ASTM F2743-11(2018)
(Guide)Standard Guide for Coating Inspection and Acute Particulate Characterization of Coated Drug-Eluting Vascular Stent Systems
Standard Guide for Coating Inspection and Acute Particulate Characterization of Coated Drug-Eluting Vascular Stent Systems
SIGNIFICANCE AND USE
5.1 The shedding of the coating from a vascular stent can alter its clinical safety and/or therapeutic benefit. Clinical performance (for example, drug elution) may be affected by particulate generation from the coated stent system and coating defects. This document provides guidance for coating inspection and acute particulate characterization of drug eluting vascular stents. Information about the potential for shedding can be gained during bench testing. The general guidelines presented here may be used for writing detailed protocols for specific products at the various stages of the product development process. Such testing may be performed during device development, design validation testing, lot-release testing, and/or stability testing although different requirements may apply at each stage. These suggested methods may represent a reasonable simulation of clinical usage. When establishing the coating inspection and acute particulate characterization testing conditions, the current clinical usage/practice (for example, post-dilation, overlapping stents) and the instructions for use (IFU), as applicable, should be considered. While methods for chronic particulate characterization and coating inspection have not been established, these suggested methods may be helpful in the development of chronic methods. Testing in accordance with recommendations in this guide will generate data that may lead to further improvements in the method and its validation, as well as possible advancements in device design and performance. See also FDA Guidance for Industry and FDA Staff and AAMI TIR42:2010.
SCOPE
1.1 This guide describes recommended in vitro test procedures for coating inspection and acute particulate characterization of coated drug-eluting vascular (balloon-expandable and self-expanding) stent systems.
1.2 Recommended practices for coating inspection and acute particulate characterization include baseline (deployment) testing and simulated use testing. This guide describes the capture and analysis of particulates. This guide describes the inspection of the coated stent surface. This guide was developed for characterization and not intended for production release testing of coated drug-eluting vascular stent systems although some sections may be appropriate.
1.3 Chronic particulate characterization and coating inspection are not included herein.
1.4 Coating systems specifically designed to degrade or otherwise intentionally separate themselves from the permanent stent structure may not be fully addressed herein.
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6 The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard.
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
- Status
- Published
- Publication Date
- 31-Oct-2018
- Technical Committee
- F04 - Medical and Surgical Materials and Devices
- Drafting Committee
- F04.30 - Cardiovascular Standards
Relations
- Effective Date
- 01-Nov-2018
- Effective Date
- 01-Nov-2018
- Effective Date
- 01-Nov-2018
Overview
ASTM F2743-11(2018), Standard Guide for Coating Inspection and Acute Particulate Characterization of Coated Drug-Eluting Vascular Stent Systems, provides essential guidance for evaluating the coatings and acute particulate generation of drug-eluting vascular stents. This standard is vital for bench testing to assess if coating shedding from vascular stents could impact clinical safety, drug delivery performance, or therapeutic benefit. By following this guide, stakeholders can develop testing protocols for product development, design validation, lot-release, and stability testing of balloon-expandable and self-expanding stent systems.
Bench testing for coating durability and particulate characterization is a critical step in ensuring vascular stent quality and performance, and helps simulate realistic clinical scenarios. While this guide focuses on acute (short-term) effects, it supports the continuous improvement of methods for future, chronic (long-term) assessments as well.
Key Topics
- Coating Inspection: Visual and microscopic examination of stent coatings to identify defects, anomalies, or damage before and after device manipulation or simulated clinical use.
- Acute Particulate Characterization: In vitro methodologies for capturing, counting, and analyzing particulates generated during stent deployment or use, particularly those shed from the coating.
- Baseline and Simulated Use Testing: Differentiates between straightforward stent deployment (baseline) and advanced testing scenarios that mimic clinical use (simulated), including tracking the stent through anatomical models and overlapping stents.
- Controlled Environments: Emphasizes stringent cleanliness and the use of appropriate controls to avoid contamination that could skew particulate analysis.
- Analytical Methods: Recommends optical and electronic techniques such as light obscuration, filtration with microscopy, and advanced analytical instrumentation (e.g., SEM, Raman spectroscopy) to provide insight into the source and nature of particulates.
- Documentation and Reporting: Protocols for recording defects, particulate data, and procedural parameters, supporting traceability and regulatory submissions.
Applications
The ASTM F2743-11(2018) standard guide is used by:
- Medical device manufacturers during the design, development, and validation of coated drug-eluting vascular stent systems to optimize coating processes and minimize risks of particulate generation.
- Quality assurance teams in evaluating and documenting the surface integrity and cleanliness of stents as part of lot-release or stability testing.
- Regulatory affairs professionals preparing submissions to agencies such as the FDA, since the processes described can align with requirements for demonstrating device safety and performance.
- Research and development for simulating and improving stent deployment in clinical scenarios, identifying potential issues that may arise during delivery, deployment, post-dilatation, or when stents overlap.
- Clinical risk assessments to understand how particulate debris or coating anomalies could impact patient outcomes.
Related Standards
ASTM F2743-11(2018) aligns with a number of relevant industry standards and guidance documents for vascular stents and particulate evaluation, including:
- USP <788> - Particulate Matter in Injections: Guidance for particulate testing in medical solutions.
- FDA Guidance for Industry and FDA Staff: Recommendations for non-clinical engineering tests of intravascular stents and their delivery systems.
- AAMI TIR42:2010: Best practices for evaluation of particulates associated with vascular medical devices.
By adhering to ASTM F2743-11(2018), organizations ensure compliance with internationally recognized testing principles and regulatory expectations, ultimately improving the quality and reliability of coated drug-eluting vascular stent systems.
Buy Documents
ASTM F2743-11(2018) - Standard Guide for Coating Inspection and Acute Particulate Characterization of Coated Drug-Eluting Vascular Stent Systems
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
ASTM F2743-11(2018) is a guide published by ASTM International. Its full title is "Standard Guide for Coating Inspection and Acute Particulate Characterization of Coated Drug-Eluting Vascular Stent Systems". This standard covers: SIGNIFICANCE AND USE 5.1 The shedding of the coating from a vascular stent can alter its clinical safety and/or therapeutic benefit. Clinical performance (for example, drug elution) may be affected by particulate generation from the coated stent system and coating defects. This document provides guidance for coating inspection and acute particulate characterization of drug eluting vascular stents. Information about the potential for shedding can be gained during bench testing. The general guidelines presented here may be used for writing detailed protocols for specific products at the various stages of the product development process. Such testing may be performed during device development, design validation testing, lot-release testing, and/or stability testing although different requirements may apply at each stage. These suggested methods may represent a reasonable simulation of clinical usage. When establishing the coating inspection and acute particulate characterization testing conditions, the current clinical usage/practice (for example, post-dilation, overlapping stents) and the instructions for use (IFU), as applicable, should be considered. While methods for chronic particulate characterization and coating inspection have not been established, these suggested methods may be helpful in the development of chronic methods. Testing in accordance with recommendations in this guide will generate data that may lead to further improvements in the method and its validation, as well as possible advancements in device design and performance. See also FDA Guidance for Industry and FDA Staff and AAMI TIR42:2010. SCOPE 1.1 This guide describes recommended in vitro test procedures for coating inspection and acute particulate characterization of coated drug-eluting vascular (balloon-expandable and self-expanding) stent systems. 1.2 Recommended practices for coating inspection and acute particulate characterization include baseline (deployment) testing and simulated use testing. This guide describes the capture and analysis of particulates. This guide describes the inspection of the coated stent surface. This guide was developed for characterization and not intended for production release testing of coated drug-eluting vascular stent systems although some sections may be appropriate. 1.3 Chronic particulate characterization and coating inspection are not included herein. 1.4 Coating systems specifically designed to degrade or otherwise intentionally separate themselves from the permanent stent structure may not be fully addressed herein. 1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.6 The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard. 1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
SIGNIFICANCE AND USE 5.1 The shedding of the coating from a vascular stent can alter its clinical safety and/or therapeutic benefit. Clinical performance (for example, drug elution) may be affected by particulate generation from the coated stent system and coating defects. This document provides guidance for coating inspection and acute particulate characterization of drug eluting vascular stents. Information about the potential for shedding can be gained during bench testing. The general guidelines presented here may be used for writing detailed protocols for specific products at the various stages of the product development process. Such testing may be performed during device development, design validation testing, lot-release testing, and/or stability testing although different requirements may apply at each stage. These suggested methods may represent a reasonable simulation of clinical usage. When establishing the coating inspection and acute particulate characterization testing conditions, the current clinical usage/practice (for example, post-dilation, overlapping stents) and the instructions for use (IFU), as applicable, should be considered. While methods for chronic particulate characterization and coating inspection have not been established, these suggested methods may be helpful in the development of chronic methods. Testing in accordance with recommendations in this guide will generate data that may lead to further improvements in the method and its validation, as well as possible advancements in device design and performance. See also FDA Guidance for Industry and FDA Staff and AAMI TIR42:2010. SCOPE 1.1 This guide describes recommended in vitro test procedures for coating inspection and acute particulate characterization of coated drug-eluting vascular (balloon-expandable and self-expanding) stent systems. 1.2 Recommended practices for coating inspection and acute particulate characterization include baseline (deployment) testing and simulated use testing. This guide describes the capture and analysis of particulates. This guide describes the inspection of the coated stent surface. This guide was developed for characterization and not intended for production release testing of coated drug-eluting vascular stent systems although some sections may be appropriate. 1.3 Chronic particulate characterization and coating inspection are not included herein. 1.4 Coating systems specifically designed to degrade or otherwise intentionally separate themselves from the permanent stent structure may not be fully addressed herein. 1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.6 The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard. 1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
ASTM F2743-11(2018) is classified under the following ICS (International Classification for Standards) categories: 11.040.40 - Implants for surgery, prosthetics and orthotics. The ICS classification helps identify the subject area and facilitates finding related standards.
ASTM F2743-11(2018) has the following relationships with other standards: It is inter standard links to ASTM F2743-11, ASTM F3036-21, ASTM F3320-18. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ASTM F2743-11(2018) is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: F2743 − 11 (Reapproved 2018)
Standard Guide for
Coating Inspection and Acute Particulate Characterization of
Coated Drug-Eluting Vascular Stent Systems
This standard is issued under the fixed designation F2743; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 2. Referenced Documents
1.1 This guide describes recommended in vitro test proce-
2.1 Other Standards:
dures for coating inspection and acute particulate characteriza-
USP <788> Particulate Matter in Injections
tion of coated drug-eluting vascular (balloon-expandable and FDA Guidance for Industry and FDA StaffNon-Clinical
self-expanding) stent systems. Engineering Tests and Recommended Labeling for Intra-
vascular Stents and Associated Delivery Systems, April
1.2 Recommended practices for coating inspection and 3
18, 2010
acute particulate characterization include baseline (deploy-
AAMITIR42:2010Evaluation of Particulates Associated
ment) testing and simulated use testing. This guide describes 4
with Vascular Medical Devices
the capture and analysis of particulates. This guide describes
the inspection of the coated stent surface. This guide was
3. Terminology
developed for characterization and not intended for production
3.1 Definitions:
release testing of coated drug-eluting vascular stent systems
3.1.1 mock vessel—physicalsimulationofthevasculatureat
although some sections may be appropriate.
the intended clinical deployment site.
1.3 Chronic particulate characterization and coating inspec-
3.1.2 stent system—a system comprised of a vascular stent
tion are not included herein.
and its delivery system.
3.1.3 tracking—navigation of a guide wire, guide catheter,
1.4 Coating systems specifically designed to degrade or
and/or stent system through either actual or simulated vascular
otherwise intentionally separate themselves from the perma-
anatomy.
nent stent structure may not be fully addressed herein.
3.1.4 tracking fixture—a model that simulates or replicates
1.5 The values stated in SI units are to be regarded as
thegeometryofarepresentativevasculaturethroughwhichthe
standard. No other units of measurement are included in this
stent system will be passed.
standard.
3.2 Definitions of Terms Specific to This Standard:
1.6 The values stated in inch-pound units are to be regarded
3.2.1 acute—a test timeframe intended to include stent
as standard. The values given in parentheses are mathematical
delivery and deployment beginning with the initial insertion of
conversions to SI units that are provided for information only
stent system until full removal of the delivery system and its
and are not considered standard.
accessory devices.
1.7 This international standard was developed in accor- 3.2.2 baseline—coating inspection and acute particulate
characterizationafterstentexpansiontothedesireddiameterin
dance with internationally recognized principles on standard-
an unconstrained environment and without tracking.
ization established in the Decision on Principles for the
Development of International Standards, Guides and Recom-
3.2.3 chronic—a test timeframe intended to mimic the
mendations issued by the World Trade Organization Technical
implantation time after full removal of the delivery system and
Barriers to Trade (TBT) Committee.
its accessory devices.
1 2
This guide is under the jurisdiction of ASTM Committee F04 on Medical and Available from U.S. Pharmacopeia (USP), 12601Twinbrook Pkwy., Rockville,
Surgical Materials and Devices and is the direct responsibility of Subcommittee MD 20852-1790, http://www.usp.org.
F04.30 on Cardiovascular Standards. Available from Food and DrugAdministration (FDA), 10903 New Hampshire
Current edition approved Nov. 1, 2018. Published November 2018. Originally Ave., Silver Spring, MD 20993-0002, http://www.fda.gov.
approved in 2011. Last previous edition approved in 2011 as F2743–11. DOI: Available from Association for the Advancement of Medical Instrumentation
10.1520/F2743-11R18. (AAMI), 4301 North Fairfax Dr., Suite 301, Arlington, VA 22203-1633.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
F2743 − 11 (2018)
3.2.4 constrained environment—a deployment site in which 4.3.1 Particles released may be captured in a collection
the stent is deployed into a mock vessel. beaker and sampled for count/size using light obscuration or
filtration/microscopy.Theneedforpost-dilatation,overlapping
3.2.5 simulated use—coating inspection and acute particu-
or to limit self-expansion may require deployment into a mock
late characterization after tracking in simulated anatomy and
vessel, or
aqueous environment. It may also include deployment in bent
4.3.2 Particles released may be acquired and continuously
configuration, deployment in overlapped configuration, post-
counted in an apparatus (for example, tube) for facilitating
dilatation,orotherscenariosthatcanreasonablybeexpectedin
flow.
clinical use.
3.2.6 unconstrained environment—a deployment site in
5. Significance and Use
which the stent is not constrained by a mock vessel. Compare
to “Constrained Environment”.
5.1 The shedding of the coating from a vascular stent can
alter its clinical safety and/or therapeutic benefit. Clinical
4. Summary of Practice
performance (for example, drug elution) may be affected by
4.1 Test Sequence and Samples—Baseline and Simulated
particulategenerationfromthecoatedstentsystemandcoating
Use Testing are conducted as two separate tests. Coating
defects. This document provides guidance for coating inspec-
inspection and acute particulate characterization may be per-
tion and acute particulate characterization of drug eluting
formed as two separate tests with independent samples.
vascular stents. Information about the potential for shedding
4.2 Baseline Testing—Asingle stent is deployed to nominal can be gained during bench testing. The general guidelines
or maximum labeled diameter. The stent is expanded in an
presented here may be used for writing detailed protocols for
unconstrained environment so as to characterize the stent only. specific products at the various stages of the product develop-
Baseline testing includes coating inspection and acute particu-
ment process. Such testing may be performed during device
late characterization of the stent. Baseline coating inspection development, design validation testing, lot-release testing,
may be conducted after deployment in air or in an aqueous
and/or stability testing although different requirements may
unconstrained environment. Baseline acute particulate charac- apply at each stage. These suggested methods may represent a
terization should be conducted in an aqueous unconstrained
reasonable simulation of clinical usage. When establishing the
environment.Thesurfacesofthestentcoatingareinspectedfor coatinginspectionandacuteparticulatecharacterizationtesting
defects or other adverse attributes caused by this procedure.
conditions, the current clinical usage/practice (for example,
Cumulative particulates released are captured or continuously post-dilation, overlapping stents) and the instructions for use
monitored, counted and classified according to size ranges. (IFU), as applicable, should be considered. While methods for
4.2.1 Particles released may be captured in a receptacle and chronic particulate characterization and coating inspection
sampled for count/size using light obscuration or filtration/ have not been established, these suggested methods may be
microscopy, or helpful in the development of chronic methods. Testing in
4.2.2 Particles released may be acquired and continuously accordance with recommendations in this guide will generate
counted in an apparatus (for example, tube) for facilitating data that may lead to further improvements in the method and
flow. its validation, as well as possible advancements in device
design and performance. See also FDA Guidance for Industry
4.3 Simulated Use Testing—The stent system is tracked in
and FDA Staff and AAMITIR42:2010.
an aqueous environment, through an appropriately clean, in
vitromodelsimulatingthevascularanatomytobenavigatedto
6. Suggested Materials and Reagents
access the targeted clinical deployment site.Accessory devices
(for example, guidewires, guide catheters, and so forth) are
6.1 Baseline Testing:
utilized as indicated in the IFU. The stent is deployed either
6.1.1 Beaker.
singly or overlapped with another stent and bent configuration
6.1.2 Filtered (for example, 1.2 µm or finer), de-ionized or
to represent worst-case clinical condition, as appropriate. A
distilled water, in general accordance with USP<788>. Other
constrained environment should be used as the deployment
solutions may be used if justified.
location. Stents should be expanded in accordance with the
6.1.3 Heating system, capable of maintaining fluid tempera-
IFU, including expansion to post-dilatation limits, as appropri-
ture at 37 6 2ºC.
ate. Cumulative particulates released from the stent(s), stent
6.1.4 Particulate filter, 1.2 µm or finer, with appropriate
coating(s), stent system(s) and accessory devices (if used)
holder
during the procedure are captured or continuously monitored,
6.1.5 Particulate analyzer, capable of detecting and count-
counted and classified according to size ranges. Particulate
ing particulates in appropriate size ranges (for example,
characterization may be necessary to aid in classifying poten-
≥10µm).
tial particulate sources, and the test developer should under-
6.1.6 Calibration standards, for particulate sizing and
stand the constituents of the coated stent system. The surfaces
counting.
ofthestentcoating(s)areinspectedfordefectsorotheradverse
attributes caused by this procedure. Analysis of particulates 6.1.7 Analytical instrumentation for particulate character-
and surface inspection may be accomplished using the same ization [for example FTIR (Fourier transform infrared)
test articles subjected to tracking and deployment, if appropri- spectroscopy, Raman Spectroscopy, Scanning Electron Micro-
ate. scope (SEM) with Energy Dispersive Spectroscopy (EDAX),
F2743 − 11 (2018)
X-ray photoelectron spectroscopy (XPS) or Time-of-flight whichwillnotaffecttheintegrityofthestudy).Themeaningful
secondary ionization mass spectroscopy (TOF-SIMS)] (if uti- characterization of size and quantity of small particulates shed
lized). by the coated stent can be significantly impacted by environ-
6.1.8 Continuous flow particulate counting system (if uti- mental contamination. Likewise, contamination on the stent
lized): surface may be misinterpreted as coating defects or may mask
6.1.8.1 Apparatus (for example, tube) for facilitating flow actual defects. Poor experimental technique and handling of
and housing the test article in an unconstrained environment. accessory devices may also be significant sources of non
6.1.8.2 Pump for controlling fluid flow. coating particulates. Physical and chemical contamination, in
6.1.8.3 Continuous flow particulate counter. addition to particulates, may impact the results of this charac-
terization.
6.2 Simulated Use:
6.2.1 Filtered (for example, 1.2 µm or finer), de-ionized or 7.2 Stent Surface Inspection—For complete
distilled water, in general accordance with USP<788>. Other
characterization, inspection of the surface of the stent may be
solutions may be used if justified. performed at different time points (for example, before
6.2.2 Tracking fixture, (see 3.1 and 7.3).
expansion,afterexpansiontothenominalormaximumlabeled
6.2.3 Heating system, capable of maintaining fluid tempera- diameter, and after simulated use). Representative photos
ture at 37 6 2ºC.
should be provided for each step and region, as described
6.2.4 Mock vessel, (see 3.1 and 7.4). further in Section 8. The location of the photographed regions
6.2.5 Continuous flow particulate counting system: should be predetermined.Alower magnification photograph(s)
6.2.5.1 Apparatus (for example, tube) for facilitating flow of the stent that includes and identifies the pre-specified
and housing the test article in a constrained environment. locations should also be provided. The “before expansion”
6.2.5.2 Pump for controlling fluid flow. inspection of the stent may be performed prior to or after the
6.2.5.3 Continuous flow particulate counter. stent is mounted on a delivery system; however, stent surface
6.2.6 Collection Beaker, (optional). inspections made prior to stent system assembly (for example,
6.2.7 Particulate filter 1.2 µm or finer, with appropriate crimping/loading) can make identifying the source of damage
holder (if utilized). (for example, crimping/loading or tracking and deployment)
6.2.8 Particulate analyzer, capable of detecting and count- difficult. Handling during the initial inspection may introduce
ing particulates in appropriate size ranges (for example, particulatesandcontamination.Inspectionofthestentmounted
≥10µm). on/in the delivery system may be useful for assessing initial
6.2.9 Calibration standards for particulate sizing and count- manufacturing quality and/or for establishing a baseline for
ing. determining when during the subsequent tracking/deployment
6.2.10 Accessory devices per IFU (for example, guide
process coating damage or particulate shedding may be occur-
catheter, guidewire, post-dilatation balloon catheter, and so ring. Individual defects may be assessed throughout usage, if
forth).
appropriate (for example, for investigative purposes). A stent
6.2.11 Analytical instrumentation for particulate character- may be inspected on all surfaces prior to loading onto the
ization [for example FTIR (Fourier transform infrared)
delivery system. Self-expanding stents are usually covered by
spectroscopy, Raman Spectroscopy, Scanning Electron Micro- an opaque sheath and may not be amenable to inspection after
scope (SEM) with Energy Dispersive Spectroscopy (EDAX),
loading onto the delivery system.
X-ray photoelectron spectroscopy (XPS) or Time-of-flight 7.2.1 Summary of inspection steps which may be per-
secondary ionization mass spectroscopy (TOF-SIMS)] (if uti-
formed:
lized). 7.2.1.1 Before stent loading (if applicable).
7.2.1.2 Before expansion.
6.3 Coating Inspection (Baseline and Simulated Use), Op-
7.2.1.3 After baseline expansion to nominal or maximum
tical microscope with appropriate lighting and camera and/or
diameter.
SEM.
7.2.1.4 After simulated use.
6.4 Test Articles—Unless otherwise justified, all samples
7.2.2 Inspections of stent surfaces may be performed by
selected for testing should be clinical or commercial quality
optical (light) microscopy, scanning electron microscopy
products(forexample,completestentsystems).Environmental
(SEM), fluorescence microscopy, Raman spectroscopy, and so
conditions(forexample,aging,shipping,storage,andsoforth)
forth. Each technique offers advantages and disadvantages:
may affect the stent system and should be considered when
7.2.2.1 Optical (Light) Microscopy—See Table 1.
assessingcoatinginspectionandacuteparticulatecharacteriza-
7.2.2.2 Scanning Electron Microscopy—See Table 2.
tion. Post-coating activities (for example, crimping,
7.2.2.3 Fluorescence Microscopy—See Table 3.
sterilization, balloon expansion) are critical for coating inspec-
7.2.2.4 Raman Spectroscopy—See Table 4.
tion and acute particulate characterization.Asufficient number
7.2.3 Examples of commonly observed surface anomalies
ofspecimensshouldbetestedtosupportanyclaimstobemade
and defects are shown in Appendix X1. See also
based on the test results.
AAMITIR42:2010.
7. Test Method Considerations
7.3 Simulated Use Tracking—Tracking during simulated
7.1 Environment—It is extremely important that all proce- use should be through a model tortuous path that simulates the
dures be performed in a controlled environment (that is, one geometry of a typical severe anatomic path through which the
F2743 − 11 (2018)
TABLE 1 Advantages and Disadvantages of Optical (Light) Microscopy
Advantages Disadvantages
A. Lower magnification (typically <200×) speeds coarse inspection A. Limited resolution
B. May allow inspection of stent system B. Smaller depth of field (focus depth) as compared to SEM
C. No stent size limitations C. Light reflections may mask features
D. Non-destructive
E. Color differentiation (if applicable)
TABLE 2 Advantages and Disadvantages of Scanning Electron Microscopy
Advantages Disadvantages
A. Greater depth of field (focus depth) compared to Optical Microscopy A. May be a destructive test
B. Higher magnification (up to 10 000× magnification) enhances view of small features B. May require sputter coating (except for environmental SEM),
precluding the use of a single test article for multi-time point
inspections
C. Slower manipulation compared to Optical Microscopy
D. Inspection of the full length stent may not be possible within the
chamber
E. Grey scale only
TABLE 3 Advantages and Disadvantages of Fluorescence Microscopy
Advantages Disadvantages
A. Signal-to-noise ratio can help discern different types of product in mixture A. High cost of capital equipment (that is, fluorescence microscope)
B. Non-destructive B. Not many materials have intrinsic fluorescence
TABLE 4 Advantages and Disadvantages of Raman Spectroscopy
Advantages Disadvantages
A. Can obtain spectroscopic data using aqueous solutions A. High cost of capital equipment (that is, Raman spectrometer or attachment)
B. Non-destructive B. Requires high sample concentration
C. Provides both chemical and structural information
stent system will be passed during clinical use. The tracking vessel.Ifusedthemockvesselshouldreasonablyrepresentthe
fixtureshouldincludeboththeportionofanatomyinwhichthe intended clinical deployment site. Critical features to be
stent system would be passed through the guide catheter and considered in selecting the appropriate mock vessel include
the portion through which it would be passed after exit from geometry (for example, inside diameter, length), mechanical
theguidecatheter,ifapplicable.Theappropriategeometrymay properties, ability to remove stent without coating damage,
be different for different intended deployment sites (for coefficient of friction of the material (for example,
example, coronary, carotid, or femoral arteries) or for different polyurethane, silicone, latex and native vessel). Any particu-
access points (for example, femoral or radial artery). Critical lates from the mock vessel will contribute to the total cumu-
features to be considered in selecting the appropriate tracking lative count unless their composition is characterized in order
fixture include lumen diameter, bend radii, bend reversals, to exclude them from the total count but should be included in
ability to clean and coefficient of friction of the tracking the test report. Means for extracting the deployed stent from
material (for example, polyurethane, silicone, Teflon, glass, the mock vessel without imposing additional damage to the
latex or native vessel). The tracking lumen for the guide coating will be necessary for the post-deployment inspection.
catheter path may be constructed of glass or other durable
7.5 Particulate Capture and Characterization:
materialswhichminimizebackgroundnoiseparticulatecounts.
7.5.1 Spiking and Recovery:
Ifjustifiable,surrogateancillarydevices(forexample,mandrel
7.5.1.1 Anappropriatespikingandrecoverystudyshouldbe
for guidewire) may be used to aid in minimizing background
performed on each test system for baseline and simu
...




Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...