet	du	présent	document	est	d’établir	des	exigences	particulières	relatives	à	la	SECURITE	DE	BASE	et
aux	PERFORMANCES	ESSENTIELLES	applicables	aux	OXYMETRES	DE	POULS	[tels	que	définis	en	201.3.217]	et	à
leurs	ACCESSOIRES.
NOTE	 Les	ACCESSOIRES	sont	inclus,	car	il	est	nécessaire	que	la	combinaison	du	MONITEUR	D’OXYMETRE	DE	POULS	et
des	ACCESSOIRES	présente	un	niveau	de	sécurité	adéquat.	Les	ACCESSOIRES	peuvent	avoir	une	incidence	significative
sur	la	SECURITE	DE	BASE	ou	les	PERFORMANCES	ESSENTIELLES	d’un	OXYMETRE	DE	POULS.
201.1.3 Normes collatérales
Le	paragraphe	1.3	de	l’IEC	60601‐1:2005+AMD1:2012	s’applique,	avec	l’ajout	suivant:
Le	présent	document	fait	référence	aux	normes	collatérales	applicables	figurant	dans	l’Article	2	de	la
norme	générale	et	l’Article	201.2	du	présent	document.
L’IEC	60601‐1‐2,	l’IEC	60601‐1‐6,	l’IEC	60601‐1‐8,	l’IEC	60601‐1‐11	et	l’IEC	60601‐1‐12	s’appliquent
[2]
tels	que	modifiés	dans	les	Articles	202,	206,	208,	211	et	212,	respectivement.	L’IEC	60601‐1‐3 	ne
s’applique	pas.	Toutes	les	autres	normes	collatérales	publiées	dans	la	série	de	normes	IEC	60601‐1
s’appliquent	telles	quelles.
201.1.4 Normes particulières
Le	paragraphe	1.4	de	l’IEC	60601‐1:2005+AMD1:2012	est	remplacé	par:
Dans	la	série	IEC	60601,	des	normes	particulières	peuvent	modifier,	remplacer	ou	supprimer	des
exigences	contenues	dans	la	norme	générale	et	dans	les	normes	collatérales,	en	fonction	de	ce	qui	est
approprié	à	l’APPAREIL	EM	spécifique	considéré	et	peuvent	ajouter	d’autres	exigences	de	SECURITE	DE	BASE
ou	de	PERFORMANCES	ESSENTIELLES.
Une	exigence	d’une	norme	particulière	prévaut	sur	l’exigence	correspondante	de	la	norme	générale	ou
des	normes	collatérales.
Par	souci	de	concision,	l’IEC	60601‐1:2005+AMD1:2012	est	appelée	«la	norme	générale»	dans	le
présent	document.	Il	est	fait	référence	aux	normes	collatérales	par	leur	numéro.
La	numérotation	des	articles	et	des	paragraphes	du	présent	document	correspond	à	celle	de	la	norme
générale	avec	le	préfixe	«201»	(par	exemple	dans	le	présent	document,	201.1	reprend	le	contenu	de
l’Article	1	de	la	norme	générale)	ou	de	la	norme	collatérale	applicable	avec	le	préfixe	«2xx»,	où	xx
correspond	aux	derniers	chiffres	du	numéro	de	document	de	la	norme	collatérale	(par	exemple	dans	le
présent	document,	202.4	reprend	le	contenu	de	l’Article	4	de	la	norme	collatérale	IEC	60601‐1‐2,	208.4
reprend	le	 contenu	de	l’Article	4	 de	la	 norme	collatérale	IEC	60601‐1‐8,	etc.).	Les	modifications
apportées	au	texte	de	la	norme	générale	sont	spécifiées	par	l’utilisation	des	mots	suivants:
«Remplacement»	signifie	que	l’article	ou	le	paragraphe	de	la	norme	générale	ou	de	la	norme	collatérale
est	remplacé	en	intégralité	par	le	texte	du	présent	document.
©	ISO	2017	–	Tous	droits	réservés
2
---------------------- Page: 10 ----------------------
ISO 80601-2-61:2017(F)
«Ajout»	signifie	que	le	texte	du	présent	document	est	ajouté	aux	exigences	de	la	norme	générale	ou	de	la
norme	collatérale	applicable.
«Amendement»	signifie	que	l’article	ou	le	paragraphe	de	la	norme	générale	ou	de	la	norme	collatérale
applicable	est	amendé	comme	indiqué	par	le	texte	du	présent	document.
Les	paragraphes	ou	les	figures	qui	ont	été	ajoutés	par	rapport	à	ceux	de	la	norme	générale	sont
numérotés	en	commençant	à	partir	de	201.101,	les	annexes	supplémentaires	sont	appelées	AA,	BB,	etc.,
et	les	points	supplémentaires	sont	appelés	aa), bb),	etc.
Les	paragraphes	ou	les	figures	qui	ont	été	ajoutés	par	rapport	à	ceux	d’une	norme	collatérale	sont
numérotés	en	commençant	à	partir	de	2xx,	où	«x»	est	le	numéro	de	la	norme	collatérale	standard,	par
exemple	202	pour	l’IEC	60601‐1‐2,	203	pour	l’IEC	60601‐1‐3,	etc.
L’expression	«présent	document»	est	utilisée	pour	faire	référence	à	la	norme	générale,	aux	normes
collatérales	applicables	et	au	présent	document	particulier	considérés	conjointement.
Si	aucun	article	ni	aucun	paragraphe	correspondant	n’existe	dans	le	présent	document,	la	section,
l’article	ou	le	paragraphe	de	la	norme	générale	ou	de	la	norme	collatérale	applicable,	même	s’il	peut	ne
pas	être	pertinent,	s’applique	sans	aucune	modification;	lorsqu’il	est	prévu	que	l’une	des	parties	de	la
norme	générale	ou	de	la	norme	collatérale	applicable	ne	s’applique	pas,	même	si	elle	peut	être
pertinente,	une	mention	est	donnée	à	cet	effet	dans	le	présent	document	particulier.
201.2 Références normatives
Les	documents	suivants	cités	dans	le	texte	constituent,	pour	tout	ou	partie	de	leur	contenu,	des
exigences	du	présent	document.	Pour	les	références	datées,	seule	l’édition	citée	s’applique.	Pour	les
références	non	datées,	la	dernière	édition	du	document	de	référence	s'applique	(y	compris	les	éventuels
amendements).
NOTE	1	 La	manière	dont	ces	documents	référencés	sont	cités	dans	les	exigences	normatives	détermine	l’étendue
(en	totalité	ou	en	partie)	de	leur	application.
NOTE	2	 La	Bibliographie	fournit	une	liste	de	références	informatives.
L’Article	2	de	l’IEC	60601‐1:2005+AMD1:2012	s’applique,	avec	l’exception	suivante:
Remplacement:
IEC	60417,	Symboles graphiques utilisables sur le matériel.
IEC	60529:2013,	Degrés de protection procurés par les enveloppes (Code IP).
IEC	60601‐1‐2:2014,	Appareils électromédicaux — Partie 1-2: Exigences générales pour la sécurité de
base et les performances essentielles — Norme collatérale: Perturbations électromagnétiques — Exigences
et essais.
IEC	60601‐1‐6:2010,	Appareils électromédicaux — Partie 1-6: Exigences générales pour la sécurité de
base et les performances essentielles — Norme collatérale: Aptitude à l’utilisation +Amendement	1:2013.
IEC	60601‐1‐8:2006,	Appareils électromédicaux — Partie 1-8: Exigences générales pour la sécurité de
base et les performances essentielles — Norme collatérale: Exigences générales, essais et guide pour les
systèmes d’alarme des appareils et des systèmes électromédicaux +Amendement	1:2012.
IEC	60825‐1:2014,	Sécurité des appareils à laser — Partie 1: Classification des matériels et exigences.
©	ISO	2017	–	Tous	droits	réservés
3
---------------------- Page: 11 ----------------------
ISO 80601-2-61:2017(F)
Ajout:
ISO	3744:2010,	Acoustique — Détermination des niveaux de puissance acoustique et des niveaux d’énergie
acoustique émis par les sources de bruit à partir de la pression acoustique — Méthodes d’expertise pour
des conditions approchant celles du champ libre sur plan réfléchissant.
ISO	4135:2001,	Matériel d’anesthésie et de réanimation respiratoire — Vocabulaire.
ISO	7000,	Symboles graphiques utilisables sur le matériel — Symboles enregistrés.
ISO	7010,	 Symboles graphiques — Couleurs de sécurité et signaux de sécurité — Signaux de sécurité
enregistrés.
ISO	14155:2011,	Investigation clinique des dispositifs médicaux pour sujets humains — Bonnes pratiques
cliniques.
ISO	14937:2009,	Stérilisation des produits de santé — Exigences générales pour la caractérisation d’un
agent stérilisant et pour la mise au point, la validation et la vérification de routine d’un processus de
stérilisation pour dispositifs médicaux.
ISO	15223‐1:2016,	 Dispositifs médicaux — Symboles à utiliser avec les étiquettes, l’étiquetage et les
informations à fournir relatifs aux dispositifs médicaux — Partie 1: Exigences générales.
IEC	60068‐2‐27:2008,	 Essais d’environnement — Partie 2-27: Essais — Essai Ea et guide: Chocs
+Amendement	1:2013.
IEC	60068‐2‐31:2008,	 Essais d’environnement — Partie 2-31: Essais — Essai Ec: Choc lié à des
manutentions brutales, essai destiné en premier lieu aux matériels.
IEC	60068‐2‐64:2008,	Essais d’environnement — Partie 2-64: Essais — Essais Fh: Vibrations aléatoires à
large bande et guide.
IEC	60601‐1‐11:2015,	Appareils électromédicaux — Partie 1-11: Exigences générales pour la sécurité de
base et les performances essentielles — Norme collatérale: Exigences pour les appareils électromédicaux et
les systèmes électromédicaux utilisés dans l’environnement des soins à domicile.
IEC	60601‐1‐12:2014,	Appareils électromédicaux — Partie 1-12: Exigences générales pour la sécurité de
base et les performances essentielles — Norme collatérale: Exigences pour les appareils électromédicaux et
les systèmes électromédicaux destinés à être utilisés dans l’environnement des services médicaux
d’urgence.
IEC	60825‐2:2004,	Sécurité des appareils à laser — Partie 2: Sécurité des systèmes de télécommunication
par fibres optiques (STFO) +	Amendement	1:2006+	Amendement	2:2012.
IEC	62366‐1:2015,	Dispositifs médicaux — Partie 1: Application de l’ingénierie de l’aptitude à l’utilisation
aux dispositifs médicaux.
©	ISO	2017	–	Tous	droits	réservés
4
---------------------- Page: 12 ----------------------
ISO 80601-2-61:2017(F)
201.3 Termes et définitions
Pour	les	besoins	du	présent	document,	les	termes	et	définitions	de	l’ISO	3744:2010,	l’ISO	4135:2001,
l’IEC	60601‐1:2005+AMD	1:2012,	 l’IEC	60601‐1‐2:2014,	 l’IEC	60601‐1‐6:2010+AMD	1:2013,
l’IEC	60601‐1‐8:2006+AMD	1:2012,	l’IEC	60601‐1‐11:2015,	l’IEC	60601‐1‐12:2014,	l’IEC	62366‐1:2015
ainsi	que	les	suivants,	s’appliquent.
L’ISO	et	l’IEC	tiennent	à	jour	des	bases	de	données	terminologiques	destinées	à	être	utilisées	en
normal
 ...
PROJET
NORME ISO/FDIS
FINAL
INTERNATIONALE80601-2-61
ISO/TC 121/SC 3
Appareils électromédicaux —
Secrétariat: ANSI
Début de vote:
Partie 2-61:
2017-09-29
Exigences particulières pour la
Vote clos le:
sécurité de base et les performances
2017-11-25
essentielles pour les oxymètres de
pouls
Medical electrical equipment —
Part 2-61: Particular requirements for basic safety and essential
performance of pulse oximeter equipment
Ce projet est soumis à un vote parallèle à ISO et à IEC.
LES DESTINATAIRES DU PRÉSENT PROJET SONT
INVITÉS À PRÉSENTER, AVEC LEURS OBSER-
VATIONS, NOTIFICATION DES DROITS DE PRO-
TRAITEMENT PARALLÈLE ISO/CEN
PRIÉTÉ DONT ILS AURAIENT ÉVENTUELLEMENT
CONNAISSANCE ET À FOURNIR UNE DOCUMEN-
TATION EXPLICATIVE.
OUTRE LE FAIT D’ÊTRE EXAMINÉS POUR
ÉTABLIR S’ILS SONT ACCEPTABLES À DES FINS
INDUSTRIELLES, TECHNOLOGIQUES ET COM-
Numéro de référence
MERCIALES, AINSI QUE DU POINT DE VUE
ISO/FDIS 80601-2-61:2017(F)
DES UTILISATEURS, LES PROJETS DE NORMES
INTERNATIONALES DOIVENT PARFOIS ÊTRE
CONSIDÉRÉS DU POINT DE VUE DE LEUR POSSI-
BILITÉ DE DEVENIR DES NORMES POUVANT
SERVIR DE RÉFÉRENCE DANS LA RÉGLEMENTA-
©
TION NATIONALE. ISO 2017
---------------------- Page: 1 ----------------------
ISO/FDIS 80601-2-61:2017(F)
DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2017, Publié en Suisse
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni utilisée
sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie, l’affichage sur
l’internet ou sur un Intranet, sans autorisation écrite préalable. Les demandes d’autorisation peuvent être adressées à l’ISO à
l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Ch. de Blandonnet 8 • CP 401
CH-1214 Vernier, Geneva, Switzerland
Tel. +41 22 749 01 11
Fax +41 22 749 09 47
copyright@iso.org
www.iso.org
ii © ISO 2017 – Tous droits réservés
---------------------- Page: 2 ----------------------
ISO/FDIS	80601‐2‐61:2017(F)
Sommaire	 Page
Avant‐propos	.	v
Introduction	.	vii
201.	1	 Domaine	d’application,	objet	et	normes	apparentées	.	1
201.	2	 Références	normatives.	3
201.	3	 Termes	et	définitions	.	5
201.	4	 Exigences	générales	.	11
201.	5	 Exigences	générales	relatives	aux	essais	des	APPAREILS	EM	.	12
201.	6	 Classification	des	APPAREILS	EM	et	SYSTEMES	EM.	12
201.	7	 Identification,	marquage	et	documentation	des	APPAREILS	EM	.	12
201.	8	 Protection	contre	les	DANGERS	d’origine	électrique	provenant	des	APPAREILS	EM	.	17
201.	9	 Protection	contre	les	DANGERS	mécaniques	des	APPAREILS	EM	et	SYSTEMES	EM	.	17
201.	10	 Protection	contre	les	DANGERS	dus	aux	rayonnements	involontaires	ou	excessifs	.	17
201.	11	 Protection	contre	les	températures	excessives	et	les	autres	DANGERS	.	17
201.	12	 Exactitude	des	commandes	et	des	appareils	de	mesure	et	protection	contre	les
caractéristiques	de	sortie	dangereuses	.	19
201.	13	 SITUATIONS	DANGEREUSES	et	conditions	de	défaut	pour	les	APPAREILS	EM	.	24
201.	14	 SYSTEMES	ELECTROMEDICAUX	PROGRAMMABLES	(SEMP)	.	24
201.	15	 Construction	des	APPAREILS	EM	.	24
201.	16	 SYSTEMES	EM	.	26
201.	17	 Compatibilité	électromagnétique	des	APPAREILS	EM	et	des	SYSTEMES	EM	.	26
201.	101	*	CAPTEURS	D’OXYMETRE	DE	POULS	ET	CABLES	DE	RACCORDEMENT	DE	CAPTEURS	.	26
201.	102	SIGNAL	D’INFORMATION	relatif	au	pouls	de	saturation	.	26
201.	103	CONNEXION	FONCTIONNELLE	.	27
202	 Perturbations	électromagnétiques	—	Exigences	et	essais	.	27
206	 Aptitude	à	l’utilisation	.	28
208	 Exigences	générales,	essais	et	guide	pour	les	systèmes	d’alarme	des	appareils	et
des	systèmes	électromédicaux	.	29
211	 Exigences	pour	les	appareils	électromédicaux	et	les	systèmes	électromédicaux
utilisés	dans	l’environnement	des	soins	à	domicile	.	30
212	 Exigences	pour	les	appareils	électromédicaux	et	les	systèmes	électromédicaux
destinés	à	être	utilisés	dans	l’environnement	des	services	médicaux	d’urgence	.	30
Annexe	C	(informative)		Guide	pour	l’application	des	exigences	relatives	au	marquage	et	à
l’étiquetage	des	APPAREILS	EM	et	SYSTEMES	EM	.	31
Annexe	D	(informative)		Symboles	utilisés	pour	le	marquage	.	35
Annexe	AA	(informative)		Recommandations	particulières	et	justifications	.	37
©	ISO	2017	–	Tous	droits	réservés
iii
---------------------- Page: 3 ----------------------
ISO/FDIS	80601‐2‐61:2017(F)
Annexe	BB	(informative)		Température	de	la	peau	au	contact	du	capteur
de	l’OXYMETRE	DE
POULS	.	47
ANNEXE	CC	(informative)		Détermination	de	l’EXACTITUDE.	52
Annexe	DD	(informative) 	Normes	d’étalonnage	.	62
Annexe	EE	(informative)		Ligne	directrice	pour	l’évaluation	et	la	documentation	des	sujets
humains	en	matière	d’EXACTITUDE	de	la	SpO	.	63
2
Annexe	FF	(informative)		Simulateurs,	dispositifs	d’étalonnage	et	DISPOSITIFS	D’ESSAI
FONCTIONNELS	pour	OXYMETRES	DE	POULS	.	71
Annexe	GG	(informative) Concepts	relatifs	au	temps	de	réponse	des	APPAREILS	EM	.	81
Annexe	HH	(normative)		Exigences	relatives	aux	interfaces	de	données	.	85
Annexe	II	(informative)		Référence	aux	PRINCIPES	ESSENTIELS	.	90
Annexe	JJ	(informative)		Terminologie	—	Index	alphabétique	des	termes	définis	.	94
Bibliographie	.	97
©	ISO	2017	–	Tous	droits	réservés
iv
---------------------- Page: 4 ----------------------
ISO/FDIS	80601‐2‐61:2017(F)
Avant‐propos
L’ISO	 (Organisation	 internationale	 de	 normalisation)	 est	 une	 fédération	 mondiale	 d’organismes
nationaux	de	normalisation	(comités	membres	de	l’ISO).	L’élaboration	des	Normes	internationales	est
en	général	confiée	aux	comités	techniques	de	l’ISO.	Chaque	comité	membre	intéressé	par	une	étude	a	le
droit	 de	 faire	 partie	 du	 comité	 technique	 créé	 à	 cet	 effet.	 Les	 organisations	 internationales,
gouvernementales	et	non	gouvernementales,	en	liaison	avec	l’ISO	participent	également	aux	travaux.
L’ISO	 collabore	 étroitement	 avec	 la	 Commission	 électrotechnique	 internationale	 (IEC)	 en	 ce	 qui
concerne	la	normalisation	électrotechnique.
Les	procédures	utilisées	pour	élaborer	le	présent	document	et	celles	destinées	à	sa	mise	à	jour	sont
décrites	dans	les	Directives	ISO/IEC,	Partie	1.	Il	convient,	en	particulier	de	prendre	note	des	différents
critères	d’approbation	requis	pour	les	différents	types	de	documents	ISO.	Le	présent	document	a	été
rédigé	 conformément	 aux	 règles	 de	 rédaction	 données	 dans	 les	 Directives	 ISO/IEC,	 Partie	2
(voir	www.iso.org/directives).
L’attention	est	attirée	sur	le	fait	que	certains	des	éléments	du	présent	document	peuvent	faire	l’objet	de
droits	de	propriété	intellectuelle	ou	de	droits	analogues.	L’ISO	ne	saurait	être	tenue	pour	responsable
de	ne	pas	avoir	identifié	de	tels	droits	de	propriété	et	averti	de	leur	existence.	Les	détails	concernant	les
références	 aux	 droits	 de	 propriété	 intellectuelle	 ou	 autres	 droits	 analogues	 identifiés	 lors	 de
l’élaboration	du	document	sont	indiqués	dans	l’Introduction	et/ou	dans	la	liste	des	déclarations	de
brevets	reçues	par	l’ISO	(voir	www.iso.org/brevets).
Les	appellations	commerciales	éventuellement	mentionnées	dans	le	présent	document	sont	données
pour	information,	par	souci	de	commodité,	à	l’intention	des	utilisateurs	et	ne	sauraient	constituer	un
engagement.
Pour	une	explication	de	la	nature	volontaire	des	normes,	la	signification	des	termes	et	expressions
spécifiques	de	l’ISO	liés	à	l’évaluation	de	la	conformité,	ou	pour	toute	information	au	sujet	de	l’adhésion
de	l’ISO	aux	principes	de	l’Organisation	mondiale	du	commerce	(OMC)	concernant	les	obstacles
techniques	au	commerce	(OTC),	voir	le	lien	suivant:	www.iso.org/iso/fr/avant‐propos.html.
Le	présent	document	a	été	élaboré	conjointement	par	le	comité	technique	ISO/TC	121,	 Matériel
d’anesthésie	et	de	réanimation	respiratoire,	sous‐comité	SC	3,	Ventilateurs	pulmonaires	et	équipements
connexes	et	le	comité	technique	IEC/TC	62,	 Équipements	 électriques	 dans	 la	 pratique	 médicale,
sous‐comité	SC	D,	Appareils	électromédicaux.	Le	présent	projet	de	document	a	été	mis	en	circulation
auprès	des	organismes	nationaux	de	l’ISO	et	de	l’IEC	en	vue	d’un	vote.
Cette	deuxième	édition	annule	et	remplace	la	première	édition	(ISO	80601‐2‐61:2011),	qui	a	fait	l’objet
d’une	révision	technique.	Il	comprend	un	alignement	avec	l’Amendement	1	de	la	troisième	édition	de
l’IEC	60601‐1	et	de	la	deuxième	édition	de	l’IEC	60601‐1‐8,	ainsi	que	de	la	quatrième	édition	de
l’IEC	60601‐1‐2,	la	troisième	édition	de	l’IEC	60601‐1‐6,	la	deuxième	édition	de	l’IEC	60601‐1‐11	et
l’IEC	60601‐1‐12.
©	ISO	2017	–	Tous	droits	réservés
v
---------------------- Page: 5 ----------------------
ISO/FDIS	80601‐2‐61:2017(F)
Les	principales	modifications	apportées	sont	les	suivantes:
 mise	à	jour	de	la	justification	(Annexe	AA)	et	des	références	relatives	aux	progrès	réalisés	dans	la
compréhension	de	l’hypoxémie,	des	dossiers	médicaux	informatisés	et	des	SYSTEMES	D’ALARME;
 modification	de	la	protection	contre	la	pénétration	de	IPX1	à	IPX2;
et	les	ajouts	suivants:
 article	211,	exigences	d’utilisation	dans	l’ENVIRONNEMENT	DES	SOINS	A	DOMICILE;
 article	212,	exigences	d’utilisation	dans	l’environnement	des	services	médicaux	d’urgence	(SMU);
 annexe	HH,	exigences	relatives	aux	interfaces	de	données.
Une	liste	de	toutes	les	parties	de	la	série	ISO/IEC	80601	se	trouve	sur	le	site	Web	de	l’ISO.
©	ISO	2017	–	Tous	droits	réservés
vi
---------------------- Page: 6 ----------------------
ISO/FDIS	80601‐2‐61:2017(F)
Introduction
La	mesure	approchée	de	la	saturation	de	l’hémoglobine	artérielle	et	de	la	fréquence	du	pouls	par	le
biais	de	l’oxymétrie	de	pouls	est	une	pratique	courante	dans	de	nombreuses	spécialités	de	la	médecine.
Le	présent	document	couvre	les	exigences	de	SECURITE	DE	BASE	et	de	PERFORMANCES	ESSENTIELLES	pouvant
être	atteintes	dans	les	limites	de	la	technologie	existante.
Les	comités	reconnaissent	la	nécessité	de	réviser	la	première	édition	du	présent	document	en	raison	de
la	publication	de	la	première	édition	de	l’IEC	60601‐1‐12,	ainsi	que	de	la	quatrième	édition	de
l’IEC	60601‐1‐2,	de	la	deuxième	édition	de	l’IEC	60601‐1‐11	et	des	premiers	Amendements	de	la
troisième	édition	de	l’IEC	60601‐1,	de	la	troisième	édition	de	l’IEC	60601‐1‐6	et	de	la	deuxième	édition
de	l’IEC	60601‐1‐8.
L’Annexe	AA	contient	une	justification	de	certaines	exigences.	Il	est	inclus	de	façon	à	fournir	un	aperçu
supplémentaire	du	raisonnement	suivi	au	sein	des	comités	et	ayant	abouti	à	une	exigence,	ainsi	que	de
l’identification	des	DANGERS	traités	par	l’exigence.
L’Annexe	BB	propose	une	enquête	bibliographique	relative	à	la	détermination	de	la	température	de
sécurité	maximale	de	l’interface	entre	le	CAPTEUR	DE	L’OXYMETRE	DE	POULS	et	le	tissu	du	PATIENT.
L’Annexe	CC	s’intéresse	à	la	fois	aux	formules	mathématiques	utilisées	pour	évaluer	l’EXACTITUDE	de	la
SpO 	des	mesurages	de	l’OXYMETRE	DE	POULS	et	aux	noms	attribués	à	ces	formules.
2
L’Annexe	DD	présente	des	recommandations	relatives	aux	situations	dans	lesquelles	un	étalonnage	de
l’OXYMETRE	DE	POULS	par	une	analyse	de	sang	in	vitro	est	nécessaire.
L’Annexe	EE	 propose	 des	 lignes	 directrices	 pour	 l’ETUDE	 DE	 DESATURATION	 CONTROLEE	 permettant
l’étalonnage	de	l’OXYMETRE	DE	POULS.
L’Annexe	FF	est	une	introduction	didactique	à	différents	types	de	dispositifs	d’essai	utilisés	dans
l’oxymétrie	de	pouls.
L’Annexe	GG	définit	les	concepts	relatifs	au	temps	de	réponse	de	l’OXYMETRE	DE	POULS.
L’Annexe	HH	décrit	les	exigences	relatives	aux	interfaces	de	données.
L’Annexe	II	 comprend	 une	 Référence	 aux	 PRINCIPES	 ESSENTIELS	qui	se	trouvait	auparavant	dans
l’Annexe	HH.
Cette	publication	a	été	rédigée	selon	les	Directives	ISO/IEC,	Partie	2.
Dans	le	présent	document,	les	types	de	polices	suivants	sont	utilisés:
 exigences	et	définitions:	caractères	cambria;
 spécifications	d’essai:	caractères	italiques;
 les	indications	de	nature	informative	apparaissant	hors	des	tableaux,	comme	les	notes,	les
exemples	et	les	références:	en	petits	caractères;	le	texte	normatif	des	tableaux	est	également	en
petits	caractères;
©	ISO	2017	–	Tous	droits	réservés
vii
---------------------- Page: 7 ----------------------
ISO/FDIS	80601‐2‐61:2017(F)
1)
 les	TERMES	DEFINIS	DANS	L’ARTICLE	3	DE	LA	NORME	GENERALE ,	DANS	LE	PRESENT	DOCUMENT	OU	EN	NOTE:	EN	PETITES
MAJUSCULES.
Concernant	la	structure	du	présent	document,	le	terme:
 «article»	désigne	l’une	des	six	sections	numérotées	dans	la	table	des	matières,	avec	toutes	ses
subdivisions	(par	exemple	l’Article	201	inclut	les	paragraphes	201.7.1	et	201.7.2);	et
 «paragraphe»	 désigne	 une	 subdivision	 numérotée	 d’un	 article	 (par	 exemple	 201.7.1,	 7,2	 et
201.7.2.1	sont	tous	des	paragraphes	de	l’Article	201.7).
Les	références	à	des	articles	du	présent	document	sont	précédées	du	terme	«Article»	suivi	du	numéro
de	l’article	concerné.	Dans	le	présent	document,	les	références	aux	paragraphes	utilisent	uniquement	le
numéro	du	paragraphe	concerné.
Dans	le	présent	document,	la	conjonction	«ou»	est	utilisée	avec	la	valeur	d’un	«ou	inclusif»;	ainsi	un
énoncé	est	vrai	si	une	combinaison	des	conditions	quelle	qu’elle	soit	est	vraie.
Les	formes	verbales	utilisées	dans	le	présent	document	sont	conformes	à	l’usage	décrit	dans	l’Annexe	H
des	directives	ISO/IEC,	Partie	2.	Pour	les	besoins	du	présent	document,	la	forme	verbale:
 «	devoir	»	signifie	que	la	satisfaction	à	une	exigence	ou	à	un	essai	est	obligatoire	pour	la	conformité
au	présent	document;
 «il	convient/il	est	recommandé»	signifie	que	la	satisfaction	à	une	exigence	ou	à	un	essai	est
recommandée,	mais	n’est	pas	obligatoire	pour	la	conformité	au	présent	document;	et
 «pouvoir»	mis	au	présent	de	l’indicatif	est	utilisé	pour	décrire	un	moyen	admissible	pour	satisfaire
à	une	exigence	ou	à	un	essai.
Lorsqu’un	astérisque	(*)	est	utilisé	comme	premier	caractère	devant	un	titre,	au	début	d’un	alinéa	ou
devant	un	titre	de	tableau,	il	indique	l’existence	d’une	recommandation	ou	d’une	justification	à
consulter	dans	l’Annexe	AA.
1)
	 La	norme	générale	est	l’IEC	60601‐1:2005	et	l’IEC	60601‐1:2005/AMD1:2012,	Appareils	électromédicaux	—
Partie	1:	Exigences	générales	pour	la	sécurité	de	base	et	les	performances	essentielles.
©	ISO	2017	–	Tous	droits	réservés
viii
---------------------- Page: 8 ----------------------
PROJET	FINAL	DE	NORME	INTERNATIONALE	 ISO/FDIS	80601‐2‐61:2017(F)
Appareils	électromédicaux	—	Partie	2‐61:	Exigences
particulières	pour	la	sécurité	de	base	et	les	performances
essentielles	pour	les	oxymètres	de	pouls
201.1	 Domaine	d’application,	objet	et	normes	apparentées
L’Article	1	de	l’IEC	60601‐1:2005+AMD1:2012	s’applique,	avec	l’exception	suivante:
201.1.1	*	Domaine	d’application
Remplacement:
Le	présent	document	s’applique	à	la	SECURITE	DE	BASE	et	aux	PERFORMANCES	ESSENTIELLES	des	OXYMETRES
DE	POULS	conçus	pour	une	utilisation	chez	les	êtres	humains,	ci‐après	dénommés	APPAREILS	EM.	Sont
inclus	tous	les	éléments	de	l’appareil	nécessaires	pour	une	UTILISATION	NORMALE,	y	compris	le	MONITEUR
DE	L’OXYMETRE	DE	POULS,	le	CAPTEUR	DE	L’OXYMETRE	DE	POULS	et	le	CABLE	DE	RACCORDEMENT	DU	CAPTEUR.
Ces	exigences	s’appliquent	de	la	même	manière	aux	OXYMETRES	DE	POULS,	y	compris	aux	MONITEURS
D’OXYMETRE	DE	POULS,	aux	CAPTEURS	D’OXYMETRE	DE	POULS	et	aux	CABLES	DE	RACCORDEMENT	DU	CAPTEUR,	qui
ont	été	REMIS	EN	ETAT.
L’utilisation	prévue	des	OXYMETRES	DE	POULS	inclut	notamment,	sans	toutefois	s’y	limiter,	l’estimation	de
la	saturation	en	oxygène	de	l’hémoglobine	artérielle	et	de	la	fréquence	du	pouls	chez	des	PATIENTS	en
établissement	de	soins	professionnel	ainsi	que	chez	les	PATIENTS	dans	l’ENVIRONNEMENT	DES	SOINS	A
DOMICILE	et	dans	l’ENVIRONNEMENT	DES	SERVICES	MEDICAUX	D’URGENCE.
Le	présent	document	ne	s’applique	pas	aux	OXYMETRES	DE	POULS	destinés	à	être	utilisés	dans	le	cadre	de
recherches	en	laboratoire,	de	même	qu’aux	oxymètres	nécessitant	un	échantillon	de	sang	du	PATIENT.
Si	 un	 article	 ou	 un	 paragraphe	 est	 spécifiquement	 destiné	 à	 être	 applicable	 uniquement	 aux
APPAREILS	EM	ou	uniquement	aux	SYSTEMES	EM,	le	titre	et	le	contenu	de	cet	article	ou	de	ce	paragraphe
l’indiquent.	Si	cela	n’est	pas	le	cas,	l’article	ou	le	paragraphe	s’applique	à	la	fois	aux	APPAREILS	EM	et	aux
SYSTEMES	EM,	selon	le	cas.
Les	DANGERS	inhérents	à	la	fonction	physiologique	prévue	de	l’APPAREIL	EM	ou	des	SYSTEMES	EM	dans	le
cadre	 du	 domaine	 d’application	 du	 présent	 document	 ne	 sont	 pas	couverts	par	des	exigences
spécifiques	contenues	dans	le	présent	document,	à	l’exception	de	201.11	et	de	7.2.13	et	8.4.1	de	la
norme	générale.
NOTE	1	 Voir	également	4.2	de	la	norme	générale.	La	norme	générale	est	l’IEC	60601‐1:2005+AMD1:2012,
Appareils	électromédicaux	—	Partie	1:	Exigences	générales	pour	la	sécurité	de	base	et	les	performances	essentielles.
Le	présent	document	peut	également	être	appliqué	aux	APPAREILS	EM	et	à	leurs	ACCESSOIRES	utilisés	pour
l’atténuation	d’une	maladie,	la	compensation	ou	l’atténuation	d’une	blessure	ou	d’une	incapacité.
Le	présent	document	ne	s’applique	pas	aux	OXYMETRES	DE	POULS	destinés	uniquement	à	la	surveillance
de	fœtus.
Le	présent	document	ne	s’applique	pas	aux	appareils	distants	ou	esclaves	(secondaires)	qui	affichent	les
valeurs	de	la	SpO 	et	qui	sont	situés	à	l’extérieur	de	l’ENVIRONNEMENT	DU	PATIENT.
2
NOTE	2	 Il	est	attendu	que	les	APPAREILS	EM	qui	permettent	une	sélection	entre	des	fonctions	de	diagnostic	et	de
surveillance	satisfassent	aux	exigences	du	document	approprié	lorsqu’ils	sont	configurés	pour	ladite	fonction.
©	ISO	2017	–	Tous	droits	réservés
1
---------------------- Page: 9 ----------------------
ISO/FDIS	80601‐2‐61:2017(F)
Le	présent	document	est	applicable	aux	OXYMETRES	DE	POULS	destinés	à	être	utilisés	dans	des	conditions
environnementales	extrêmes	ou	non	maîtrisées	à	l’extérieur	de	l’environnement	hospitalier	ou	d’un
cabinet	de	médecin,	telles	que	dans	des	ambulances	ou	dans	les	transports	aériens.	Des	normes
supplémentaires	peuvent	s’appliquer	aux	OXYMETRES	DE	POULS	pour	ces	environnements	d’utilisation.
Le	 présent	 document	 est	 une	 norme	 particulière	 des	 séries	 de	 normes	 de	 l’IEC	60601‐1	 et	 de
l’ISO/IEC	80601.
201.1.2	Objet
Le	paragraphe	1.2	de	la	norme	générale	est	remplacé	par:
L’objet	du	présent	document	est	d’établir	des	exigences	particulières	relatives	à	la	SECURITE	DE	BASE	et
aux	PERFORMANCES	ESSENTIELLES	applicables	aux	OXYMETRES	DE	POULS	[tels	que	définis	en	201.3.217]	et	à
leurs	ACCESSOIRES.
NOTE	 Les	ACCESSOIRES	sont	inclus,	car	il	est	nécessaire	que	la	combinaison	du	MONITEUR	D’OXYMETRE	DE	POULS	et
des	ACCESSOIRES	présente	un	niveau	de	sécurité	adéquat.	Les	ACCESSOIRES	peuvent	avoir	une	incidence	significative
sur	la	SECURITE	DE	BASE	ou	les	PERFORMANCES	ESSENTIELLES	d’un	OXYMETRE	DE	POULS.
201.1.3	Normes	collatérales
Le	paragraphe	1.3	de	l’IEC	60601‐1:2005+AMD1:2012	s’applique,	avec	l’ajout	suivant:
Le	présent	document	fait	référence	aux	normes	collatérales	applicables	figurant	dans	l’Article	2	de	la
norme	générale	et	l’Article	201.2	du	présent	document.
L’IEC	60601‐1‐2,	l’IEC	60601‐1‐6,	l’IEC	60601‐1‐8,	l’IEC	60601‐1‐11	et	l’IEC	60601‐1‐12	s’appliquent
[2]
tels	que	modifiés	dans	les	Articles	202,	206,	208,	211	et	212,	respectivement.	L’IEC	60601‐1‐3 	ne
s’applique	pas.	Toutes	les	autres	normes	collatérales	publiées	dans	la	série	de	normes	IEC	60601‐1
s’appliquent	telles	quelles.
201.1.4	Normes	particulières
Le	paragraphe	1.4	de	l’IEC	60601‐1:2005+AMD1:2012	est	remplacé	par:
Dans	la	série	IEC	60601,	des	normes	particulières	peuvent	modifier,	remplacer	ou	supprimer	des
exigences	contenues	dans	la	norme	générale	et	dans	les	normes	collatérales,	en	fonction	de	ce	qui	est
approprié	à	l’APPAREIL	EM	spécifique	considéré	et	peuvent	ajouter	d’autres	exigences	de	SECURITE	DE	BASE
ou	de	PERFORMANCES	ESSENTIELLES.
Une	exigence	d’une	norme	particulière	prévaut	sur	l’exigence	correspondante	de	la	norme	générale	ou
des	normes	collatérales.
Par	souci	de	concision,	l’IEC	60601‐1:2005+AMD1:2012	est	appelée	«la	norme	générale»	dans	le
présent	document.	Il	est	fait	référence	aux	normes	collatérales	par	leur	numéro.
La	numérotation	des	articles	et	des	paragraphes	du	présent	document	correspond	à	celle	de	la	norme
générale	avec	le	préfixe	«201»	(par	exemple	dans	le	présent	document,	201.1	reprend	le	contenu	de
l’Article	1	de	la	norme	générale)	ou	de	la	norme	collatérale	applicable	avec	le	préfixe	«2xx»,	où	xx
correspond	aux	derniers	chiffres	du	numéro	de	document	de	la	norme	collatérale	(par	exemple	dans	le
présent	document,	202.4	reprend	le	contenu	de	l’Article	4	de	la	norme	collatérale	IEC	60601‐1‐2,	208.4
reprend	le	 contenu	de	l’Article	4	 de	la	 norme	collatérale	IEC	60601‐1‐8,	etc.).	Les	modifications
apportées	au	texte	de	la	norme	générale	sont	spécifiées	par	l’utilisation	des	mots	suivants:
«Remplacement»	signifie	que	l’article	ou	le	paragraphe	de	la	norme	générale	ou	de	la	norme	collatérale
est	remplacé	en	intégralité	par	le	texte	du	présent	document.
©	ISO	2017	–	Tous	droits	réservés
2
---------------------- Page: 10 ----------------------
ISO/FDIS	80601‐2‐61:2017(F)
«Ajout»	signifie	que	le	texte	du	présent	document	est	ajouté	aux	exigences	de	la	norme	générale	ou	de	la
norme	collatérale	applicable.
«Amendement»	signifie	que	l’article	ou	le	paragraphe	de	la	norme	générale	ou	de	la	norme	collatérale
applicable	est	amendé	comme	indiqué	par	le	texte	du	présent	document.
Les	paragraphes	ou	les	figures	qui	ont	été	ajoutés	par	rapport	à	ceux	de	la	norme	générale	sont
numérotés	en	commençant	à	partir	de	201.101,	les	annexes	supplémentaires	sont	appelées	AA,	BB,	etc.,
et	les	points	supplémentaires	sont	appelés	aa),	bb),	etc.
Les	paragraphes	ou	les	figures	qui	ont	été	ajoutés	par	rapport	à	ceux	d’une	norme	collatérale	sont
numérotés	en	commençant	à	partir	de	2xx,	où	«x»	est	le	numéro	de	la	norme	collatérale	standard,	par
exemple	202	pour	l’IEC	60601‐1‐2,	203	pour	l’IEC	60601‐1‐3,	etc.
L’expression	«présent	document»	est	utilisée	pour	faire	référence	à	la	norme	générale,	aux	normes
collatérales	applicables	et	au	présent	document	particulier	considérés	conjointement.
Si	aucun	article	ni	aucun	paragraphe	correspondant	n’existe	dans	le	présent	document,	la	section,
l’article	ou	le	paragraphe	de	la	norme	générale	ou	de	la	norme	collatérale	applicable,	même	s’il	peut	ne
pas	être	pertinent,	s’applique	sans	aucune	modification;	lorsqu’il	est	prévu	que	l’une	des	parties	de	la
norme	générale	ou	de	la	norme	collatérale	applicable	ne	s’applique	pas,	même	si	elle	peut	être
pertinente,	une	mention	est	donnée	à	cet	effet	dans	le	présent	document	particulier.
201.2	 Références	normatives
Les	documents	suivants	cités	dans	le	texte	constituent,	pour	tout	ou	partie	de	leur	contenu,	des
exigences	du	présent	document.	Pour	les	références	datées,	seule	l’édition	citée	s’applique.	Pour	les
références	non	datées,	la	dernière	édition	du	document	de	référence	s'applique	(y	compris	les	éventuels
amendements).
NOTE	1	 La	manière	dont	ces	documents	référencés	sont	cités	dans	les	exigences	normatives	détermine	l’étendue
(en	totalité	ou	en	partie)	de	leur	application.
NOTE	2	 La	Bibliographie	fournit	une	liste	de	références	informatives.
L’Article	2	de	l’IEC	60601‐1:2005+AMD1:2012	s’applique,	avec	l’exception	suivante:
Remplacement:
IEC	60417,	Symboles	graphiques	utilisables	sur	le	matériel.
IEC	60529:2013,	Degrés	de	protection	procurés	par	les	enveloppes	(Code	IP).
IEC	60601‐1‐2:2014,	Appareils	électromédicaux	—	Partie	1‐2:	Exigences	générales	pour	la	sécurité	de
base	et	les	performances	essentielles	—	Norme	collatérale:	Perturbations	électromagnétiques	—	Exigences
et	essais.
IEC	60601‐1‐6:2010,	Appareils	électromédicaux	—	Partie	1‐6:	Exigences	générales	pour	la	sécurité	de
base	et	les	performances	essentielles	—	Norme	collatérale:	Aptitude	à	l’utilisation	+Amendement	1:2013.
IEC	60601‐1‐8:2006,	Appareils	électromédicaux	—	Partie	1‐8:	Exigences	générales	pour	la	sécurité	de
base	et	les	performances	essentielles	—	Norme	collatérale:	Exigences	générales,	essais	et	guide	pour	les
systèmes	d’alarme	des	appareils	et	des	systèmes	électromédicaux	+Amendement	1:2012.
IEC	60825‐1:2014,	Sécurité	des	appareils	à	laser	—	Partie	1:	Classification	des	matériels	et	exigences.
©	ISO	2017	–	Tous	droits	réservés
3
---------------------- Page: 11 ----------------------
ISO/FDIS	80601‐2‐61:2017(F)
Ajout:
ISO	3744:2010,	Acoustique	—	Détermination	des	niveaux	de	puissance	acoustique	et	des	niveaux	d’énergie
acoustique	émis	par	les	sources	de	bruit	à	partir	de	la	pression	acoustique	—	Méthodes	d’expertise	pour
des	conditions	approchant	celles	du	champ	libre	sur	plan	réfléchissant.
ISO	4135:2001,	Matériel	d’anesthésie	et	de	réanimation	respiratoire	—	Vocabulaire.
ISO	7000,	Symboles	graphiques	utilisables	sur	le	matériel	—	Symboles	enregistrés.
ISO	7010,	Symboles	graphiques	—	Couleurs	de	sécurité	et	signaux	de	sécurité	—	Signaux	de	sécurité
enregistrés.
ISO	14155:2011,	Investigation	clinique	des	dispositifs	médicaux	pour	sujets	humains	—	Bonnes	pratiques
cliniques.
ISO	14937:2009,	Stérilisation	des	produits	de	santé	—	Exigences	générales	pour	la	caractérisation	d’un
agent	stérilisant	et	pour	la	mise	au	point,	la	validation	et	la	vérification	de	routine	d’un	processus	de
stérilisation	pour	dispositifs	médicaux.
ISO	15223‐1:2016,	 Dispositifs	 médicaux	—	 Symboles	 à	 utiliser	 avec	 les	 étiquettes,	 l’étiquetage	 et	 les
informations	à	fournir	relatifs	aux	dispositifs	médicaux	—	Partie	1:	Exigences	générales.
IEC	60068‐2‐27:2008,	 Essais	 d’environnement	—	 Partie	2‐27:	 Essais	—	 Essai	 Ea	 et	 guide:	 Chocs
+Amendement	1:2013.
IEC	60068‐2‐31:2008,	 Essais	 d’environnement	—	 Partie	2‐31:	 Essais	—	 Essai	 Ec:	 Choc	 lié	 à	 des
manutentions	brutales,	essai	destiné	en	premier	lieu	aux	matériels.
IEC	60068‐2‐64:2008,	Essais	d’environnement	—	Partie	2‐64:	Essais	—	Essais	Fh:	Vibrations	aléatoires	à
large	bande	et	guide.
IEC	60601‐1‐11:2015,	Appareils	électromédicaux	—	Partie	1‐11:	Exigences	générales	pour	la	sécurité	de
base	et	les	performances	essentielles	—	Norme	collatérale:	Exigences	pour	les	appareils	électromédicaux	et
les	systèmes	électromédicaux	utilisés	dans	l’environnement	des	soins	à	domicile.
IEC	60601‐1‐12:2014,	Appareils	électromédicaux	—	Partie	1‐12:	Exigences	générales	pour	la	sécurité	de
base	et	les	performances	essentielles	—	Norme	collatérale:	Exigences	pour	les	appareils	électromédicaux	et
les	 systèmes	 électromédicaux	 destinés	 à	 être	 utilisés	 dans	 l’environnement	 des	 services	 médicaux
d’urgence.
IEC	60825‐2:2004,	Sécurité	des	appareils	à	laser	—	Partie	2:	Sécurité	des	systèmes	de	télécommunication
par	fibres	optiques	(STFO)	+	Amendement	1:2006+	Amendement	2:20
 ...