IEC 60601-1-10
®

Edition 1.0 2013-11
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
AMENDMENT 1
AMENDEMENT 1

Medical electrical equipment –
Part 1-10: General requirements for basic safety and essential performance –
Collateral Standard: Requirements for the development of physiologic
closed-loop controllers

Appareils électromédicaux –
Partie 1-10: Exigences générales pour la sécurité de base et les performances
essentielles – Norme collatérale: Exigences pour le développement des
régulateurs physiologiques en boucle fermée

IEC 60601-1-10:2007/A1:2013

---------------------- Page: 1 ----------------------
THIS PUBLICATION IS COPYRIGHT PROTECTED
Copyright © 2013 IEC, Geneva, Switzerland

All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester.
If you have any questions about IEC copyright or have an enquiry about obtaining additional rights to this publication,
please contact the address below or your local IEC member National Committee for further information.


Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni
utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et les
microfilms, sans l'accord écrit de la CEI ou du Comité national de la CEI du pays du demandeur.
Si vous avez des questions sur le copyright de la CEI ou si vous désirez obtenir des droits supplémentaires sur cette
publication, utilisez les coordonnées ci-après ou contactez le Comité national de la CEI de votre pays de résidence.

IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé Fax: +41 22 919 03 00
CH-1211 Geneva 20 info@iec.ch
Switzerland www.iec.ch

About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.

About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.

Useful links:

IEC publications search - www.iec.ch/searchpub Electropedia - www.electropedia.org
The advanced search enables you to find IEC publications The world's leading online dictionary of electronic and
by a variety of criteria (reference number, text, technical electrical terms containing more than 30 000 terms and
committee,…). definitions in English and French, with equivalent terms in
It also gives information on projects, replaced and additional languages. Also known as the International
withdrawn publications. Electrotechnical Vocabulary (IEV) on-line.

IEC Just Published - webstore.iec.ch/justpublished Customer Service Centre - webstore.iec.ch/csc
Stay up to date on all new IEC publications. Just Published If you wish to give us your feedback on this publication
details all new publications released. Available on-line and or need further assistance, please contact the
also once a month by email. Customer Service Centre: csc@iec.ch.


A propos de la CEI
La Commission Electrotechnique Internationale (CEI) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.

A propos des publications CEI
Le contenu technique des publications de la CEI est constamment revu. Veuillez vous assurer que vous possédez
l’édition la plus récente, un corrigendum ou amendement peut avoir été publié.

Liens utiles:

Recherche de publications CEI - www.iec.ch/searchpub Electropedia - www.electropedia.org
La recherche avancée vous permet de trouver des Le premier dictionnaire en ligne au monde de termes
publications CEI en utilisant différents critères (numéro de électroniques et électriques. Il contient plus de 30 000
référence, texte, comité d’études,…). termes et définitions en anglais et en français, ainsi que
Elle donne aussi des informations sur les projets et les les termes équivalents dans les langues additionnelles.
publications remplacées ou retirées. Egalement appelé Vocabulaire Electrotechnique
International (VEI) en ligne.
Just Published CEI - webstore.iec.ch/justpublished
Service Clients - webstore.iec.ch/csc
Restez informé sur les nouvelles publications de la CEI.
Just Published détaille les nouvelles publications parues. Si vous désirez nous donner des commentaires sur
Disponible en ligne et aussi une fois par mois par email. cette publication ou si vous avez des questions
contactez-nous: csc@iec.ch.

---------------------- Page: 2 ----------------------
IEC 60601-1-10

®


Edition 1.0 2013-11




INTERNATIONAL



STANDARD




NORME



INTERNATIONALE




AMENDMENT 1

AMENDEMENT 1





Medical electrical equipment –

Part 1-10: General requirements for basic safety and essential performance –

Collateral Standard: Requirements for the development of physiologic


closed-loop controllers



Appareils électromédicaux –


Partie 1-10: Exigences générales pour la sécurité de base et les performances

essentielles – Norme collatérale: Exigences pour le développement des

régulateurs physiologiques en boucle fermée










INTERNATIONAL

ELECTROTECHNICAL

COMMISSION


COMMISSION

ELECTROTECHNIQUE

PRICE CODE
INTERNATIONALE

E
CODE PRIX


ICS 11.040 ISBN 978-2-83221-133-5



Warning! Make sure that you obtained this publication from an authorized distributor.

Attention! Veuillez vous assurer que vous avez obtenu cette publication via un distributeur agréé.

® Registered trademark of the International Electrotechnical Commission
Marque déposée de la Commission Electrotechnique Internationale

---------------------- Page: 3 ----------------------
– 2 – 60601-1-10 Amend. 1 © IEC:2013
FOREWORD
This amendment has been prepared by subcommittee 62A: Common aspects of electrical
equipment used in medical practice, of IEC technical committee 62: Electrical equipment in
medical practice, and ISO subcommittees SC1: Breathing attachments and anaesthetic
machines, and SC3: Lung ventilators and related devices of ISO technical committee 121:
Anaesthetic and respiratory equipment.
The text of this amendment is based on the following documents:
FDIS Report on voting
62A/888/FDIS 62A/896/RVD

Full information on the voting for the approval of this amendment can be found in the report
on voting indicated in the above table. In ISO, the standard has been approved by 15 P-
members out of 15 having cast a vote.
The committee has decided that the contents of this amendment and the base publication will
remain unchanged until the stability date indicated on the IEC web site under
"http://webstore.iec.ch" in the data related to the specific publication. At this date, the
publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
_____________

INTRODUCTION TO THE AMENDMENT
The first edition of IEC 60601-1-10 was published in 2007. This amendment is intended to
update the references to IEC 60601-1:2005 to include Amendment 1:2012,
to update IEC 60601-1-6:2006 to IEC 60601-1-6:2010, including its Amendment 1 and
to update references to IEC 60601-1-8:2006 to include its Amendment 1:2012. This
amendment also removes the normative reference to IEC 62304:2006. This collateral
standard made reference to IEC 62304 because elements of the software process were not
fully covered by Clause 14 of IEC 60601-1:2005. Amendment 1 to IEC 60601-1:2005
incorporates the needed software process requirement into Clause 14. Therefore, it is
redundant and potentially confusing to have IEC 62304 explicitly called out in this collateral
standard.
FOREWORD
Add the following note at the end of the Foreword:
NOTE The attention of National Committees is drawn to the fact that equipment manufacturers and testing
organizations may need a transitional period following publication of a new, amended or revised IEC or
ISO publication in which to make products in accordance with the new requirements and to equip themselves for
conducting new or revised tests. It is the recommendation of the committee that the content of this publication be
adopted for implementation nationally not earlier than 3 years from the date of publication.

---------------------- Page: 4 ----------------------
60601-1-10 Amend. 1 © IEC:2013 – 3 –
1 Scope, object and related standards
1.3.1 IEC 60601-1
Replace the existing first dashed item with:
– "the general standard" designates IEC 60601-1 alone (IEC 60601-1:2005+A1:2012);
Replace the existing second dashed item with:
– "this collateral standard" designates IEC 60601-1-10 alone (IEC 60601-1-
10:2007+A1:2013)
2 Normative references
In the existing introductory paragraph, replace the first sentence with:
The following documents, in whole or in part, are normatively referenced in this document and
are indispensable for its application.
Replace the existing references to IEC 60601-1, IEC 60601-1-6 and IEC 60601-1-8 by the
following new references:
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic
safety and essential performance
Amendment 1:2012
IEC 60601-1-6:2010, Medical electrical equipment – Part 1-6: General requirements for basic
safety and essential performance – Collateral standard: Usability
Amendment 1:2013
IEC 60601-1-8:2006, Medical electrical equipment – Part 1-8: General requirements for basic
safety and essential performance – Collateral standard: General requirements, tests and
guidance for alarm systems in medical electrical equipment and medical electrical systems
Amendment 1:2012
Delete the following normative reference:
IEC 62304:2006, Medical device software – Software life cycle processes
Add the following normative reference:
IEC 62366:2007, Medical devices – Application of usability engineering to medical devices
3 Terms and definitions
Replace the existing introductory paragraph with:
For the purposes of this document, the terms and definitions given in
IEC 60601-1:2005+A1:2012, IEC 60601-1-6:2010+ A1:2013, IEC 60601-1-8:2006+A1:2012,
IEC 62366:2007 and the following apply.
7 * PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS)
Replace the existing text of the clause with the following:

---------------------- Page: 5 ----------------------
– 4 – 60601-1-10 Amend. 1 © IEC:2013
For ME EQUIPMENT and ME SYSTEMS that incorporate a PCLC and incorporate PEMS, the
requirements of Clause 14 of the general standard apply.
Annex A – General guidance and rationale
A.2 Rationale for particular clauses and subclauses
Subclause 5.1 – Instructions for use
In the existing first paragraph, replace "TRAINING (see IEC 60601-1-6)" with "training (see
IEC 62366)".
Subclause 6.1 – Usability
In the existing first paragraph, in the first sentence replace "IEC 60601-1-6" with "IEC 62366",
and in the penultimate sentence place "TRAINING" in normal font.
Clause 7 – PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS)
Replace the existing text of the rationale with the following:
Clause 7 of the first edition of this collateral standard applied the requirements of
IEC 62304:2006 to the software elements of the PEMS. This was in addition to the requirement
of Clause 14 of the general standard. Amendment 1 to the general standard incorporates the
software PROCESS requirements into Clause 14. Therefore, the additional requirements in
Clause 7 of this collateral standard are redundant and can, in effect, be deleted.
Subclause 8.1 – General
In the existing first paragraph, in the third sentence replace IEC 60601-1" with "the general
standard", and in the last sentence replace "IEC 60601-1" with "general standard", and delete
"or IEC 62304".
Subclause 8.2.2.4 – Specification of operating conditions
In the existing second dashed item, place "TRAINING" in normal font.
Index of defined terms used with this collateral standard
Replace the references to the following defined terms with:
ALARM CONDITION . IEC 60601-1
...