Status
Not Published
Publication Date
14-Dec-2027
Current Stage
4060 - Enquiry results established and sent to TC, SR, BTTF - Enquiry
Start Date
02-Oct-2026
Due Date
02-Oct-2026
Completion Date
02-Oct-2026

Buy Documents

Draft

prEN IEC 61757-2-2:2026 - BARVE

English language (33 pages)
Preview
Preview
e-Library read for
×1 day

prEN IEC 61757-2-2:2026 is a draft standard from CLC for fibre optic sensors, focused on temperature measurement by distributed sensing. It specifies the terminology, test setups, measurement procedures, and performance parameters for distributed fibre optic temperature sensing systems (DTS), so manufacturers, test laboratories, and users can measure and report DTS behaviour in a consistent way.

What does prEN IEC 61757-2-2:2026 specify?

prEN IEC 61757-2-2:2026 specifies how to characterize distributed fibre optic temperature sensing systems. The document covers Raman scattering, Brillouin scattering, and Rayleigh scattering based measurements, and it defines how to test and report the main DTS performance parameters.

The standard is organized around a general test setup in Clause 4 and measurement procedures in Clause 5. Annex A provides a reporting table for measured performance values, and Annex B gives an informative method for studying point defect effects.

AnnexWhat it covers
Annex ABlank measurement parameter performance table for recording and quoting DTS results
Annex BInformative procedure for point defect effects, including temperature dead zone and point defect temperature offset

What are the key requirements of prEN IEC 61757-2-2:2026?

The core of prEN IEC 61757-2-2:2026 is a set of controlled test methods for DTS performance. Clause 4 defines the general setup, and Clause 5 turns that setup into repeatable procedures for the main parameters that matter in practice: error, resolution, repeatability, stability, warm-up, and attenuation range.

Clause 4 - use a controlled and documented test setup

Clause 4 requires a general setup for single-ended and loop configurations, with thermal enclosures, traceable calibrated reference temperature sensors, and low-loss fibre connections. The fibre should be the type recommended by the manufacturer unless otherwise agreed. In practice, this means the test arrangement is not left to ad hoc laboratory practice - the setup, fibre lengths, operating wavelength, sample spacing, spatial resolution setting, and operating temperature limits all have to be recorded.

Clause 4 also says the DTS device should be calibrated according to the manufacturer's recommendations before measurements begin. The reference temperature measurement uncertainty must be at least five times smaller than the temperature measurement error being assessed. That matters because the reference has to be clearly better than the instrument under test.

The main performance parameters are measured by specific trace-based methods

Clause 5.1 defines temperature measurement error using 20 traces at three fibre temperatures and three measurement times. The result is based on a centred, uniformly weighted moving average over 51 points, and the worst absolute average error is reported. This tells a buyer or user how far the DTS output can drift from the true fibre temperature under controlled conditions.

Clause 5.2 defines spatial resolution by testing three fibre lengths relative to the claimed resolution. The sample spacing must be no more than half the claimed spatial resolution, and the result is the fibre length that satisfies the response criteria. In practice, this shows the smallest hot spot the DTS device can resolve.

Clause 5.3 and 5.4 use the same 20-trace data sets to quantify temperature repeatability and spatial temperature uncertainty. Repeatability is based on twice the standard deviation over time, while spatial temperature uncertainty is based on twice the standard deviation across adjacent points in a trace, then averaged over traces. These parameters matter when you need to know whether the instrument gives stable results and how sharply it localizes temperature along the fibre.

Environmental behaviour is tested across the operating range

Clause 5.5 evaluates environmental temperature stability by cycling the DTS device through low, middle, and high instrument temperatures. The clause defines worst-case environmental temperature effect, low and high environmental temperature effects, and environmental temperature repeatability. For practical use, this shows whether the DTS device keeps its readings stable as the instrument itself heats up or cools down.

Clause 5.6 measures warm-up time. The DTS device is powered off, stabilized, powered on, and then timed until the main performance parameters reach their claimed values. Warm-up time is reported separately for nominal, minimum, and maximum operating temperatures. This is useful when the system must be ready for a job quickly after start-up or after an environmental change.

Clause 5.7 checks attenuation range

Clause 5.7 defines attenuation range as the amount of total loss the DTS system can tolerate while still meeting performance, using a variable optical attenuator (VOA). The test evaluates temperature repeatability and, by the same method, spatial temperature uncertainty at 90 % of the total fibre length. In practice, this helps users judge how much connector, splice, fibre, or attenuator loss a system can handle before performance degrades.

The document also requires structured reporting

Clause 4.2 lists the information that should be documented with the results, including the setup configuration, fibre details, wavelengths, distance measurement range, sample spacing, spatial resolution setting, enclosure temperatures, and operating limits. That is important because DTS performance depends strongly on the exact fibre path and test conditions, so quoted results are only meaningful when those conditions are recorded.

What terms does prEN IEC 61757-2-2:2026 define?

  • Distributed fibre optic temperature sensing system (DTS) - the measurement set-up consisting of a distributed fibre optic sensor and a DTS device, with processor, data archive, and user interface, that provides spatially resolved temperature measurement.
  • Attenuation range - the total cumulated optical loss tolerated by the DTS system without degrading specified measurement performance beyond a given factor.
  • Distance measurement range - the maximum distance along the fibre within which the instrument meets its specified performance under defined conditions.
  • Spatial resolution - the smallest heated length of fibre for which the DTS device can measure the hot spot within the specified temperature measurement error.
  • Temperature measurement error - the maximum difference between the measured temperature, after the specified smoothing method, and the reference temperature over the full operating range and acquisition times.
  • Temperature repeatability - the precision of temperature data from repeated traces at a given location, expressed as twice the standard deviation.
  • Warm-up time - the time needed after power-on for the DTS device to reach its claimed performance at a given operating temperature.
  • Point defect temperature offset - the difference between the average temperatures measured on the two sides of a point defect when the actual fibre temperature is the same.

Who uses prEN IEC 61757-2-2:2026?

prEN IEC 61757-2-2:2026 is used by manufacturers of DTS devices, calibration and test laboratories, and organizations that specify or purchase fibre optic temperature sensing systems. It is also relevant to quality managers and engineers who need to compare quoted performance with measured performance.

In practice, these users apply it to:

  • design a valid DTS test setup
  • verify claimed temperature performance
  • compare single-ended and loop configurations
  • document performance data for procurement, qualification, or internal validation
  • investigate the effect of attenuation or point defects on measurement quality

What changed in prEN IEC 61757-2-2:2026 from the previous edition?

prEN IEC 61757-2-2:2026 cancels and replaces the first edition published in 2016. The document states that this edition is a technical revision with these significant changes:

  • updated terms and definitions
  • addition of a symbol list
  • updates to Figures 1, 2, 6, 9, 10, and B.1
  • addition of Table 1

Which standards are used with prEN IEC 61757-2-2:2026?

  • IEC 60050 (all parts), International Electrotechnical Vocabulary - provides common electrotechnical terminology used in the document.
  • IEC 61757:2026, Fibre optic sensors - Generic specification - provides the generic framework that this part builds on.
  • IEC TR 61931, Fibre optic - Terminology - supports the fibre optic terminology used for DTS.
  • ISO/IEC Guide 99, International vocabulary of metrology - Basic and general concepts and associated terms (VIM) - supports metrology terms such as uncertainty and traceability.

Clause 4.1 also cites SEAFOM-MSP-01:2016 as further information on general test setups and measurement procedures.

What does the prEN IEC 61757-2-2:2026 document contain?

The document contains two general test setup figures, one for single-ended configuration and one for loop configuration, plus figures that illustrate the calculation methods for error, resolution, repeatability, uncertainty, warm-up time, and point defect effects. It also contains Table 1 for the environmental temperature cycle and Table A.1 for recording performance results.

Clause 5 gives the step-by-step procedures and the calculation methods, including moving averages, standard deviations, and peak-to-peak comparisons. Annex A provides a ready-made reporting table for quoted or measured values, which is useful for testing, certification, and technical comparison of devices. Annex B adds a controlled method for examining how a point defect changes the measurement trace, including temperature dead zone and offset.

Relations

Effective Date
02-Dec-2025

Buy Documents

Draft

prEN IEC 61757-2-2:2026 - BARVE

English language (33 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

ANCE

Mexican certification and testing association.

EMA Mexico Verified

Intertek Slovenia

Intertek testing, inspection, and certification services in Slovenia.

UKAS Slovenia Verified

LNE (Laboratoire National de Métrologie et d'Essais)

French national laboratory for metrology and testing.

COFRAC France Verified

Sponsored listings

Frequently Asked Questions

prEN IEC 61757-2-2:2026 is a draft published by CLC. Its full title is "Fibre optic sensors - Part 2-2: Temperature measurement - Distributed sensing". This standard covers: Fibre optic sensors - Part 2-2: Temperature measurement - Distributed sensing

Fibre optic sensors - Part 2-2: Temperature measurement - Distributed sensing

prEN IEC 61757-2-2:2026 is classified under the following ICS (International Classification for Standards) categories: 33.180.99 - Other fibre optic equipment. The ICS classification helps identify the subject area and facilitates finding related standards.

prEN IEC 61757-2-2:2026 has the following relationships with other standards: It is inter standard links to EN 61757-2-2:2017. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

prEN IEC 61757-2-2:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • 2 FOREWORD
  • 3 INTRODUCTION
  • 4 1 Scope
  • 5 2 Normative references
  • 6 3 Terms, definitions, abbreviated terms, and symbols
  • 7 3.1 Terms and definitions
  • 8 3.2 Abbreviated terms
  • 9 3.3 Symbols
  • 10 4 General test setups for measurement of performance parameters
  • 11 4.1 General and test setup requirements
  • 12 4.2 General required information to be documented
  • 13 5 Measurement procedures for performance parameters
  • 14 5.1 Temperature measurement error
  • 15 5.1.1 Test procedure and conditions
  • 16 5.1.2 Parameter calculation
  • 17 5.2 Spatial resolution
  • 18 5.2.1 Test procedure and conditions
  • 19 5.2.2 Parameter calculation
  • 20 5.3 Temperature repeatability
  • 21 5.3.1 Test procedure and conditions
  • 22 5.3.2 Parameter calculation
  • 23 5.3.3 Formulas
  • 24 5.4 Spatial temperature uncertainty
  • 25 5.4.1 Test procedure and conditions
  • 26 5.4.2 Parameter calculation
  • 27 5.5 Environmental temperature stability
  • 28 5.5.1 Test procedure and conditions
  • 29 5.5.2 Parameter calculation
  • 30 5.6 Warm-up time
  • 31 5.6.1 Test procedure and conditions
  • 32 5.6.2 Parameter calculation
  • 33 5.7 Attenuation range
  • 34 5.7.1 Test procedure and conditions
  • 35 5.7.2 Parameter calculation
  • 36 Annex A (informative) Measurement parameter performance table
  • 37 Annex B (informative) Point defect effects
  • 38 B.1 General
  • 39 B.2 Point defect
  • 40 B.3 Test procedures and conditions
  • 41 Bibliography
  • 43 Figure 1 – General test setup: single-ended configuration
  • 44 Figure 2 – General test setup: loop configuration
  • 45 Figure 3 – Temperature measurement error calculation: step a)
  • 46 Figure 4 – Temperature measurement error calculation: steps b) and c)

134 INTRODUCTION 135 This document is part of the IEC 61757:2026, which is dedicated to fibre optic sensors. Generic 136 specifications for fibre optic sensors are defined in IEC 61757:2026.

137 The individual parts of the IEC 61757:2026 are numbered as IEC 61757:2026-M-T, where M 138 denotes the measurand and T the technology of the fibre optic sensor. The IEC 61757-2-T 139 series is concerned with temperature measurements.

140 1 Scope 141 This part of IEC 61757 defines the terminology, structure, and measurement methods of 142 distributed fibre optic sensors for temperature sensing (DTS). DTS includes the use of Raman 143 scattering, Brillouin scattering and Rayleigh scattering effects. In addition, Raman scattering 144 and Rayleigh scattering based measurements are performed with a single-ended fibre 145 configuration only. Brillouin scattering based measurements are performed with a single -ended 146 fibre or fibre loop configuration. The technique accessible from both sides at same time (e. g. 147 Brillouin optical time domain analysis, BOTDA) is referred to here as a loop configuration. 148 This document specifies the most important DTS performance parameters and defines the 149 procedures for their determination. In addition to the group of performance parameters, a list of 150 additional parameters has been defined to support the definition of the measurement 151 specifications and their associated test procedures. The definitions of these additional 152 parameters are provided for informational purposes and should be included with the sets of 153 performance parameters.

154 A general test setup is defined in which all parameters can be gathered through a set of tests. 155 The specific tests are described within the clause for each measurement parameter. This 156 general test setup is depicted and described in Clause 4 along with a list of general information 157 that will be documented based upon the specific DTS device and test setup used to measure 158 these parameters in accordance with this standard.

159 Annex A provides a blank performance parameter table which should be used to record the 160 performance parameter values for a given DTS device and chosen optical test setup 161 configuration.

162 2 Normative references 163 The following documents are referred to in the text in such a way that some or all of their content 164 constitutes requirements of this document. For dated references, only the edition cited applies. 165 For undated references, the latest edition of the referenced document (including any 166 amendments) applies.

167 IEC 60050 (all parts), International Electrotechnical Vocabulary 168 IEC 61757:2026, Fibre optic sensors - Generic specification 169 IEC TR 61931, Fibre optic - Terminology 170 ISO/IEC Guide 99, International vocabulary of metrology - Basic and general concepts and 171 associated terms (VIM) 172 3 Terms, definitions, abbreviated terms, and symbols 173 3.1 Terms and definitions 174 For the purposes of this document, the terms and definitions given in IEC 61757:2026, IEC 175 60050 (all parts), IEC TR 61931, ISO/IEC Guide 99 (VIM) and the following apply. 176 ISO and IEC maintain terminology databases for use in standardization at the following 177 addresses:

178 – IEC Electropedia: available at https://www.electropedia.org/ 179 – ISO Online browsing platform: available at https://www.iso.org/obp 181 attenuation range 182 total cumulated optical loss (one way loss) tolerated by the DTS system without affecting the 183 specified measurement performance more than a given factor at a given location, spatial 184 resolution, and measurement time 185 Note 1 to entry: Part of the total cumulative loss can be the fibre attenuation, point defect losses introduced by 186 components such as connectors, splices, kink in the fibre, attenuators.

187 Note 2 to entry: The attenuation range is expressed in decibels (dB).

188 3.1.2 189 distributed fibre optic temperature sensing system 190 DTS 191 measurement set-up consisting of a distributed fibre optic sensor connected to a DTS device, 192 including processor, data archive, and user interface, which provides a spatially resolved 193 temperature measurement 194 Note 1 to entry: Synonyms for DTS device include DTS unit, DTS interrogator, DTS controller, and DTS instrument. 195 3.1.3 196 distance measurement range 197 maximum distance from the DTS device output connector along the fibre optic sensor within 198 which the instrument measures a temperature with specified measurement performance under 199 defined conditions 200 Note 1 to entry: Defined conditions are spatial resolution, spatial temperature uncertainty, and measurement time. 201 Note 2 to entry: This supporting parameter is closely related to the total accumulated optical loss (one way) tolerated 202 by the interrogation unit without affecting specified measurement performance. In test cases used to prove or verify 203 the reported specifications, the total fibre length is equal to or greater than the specified distance measurement 204 range, for the tolerated total accumulated optical loss.

205 Note 3 to entry: The distance measurement range is expressed in length units (m or km). 206 3.1.4 207 environmental temperature repeatability 208 difference of the measured constant fibre optic sensor temperature at a specified DTS device 209 temperature (e. g. nominal operating temperature) before and after temperature cycling of the 210 DTS device across the entire DTS device operating temperature range 211 Note 1 to entry: This parameter is derived from environmental temperature stability. 212 3.1.5 213 environmental temperature stability 214 difference of the measured constant fibre optic sensor temperature before, during and after 215 temperature cycling of the DTS device across the entire DTS device operating temperature 216 range 217 Note 1 to entry: Worst case environmental temperature effect, high/low environmental temperature effect, and 218 environmental temperature repeatability are derived from this definition.

219 3.1.6 220 high/low environmental temperature effect 221 difference of the measured constant fibre optic sensor temperature at the high and low 222 temperature limit of the DTS device temperature operating range 223 Note 1 to entry: This parameter is derived from environmental temperature stability. 224 3.1.7 225 hot spot 226 short length of fibre optic sensor (L ) which is exposed by a measurable temperature change HS 227 (ΔT) which is significantly bigger than the DTS device temperature repeatability and which is 228 confirmed by reference temperature devices in two thermal chambers 229 Note 1 to entry: See Clause 4 and Figure 6.

230 3.1.8 231 location 232 L 233 optical distance (specified in length units) from the DTS device output connector to a desired 234 temperature sample point along the fibre optic sensor 235 Note 1 to entry: The furthest location from DTS device output connector for the particular test is quantified as L m Z 236 and is often chosen to be the same as the distance measurement range for purposes of comparing the measurement 237 results with quoted specifications.

238 3.1.9 239 measurement time 240 time between independent temperature measurements when making successive measurements 241 on a single fibre optic sensor 242 Note 1 to entry: This parameter includes acquisition time and processing time for measured data. This parameter 243 is selectable by the user typically in some limited fashion. Multiple independent temperature measurements may be 244 averaged together to provide an overall measurement time.

245 Note 2 to entry: Equivalently, it is the time interval between successive temperature trace timestamps under these 246 conditions.

247 3.1.10 248 point defect 249 local variation in the optical properties of a fibre optic sensor, occurring at a single location or 250 over a length substantially shorter than the DTS device spatial resolution 251 Note 1 to entry: The definition of a point defect covers a variety of possibilities that can have similar effects on the 252 temperature measurement. Examples include 253 – a poor fibre splice or a splice between two fibres of different core diameter; 254 – a back reflection from a fibre connection; 255 – a local fibre position with high loss, e.g., a bend or kink.

256 3.1.11 257 point defect temperature offset 258 difference between the average values of the temperature sample points in two zones on the 259 temperature trace, one each side of a point defect, where the actual fibre optic sensor 260 temperatures are the same 261 Note 1 to entry: The point defect temperature offset can be positive, negative or zero. 262 3.1.12 263 sample spacing 264 distance between two consecutive temperature sample points in a single temperature trace 265 Note 1 to entry: See Example of a temperature trace with temperature sample points:

267 Note 2 to entry: Sample spacing can be a user-selectable DTS device parameter. 268 Note 3 to entry: The sample spacing is expressed in length units.

269 3.1.13 270 spatial resolution 271 smallest length of a temperature-affected fibre optic sensor for which a DTS device can 272 measure the reference temperature of the hot spot fibre condition within the specified 273 temperature measurement error of the DTS device 274 3.1.14 275 spatial temperature uncertainty 276 uncertainty of location of temperature data in a single temperature trace expressed by twice the 277 standard deviation of a specified number of adjacent temperature sample points, with the fibre 278 optic sensor held at constant temperature 279 3.1.15 280 temperature dead zone 281 limited zone of a temperature trace, where the temperature sample points deviate from the 282 undisturbed parts of the trace by a specified limit due to a point defect 283 3.1.16 284 temperature measurement error 285 maximum difference between a centred and uniformly weighted moving average of the 286 measured temperature and a reference temperature for all data points of the fibre optic sensor 287 over the full operating temperature range and all acquisition times 288 Note 1 to entry: Single value (worst case) is specified in temperature units (e.g. ± 0,8 °C). 289 Note 2 to entry: The number of elements used for the moving average is defined in Clause 5. In practical 290 applications, other methods of smoothing might be applicable.

291 3.1.17 292 temperature repeatability 293 precision of temperature data based on repeated temperature traces at a given location 294 expressed by twice the standard deviation of corresponding temperature sample points in each 295 temperature trace, with the fibre optic sensor held at constant temperature 297 temperature sample point 298 measured temperature value associated with a single point at a known location along a fibre 299 optic sensor 300 Note 1 to entry: Due to thermodynamic effects, the measured value represents the temperature along a very small 301 section of the fibre optic sensor that includes the point.

302 3.1.19 303 temperature trace 304 set of temperature sample points distributed along a fibre optic sensor and spaced by the 305 sample spacing 306 Note 1 to entry: All the sample points are associated with a common time of measurement, often called the trace 307 timestamp. The measured values represent the temperature during a period that includes the timestamp. 308 Note 2 to entry: All the sample points in a temperature trace are measured values produced by the DTS device, 309 and not interpolated or smoothed values produced by subsequent processing outside the DTS device. 310 3.1.20 311 total fibre length 312 L

Z 313 distance from the DTS device output connector to the final end of the fibre optic sensor 314 Note 1 to entry: Final end of the fibre optic sensor can either be a purposefully cut or terminated end of the fibre 315 physically far from the DTS device (in a single-ended configuration), or the end of a loop consisting of a connector 316 that is connected to the same DTS device (in a loop configuration).

317 Note 2 to entry: This parameter is either equal to or greater than the distance measurement range. 318 Note 3 to entry: The distance measurement range is expressed in length units. 319 3.1.21 320 worst case environmental temperature effect 321 maximum difference of the measured constant fibre optic sensor temperature at different 322 locations along the sensor during a complete temperature cycling of the DTS device across the 323 entire instrument operating temperature range 324 Note 1 to entry: This parameter is derived from environmental temperature stability. 325 3.2 Abbreviated terms

DTS distributed fibre optic temperature sensing system TE thermal enclosures (temperature-controlled) x

VOA variable optical attenuator 326 3.3 Symbols th

A(i) spatial temperature uncertainty at i sample point location L location

L fibre sensor length (loose and strain-free wound) < spatial resolution A

L fibre sensor length (loose and strain-free wound) = claimed spatial resolution B

L fibre sensor length (loose and strain-free wound) > 4 times spatial resolution C

L fibre sensor length (loose and strain-free wound) D

L short length of fibre optic sensor which is exposed by a measurable HS temperature change (∆T)

L ,L fibre length (normal spool)

W W/2

L total fibre length

Z

N number of traces n is the number of data points for calculation

S point centred and uniformly weighted moving average curve i th

S (i) standard deviation for each data point (i sample point location within a trace) x over time (N consecutive traces) th

S(i) temperature repeatability at i sample point location T constant optical fibre temperature const ∆T measurable temperature range

T designated DTS device operating temperature op

T high environmental temperature effect of the DTS device eh

T low environmental temperature effect of the DTS device el

T environmental temperature repeatability er

T worst case environmental temperature effect of the DTS device ew th th

T (i) temperature data at i sample point location of j trace j

T minimal DTS device operating temperature op-min

T maximal DTS device operating temperature op-max

T , T , lowest, middle and highest temperature of the fibre optic sensor temperature low mid range

T high

T traceable calibrated reference temperature ref t ,t ,t shortest, middle and longest possible DTS device measurement time min mid max σ temperature standard deviation 327 4 General test setups for measurement of performance parameters 328 4.1 General and test setup requirements 329 General test setups for single and loop configurations are schematically shown in Figure 1 and 330 Figure 2 respectively. Their aim is to provide a common base for determining the measurement 331 specifications while at the same time minimizing complexity, cost, reconfiguration requirements, 332 and test execution time. Further information about general test setups and measurement 333 procedures can be found in SEAFOM-MSP-01:2016 [1].

335 Key 1 thermal enclosure (temperature-controlled) TE DEV 2 DTS device 3 DTS device output connector 4 fibre connection 5 optional variable optical attenuator 6 fibre sensor length L (loose and strain-free wound) D 7 fibre length L or L (normal spool)

W W/2 8 fibre sensor length L (loose and strain-free wound) D 9 thermal enclosure (temperature-controlled) TE D 10 fibre sensor length L (loose and strain-free wound) > 4 × spatial resolution C 11 fibre sensor length L (loose and strain-free wound) < spatial resolution A 12 fibre sensor length L (loose and strain-free wound) = claimed spatial resolution B 13 thermal enclosure (temperature-controlled) TE C 14 thermal enclosure (temperature-controlled) TE ABC 15 fibre termination 16 traceable calibrated reference temperature sensors 336 Figure 1 – General test setup: single-ended configuration 338 Key 1 thermal enclosure (temperature-controlled) TE DEV 2 DTS device 3 DTS device output connector 4 fibre connection 5 optional variable optical attenuator 6 fibre sensor length L (loose and strain-free wound) D 7 fibre length L or L (normal spool)

W W/2 8 fibre sensor length L (loose and strain-free wound) D 9 thermal enclosure (temperature-controlled) TE D 10 fibre sensor length L (loose and strain-free wound) > 4 × spatial resolution C 11 fibre sensor length L (loose and strain-free wound) < spatial resolution A 12 fibre sensor length L (loose and strain-free wound) = claimed spatial resolution B 13 thermal enclosure (temperature-controlled) TE C 14 thermal enclosure (temperature-controlled) TE ABC 15 fibre length L or L (normal spool)

W W/2 16 DTS interrogation unit input connector 17 traceable calibrated reference temperature sensors 339 Figure 2 – General test setup: loop configuration 340 Individual evaluation procedures may be performed with a modified setup providing the required 341 measurement conditions. In this case, a detailed setup description and documentation is 342 required.

343 The DTS device shall be tested with a fibre optic sensor recommended by the DTS device 344 manufacturer, unless otherwise specified or agreed. The fibre lengths L , L , L of the fibre A B C 345 coils in the thermal enclosures TE and TE shall be selected based upon the expected ABC C 346 spatial resolution of the DTS device being tested. The fibre lengths L and L within and outside D W 347 TE shall be chosen to make the total fibre lengthL match the claimed distance measurement D Z 348 range of the DTS device being tested. In single ended configurations (see Figure 1) the total 349 fibre lengthL shall be equal to the total length of fibre from the DTS device output connector Z 350 up to the terminated end of the spatial resolution fibre section represented by fibre lengths L , A 351 L , L .

B C 352 NOTE 1 The use of fibre length L , located before and after a long length L of fibre (which makes up the total D W 353 fibre lengthL ), provides a test setup capable of accommodating various DTS device with different distance Z 354 measurement ranges.

355 Fibre length L shall be equal to 10 % of the total fibre lengthL .

D Z 356 In loop configuration (Figure 2) the total fibre lengthL shall be equal to the total length of fibre Z 357 from the DTS device output connector up to the DTS device input connector. In case of a DTS 358 device comparison with a single-ended configuration, the length of the normal spools shall be 359 (2 × L ). This guarantees the same overall attenuation. In all other cases the length of the W/2 360 normal spools shall be (2 × L ).

W 361 Low insertion loss and back-reflections should be accomplished when connecting the fibres by 362 connectors or by fusion splices.

363 The fibres in the thermal enclosures shall be coiled in such a way (loose wound) that the fibre 364 is completely exposed by the surrounding temperature, and that there is no fibre strain. Normal 365 spool in this case means a fibre spool as delivered from the fibre supplier. 366 The realization of the required thermal enclosures can differ in certain respects, such as using 367 climate chambers or liquid filled calibration baths. It should be assured that the TE and TE C ABC 368 provide a large and sharp enough temperature difference between the coils (at least 20 °C 369 occurring over no longer than 50 % of the rated spatial resolution). Traceable calibrated 370 reference temperature sensors shall be used to monitor the required coil temperatures. The 371 reference temperature measurement uncertainty shall be at least a factor of 5 smaller than the 372 temperature measurement error that is being assessed.

373 NOTE 2 Further information on how to use a crushed-ice bath for metrological comparison of Raman-distributed 374 temperature sensors is provided in Failleau, G. et al [2].

...