EN IEC 61290-3-2:2026
(Main)Optical amplifiers - Test methods - Part 3-2: Noise figure parameters - Electrical spectrum analyzer method
- Status
- Not Published
- Publication Date
- 03-Dec-2026
- Technical Committee
- CLC/SR 86C - Fibre optic systems and active devices
- Current Stage
- 6055 - Ratification Completed (DOR) - Publishing
- Start Date
- 12-Nov-2026
- Due Date
- 12-Nov-2026
- Completion Date
- 12-Nov-2026
EN IEC 61290-3-2:2026 is the CLC standard for measuring noise figure parameters of optical amplifiers by the electrical spectrum analyzer (ESA) method. It applies to commercially available optical amplifiers (OAs), including optical fibre amplifiers (OFAs), semiconductor optical amplifiers (SOAs), and planar waveguide optical amplifiers (PWOAs). For test labs, manufacturers, and quality teams, it sets out the apparatus, calibration, measurement, calculation, and reporting steps needed to obtain total noise figure results in a repeatable way.
What does EN IEC 61290-3-2:2026 specify?
EN IEC 61290-3-2:2026 specifies how to measure the noise figure parameters of OAs using an ESA. The document is organized in clauses 1 to 9, covering scope, normative references, terms, general principles, apparatus, test specimen, procedure, calculation, and test results.
The standard is focused on accurate and reliable electrical noise measurement. It mentions an optical spectrum analysis approach as an alternative for other noise parameters, but that method is outside the scope of this document.
What are the key requirements of EN IEC 61290-3-2:2026?
EN IEC 61290-3-2:2026 uses direct electrical noise measurement to determine the total noise figure, including relevant noise contributions. This matters because it can be applied to all types of OAs, including devices where noise is not dominated only by amplified spontaneous emission.
- Clause 4 states that the method measures total noise figure directly and can be used for all OAs, including SOAs and Raman amplifiers. In practice, this tells the user when the ESA method is appropriate and when a different noise-parameter method may be needed.
- Clause 4 and Clause 5 set suggested operating values and performance limits for the source and receiver chain, including modulation, linewidth, RIN, attenuation, and detector behavior. These values are there to keep the measurement in a valid range and reduce uncertainty.
- Clause 5 requires a laser source with a single-line spectrum, sinusoidal modulation, an isolator, an input VOA, and a polarization controller. This lets the lab control input power and polarization without changing the source spectrum or noise behavior.
- Clause 5 also defines the receiver chain, including an output VOA, optical detector, electrical amplifier, and ESA with a suitable frequency range. In practice, this prevents the measurement chain from hiding the amplifier noise that is being measured.
- Clause 6 requires the OA to be tested at nominal operating conditions and calls for care with polarization state. That matters because polarization changes can alter input power and multiple path interference effects, which directly affect the measured noise figure.
- Clause 7 provides two routes: the frequency-scanning technique and the selected-frequency technique. Use frequency scanning when the OA noise variation is unknown or non-monotonic, and use selected frequency when the noise is essentially flat or falls monotonically with frequency.
- Clause 7.5 says good accuracy depends on the receiver thermal noise being at least 3 dB below shot noise, and on source RIN being below shot noise at the highest input power. This is a practical check on whether the setup is quiet enough to make a meaningful measurement.
- Clause 8 converts the measured data into noise factor, noise figure, and, where needed, MPI figure of merit and signal-spontaneous noise figure. This is the reporting step that turns raw ESA data into results that can be compared across devices and test houses.
What terms does EN IEC 61290-3-2:2026 define?
- Electrical spectrum analyzer (ESA) - the instrument used to measure the electrical noise power versus baseband frequency.
- Optical amplifier (OA) - the device under test, covering the amplifier types listed in the scope.
- Optical fibre amplifier (OFA) - an OA using optically pumped fibres, including rare-earth doped fibres and Raman-effect designs.
- Semiconductor optical amplifier (SOA) - an OA based on a semiconductor gain medium.
- Planar waveguide optical amplifier (PWOA) - an OA implemented in a planar waveguide structure.
- Relative intensity noise (RIN) - the source’s power fluctuation noise, expressed per hertz.
- Multiple path interference (MPI) - interference-related noise contribution that can make the noise figure vary with frequency.
- Variable optical attenuator (VOA) - an attenuator used to change optical power while keeping the source conditions controlled.
Who uses EN IEC 61290-3-2:2026?
EN IEC 61290-3-2:2026 is used by optical amplifier manufacturers, test laboratories, and quality assurance teams that need a standard ESA method for noise figure measurement. It is also relevant to engineers qualifying OFAs, SOAs, and PWOAs, and to organizations that need to compare amplifier performance using the same test method.
In practice, users apply it when setting up measurement equipment, calibrating the source and receiver path, checking measurement uncertainty, and reporting noise figure results for product development, verification, or incoming inspection.
What changed in EN IEC 61290-3-2:2026 from the previous edition?
EN IEC 61290-3-2:2026 is the third edition and replaces the second edition published in 2008. The foreword states these technical changes:
- Clause 1 scope shortened and clarified
- Clause 2 normative references updated
- Duplicate instructions removed from 7.2 for measuring time-average input power
- Unused symbols removed from 3.3 and one new symbol added
Which standards are used with EN IEC 61290-3-2:2026?
- IEC 61291-1 - the normative reference for generic optical amplifier specification and the terms and definitions used by this standard.
- IEC 61290-3 - the companion part that gives the general aspects of noise figure test methods.
- IEC TR 61292-2 - the technical background for the noise figure calculations used in Clause 8.
- IEC 60728-6 - provides a more detailed description of the self-heterodyning method used for linewidth measurement.
What does the EN IEC 61290-3-2:2026 document contain?
EN IEC 61290-3-2:2026 includes a schematic of a possible measurement set-up in Figure 1, then gives the full measurement flow in Clauses 7 and 8. Clause 7 covers calibration and measurement by both the frequency-scanning technique and the selected-frequency technique. Clause 8 gives the calculation steps for calibration results, test results, MPI figure of merit, and optional signal-spontaneous noise figure.
Clause 9 lists the information that should be reported, including the test arrangement, wavelength, source linewidth, RIN, modulation conditions, input power, ESA resolution bandwidth, and the resulting noise figure values.
Relations
- Effective Date
- 17-Nov-2026
- Effective Date
- 13-May-2025
Get Certified
Connect with accredited certification bodies for this standard

ANCE
Mexican certification and testing association.

Intertek Slovenia
Intertek testing, inspection, and certification services in Slovenia.

LNE (Laboratoire National de Métrologie et d'Essais)
French national laboratory for metrology and testing.
Sponsored listings
Frequently Asked Questions
EN IEC 61290-3-2:2026 is a draft published by CLC. Its full title is "Optical amplifiers - Test methods - Part 3-2: Noise figure parameters - Electrical spectrum analyzer method". This standard covers: Optical amplifiers - Test methods - Part 3-2: Noise figure parameters - Electrical spectrum analyzer method
Optical amplifiers - Test methods - Part 3-2: Noise figure parameters - Electrical spectrum analyzer method
EN IEC 61290-3-2:2026 is classified under the following ICS (International Classification for Standards) categories: 33.180.30 - Optic amplifiers. The ICS classification helps identify the subject area and facilitates finding related standards.
EN IEC 61290-3-2:2026 has the following relationships with other standards: It is inter standard links to EN IEC 61291-1:2018, EN 61290-3-2:2008. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
EN IEC 61290-3-2:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- 1 Scope
- 2 Normative references
- 3 Terms, definitions, abbreviated terms, and symbols
- 3.1 Terms and definitions
- 3.2 Abbreviated terms
- 3.3 Symbols
- 4 General
- 5 Apparatus
- 10 MHz to 2 GHz (‡) and the relative modulation amplitude of the source module from 2 to
- 6 Test specimen
- 7 Procedure
- 7.1 General remark
- 7.2 Frequency-scanning technique: calibration
- 7.3 Frequency-scanning technique: measurement
- 7.4 Selected-frequency technique: calibration and measurement
- 7.5 Measurement accuracy limitations
- 8 Calculation
- 8.1 General
- 8.2 Calculation of calibration results
- 8.3 Calculation of test results for the frequency-scanning technique
- 8.4 Calculation of test results for the selected-frequency technique
- 9 Test results
- Bibliography
INTRODUCTION This part of IEC 61290 is devoted to noise figure measurements of optical amplifiers using the electrical spectrum analyzer method. The technology of optical amplifiers is still rapidly evolving, hence amendments and new additions to this standard can be expected. This document should be read in conjunction with IEC 61290-3 [1] and IEC 61291-1. Each symbol and abbreviation introduced in this document is generally explained in the text the first time it appears. However, for an easier understanding of the whole text, a list of all symbols and abbreviations used in this document is given in Clause 3.
oSIST prEN IEC 61290-3-2:2025
1 Scope
This part of IEC 61290 applies to all commercially available optical amplifiers (OAs), including OAs using optically pumped fibres (i.e. optical fibre amplifiers (OFAs) based on either rareearth doped fibres or on the Raman effect), semiconductor optical amplifiers (SOAs), and planar waveguide optical amplifiers (PWOAs).
This document establishes requirements for accurate and reliable measurements of the noise figure parameters of OAs by means of the electrical spectrum analyzer (ESA) method. The noise figure parameters of OAs are defined in IEC 61291-1.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
IEC 61291-1, Optical amplifiers - Part 1: Generic specification
3 Terms, definitions, abbreviated terms, and symbols
3.1 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 61291-1 apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- IEC Electropedia: available at https://www.electropedia.org/
- ISO Online browsing platform: available at https://www.iso.org/obp
3.2 Abbreviated terms
CW continuous wave DFB distributed feedback ESA electrical spectrum analyzer FWHM full width at half maximum MPI multiple path interference OA optical amplifier OFA optical fibre amplifier PWOA planar waveguide optical amplifier −1 RIN relative intensity noise (of the source, expressed in Hz ) RMS root mean square SOA semiconductor optical amplifier VOA variable optical attenuator
3.3 Symbols
B calibrated, noise equivalent ESA electrical bandwidth (not necessarily the e resolution bandwidth) c speed of light in vacuum e electron charge
oSIST prEN IEC 61290-3-2:2025
f baseband frequency F (total) noise factor F frequency-independent contribution to total noise factor non-mpi G OA optical signal gain h Planck's constant k optical power reduction factor (default k = 0,5); it can be obtained by taking the square root of the electrical power reduction factor v optical frequency = c/λ ∆v source FWHM linewidth with modulation on −1 2 H ,H (f) transfer function of receiver in watts (= S /∆P ) 0 0 esa in I multiple path interference figure of merit, the noise factor contribution caused mpi by multiple path interference integrated over all baseband frequencies (0 to infinity) I photodetector current pd λ wavelength in vacuum m relative modulation amplitude (the ratio of RMS optical power modulation amplitude to average optical power) NF(f) (total) noise figure NF signal-spontaneous noise figure sig-sp N (f) (frequency-dependent) ESA noise contribution caused by the laser relative rin,0 intensity noise, at calibration conditions N (frequency-independent) shot noise caused by the optical input power, at shot,0 calibration conditions, measured with ESA N thermal noise level as measured with ESA (optical input port of receiver thermal module closed) N (f) (frequency-dependent) noise power measured with ESA with input and output
VOA set to 0 dB, thermal noise level subtracted, without OA test device N '(f) (frequency-dependent) noise power measured with ESA with input VOA set to
3 dB (default) and output VOA set to 0 dB, thermal noise level subtracted, without OA test device N (f) frequency-dependent noise power, with OA inserted, thermal noise level
subtracted, measured with ESA P time-averaged optical input power = T P (with modulation on); optical in in in,0 power radiated from the end of the input jumper cable P time-averaged optical input power at 0 dB setting of input VOA (with in,0 modulation on) ∆P RMS optical power amplitude in,rms P total optical power radiated from the output port of the OA, including the ASE out r ,r (f) effective photodetector responsivity through output VOA at 0 dB setting 0 0 R (f) source relative intensity noise; generally, the square of the RMS optical power RIN,source fluctuation divided by the (baseband) bandwidth and the square of the CW power S electrical power of the modulation signal at T = 1, measured with ESA, 0 in without OA inserted S electrical power of the modulation signal, with OA inserted, measured with
ESA
oSIST prEN IEC 61290-3-2:2025
T transmission factor of input VOA relative to transmission at 0 dB setting, in expressed in linear units T transmission factor of output attenuator relative to transmission at 0 dB out setting, expressed in linear units
4 General
The present test method is based on direct electrical noise measurement, and it is directly related to its definition including all relevant noise contributions. Therefore, this method can be used for all types of optical amplifiers, including SOA and Raman amplifiers which can have significant contributions besides amplified spontaneous emission, because it measures the total noise figure. For further details of applicability, see the IEC 61290-3 series. An alternative test method based on optical spectrum analysis can be used, particularly for different noise parameters (such as the signal-spontaneous noise factor) but it is not included in the scope of this standard.
All numerical values followed by (‡) are suggested values for which the measurement is assured. Other values can be acceptable but should be verified. A measurement accuracy for the average noise factor of ±20% (‡), respectively ±1 dB, should be attainable with this method (see Clause 7).
NOTE General aspects of noise figure test methods are described in IEC 61290-3.
5 Apparatus
A schematic diagram of a possible implementation of the measurement set-up is shown in Figure 1.
The test equipment listed below, with the required characteristics, is needed. a) A source module with the following components.
1) A laser source with a single-line spectrum, for example, a distributed feedback (DFB) laser diode. The laser source shall be amplitude modulated with a sinusoidal signal of a single frequency that is higher than the linewidth of the laser source. A modulation frequency at least 3 times higher than the linewidth is recommended. The relative modulation amplitude, m (that is, the ratio of root mean square (RMS) of the optical power modulation amplitude to the average optical power) shall be small enough to ensure operation in the linear region. A value of m between 2 % and 10 %(‡) is considered adequate. Direct or external modulation of the laser source may be used. The average output power, P , of the laser source should be at least 0 dBm, to generate in,0 the desired OA saturation state.
The linewidth FWHM (full width at half maximum) under modulation shall be between 20 MHz(‡) and 100 MHz(‡). This is considered the best range for accurate determination of the noise contribution from multiple path interference, because it closely reflects the typical linewidths of DFB lasers, which is the typical laser source used in conjunction with OAs. A linewidth of 20 MHz is adequate for a minimum spacing of 7,5 m between the OA internal reflection points. Using narrower linewidths can lead to the undesired situation that the internal reflections in the OA interfere coherently and that substantially different noise figure results are obtained. A linewidth of more than 100 MHz will cause OA noise contributions at frequencies that are higher than the high end of the ESA bandwidth. The relative intensity noise (RIN) of the laser source shall be less than −150 dB/Hz(‡) within the frequency range of interest (for example, 10 MHz to 2 GHz). The spontaneous emission power, relative to the signal power, shall be less than −40 dB/nm(‡) to avoid large noise contributions from spontaneous-spontaneous mixing of the source spontaneous emission.
oSIST prEN IEC 61290-3-2:2025
2) A built-in or external isolator, so that external reflections have no influence on the laser spectrum and on the laser relative intensity noise. The isolator shall have an optical isolation of larger than 60 dB(‡). The reflectance at the isolator output port shall be less than −50 dB(‡).
3) An input VOA, providing >40 dB attenuation range with ±0,05 dB(‡) linearity, and external and internal reflectances of less than −50 dB(‡). This VOA is used to change the source output power without changing its spectrum, RIN, or state of polarization. The purpose of this VOA is to control the input power and to allow a distinction of shot noise from other noise sources during calibration. A simpler VOA with no linearity requirements may be used if the change in attenuation is measured with the electrical spectrum analyzer.
4) A polarization controller acts generating of all possible output polarization states from an arbitrary input polarization state. The optical output power on output polarization state shall be within ±0,01 dB(‡), and the reflectance shall be less than –50 dB(‡).
Figure 1 – Scheme of a measurement set-up b) A modulation source (i.e., signal generator) capable of generating the frequency range from
10 MHz to 2 GHz (‡) and the relative modulation amplitude of the source module from 2 to
10% (‡).
c) An optical power meter with the following capabilities:
...