Electrical apparatus for the detection and measurement of toxic and combustible gases in car parks and tunnels - Part 1: General performance requirements and test methods for the detection and measurement of carbon monoxide and nitrogen oxides

This European Standard applies to apparatus for the detection and/or the measurement of carbon monoxide (CO), nitrogen monoxide (NO) and nitrogen dioxide (NO2) intended to control a ventilation system and/or to give an indication, alarm or any other signal to warn of a toxic hazard. These three gases are generically called “target gases” for the purpose of this European Standard. National and local regulations might not require detection of NO or NO2 and might require detection of other gases or vapours. This European Standard includes requirements for remote gas sensors (RGS) to be used in car parks and tunnels and requirements for the control unit (CU) to be used in car parks. This European Standard specifies general requirements for construction and testing and describes the test methods that apply to fixed apparatus for the detection and/or the measurement of the concentration of the target gases in car parks and tunnels. This European Standard may also be applied to similar applications where the concentration of the target gases could lead to a risk to health, for example loading areas for trucks and underground bus stations. This European Standard also applies when an apparatus manufacturer makes any claims regarding superior performance that exceeds these minimum requirements. This European Standard applies to apparatus, including the sampling system if applicable. This European Standard does not specify requirements for apparatus to be installed in hazardous areas. This European Standard does not apply for applications already covered by the following standards: – domestic premises, covered by EN 50291-1; – boats, craft, caravans or mobile homes, covered by EN 50291-2; – workplace atmospheres, covered by EN 45544 series; – emissions of heaters, covered by EN 50379 series; – motor vehicles emissions, covered by ISO/PAS 3930; – monitoring of the LEL level of combustible gases, covered by EN 60079-29-1. This European Standard does not apply for the following applications and technologies: – confined spaces not accessible to people; – laboratory or analytical equipment; – apparatus used to control industrial processes; – portable and transportable apparatus; – open path gas detection; – tunnel construction; – monitoring of particulates and dust; – monitoring of combustible gases; – CO monitoring for fire detection.

Elektrische Geräte für die Detektion und Messung von toxischen (und brennbaren) Gasen in Tiefgaragen und Tunneln - Teil 1: Allgemeine Anforderungen an das Betriebsverhalten sowie Prüfverfahren für die Detektion und Messung von Kohlenmonoxid und Stickoxiden

Diese Europäische Norm ist anwendbar für Geräte zur Detektion und / oder Messung von Kohlenmonoxid (CO), Stickstoffmonoxid (NO) und Stickstoffdioxid (NO2), die ein Lüftungssystem steuern und/oder mittels Anzeige, Alarm oder einer anderen Form von Signal vor der Gefahr einer Vergiftung warnen. Diese drei Gase werden zum Zweck dieser Europäischen Norm grundsätzlich als Zielgase bezeichnet. Nationale und örtliche Vorschriften können möglicherweise auf die Detektion von NO oder NO2 verzichten, dafür aber die Detektion anderer Gase oder Dämpfe verlangen. Diese Europäische Norm enthält Anforderungen für externe Sensoren (RGS) zum Einsatz in Tiefgaragen und Tunneln sowie Anforderungen an Steuereinheiten (CU) zum Einsatz in Tiefgaragen. Diese Europäische Norm enthält die allgemeinen Anforderungen an Bauweise, Prüfung und Betriebs-verhalten und beschreibt die Prüfverfahren für ortsfeste Geräte für die Detektion und / oder Messung der Konzentration der Zielgase in Tiefgaragen und Tunneln. Diese Europäische Norm darf in ähnlichen Einsatzbereichen angewendet werden, wo die Konzentration der Zielgase zu einem Gesundheitsrisiko führen kann zum Beispiel in LKW-Ladezonen oder Bushaltestellen in Unterführungen. Diese Europäische Norm gilt ebenfalls, wenn ein Gerätehersteller irgendwelche Angaben zu speziellen Eigenschaften der Bauart oder zu besserem Betriebsverhalten macht, welche die Mindestanforderungen übersteigen. Diese Europäische Norm gilt für Geräte einschließlich des Probenahmesystems, falls anwendbar. Diese Europäische Norm enthält keine Anforderungen an Geräte, damit sie in explosionsgefährdeten Bereichen montiert werden dürfen. Diese Europäische Norm gilt nicht für die Anwendungen, die bereits durch die folgenden Normen abgedeckt werden: – Wohnhäuser, siehe EN 50291–1; – Boote, Caravans und Wohnmobile, siehe EN 50291–2; – Arbeitsplatzumgebungen, siehe EN 45544; – Emissionen von Heizungsanlagen, siehe EN 50379; – Emissionen von Motorfahrzeugen, siehe ISO/PAS 3930; – Überwachung der UEG bei brennbaren Gasen, siehe EN 60079–29–1. Diese Europäische Norm gilt nicht für die folgenden Anwendungen und Technologien: – Enge für Personen nicht zugängliche Hohlräume; – Laborgeräte oder Analysatoren; – Geräte zur industriellen Prozesssteuerung; – Tragbare und transportable Geräte; – Geräte mit offener Messstrecke; – die Bauphase bei Tunneln; – Überwachung von Partikeln und Staub; – Überwachung von brennbaren Gasen; – CO-Überwachung zur Brandmeldung.

Appareil électrique de détection de mesure de gaz combustible et toxique dans les parcs de stationnement et les tunnels – Partie 1: Exigences de performance générales et méthodes pour la détection et la mesure du monoxyde de carbone et d’oxyde d’azote

La présente Norme européenne s’applique aux appareils de détection et/ou de mesure du monoxyde de carbone (CO), du monoxyde d’azote (NO) et du dioxyde d’azote (NO2), destinés à réguler un système de ventilation et/ou à fournir une indication, une alarme ou tout autre signal d’avertissement d’un danger de toxicité. Pour les besoins de la présente Norme européenne, ces trois gaz sont désignés de manière générique “gaz cibles”. Les réglementations nationales et locales peuvent ne pas exiger la détection du NO ou du NO2 mais plutôt la détection d’autres gaz ou vapeurs. La présente Norme européenne comprend des exigences relatives aux capteurs à distance (RGS) à utiliser pour les parcs de stationnement et les tunnels ainsi que des exigences applicables à l’unité de contrôle/commande (CU) à utiliser pour les parcs de stationnement. La présente Norme européenne spécifie les exigences générales relatives à la construction et aux essais et décrit les méthodes d’essai applicables aux appareils fixes de détection et/ou de mesure de la concentration des gaz cibles dans les parcs de stationnement et les tunnels. La présente Norme européenne peut également concerner des applications similaires dans le cadre desquelles la concentration des gaz cibles est susceptible d’engendrer un risque pour la santé, par exemple dans les zones de chargement des gares routières et d’autobus souterraines. La présente Norme européenne s’applique également dans le cas où un constructeur d’appareil formule des revendications en matière de performances ou d’aptitude à la fonction supérieures aux exigences minimales considérées. La présente Norme européenne s’applique aux appareils, y compris le système d’échantillonnage le cas échéant. La présente Norme européenne ne spécifie aucune exigence relative aux appareils à installer dans des zones dangereuses. La présente Norme européenne ne concerne pas les applications déjà couvertes par les normes suivantes: – locaux à usage domestique, couverts par l’EN 50291-1; – véhicules de loisir et locaux similaires (caravanes, mobiles homes) incluant les embarcations de loisir, couverts par l’EN 50291-2; – atmosphères des lieux de travail, couvertes par la série EN 45544; – conduits d’évacuation des appareils de chauffage, couverts par la série EN 50379; – gaz d'échappement des véhicules, couverts par l’ISO/PAS 3930; – surveillance du niveau LIE des gaz combustibles, couverte par l’EN 60079-29-1. La présente Norme européenne ne concerne pas les applications et technologies suivantes: – espaces confinés non accessibles au public; – laboratoire ou équipements d’analyse; – appareils utilisés pour le contrôle des procédés industriels; – appareils portables et transportables; – détection de gaz à chemin optique; – construction de tunnel; – surveillance des matières particulaires et des poussières; – surveillance des gaz combustibles; – surveillance du CO en matière de détection incendie.

Električne naprave za odkrivanje in merjenje strupenih in gorljivih plinov na avtomobilskih parkiriščih in v predorih - 1. del: Splošne tehnične zahteve in preskusne metode za zaznavanje in merjenje ogljikovega monoksida in dušikovih oksidov

Ta mednarodni standard velja za naprave za odkrivanje in/ali merjenje ogljikovega monoksida (CO), dušikovega monoksida (NO) in dušikovega dioksida (NO2), ki se uporabljajo za nadzor prezračevalnega sistema in/ali za prikaz, alarm ali drug signal, ki opozarja na nevarnost strupov. Ti trije plini se za ta evropski standard na splošno imenujejo »ciljni plini«. Nacionalne in lokalne zahteve morda ne zahtevajo odkrivanja NO ali NO2, lahko pa zahtevajo odkrivanje drugih plinov ali hlapov.
Ta evropski standard vključuje zahteve za oddaljene plinske senzorje (RGS), ki se uporabljajo na avtomobilskih parkiriščih in v predorih, in zahteve za nadzorno enoto (CU), ki se uporablja na avtomobilskih parkiriščih.
Ta evropski standard določa splošne zahteve za gradnjo in preskušanje ter opisuje preskusne metode, ki veljajo za fiksne naprave za odkrivanje in/ali merjenje koncentracij ciljnih plinov na avtomobilskih parkiriščih in v predorih. Ta evropski standard se lahko uporabi tudi pri podobnih aplikacijah, kjer koncentracija ciljnih plinov lahko privede do tveganja zdravja, na primer na območjih za nakladanje tovornjakov in podzemnih avtobusnih postajah.
Ta evropski standard velja tudi pri izjavah proizvajalca naprave o boljši zmogljivosti, ki presega minimalne zahteve.
Ta evropski standard velja za naprave, vključno s sistemom za vzorčenje, če je primerno.
Ta evropski standard ne določa zahtev za naprave, nameščene na nevarnih območjih.
Ta evropski standard ne velja za aplikacije, ki jih že zajemajo naslednji standardi:
– gospodinjstva, ki jih zajema EN 50291-1,
– čolni, plovila, prikolice ali avtodomi, ki jih zajema EN 50291-2,
– atmosfere na delovnem mestu, ki jih zajema serija EN 45544,
– emisije grelnikov, ki jih zajema serija EN 50379,
– emisije motornih vozil, ki jih zajema ISO/PAS 3930,
– nadzor ravni vnetljivih plinov LEL, ki ga zajema EN 60079-29-1.
Ta evropski standard ne velja za naslednje aplikacije in tehnologije:
– omejeni prostori, nedostopni za ljudi,
– laboratorijska ali analitska oprema,
– naprave za nadzor industrijskih procesov,
– prenosne in premične naprave,
– odkrivanje plinov na odprtih poteh,
– gradnja predorov,
– nadzor delcev in prahu,
– nadzor vnetljivih plinov,
– nadzor CO za odkrivanje požarov.

General Information

Status
Published
Publication Date
15-Sep-2011
Withdrawal Date
14-Aug-2014
Current Stage
9093 - Decision to confirm - Review Enquiry
Start Date
18-Jan-2023
Completion Date
23-Sep-2025

Relations

Standard
EN 50545-1:2011
English language
31 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI STANDARD
01-december-2011
(OHNWULþQHQDSUDYH]DRGNULYDQMHLQPHUMHQMHVWUXSHQLKLQJRUOMLYLKSOLQRYQD
DYWRPRELOVNLKSDUNLULãþLKLQYSUHGRULKGHO6SORãQHWHKQLþQH]DKWHYHLQ
SUHVNXVQHPHWRGH]D]D]QDYDQMHLQPHUMHQMHRJOMLNRYHJDPRQRNVLGDLQGXãLNRYLK
RNVLGRY
Electrical apparatus for the detection and measurement of toxic and combustible gases
in car parks and tunnels - Part 1: General performance requirements and test methods
for the detection and measurement of carbon monoxide and nitrogen oxides
Elektrische Geräte für die Detektion und Messung von toxischen (und brennbaren)
Gasen in Tiefgaragen und Tunneln - Teil 1: Allgemeine Anforderungen an das
Betriebsverhalten sowie Prüfverfahren für die Detektion und Messung von
Kohlenmonoxid und Stickoxiden
Appareil électrique de détection de mesure de gaz combustible et toxique dans les parcs
de stationnement et les tunnels - Partie 1: Exigences de performance générales et
méthodes pour la détection et la mesure du monoxyde de carbone et d’oxyde d’azote
Ta slovenski standard je istoveten z: EN 50545-1:2011
ICS:
13.040.50 Emisije izpušnih plinov v Transport exhaust emissions
prometu
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EUROPEAN STANDARD
EN 50545-1
NORME EUROPÉENNE
September 2011
EUROPÄISCHE NORM
ICS 13.040.50
English version
Electrical apparatus for the detection and measurement of toxic and
combustible gases in car parks and tunnels -
Part 1: General performance requirements and test methods for the
detection and measurement of carbon monoxide and nitrogen oxides

Appareil électrique de détection de Elektrische Geräte für die Detektion und
mesure de gaz combustible et toxique Messung von toxischen (und brennbaren)
dans les parcs de stationnement et les Gasen in Tiefgaragen und Tunneln -
tunnels – Partie 1: Exigences de Teil 1: Allgemeine Anforderungen an das
performance générales et méthodes pour Betriebsverhalten sowie Prüfverfahren für
la détection et la mesure du monoxyde de die Detektion und Messung von
carbone et d’oxyde d’azote Kohlenmonoxid und Stickoxiden

This European Standard was approved by CENELEC on 2011-08-15. CENELEC members are bound to comply
with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard
the status of a national standard without any alteration.

Up-to-date lists and bibliographical references concerning such national standards may be obtained on
application to the Central Secretariat or to any CENELEC member.

This European Standard exists in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CENELEC member into its own language and notified
to the Central Secretariat has the same status as the official versions.

CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus,
the Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy,
Latvia, Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia,
Spain, Sweden, Switzerland and the United Kingdom.

CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung

Management Centre: Avenue Marnix 17, B - 1000 Brussels

© 2011 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
Ref. No. EN 50545-1:2011 E
Contents
Foreword . 4
Introduction . 5
1 Scope . 6
2 Normative references . 7
3 Terms and definitions . 7
3.1 Gas properties . 7
3.2 Types of apparatus and components . 8
3.3 Sensors. 8
3.4 Signals and alarms . 9
3.5 Time . 9
3.6 Other .11
4 General requirements .11
4.1 Introduction .11
4.2 Construction requirements .11
4.3 Operational requirements .14
4.4 Access level .16
4.5 Information for the user .17
5 Test requirements .18
5.1 General .18
5.2 Sequence of tests .18
5.3 Preparation of apparatus before testing .19
5.4 Test gas application .19
5.5 Normal conditions for test .20
6 Test methods .21
6.1 General .21
6.2 Unpowered storage .21
6.3 Linearity .21
6.4 Alarm set points and outputs (car parks only) .22
6.5 Repeatability .23
6.6 Temperature .23
6.7 Humidity .24
6.8 Air velocity (for diffusion apparatus only) .24
6.9 Flow rate (for aspirated apparatus only) .24
6.10 Interfering gases .24
6.11 Recovery from high gas concentrations .25
6.12 Mechanical strength .25
6.13 Warm-up time .25
6.14 Response time .25
6.15 Power supply variations .26
6.16 Long-term stability .26
6.17 Field calibration, or field verification kit .26
6.18 Calibration mask .26
6.19 Testing of multiple point selector .26
6.20 Vibration.27
6.21 Electromagnetic compatibility .27
Annex A (informative) Using NO and NO standard test gases for standard and
humidity testing .28
A.1 General .28
A.2 NO gas testing .28
A.3 Humidity testing .28
Annex B (informative) Clarification regarding “6.4 Alarm set points and outputs
(car parks only)” .29
Bibliography .31

– 3 – EN 50545-1:2011
Figures
Figure 1 – Warm-up time in clean air (typical). 10
Figure 2 – Warm-up time in standard test gas (typical) . 10
Figure B.1 – Output of the sensor . 29
Figure B.2 – Time-weighted measured gas concentration . 29
Figure B.3 – Missing areas for TWA calculation . 30

Tables
Table 1 – Measuring range. 14
Table 2 – Alarm levels . 14
Table 3 – Maximum uncertainty and concentration of standard test gases . 20
Table 4 – Tolerance for linearity . 22
Table 5 – ppm.min calculated from Table 2 and alarms 1,2 test gas concentrations . 22
Table 6 – Tolerances for alarm 1 and alarm 2 activation . 23
Table 7 – Tolerances for zero deviation . 23
Table 8 – Interfering gases table: test results to be inserted into the blank cells. 24
Table 9 – Gas concentration for recovery test . 25

Foreword
This document (EN 50545-1:2011) has been prepared by CLC Technical Body 216 "Gas detectors".

The following dates are fixed:

• latest date by which this document has (dop) 2012-08-15
to be implemented at national level by
publication of an identical national
standard or by endorsement
(dow) 2014-08-15
• latest date by which the national
standards conflicting with this
document have to be withdrawn
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC [and/or CEN] shall not be held responsible for identifying any or all such
patent rights.
– 5 – EN 50545-1:2011
Introduction
This European Standard does not give guidance on installation of a gas detection system.
This European Standard specifies unique ranges and alarm levels for type testing, specifying
minimum requirements. All alarm levels are variable and may be adapted to national and/or local
regulations.
Separate type testing of remote gas sensors and control units is permitted. It is common practice to
use an integrated tunnel control system that includes processing of toxic gas measurement along with
other monitoring functions. When remote gas detectors and control units are type tested separately, it
is the responsibility of the manufacturer to ensure that the assembled system complies with the
requirements of this European Standard. The maximum capacity of the assembled system should not
exceed the capacity of the system that has been type tested.
Engineering companies or installers who buy equipment from different manufacturers are responsible
for the proper integration of the system.

1 Scope
This European Standard applies to apparatus for the detection and/or the measurement of carbon
monoxide (CO), nitrogen monoxide (NO) and nitrogen dioxide (NO ) intended to control a ventilation
system and/or to give an indication, alarm or any other signal to warn of a toxic hazard. These three
gases are generically called “target gases” for the purpose of this European Standard.
National and local regulations might not require detection of NO or NO and might require detection of
other gases or vapours.
This European Standard includes requirements for remote gas sensors (RGS) to be used in car parks
and tunnels and requirements for the control unit (CU) to be used in car parks.
This European Standard specifies general requirements for construction and testing and describes the
test methods that apply to fixed apparatus for the detection and/or the measurement of the
concentration of the target gases in car parks and tunnels. This European Standard may also be
applied to similar applications where the concentration of the target gases could lead to a risk to
health, for example loading areas for trucks and underground bus stations.
This European Standard also applies when an apparatus manufacturer makes any claims regarding
superior performance that exceeds these minimum requirements.
This European Standard applies to apparatus, including the sampling system if applicable.
This European Standard does not specify requirements for apparatus to be installed in hazardous areas.
This European Standard does not apply for applications already covered by the following standards:
– domestic premises, covered by EN 50291-1;
– boats, craft, caravans or mobile homes, covered by EN 50291-2;
– workplace atmospheres, covered by EN 45544 series;
– emissions of heaters, covered by EN 50379 series;
– motor vehicles emissions, covered by ISO/PAS 3930;
– monitoring of the LEL level of combustible gases, covered by EN 60079-29-1.
This European Standard does not apply for the following applications and technologies:
– confined spaces not accessible to people;
– laboratory or analytical equipment;
– apparatus used to control industrial processes;
– portable and transportable apparatus;
– open path gas detection;
– tunnel construction;
– monitoring of particulates and dust;
– monitoring of combustible gases;
– CO monitoring for fire detection.

– 7 – EN 50545-1:2011
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
EN 50270, Electromagnetic compatibility – Electrical apparatus for the detection and measurement of
combustible gases, toxic gases or oxygen
EN 50271, Electrical apparatus for the detection and measurement of combustible gases, toxic gases
or oxygen – Requirements and tests for apparatus using software and/or digital technologies
EN 60073, Basic and safety principles for man-machine interface, marking and identification – Coding
principles for indicators and actuators (IEC 60073)
EN 60335-1:2002 + corr. Jul.2009 + corr. May.2010 + A1:2004 + corr. Jan.2007 + A2:2006 +
A11:2004 + A12:2006 + corr. Feb.2007 + A13:2008 + A14:2010, Household and similar electrical
appliances – Safety – Part 1: General requirements (IEC 60335-1:2001 + A1:2004 + A2:2006 +
corrigendum Aug. 2006)
EN 60529, Degrees of protection provided by enclosures (IP Code) (IEC 60529)
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply:
3.1 Gas properties
3.1.1
clean air
air that is free of target gas, interfering gases or contaminating substances
3.1.2
hazardous area (potentially explosive atmosphere)
an atmosphere which could become explosive due to local and operational conditions
NOTE Operation in hazardous areas may require legislative measures for the approval, installation and
construction requirements of the apparatus.
3.1.3
ppm.min
gas concentration multiplied by the time, in minutes, of gas application
NOTE Used for testing Time-weighted Average (TWA).
3.1.4
volume ratio
V/V
ratio of one component in a mixture divided by the sum of the volumes of all the components before
they are mixed at specified temperature and pressure conditions
[ISO 7504]
NOTE 1 This is also known as concentration.
NOTE 2 Assuming ideal behaviour or gases, the volume ratio coincides with the molar ratio (mol/mol). The ppm units are
-6
equivalent to the International System units 10 vol/vol.

3.2 Types of apparatus and components
3.2.1
calibration mask
device that when applied to the sensor provides the continuous and renovated gas supply to the
sensor such that the gas concentration reaching the sensing element remains unchanged
3.2.2
control unit
CU
part of the apparatus that receives and processes the analogue or digital signal generated by the
sensors and also generates alarms and output functions
3.2.3
fixed apparatus
apparatus that is intended to have all parts permanently installed
3.2.4
power supply
device that, by means of mains electricity or one or more rechargeable batteries, provides the suitable
power to the system or part of it
3.2.5
remote gas sensor
RGS
sensor that is not integral to the control unit of the apparatus
3.2.6
sample line
tubing and connection components by means of which the gas being sampled is conveyed to the
sensor
3.2.7
special tool
tool required to gain access to, or to adjust controls
NOTE The design of the tool is intended to discourage unauthorized interference with the apparatus.
3.3 Sensors
3.3.1
aspirated sensor
sensor that accesses the gas by drawing it to the sensing element with a pump or by other means. If
the device samples more than one point, it cycles through all related points
3.3.2
diffusion sensor
sensor that accesses the gas by convection and diffusion from the atmosphere to the sensing
element, i.e. under conditions in which there is no aspirated flow
3.3.3
sensing element
device with an electric output signal that changes when it interacts with the target gas
3.3.4
sensor
assembly in which the sensing element is housed and that may also contain associated circuit
components
NOTE The sensor may be integral with either the CU or remote (RGS).

– 9 – EN 50545-1:2011
3.4 Signals and alarms
3.4.1
activation levels
sensor signal levels that trigger outputs and/or an alarm indications
3.4.2
alarm indication
audible or visible indication activated when a pre-set concentration level is reached
3.4.3
alarm set point
adjustable setting of the apparatus that sets the level of concentration over a period of time at which
the apparatus will automatically initiate an alarm indication and/or output
3.4.4
fault
dysfunction of a part of the system that does not disable the rest of the measure and control functions
3.4.5
fault signal
audible, visible or other type of output different from the alarm signal, signalling, directly or indirectly, a
warning or indication that the apparatus is not working satisfactorily
3.4.6
indicators
means to provide visible and/or audible information
3.4.7
latching alarm
alarm that, once activated, requires deliberate action for deactivation
3.4.8
system fault
dysfunction of the system that disables the control function
3.4.9
output
physical interface that drives external actuators
EXAMPLES Air extractors, remote warning and alarm devices.
3.4.10
quiescent condition
normal operative condition where no alarm or fault or other abnormal condition occurs
3.5 Time
3.5.1
final indication
sensor signal indication given by the apparatus after stabilization
3.5.2
stabilized apparatus
state when three successive readings of an apparatus, taken at five-minute intervals, indicate no
changes greater than ± 1 % of the measuring range

3.5.3
time of response
t
x
time interval, with the apparatus in a warmed-up condition, between the time when an instantaneous
variation in volume ratio is produced at the apparatus inlet and the time when the response reaches a
stated percentage (x) of the final indication
NOTE This European Standard refers to t as the time of response.
3.5.4
warm up time
time interval, with the apparatus in a stated atmosphere, between the time when the apparatus is
switched on and the time when the indication reaches and remains within the stated tolerances
NOTE See Figures 1 and 2.
Key
1 indication
2 power off in clean air
3 power on in clean air
4 apparatus zero
5 specified tolerance band on zero indication
6 time
7 warm-up time
Figure 1 – Warm-up time in clean air (typical)

Key
1 indication
2 power off in standard test gas
3 power on in standard test gas
4 volume ratio of standard test gas
5 specified tolerances of the indication
6 time
7 zero in clean air
8 warm-up time
Figure 2 – Warm-up time in standard test gas (typical)

– 11 – EN 50545-1:2011
3.6 Other
3.6.1
systematic error of deviation
difference between the arithmetic average value of n consecutive measurements of a reference gas,
carried out at reference conditions, and the accepted reference value
3.6.2
zone
geographical subdivision of the surveyed premises in which one or more sample points or RGS are
installed and for which a common output and/or indication is provided
4 General requirements
4.1 Introduction
4.1.1 Purpose
The apparatus shall detect the presence of target gases in fixed installations under the stated application
conditions. The sensor shall produce an analogue or digital signal (related to the concentration of the
target gases) that is processed in the CU and produces an output signal and/or an alarm indication as
defined in 4.3. Its function is to prevent and warn about the potential toxic gas risks.
4.1.2 Functionality
Apparatus for the detection and measurement of target gases in car parks and tunnels shall include
the following functionalities:
– sensors (RGS or internal in the apparatus);
– control unit / control indicating equipment (CU);
– power supply;
– indicators (visible and audible);
– output functions.
NOTE Functionalities may be in a single housing or distributed in multiple housings, and may be combined with other
functional modules. Type testing may be carried out for a complete apparatus or separately for RGS and/or CU.
4.1.3 Software-controlled apparatus
Software-controlled apparatus shall conform to EN 50271.
4.2 Construction requirements
4.2.1 General
The apparatus or parts thereof (e.g. RGS) specifically intended for use in the presence of corrosive
vapours or gases, or which might produce corrosive products as a result of the detection process shall
be constructed of materials known to be resistant to corrosion by such substances.
The apparatus shall be constructed to facilitate regular operation and performance checks.
All adjustment devices shall be designed so as to discourage unauthorized or inadvertent interference
with the apparatus.
The RGS shall be constructed to provide the degree of protection, as defined in EN 60529, in all
orientations of installation specified by the manufacturer as defined in 4.5.2. The degree of protection
shall be at least IP54.
Aspirated apparatus and control units, when separated from the RGS, shall be constructed to provide
a degree of protection of at least IP20.

4.2.2 Interface requirements
4.2.2.1 Interface requirements for separately tested RGS
A specification shall be supplied with the RGS that describes the relationship between the gas
concentration and the corresponding output signal (transfer function). The specification shall be
detailed to the extent that the accuracy of the transfer function can be verified. As a minimum, data
shall be provided showing the relationship between the output signal and the gas concentrations
corresponding to 0 %, 10 %, 25 %, 50 %, 75 % and 100 % of full-scale indication. Over-range and
status signals (e.g. fault or inhibit) shall also be specified.
Where necessary, equipment shall be provided to interpret the output signal or indication, which will
enable the accuracy of the transfer function to be verified.
4.2.2.2 Interface requirements for separately tested CU
A specification shall be supplied with the apparatus that describes the relationship between the input
signal and the calculated gas concentration (transfer function). The specification shall be detailed to
the extent that the accuracy of the transfer function can be verified. As a minimum, data shall be
provided showing the relationship between the input signal and the gas concentrations corresponding
to 0 %, 10 %, 25 %, 50 %, 75 % and 100 % of full-scale output indication. Required inputs for over-
range indications and status signals (e.g. fault or inhibit) shall also be specified.
Where necessary, equipment shall be provided to generate any required input signal, which will
enable the accuracy of the transfer function to be verified.
4.2.3 Measuring range and alarm levels
The apparatus shall measure the target gas volumetric concentrations with alarm levels and
measuring range as specified in 4.3.2.2.
For car parks, the apparatus shall identify at least three threshold conditions.
4.2.4 Indicating devices on the CU
4.2.4.1 General
The apparatus shall be equipped with the devices listed below, in order to give operating and
maintenance personnel the necessary information about the functioning conditions and the installation
it controls. These indications shall be visible and/or audible without manipulation at any access level
(as defined in 4.4).
NOTE Remote indicating devices are not covered by this European Standard; they are specified by local regulations.
4.2.4.2 Audible indications
The audible indications of the CU shall be the following:
– alarm indication (at least for alarm 3);
– fault indication.
The audible indication for the gas alarm condition may be the same as that for the fault warning
condition. If they are different, the gas alarm indication shall have priority.
The audible indications may be substituted by a system where the CU forwards faults and alarms.
4.2.4.3 Visual indications
The visual indications shall be the following:
– power indication;
– mains fault indication (when an emergency battery is part of the apparatus);
– fault indication;
– 13 – EN 50545-1:2011
– warm-up indication (when the function is provided);
– low battery indication (when an emergency battery is part of the apparatus);
– alarm indication;
– low flow rate indication (for aspirated apparatus);
– test condition (maintenance mode);
– disabled condition (part of the system).
A display shall be provided to indicate the measured gas concentration for each detection point. It
shall show either the instant or the average measurement.
When displaying different target gases, the concentrations shall be displayed unambiguously.
Indications relating to different status and alarm conditions may be displayed at the same time.
Indicator lights shall be adequately labelled to show their functions and shall be coloured as follows (in
accordance with EN 60073):
– alarms indicating the presence of a gas concentration above an alarm set point shall be coloured
RED;
– equipment fault indicators shall be coloured YELLOW;
– power supply indicators shall be coloured GREEN.
4.2.4.4 Priority of indications
If several fault or alarm indications are generated simultaneously by the apparatus, it shall be possible
to identify separately each indication.
A warning shall be shown if not all indications can be displayed simultaneously. In this case, the
indication with the highest alarm priority at the given time shall be displayed or activated. It shall be
possible to interrogate the indications that are not currently shown or activated.
Priority of alarm indications shall be defined and documented in the instruction manual.
It is sufficient to indicate the higher priority alarm only.
EXAMPLE Exceeding the second threshold value includes exceeding of the first threshold value.
After cancelling the higher order indication, any lower priority indication shall be shown if the reason
for its activation still exists.
4.2.5 Output of CU
For each zone, the apparatus shall provide at least one output for each gas alarm condition (as
specified in 4.3.2.3) to initiate indicating devices or control actions.
One output shall be provided for fault signalling.
4.2.6 Digital components
The apparatus (RGS, if any, and CU) including the digital interface shall conform to EN 50271 for
digital components.
4.2.7 User interface
The apparatus shall be supplied with all the items such as special tools, keyboards, displays, terminals
and software for configuration to make all necessary field adjustments or maintenance operations.

4.2.8 Calibration mask
For diffusion apparatus, the manufacturer shall provide a calibration mask adapted to the apparatus
that shall be verified as being appropriate for the calibration and type testing (see 6.18).
The verified calibration mask shall be used for type testing of the sensor.
When applying it, removal of any protection cover that could be mounted on the RGS is allowed.
4.3 Operational requirements
4.3.1 Introduction
The apparatus shall have two operational modes:
– measuring mode, which includes quiescent mode and gas alarm condition;
– special state mode, which includes fault condition, disabled RGS, warm-up and maintenance.
4.3.2 Measuring mode
4.3.2.1 Measuring range
The measuring range for each target gas shall be as specified in Table 1.
Table 1 – Measuring range
Target gas Range
ppm
CO 0 … 300
NO 0 … 30
NO 0 … 100
4.3.2.2 Alarm levels
The apparatus shall be type tested with the target gas volumetric concentrations and alarm levels as
defined in Table 2.
To fulfil different local regulations, alarm set points shall be adjustable throughout the measuring
ranges and averaging time periods shall be adjustable from 5 min to 60 min.
The resolution for alarm setting shall be 1 % of the measuring range or better.
Table 2 – Alarm levels
All values in ppm
Target gas Alarm 1 Alarm 2 Alarm 3
CO 30 60 150
NO 3 6 15
NO 10 20 50
Alarms 1 and 2 are 15 min Time-weighted Averages (TWA).
Alarm 3 shall be triggered when any RGS or sample point is above the specified level continuously for
a period of one minute. The period shall be adjustable from 1 min to 5 min.

– 15 – EN 50545-1:2011
4.3.2.3 Alarm conditions and outputs
The following gas concentration measurements shall apply for sampling points or RGS:
The apparatus shall enter an alarm condition when one or more measurements exceed the alarm set
points.
Each measurement shall be processed independently, without interference from any other
measurement in the same or any other zone.
For each zone, the alarm condition (alarm 1, alarm 2, alarm 3) shall be determined by the highest of
its measurements.
The alarm 2 and alarm 3 conditions comprise and maintain (or initiate when not yet active) its related
lower alarm conditions (alarm 1, alarm 2).
For each zone, the outputs related to the alarm conditions shall be activated.
The audible indication of alarm conditions within the CU may be silenced.
4.3.3 Special state
4.3.3.1 Fault condition
If the apparatus or some of its parts are not working properly then the fault condition shall be switched
on. This shall be overridden by
– any gas alarm, which has priority, and/or
– disablement of the corresponding sensor or function, and/or
– testing of a corresponding sensor or function.
The fault indication shall be switched on if the apparatus or some of its parts are not working properly,
including, for example, a short circuit or interruption of a connection to the RGS or any other
component contained in a different cabinet than the CU.
The following faults shall be indicated separately by means of light emitting indicators and optionally
by an alphanumeric display. The indications may be suppressed during the gas alarm condition.
a) An indication for each RGS or zone in which the transmission of signals to the CU is affected by
– short circuit or interruption in a detection circuit,
– removal of a RGS or a sensor.
b) At least one common indication of any power supply fault resulting from
– short circuit or an interruption in a transmission path to a power supply, when the power
supply is contained in a different cabinet from that of the CU,
– failure of either AC or DC power supply when also provided with a back-up battery.
c) Low battery condition for a back-up battery if supplied.
d) An indication of any short circuit or interruption, at least common to all transmission paths
between parts of the CU contained in more than one mechanical cabinet, which is capable of
affecting a mandatory function, and which is not otherwise indicated as a fault of a supervised
function.
e) Flow fault indication, for aspirated apparatus only.
The audible indication of fault conditions may be silenced.
In the case of software-controlled apparatus, a system fault shall be visibly and audibly indicated, as
specified in EN 50271. This fault state shall not be suppressed by any other functional condition of the
apparatus and shall remain until manually reset and/or another manual operation.

4.3.3.2 Disabled RGS (when this function is implemented)
The following functions shall be independently disabled and re-enabled:
– each RGS (or zone);
– output signals and/or transmission paths to automatic controls or alarm devices.
The disablement shall be indicated by means of separate light emitting indicators and/or an
alphanumeric display. The indications shall not be suppressed during the gas alarm condition.
4.3.3.3 Warm-up (when this function is implemented)
When the RGS is waiting for stabilization of the sensor, an initial warm-up time during which the RGS
monitoring activity is disabled is acceptable, provided that during such time the output function of fault
or warm-up disablement is active.
If an indicator is provided, signalling that the RGS is not in measuring mode during warm-up, it shall
be different to the indicators of 4.3.3.2.
4.3.3.4 Maintenance
Maintenance mode includes
– commissioning,
– calibration,
– test (when this function is implemented),
– repair.
Maintenance mode for a single RGS, a group of RGSs or the entire system may be entered. When
maintenance mode is entered, the related outputs may be re-set to the non-alarm condition. During
maintenance, new related alarms shall not trigger the outputs.
4.3.4 Multiple point selector (aspirated only)
If the apparatus samples sequentially at multiple points, the maximum cycle time shall be 3 min for
measuring all related measuring points to ensure that at least five measurements are used in the TWA
calculation as specified in 4.3.2.2.
4.3.5 Display of CU
This read-out device is intended for the presentation of the measured gas concentrations. When the
apparatus has more than one sampling point or RGS and only one indication device, this shall show
on demand the value of each measuring point(s), distinguishing between the instant and the average
value(s). Different indication modes may be selected on user request.
4.4 Access level
The apparatus shall be designed in order to avoid manipulation by persons without specific skills or
authorization of the protected functions.
NOTE This could be done, for example, by mechanically restricting access to the apparatus with keys or special tools or by
software that may have different access levels.
The manufacturer’s established access levels shall be explained clearly.
The protected functions are
– modification of configuration,
– re-setting of latched alarms if fitted,
– maintenance including test and calibration,
– replacement of components.
– 17 – EN 50545-1:2011
4.5 Information for the user
4.5.1 Labelling and marking
The RGS shall carry a label placed onto the case of the device that is easily accessible to the installer.
The label is not required to be visible in normal operation.
The CU shall carry a label placed onto the case of the device that is easily accessible during
operation.
The labels shall be marked legibly and indelibly with the following minimum requirements:
– name of the manufacturer;
– CE certification marking;
– designation of series or type;
– serial number;
– reference to this European Standard.
4.5.2 Instruction manual
Each apparatus shall be provided with an instruction manual that includes the following information:
a) complete instructions, drawings and diagrams for safe and proper operation, installation and
servicing of the apparatus;
b) operating instructions and adjustment procedures;
c) recommendations for initial checking and calibration of the apparatus;
d) complete instructions for checking, calibrating and function tests of the apparatus on a routine
basis, including any necessary information, especially
– concentrations of calibration gases,
– flow rate,
– application method,
– calibration mask, including instructions for the use of the field calibration kit, if provided;
e) checking and calibration intervals shall be specified. Calibration interval shall be a maximum of
12 months;
f) details of operational limitations including, where applicable, the following:
– gases for which the apparatus is suitable;
– information that describes the cross interference of other gases to which the apparatus is
responsive;
– response times (t ) for the target gas(es);
– temperature range;
– humidity range;
– pressure range;
– supply voltage range;
– maximum power consumption;
– relevant characteristics and construction details of required interconnecting cables and
interface specification;
– back-up battery information (if fitted);
– sample flow rate;
– warm-up time;
– stabilization time;
g) details of storage life and limitations for the apparatus, replacement parts and accessories,
including, where applicable, the following:
– temperature;
– humidity;
– pressure;
– time;
h) for aspirated apparatus, indication of the minimum and maximum flow rates and pressure; also:
tubing type, maximum length and size for proper operation;
i) for aspirated apparatus, instructions for ensuring that the sample lines are intact and that proper
flow is established;
j) for the CU, statements of the nature and significance of all alarms and fault signals, the duration
of such alarms and signals (if time-limited or non-latching), and any provisions that can be made
for silencing or resetting such alarms and signals, as applicable;
k) details of any method for the determination of the possible sources of a malfunction and any
corrective procedures (i.e. trouble-shooting procedures);
l) for the CU, statement that alarm devices, outputs or contacts are of the non-latching types, where
applicable;
m) recommended replacement parts list.
5 Test requirements
5.1 General
For the purpose of type testing, the tests shall be carried out on one apparatus. Additional apparatus
may be used for each of the long-term stability, impact and vibration tests.
In addition to the performance testing in Clause 6, other testing is required.
– Verifications shall be carried out to ensure that the apparatus conforms to the construction
requirements of 4.2. The requirements of these tests are in general obvious, except the
requirement for short circuits of 4.3.3.1, in which resistances shall replace each wire that connects
the instrument to a RGS. The values of these resistances shall be the maximum resistance of the
connecting wires as specified in the instruction manual, which allows compliance with this
European Standard. The device used for the short circuit shall be of negligible resistance and
shall be applied at the ends of the resistances, close to the CU.
– The instruction manual shall be inspected for conformity to the requirements of 4.5.2.
– When the manufacturer makes any claims regarding any special features of construction or
superior performance that exceed these minimum requirements, additional tests shall be
performed to verify the claimed features or superior
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...