ISO 18153:2003
(Main)In vitro diagnostic medical devices — Measurement of quantities in biological samples — Metrological traceability of values for catalytic concentration of enzymes assigned calibrators and control materials
In vitro diagnostic medical devices — Measurement of quantities in biological samples — Metrological traceability of values for catalytic concentration of enzymes assigned calibrators and control materials
ISO 18153:2003 specifies how to assure the metrological traceability of values assigned to calibrators and control materials intended to establish or verify trueness of measurement of the catalytic concentration of enzymes. The calibrators and control materials are those provided by the manufacturers as part of, or to be used together with, in vitro diagnostic medical devices. The following subjects are outside the scope of ISO 18153:2003: requirements for the design or selection of a reference measurement procedure; quantities involving mass of enzyme or immunoreactivity of enzymes; control materials that do not have an assigned value and are used only for assessing the precision of a measurement procedure, either its repeatability or reproducibility (precision control materials); control materials intended for intralaboratory quality control purposes and supplied with intervals of suggested acceptable values, each interval obtained by interlaboratory consensus with respect to one specified measurement procedure, and with limiting values that are not metrologically traceable; metrological traceability of routine results to the product calibrator and their relations to any medical discrimination limit; properties involving nominal and ordinal scales.
Dispositifs médicaux de diagnostic in vitro — Mesurage des grandeurs dans des échantillons d'origine biologique — Traçabilité métrologique des valeurs de concentration catalytique des enzymes attribuées aux agents d'étalonnage et aux matériaux de contrôle
General Information
Relations
Standards Content (Sample)
INTERNATIONAL ISO
STANDARD 18153
First edition
2003-08-15
In vitro diagnostic medical devices —
Measurement of quantities in biological
samples — Metrological traceability of
values for catalytic concentration of
enzymes assigned to calibrators and
control materials
Dispositifs médicaux de diagnostic in vitro — Mesurage des grandeurs
dans des échantillons d’origine biologique — Traçabilité métrologique
des valeurs de concentration catalytique des enzymes attribuées aux
agents d’étalonnage et aux matériaux de contrôle
Reference number
©
ISO 2003
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.
© ISO 2003
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2003 – All rights reserved
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 18153 was prepared by the European Committee for Standardization (CEN) in collaboration with
Technical Committee ISO/TC 212, Clinical laboratory testing and in vitro diagnostic test systems, in
accordance with the Agreement on technical cooperation between ISO and CEN (Vienna Agreement).
Throughout the text of this document, read ".this European Standard." to mean ".this International
Standard.".
For the purposes of this International Standard, the CEN annex regarding fulfilment of European Council
Directives has been removed.
Contents
page
Foreword.v
Introduction .vi
1 Scope .1
2 Normative references .1
3 Terms and definitions.1
4 Metrological traceability chain and calibration hierarchy .3
4.1 Principles.3
4.2 Structure .4
5 Validation of metrologically traceable calibration.6
5.1 Principles.6
5.2 Analytical specificity of measurement procedures.6
5.3 Commutability of calibrators .7
5.4 Commutability of control materials.7
Annex A (informative) List of IFCC primary reference measurement procedures .8
Annex B (informative) List of certified reference materials (CRM).9
Bibliography .10
iv © ISO 2003 – All rights reserved
Foreword
This document (EN ISO 18153:2003) has been prepared by Technical Committee CEN/TC 140 "In vitro diagnostic
medical devices", the secretariat of which is held by DIN, in collaboration with Technical Committee ISO/TC 212
"Clinical laboratory testing and in vitro diagnostic test systems".
This European Standard EN ISO 18153:2003 including the Amendment shall be given the status of a national
standard, either by publication of an identical text or by endorsement, at the latest by February 2004, and
conflicting national standards shall be withdrawn at the latest by February 2004.
This document has been prepared under a mandate given to CEN by the European Commission and the European
Free Trade Association, and supports essential requirements of EU Directive(s).
For relationship with EU Directive(s), see informative annex ZA, which is an integral part of this document.
The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC), the European Confederation of
Laboratory Medicine (ECLM), and the European Diagnostic Manufacturers Association (EDMA) have contributed to
its preparation.
This standard includes a Bibliography.
Annexes A and B are informative.
According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following
countries are bound to implement this European Standard: Austria, Belgium, Czech Republic, Denmark, Finland,
France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Luxembourg, Malta, Netherlands, Norway, Portugal,
Slovakia, Spain, Sweden, Switzerland and the United Kingdom.
Introduction
The Directive 98/79/EC on in vitro diagnostic medical devices requires that the metrologically traceability of values
assigned to calibrators and control materials be assured through available reference measurement materials and
reference measurement procedures of higher order. Following this concept, the European Standard
prEN ISO 17511 on "traceability" has been elaborated which describes a hierarchical order of measurement
procedures and calibration materials. The general rules expressed in that standard also apply to quantities
involving catalytic activity. Whenever possible, metrological traceability should be demonstrated to the SI unit which
forms the top of the calibration hierarchy.
For the measurement of the catalytic activity concentration of enzymes (hereafter called 'catalytic concentration'), a
hierarchy of calibrators and measurement procedures is described in the present standard. For enzyme
measurements, the definition of the derived coherent SI unit "mole per second cubic metre", given the special
name "katal per cubic metre" by the General Conference on Weights and Measures, is the top of the hierarchy
followed by a primary reference measurement procedure to which lower level measurement procedures,
calibrators, and control materials should be traced whenever possible.
Enzymes in blood or other biological fluids can be measured for diagnostic purposes in terms of their catalytic
concentrations. The analytical principle of the measurement of the catalytic rate of conversion of substrate has
considerable advantages of speed, low limit of detection, analytical specificity, and low cost. Results of catalytic
concentration measurements are only comparable if the enzyme activities are measured under the same
conditions. Therefore, an enzyme measurand cannot be described only by kind-of-quantity (e.g. catalytic
concentration), name of enzyme and of system, but requires also the specified measurement procedure and
especially the indicator component of the measured reaction. At the top of the calibration hierarchy, the
measurement procedure should be internationally agreed, e.g. 'creatine kinase measured by the conversion rate of
NADH in the IFCC reference measurement procedure'.
Thus, the primary reference measurement procedure is an integral part of the definition of the measurand and has
to be followed in all detail, e.g. as concerns:
� kind of substrate (where the specificity of the enzyme allows this to be varied) and its concentration,
� activators and their concentrations,
� direction of catalysed reaction,
� indicator component,
� buffer system and pH,
� temperature,
� pre-incubation time,
� material used for starting the reaction,
� lag time,
� reaction time.
vi © ISO 2003 – All rights reserved
The disadvantage of the procedure-dependence of the definition of the enzyme measurand and therefore of the
results of the measurements are well known: problems are caused in external quality assessment (EQA) and in
assessing the transferability of methods; a multiplicity of biological reference intervals exists with the consequent
risk of clinical misinterpretation of enzyme results. The standardization of routine enzyme measurements is
important to laboratory medicine, to improve the clinical utility and comparability of results through the elimination of
existing differences in biological reference intervals.
Two approaches can be considered:
a) the exclusive routine use of a recommended or standardized procedure for each enzyme;
b) calibration of one or more routine procedures by commutable enzyme calibration materials with values
assigned by a chosen reference measurement procedure.
The "recommended procedure" approach (a)) has been pursued vigorously for more than twenty years. It has had
considerable success in improving the quality and comparability of enzyme measurements and in discouraging the
use of analytically unsatisfactory procedures. However, the recommended-procedure-approach to standardization
appears to have reached the limits of its usefulness. Its disadvantages include: absence of a consensus of choice
among a number of differing recommendations; intentional or unintentional modification of recommended
procedures in routine use; unresponsiveness of recommended procedures to analytical and technical
improvement; and partly non-adaptability of recommended procedures to preferred automation. As a change in
routine enzyme procedures, whether recommended or not, inevitably entails a change of biological reference
values, it is understandably unwelcome to clinicians.
Improvement of the design and analytical performance of enzyme measurements will, and should, continue.
However, this should follow the normal practice of development and dissemination of scientific advances. Attempts
to develop and promote further standardized procedures for universal use are neither practicable nor desirable.
The "reference measurement procedure and calibration material" approach (b)) has, in contrast, received relatively
little attention. Among the objections that have been raised are:
1. lack of stable enzyme reference materials in appropriate matrices to serve as calibrators;
2. dissimilarity between candidate enzyme calibrators and the analyte enzymes in human samples, including
differences in isoforms;
3. absence of a constant inter-procedure ratio between a calibrating (reference) procedure and calibrated
(routine) procedure(s), for both the enzyme calibrator and patients' samples containing the analyte enzyme
(also described as a lack of commutability).
The converse of these objections constitutes a list of specifications, both for higher order enzyme reference
materials and for families of measurement procedures between which calibration is proposed. The calibrator should
be stable and have an analyte enzyme that is close in its catalytic properties within its matrix to those of the analyte
enzyme in the routine samples. The procedures themselves should have the same specificity for the catalytic
activity of the target enzyme.
Harmonization of the results of routine enzyme measurements can thus be achieved by selecting a reference
measurement procedure and identifying a family of related procedures for each clinically important enzyme.
Results obtained by any procedure included within such a family will be metrologically traceable to the chosen
reference measurement procedure.
1 Scope
This European Standard specifies how to assure the metrological traceability of values assigned to calibrators and
control materials intended to establish or verify trueness of measurement of the catalytic concentration of enzymes.
The calibrators and control materials are those provided by the manufacturers as part of, or to be used together
with, in vitro diagnostic medical devices.
The following subjects are outside the scope of this standard:
a) requirements for the design or selection of a reference measurement procedure;
b) quantities involving mass of enzyme or immunoreactivity of enzymes;
c) control materials that do not have an assigned value and are used only for assessing the precision of a
measurement procedure, either its repeatability or reproducibility (precision control materials);
d) control materials intended for intralaboratory quality control purposes and supplied with intervals of suggested
acceptable values, each interval obtained by interlaboratory consensus with respect to one specified
measurement procedure, and with limiting values that are not metrologically traceable;
e) metrological traceability of routine results to the product calibrator and their relations to any medical
discrimination limit;
f) properties involving nominal and ordinal scales.
2 Normative references
This European Standard incorporates by dated or undated reference, provisions from other publications. These
normative references are cited at the appropriate places in the text, and the publications are listed hereafter. For
dated references, subsequent amendments to or revisions of any of these publications apply to this European
Standard only when incorporated in it by amendment or revision. For undated references the latest edition of the
publication referred to applies (including amendments).
prEN ISO 17511, In vitro diagnostic medical devices - Measurement of quantities in biological samples -
Metrological traceability of values assigned to calibrators and control materials: (ISO/FDIS 17511:2002)
nd )
International Vocabulary of Basic and General Terms in Metrology, 2 edition, Geneva: ISO, 1993
) )
2 3
Guide to the Expression of Uncertainty in Measurement, 1st edition, Geneva:ISO, 1993
3 Terms and definitions
For the purposes of this European Standard, the following terms and definitions apply.
3.1
analyte
component indicated in the name of a measurable quantity
1) The abbreviation VIM:1993 is used in this standard
2) This monograph has been prepared simultaneously in English and French by a joint working group consisting of experts appointed by:
BIPM (International Bureau of Weights and Measures), IEC (International Electrotechnical Commission), IFCC (International Federation of
Clinical Chemistry and Laboratory Medicine), ISO (International Organization for Standardization), IUPAC (International Union of Pure and
Applied Chemistry), IUPAP (International Union of Pure and Applied Physics), OIML (International Organization of Legal Metrology)
3) The abbreviation GUM:1993 is used in this standard
EXAMPLE In the type of quantity "catalytic concentration of lactate dehydrogenase isoenzyme 1 in plasma", "lactate
dehydrogenase isoenzyme 1" is the analyte. The long phrase designates the measurand (see 3.5).
3.2
catalytic activity
z
E
property of a component corresponding to the catalysed substance rate of conversion of a specified chemical
reaction, in a specified measurement system
NOTE 1 Adapted from IUPAC/IFCC 1995:9.101.3.
NOTE 2 In this standard the "component" is an enzyme.
NOTE 3 The quantity "catalytic activity" relates to an amount of active enzyme, not its concentration, see 3.3.
-1
NOTE 4 The coherent derived SI unit is "katal" (kat), equal to "mole per second" (mol s ).
NOTE 5 The measurement procedure is an essential element of the definition of the measurand.
NOTE 6 In many instances, instead of the conversion rate of the substrate ascribed in the short name of the enzyme analyte,
e.g. "creatine" in "creatine kinase", the conversion rate of an indicator substance as substrate of a combined reaction is
measured. Then the measurand should be defined as 'catalytic activity of the enzyme as measured by the conversion rate of an
indicator substance in a specified system according to a given measurement procedure', e.g. 'catalytic activity of creatine kinase
+
as measured by the rate of conversion of NADP in the IFCC reference procedure in human serum'.
3.3
catalytic-activity concentration
catalytic concentration
b
E
catalytic activity of a component divided by volume of the original system
NOTE 1 Adapted from IUPAC/IFCC 1995:9.104.2.
-3
NOTE 2 The coherent derived SI unit is "katal per cubic metre" or "mole per second cubic metre" (kat m
-1 -3
= mol s m ). In laboratory medicine, the unit of volume can be chosen to be "litre" (l).
NOTE 3 In this standard the "component" is an enzyme and the "original system" can be, e.g., the plasma of a blood sample.
3.4
commutability of a material
closeness of agreement between the mathematical relationship of the measurement results obtained by two
measurement procedures for a stated quantity in a given material, and the mathematical relationship obtained for
the quantity in routine samples
3.5
measurand
particular quantity subject to measurement
[VIM:1993, 2.6]
NOTE See 3.1, Example.
3.6
metrological traceability
property of the result of a measurement or the value of a standard whereby it can be related to stated references,
usually national or international standards, through an unbroken chain of comparisons all having stated
uncertainties
[VIM:1993, 6.10]
NOTE 1 Each comparison is achieved by a (reference) measurement procedure defined in a calibration transfer protocol.
NOTE 2 There are several types of traceability. Therefore the term 'metrological traceability' is used in the present text.
2 © ISO 2003 – All rights reserved
4 Metrological traceability chain and calibration hierarchy
4.1 Principles
4.1.1 The nomenclature and basic principles of calibration and of metrological traceability of values obtained by
measurement of quantities in biological samples as given in prEN ISO 17511 shall also apply when the analyte is
an enzyme and the measurand is the derived kind-of-quantity "catalytic activity" (or a further derived kind-of-
quantity, e.g., "catalytic concentration" or "catalytic content"). A typical number of levels in a calibration hierarchy is
shown in Figure 1. The primary reference measurement procedure shall assign a value to a primary calibrator
which is used for calibration of the next lower measurement procedure and so on to the results obtained by the
end-user for a routine sample.
NOTE The term 'primary reference measurement procedure' as used here refers to a fully detailed set of measurement
instructions whereas the term 'primary method of measurement' as defined by the Consultative Committee for Amount of
Substance (CCQM) is a generic description of a measurement principle or a measurement method covering various procedures.
4.1.2 It is a prerequisite for the applicability of such a transfer protocol that the measurement procedures used
in descending order in the hierarchical scheme measure the same quantity. Therefore, it shall be demonstrated
that procedures subordinate
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...