ISO 21291:2019
(Main)Traditional Chinese medicine -Therapeutic fumigation devices
Traditional Chinese medicine -Therapeutic fumigation devices
This document specifies the requirements and test methods for therapeutic fumigation devices, hereafter referred to as the device. This document does not apply to decoction pieces.
Médecine traditionnelle chinoise — Dispositifs de fumigation thérapeutique
General Information
- Status
- Published
- Publication Date
- 26-May-2019
- Technical Committee
- ISO/TC 249 - Traditional Chinese medicine
- Current Stage
- 9093 - International Standard confirmed
- Start Date
- 13-Jan-2025
- Completion Date
- 13-Dec-2025
Overview
ISO 21291:2019 - Traditional Chinese medicine - Therapeutic fumigation devices defines safety, performance and test requirements for therapeutic fumigation devices used in Traditional Chinese Medicine (TCM) steam therapies. The standard covers device classification (open vs closed; local vs whole‑body), required components (heating unit, decoction container, steam pathway, control and sensing units, treatment capsule/support for closed devices) and the test methods to verify conformity. It explicitly excludes decoction pieces themselves.
Key topics and technical requirements
- Device classification & components
- Open devices (exposed steam) and closed devices (treatment capsule).
- Components: heating unit, decoction container, steam outlet, temperature‑sensing elements (skin or capsule), control unit, steam pathway and patient support.
- Temperature protection
- Over‑temperature protection at steam outlet and at fumigated body areas: heating must stop automatically and alarm (auditory/visual) per IEC 60601‑1‑8.
- Open device body area limit and closed capsule limit: exceed 45 °C triggers automatic stop and alarm.
- Temperature control and display
- Set function: error between set and actual steam temperature ≤ ±2 °C.
- Real‑time display: sensor display error ≤ ±1 °C (skin or capsule sensors).
- Closed‑device temperature consistency within capsule: uniformity error ≤ ±3 °C across measurement points.
- Timing, safety and mechanical tests
- Treatment timing accuracy: ±30 s.
- Emergency stop: manual, non‑self‑resetting actuator (red, visible, accessible).
- Empty‑container heating prevention: automatic stop and alert.
- Patient support mechanical strength: withstand specified mass loading (135 kg unless otherwise specified).
- Instructions & biological safety
- Instructions must state decoction requirements, contraindications and supervisory use guidance.
- Normative references include ISO 10993‑1 (biocompatibility) and IEC 60601 series for electrical and electromagnetic safety and alarm performance.
- Test methods
- Typical tests use constant‑temperature tanks and specified measurement points and stabilization times to verify alarms, sensors and temperature consistency.
Applications and who uses this standard
- Manufacturers of TCM therapeutic fumigation devices - to design compliant products and prepare technical documentation and testing programs.
- Test laboratories and conformity assessors - for executing the specified test methods and validating safety/performance claims.
- Healthcare providers and clinics - to understand device safety features, operational limits and supervision requirements for clinical use.
- Regulators and procurement officers - to set acquisition specifications and evaluate compliance for medical/therapeutic steam devices.
Related standards
- ISO 10993‑1 (biological evaluation of medical devices)
- IEC 60601‑1 and collateral standards (electrical safety, EMC, alarms)
- IEC 60601‑2‑52 (medical beds - relevant for whole‑body capsule support)
Keywords: ISO 21291:2019, therapeutic fumigation devices, Traditional Chinese Medicine, TCM steam therapy, temperature protection, device testing, medical device safety.
Frequently Asked Questions
ISO 21291:2019 is a standard published by the International Organization for Standardization (ISO). Its full title is "Traditional Chinese medicine -Therapeutic fumigation devices". This standard covers: This document specifies the requirements and test methods for therapeutic fumigation devices, hereafter referred to as the device. This document does not apply to decoction pieces.
This document specifies the requirements and test methods for therapeutic fumigation devices, hereafter referred to as the device. This document does not apply to decoction pieces.
ISO 21291:2019 is classified under the following ICS (International Classification for Standards) categories: 11.040.60 - Therapy equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
You can purchase ISO 21291:2019 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of ISO standards.
Standards Content (Sample)
INTERNATIONAL ISO
STANDARD 21291
First edition
2019-05
Traditional Chinese medicine —
Therapeutic fumigation devices
Médecine traditionnelle chinoise — Dispositifs de fumigation
thérapeutique
Reference number
©
ISO 2019
© ISO 2019
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting
on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address
below or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Fax: +41 22 749 09 47
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii © ISO 2019 – All rights reserved
Contents Page
Foreword .iv
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Classification and components . 2
4.1 Classification . 2
4.2 Components . 2
5 Technological requirements and test methods . 5
5.1 Temperature protection . 5
5.1.1 Temperature protection at steam outlet . 5
5.1.2 Temperature protection for fumigated body areas . 6
5.2 Temperature set function . 6
5.3 Temperature display function . 6
5.3.1 Open devices — skin temperature-sensing unit . 6
5.3.2 Closed devices — capsule temperature-sensing unit . 6
5.4 Temperature consistency in steam capsule . 7
5.5 Treatment timing . 7
5.6 Emergency stop function . 7
5.7 Heating prevention against empty decoction container . 7
5.8 Mechanical strength of patient support system . 7
5.9 Instructions for device use . 8
5.10 Electrical safety . 8
5.11 Electromagnet disturbances . 8
5.12 Biological evaluation . 8
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www .iso .org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www .iso .org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to the
World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT) see www .iso
.org/iso/foreword .html.
This document was prepared by Technical Committee ISO/TC 249, Traditional Chinese medicine.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www .iso .org/members .html.
iv © ISO 2019 – All rights reserved
INTERNATIONAL STANDARD ISO 21291:2019(E)
Traditional Chinese medicine — Therapeutic fumigation
devices
1 Scope
This document specifies the requirements and test methods for therapeutic fumigation devices,
hereafter referred to as the device.
This document does not apply to decoction pieces.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 10993-1, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk
management process
IEC 60601-1:2005/AMD1: 2012, Medical electrical equipment — Part 1: General requirements for basic
safety and essential performance
IEC 60601-1-2, Medical electrical equipment — Part 1-2: General requirements for basic safety and essential
performance — Collateral standard: Electromagnetic disturbances — Requirements and tests
IEC 60601-2-52, Medical electrical equipment — Part 2-52: Particular requirements for the basic safety
and essential performance of medical beds
IEC 60601-1-8, Medical electrical equipment — Part 1-8: General requirements for basic safety and
essential performance — Collateral standard: General requirements, tests and guidance for alarm systems
in medical electrical equipment and medical electrical systems
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https: //www .iso .org/obp
— IECElectropedia: available at http: //www .electropedia .org/
3.1
therapeutic fumigation device
device that heats liquid decoction pieces to form steam
Note 1 to entry: The steam is used therapeutically.
3.2
closed therapeutic fumigation device
therapeutic fumigation device which uses an enclosed capsule for treatment
Note 1 to entry: During treatment, the steam and areas of the patient’s body undergoing therapeutic fumigation
are predominantly isolated from the surrounding environment.
3.3
open therapeutic fumigation device
therapeutic fumigation device which does not use an enclosed capsule for treatment
Note 1 to entry: During treatment, both the steam and areas of the patient’s body undergoing therapeutic
fumigation are exposed to the surrounding environment.
4 Classification and components
4.1 Classification
The device is classified according to the body area being treated and the fumigation method being
applied.
a) Local type and full-body type can be classified by body area being treated (see Figure 1 and
Figure 3).
b) Open therapeutic fumigation devices and closed therapeutic fumigation devices, hereinafter
referred to as open devices
...










Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...