Standard Practice for Sampling, Preservation and Mitigating Interferences in Water Samples for Analysis of Cyanide

SIGNIFICANCE AND USE
Cyanide is routinely analyzed in water samples, often to demonstrate regulatory compliance; however, improper sample collection or pretreatment can result in significant positive or negative bias potentially resulting in unnecessary permit violations or undetected cyanide releases into the environment.
SCOPE
1.1 This practice is applicable for the collection and preservation of water samples for the analysis of cyanide. This practice addresses the mitigation of known interferences prior to the analysis of cyanide. Responsibilities of field sampling personnel and the laboratory are indicated.
1.2 The sampling, preservation and mitigation of interference procedures described in this practice are recommended for the analysis of total cyanide, available cyanide, weak acid dissociable cyanide, and free cyanide by Test Methods D 2036, D 4282, D 4374, D 6888, D 6994, D 7237, D 7284, and D 7511. The information supplied in this practice can also be applied to other analytical methods for cyanide, for example, EPA Method 335.4.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
30-Sep-2009
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM D7365-09a - Standard Practice for Sampling, Preservation and Mitigating Interferences in Water Samples for Analysis of Cyanide
English language
11 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM D7365-09a - Standard Practice for Sampling, Preservation and Mitigating Interferences in Water Samples for Analysis of Cyanide
English language
11 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: D7365 − 09a
StandardPractice for
Sampling, Preservation and Mitigating Interferences in
1
Water Samples for Analysis of Cyanide
This standard is issued under the fixed designation D7365; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope D3856 Guide for Management Systems in Laboratories
Engaged in Analysis of Water
1.1 This practice is applicable for the collection and pres-
D4282 Test Method for Determination of Free Cyanide in
ervation of water samples for the analysis of cyanide. This
Water and Wastewater by Microdiffusion
practice addresses the mitigation of known interferences prior
D4374 Test Methods for Cyanides in Water—Automated
to the analysis of cyanide. Responsibilities of field sampling
Methods for Total Cyanide, Weak Acid Dissociable
personnel and the laboratory are indicated.
3
Cyanide, and Thiocyanate (Withdrawn 2012)
1.2 The sampling, preservation and mitigation of interfer-
D4411 Guide for Sampling Fluvial Sediment in Motion
ence procedures described in this practice are recommended
D4840 Guide for Sample Chain-of-Custody Procedures
for the analysis of total cyanide, available cyanide, weak acid
D4841 Practice for Estimation of Holding Time for Water
dissociable cyanide, and free cyanide by Test Methods D2036,
Samples Containing Organic and Inorganic Constituents
D4282, D4374, D6888, D6994, D7237, D7284, and D7511.
D5847 Practice for Writing Quality Control Specifications
The information supplied in this practice can also be applied to
for Standard Test Methods for Water Analysis
other analytical methods for cyanide, for example, EPA
Method 335.4. D6888 Test Method for Available Cyanide with Ligand
Displacement and Flow InjectionAnalysis (FIA) Utilizing
1.3 The values stated in SI units are to be regarded as
Gas Diffusion Separation and Amperometric Detection
standard. No other units of measurement are included in this
D6994 Test Method for Determination of Metal Cyanide
standard.
Complexes in Wastewater, Surface Water, Groundwater
1.4 This standard does not purport to address all of the
and Drinking Water Using Anion Exchange Chromatog-
safety concerns, if any, associated with its use. It is the
raphy with UV Detection
responsibility of the user of this standard to establish appro-
D6696 Guide for Understanding Cyanide Species
priate safety and health practices and determine the applica-
D7237 Test Method for Free Cyanide with Flow Injection
bility of regulatory limitations prior to use.
Analysis (FIA) Utilizing Gas Diffusion Separation and
Amperometric Detection
2. Referenced Documents
D7284 Test Method for Total Cyanide in Water by Micro
2
2.1 ASTM Standards:
Distillation followed by Flow InjectionAnalysis with Gas
D1129 Terminology Relating to Water
Diffusion Separation and Amperometric Detection
D1193 Specification for Reagent Water
D7511 Test Method for Total Cyanide by Segmented Flow
D2036 Test Methods for Cyanides in Water
InjectionAnalysis, In-Line Ultraviolet Digestion andAm-
D3370 Practices for Sampling Water from Closed Conduits
perometric Detection
D3694 Practices for Preparation of Sample Containers and
4
for Preservation of Organic Constituents
2.2 U.S. EPA Methods:
EPA OIA-1677
EPA Method 335.2
1
This practice is under the jurisdiction ofASTM Committee D19 on Water and
EPA Method 335.4
is the direct responsibility of Subcommittee D19.06 on Methods for Analysis for
Organic Substances in Water.
Current edition approved Oct. 1, 2009. Published October 2009. Originally
approved in 2007. Last previous edition approved in 2009 as D7365 – 09. DOI:
3
10.1520/D7365-09A. The last approved version of this historical standard is referenced on
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or www.astm.org.
4
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Available from United States Environmental Protection Agency (EPA), Ariel
Standards volume information, refer to the standard’s Document Summary page on Rios Bldg., 1200 Pennsylvania Ave., NW, Washington, DC 20460, http://
the ASTM website. www.epa.gov.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
D7365 − 09a
5
2.3 USGS Methods: 6.6 pH Indicator Test Strips—pH indicator test strips ca-
USGS I-3300-85 pable of changing color at 0.5 pH units in the range of pH 2 to
USGS I-4302-85 12. More than one test strip may be necessary to cover this
range.
3. Terminology
6.7 Sodium Hydroxide Solution (1 M)—In a 1-L volumetric
3.1 Definitions:
fla
...

This document is not anASTM standard and is intended only to provide the user of anASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation:D7365–09 Designation:D7365–09a
Standard Practice for
Sampling, Preservation and Mitigating Interferences in
1
Water Samples for Analysis of Cyanide
This standard is issued under the fixed designation D 7365; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This practice is applicable for the collection and preservation of water samples for the analysis of cyanide. This practice
addresses the mitigation of known interferences prior to the analysis of cyanide. Responsibilities of field sampling personnel and
the laboratory are indicated.
1.2 The sampling, preservation and mitigation of interference procedures described in this practice are recommended for the
analysis of total cyanide, available cyanide, weak acid dissociable cyanide, and free cyanide by Test Methods D 2036, D 4282,
D 4374, D 6888, D 6994, D 7237, D 7284, and D 7511. The information supplied in this practice can also be applied to other
analytical methods for cyanide, for example, EPA Method 335.4.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
2
2.1 ASTM Standards:
D 1129 Terminology Relating to Water
D 1193 Specification for Reagent Water
D 2036 Test Methods for Cyanides in Water
D 3370 Practices for Sampling Water from Closed Conduits
D 3694 Practices for Preparation of Sample Containers and for Preservation of Organic Constituents
D 3856 Guide for Good Laboratory Practices in Laboratories Engaged in Sampling and Analysis of Water
D 4282 Test Method for Determination of Free Cyanide in Water and Wastewater by Microdiffusion
D 4374 Test Methods for Cyanides in WaterAutomated Methods for Total Cyanide, Weak Acid Dissociable Cyanide, and
Thiocyanate
D 4411 Guide for Sampling Fluvial Sediment in Motion
D 4840 Guide for Sample Chain-of-Custody Procedures
D 4841 Practice for Estimation of Holding Time for Water Samples Containing Organic and Inorganic Constituents
D 5847 Practice for Writing Quality Control Specifications for Standard Test Methods for Water Analysis
D 6888 Test Method for Available Cyanide with Ligand Displacement and Flow Injection Analysis (FIA) Utilizing Gas
Diffusion Separation and Amperometric Detection
D 6994 Test Method for Determination of Metal Cyanide Complexes inWastewater, SurfaceWater, Groundwater and Drinking
Water Using Anion Exchange Chromatography with UV Detection
D 6696 Guide for Understanding Cyanide Species
D 7237 Test Method for Aquatic Free Cyanide with Flow Injection Analysis (FIA) Utilizing Gas Diffusion Separation and
Amperometric Detection
D 7284 Test Method for Total Cyanide in Water by Micro Distillation followed by Flow InjectionAnalysis with Gas Diffusion
Separation and Amperometric Detection
D 7511 Total Cyanide by Segmented Flow Injection Analysis, In-Line Ultraviolet Digestion and Amperometric Detection
1
This practice is under the jurisdiction ofASTM Committee D19 on Water and is the direct responsibility of Subcommittee D19.06 on Methods forAnalysis for Organic
Substances in Water.
Current edition approved Feb. 15,Oct. 1, 2009. Published MarchOctober 2009. Originally approved in 2007. Last previous edition approved in 20072009 as D 7365 – 079.
2
For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1

---------------------- Page: 1 ----------------------
D7365–09a
3
2.2 U.S. EPA Methods:
EPA OIA-1677
EPA Method 335.2
EPA Method 335.4
2.3 APHA Standard:
Standard Methods 4500-CN Methods C, D, E, F, G, and I
4
2.4 USGS Methods:
USGS I-3300-85
USGS I-4302-85
3. Terminology
3.1 Definitions:
Definitions: For definitions of terms used in this practice, refer to Terminology D 1129 and Guide D 6696.
3.2In this practice,
...