ASTM E1249-15(2021)
(Practice)Standard Practice for Minimizing Dosimetry Errors in Radiation Hardness Testing of Silicon Electronic Devices Using Co-60 Sources
Standard Practice for Minimizing Dosimetry Errors in Radiation Hardness Testing of Silicon Electronic Devices Using Co-60 Sources
SIGNIFICANCE AND USE
4.1 Division of the Co-60 Hardness Testing into Five Parts:
4.1.1 The equilibrium absorbed dose shall be measured with a dosimeter, such as a TLD, located adjacent to the device under test. Alternatively, a dosimeter may be irradiated in the position of the device before or after irradiation of the device.
4.1.2 This absorbed dose measured by the dosimeter shall be converted to the equilibrium absorbed dose in the material of interest within the critical region within the device under test, for example the SiO2 gate oxide of an MOS device.
4.1.3 A correction for absorbed-dose enhancement effects shall be considered. This correction is dependent upon the photon energy that strikes the device under test.
4.1.4 A correlation should be made between the absorbed dose in the critical region (for example, the gate oxide mentioned in 4.1.2) and some electrically important effect (such as charge trapped at the Si/SiO2 interface as manifested by a shift in threshold voltage).
4.1.5 An extrapolation should then be made from the results of the test to the results that would be expected for the device under test under actual operating conditions.
Note 5: The parts of a test discussed in 4.1.2 and 4.1.3 are the subject of this practice. The subject of 4.1.1 is covered and referenced in other standards such as Practice E668 and ICRU Report 14. The parts of a test discussed in 4.1.4 and 4.1.5 are outside the scope of this practice.
4.2 Low-Energy Components in the Spectrum—Some of the primary Co-60 gamma rays (1.17 and 1.33 MeV) produce lower energy photons by Compton scattering within the Co-60 source structure, within materials that lie between the source and the device under test, and within materials that lie beyond the device but contribute to backscattering. As a result of the complexity of these effects, the photon energy spectrum striking the device usually is not well known. This point is further discussed in Section 5 and Appendix X1. The presence of ...
SCOPE
1.1 This practice covers recommended procedures for the use of dosimeters, such as thermoluminescent dosimeters (TLD's), to determine the absorbed dose in a region of interest within an electronic device irradiated using a Co-60 source. Co-60 sources are commonly used for the absorbed dose testing of silicon electronic devices.
Note 1: This absorbed-dose testing is sometimes called “total dose testing” to distinguish it from “dose rate testing.”
Note 2: The effects of ionizing radiation on some types of electronic devices may depend on both the absorbed dose and the absorbed dose rate; that is, the effects may be different if the device is irradiated to the same absorbed-dose level at different absorbed-dose rates. Absorbed-dose rate effects are not covered in this practice but should be considered in radiation hardness testing.
1.2 The principal potential error for the measurement of absorbed dose in electronic devices arises from non-equilibrium energy deposition effects in the vicinity of material interfaces.
1.3 Information is given about absorbed-dose enhancement effects in the vicinity of material interfaces. The sensitivity of such effects to low energy components in the Co-60 photon energy spectrum is emphasized.
1.4 A brief description is given of typical Co-60 sources with special emphasis on the presence of low energy components in the photon energy spectrum output from such sources.
1.5 Procedures are given for minimizing the low energy components of the photon energy spectrum from Co-60 sources, using filtration. The use of a filter box to achieve such filtration is recommended.
1.6 Information is given on absorbed-dose enhancement effects that are dependent on the device orientation with respect to the Co-60 source.
1.7 The use of spectrum filtration and appropriate device orientation provides a radiation environment whereby the absorbed dose in the sensitive region of an elect...
General Information
Relations
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E1249 − 15 (Reapproved 2021)
Standard Practice for
Minimizing Dosimetry Errors in Radiation Hardness Testing
of Silicon Electronic Devices Using Co-60 Sources
This standard is issued under the fixed designation E1249; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the U.S. Department of Defense.
1. Scope 1.7 The use of spectrum filtration and appropriate device
orientation provides a radiation environment whereby the
1.1 This practice covers recommended procedures for the
absorbed dose in the sensitive region of an electronic device
use of dosimeters, such as thermoluminescent dosimeters
can be calculated within defined error limits without detailed
(TLD’s),todetermine the absorbed dose in a regionofinterest
knowledge of either the device structure or of the photon
within an electronic device irradiated using a Co-60 source.
energyspectrumofthesource,andhence,withoutknowingthe
Co-60 sources are commonly used for the absorbed dose
details of the absorbed-dose enhancement effects.
testing of silicon electronic devices.
1.8 The recommendations of this practice are primarily
NOTE 1—This absorbed-dose testing is sometimes called “total dose
applicable to piece-part testing of electronic devices. Elec-
testing” to distinguish it from “dose rate testing.”
tronic circuit board and electronic system testing may intro-
NOTE 2—The effects of ionizing radiation on some types of electronic
devicesmaydependonboththeabsorbeddoseandtheabsorbeddoserate;
duce problems that are not adequately treated by the methods
that is, the effects may be different if the device is irradiated to the same
recommended here.
absorbed-dose level at different absorbed-dose rates. Absorbed-dose rate
1.9 This standard does not purport to address all of the
effects are not covered in this practice but should be considered in
radiation hardness testing. safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
1.2 The principal potential error for the measurement of
priate safety, health, and environmental practices and deter-
absorbed dose in electronic devices arises from non-
mine the applicability of regulatory limitations prior to use.
equilibriumenergydepositioneffectsinthevicinityofmaterial
1.10 This international standard was developed in accor-
interfaces.
dance with internationally recognized principles on standard-
1.3 Information is given about absorbed-dose enhancement
ization established in the Decision on Principles for the
effects in the vicinity of material interfaces. The sensitivity of
Development of International Standards, Guides and Recom-
such effects to low energy components in the Co-60 photon
mendations issued by the World Trade Organization Technical
energy spectrum is emphasized.
Barriers to Trade (TBT) Committee.
1.4 A brief description is given of typical Co-60 sources
with special emphasis on the presence of low energy compo-
2. Referenced Documents
nentsinthephotonenergy spectrum output from suchsources. 2
2.1 ASTM Standards:
1.5 Procedures are given for minimizing the low energy E170Terminology Relating to Radiation Measurements and
components of the photon energy spectrum from Co-60 Dosimetry
sources, using filtration.The use of a filter box to achieve such E666Practice for CalculatingAbsorbed Dose From Gamma
filtration is recommended. or X Radiation
E668Practice for Application of Thermoluminescence-
1.6 Information is given on absorbed-dose enhancement
Dosimetry (TLD) Systems for Determining Absorbed
effectsthataredependentonthedeviceorientationwithrespect
DoseinRadiation-HardnessTestingofElectronicDevices
to the Co-60 source.
E1250Test Method forApplication of Ionization Chambers
to Assess the Low Energy Gamma Component of
This practice is under the jurisdiction of ASTM Committee E10 on Nuclear
Technology and Applicationsand is the direct responsibility of Subcommittee
E10.07 on Radiation Dosimetry for Radiation Effects on Materials and Devices. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Feb. 1, 2021. Published February 2021. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 1988. Last previous edition approved in 2015 as E1249–15. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/E1249-15R21. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E1249 − 15 (2021)
Cobalt-60 Irradiators Used in Radiation-Hardness Testing dosimeter, or both, for the purpose of minimizing low energy
of Silicon Electronic Devices components of the incident photon energy spectrum.
2.2 International Commission on Radiation Units and Mea- 3.10 spectrum filter—material layer intercepting photons on
surements Reports: their path between the Co-60 source and the device under test.
ICRUReport14 Radiation Dosimetry: X-Rays andGamma Thepurposeofthefilteristoreducelowenergycomponentsof
Rays With Maximum Photon Energies Between 0.6 and
the photon energy spectrum.
50 MeV
3.11 spectrum hardening—process by which the fraction of
ICRUReport18SpecificationofHighActivityGamma-Ray
low energy components of the photon energy spectrum is
Sources
reduced.
3.12 spectrum softening—process by which the fraction of
3. Terminology
low energy components of the photon energy spectrum is
3.1 absorber—material that reduces the photon fluence rate
increased.
from a Co-60 source by any interaction mechanism.
4. Significance and Use
3.2 absorbed-dose enhancement—increase (or decrease) in
the absorbed dose (as compared to the equilibrium absorbed
4.1 Division of the Co-60 Hardness Testing into Five Parts:
dose) at a point in a material of interest. This can be expected
4.1.1 Theequilibriumabsorbeddoseshallbemeasuredwith
to occur near an interface with a material of higher or lower
a dosimeter, such as a TLD, located adjacent to the device
atomic number.
under test. Alternatively, a dosimeter may be irradiated in the
position of the device before or after irradiation of the device.
3.3 absorbed-dose enhancement factor— ratio of the ab-
sorbed dose at a point in a material of interest to the 4.1.2 This absorbed dose measured by the dosimeter shall
be converted to the equilibrium absorbed dose in the material
equilibrium absorbed dose in that same material.
of interest within the critical region within the device under
3.4 average absorbed dose—mass weighted mean of the
test, for example the SiO gate oxide of an MOS device.
absorbed dose over a region of interest.
4.1.3 A correction for absorbed-dose enhancement effects
3.5 average absorbed-dose enhancement factor—ratio of
shall be considered. This correction is dependent upon the
the average absorbed dose in a region of interest to the
photon energy that strikes the device under test.
equilibrium absorbed dose (1).
4.1.4 A correlation should be made between the absorbed
dose in the critical region (for example, the gate oxide
NOTE 3—For a description of the necessary conditions for measuring
equilibrium absorbed dose, see 6.3.1 and the term charged particle mentioned in 4.1.2) and some electrically important effect
equilibriuminTerminologyE170,whichprovidesdefinitionsanddescrip-
(such as charge trapped at the Si/SiO interface as manifested
tions of other applicable terms of this practice.
by a shift in threshold voltage).
3.6 beam trap—absorber that is designed to remove the 4.1.5 Anextrapolationshouldthenbemadefromtheresults
beam that has been transmitted through the device under test.
of the test to the results that would be expected for the device
Its purpose is to eliminate the scattering of the transmitted under test under actual operating conditions.
beam back into the device under test.
NOTE 5—The parts of a test discussed in 4.1.2 and 4.1.3 are the subject
3.7 clean spectrum—onethatisrelativelyfreeoflowenergy of this practice. The subject of 4.1.1 is covered and referenced in other
standards such as Practice E668 and ICRU Report 14. The parts of a test
componentsinthephotonenergyspectrum.Forexample,fora
discussed in 4.1.4 and 4.1.5 are outside the scope of this practice.
Co-60sourceanideallycleanspectrumwouldcontainonlythe
4.2 Low-Energy Components in the Spectrum—Some of the
primary 1.17 and 1.33 MeV photons of Co-60 decay.
primary Co-60 gamma rays (1.17 and 1.33 MeV) produce
3.8 equilibrium absorbed dose—absorbed dose at some
lower energy photons by Compton scattering within the Co-60
incremental volume within the material in which the condition
source structure, within materials that lie between the source
of charged particle equilibrium (the energies, number, and
and the device under test, and within materials that lie beyond
direction of charged particles induced by the radiation are
the device but contribute to backscattering. As a result of the
constant throughout the volume) exists (see Terminology
complexity of these effects, the photon energy spectrum
E170).
striking the device usually is not well known. This point is
NOTE 4—For practical purposes the equilibrium absorbed dose is the
further discussed in Section 5 and Appendix X1.The presence
absorbed dose value that exists in a material at a distance from any
of low-energy photons in the incident spectrum can result in
interface with another material, greater than the range of the maximum
dosimetry errors. This practice defines test procedures that
energy secondary electrons generated by the incident photons.
shouldminimizedosimetryerrorswithouttheneedtoknowthe
3.9 filter box—container, made of one or more layers of
spectrum. These recommended procedures are discussed in
different materials, surrounding a device under test or a
4.5, 4.6, Section 7, and Appendix X5.
4.3 Conversion to Equilibrium Absorbed Dose in the Device
Material—Theconversionfromthemeasuredabsorbeddosein
AvailablefromInternationalCommissiononRadiationUnits,7910Woodmont
thematerialofthedosimeter(suchastheCaF ofaTLD)tothe
Ave., Washington, DC 20014.
equivalentabsorbeddoseinthematerialofinterest(suchasthe
The boldface numbers in parentheses refer to the list of references appended to
this practice. SiO ofthegateoxideofadevice)isdependentontheincident
E1249 − 15 (2021)
photon energy spectrum. However, if the simplifying assump- Finally,shieldingmaterialsoftungsten,lead,concrete,orwater
tion is made that all incident photons have the energies of the are often present. Therefore, a significant fraction of the
primaryCo-60gammarays,thentheconversionfromabsorbed photons incident on the device under test are the result of
dose in the dosimeter to that in the device under test can be Compton scattering that produces low energy components in
made using tabulated values for the energy absorption coeffi- the source output photon energy spectrum (see ICRU Report
cients for the dosimeter and device materials. Where this 18 for additional discussion of gamma-ray sources).
simplification is appropriate, the error incurred by its use to
NOTE 6—As an example, the energy spectrum from even a relatively
determine equilibrium absorbed dose is usually less than 5%
clean Co-60 source has about 35% of its total number of photons with
(see 6.3).
energies of less than 1 MeV (see Ref (2) and Appendix X1).
4.4 Absorbed-Dose Enhancement Effects— If a higher
5.2 Evenforagivensource,aconsiderablevariabilityexists
atomicnumbermaterialliesadjacenttoaloweratomicnumber
in the output energy spectrum depending on the geometry and
material, the energy deposition in the region adjacent to the
positionofirradiation.Thespectrumatanypositionisaffected
interface is a complex function of the incident photon energy
by scattering from walls, floor, and ceiling and by scattering
spectrum, the material composition, and the spatial arrange-
from material located nearby.
mentofthesourceandabsorbers.Theabsorbeddosenearsuch
NOTE 7—A qualitative estimate of the spectrum hardness for a given
an interface cannot be adequately determined using the proce-
source can be obtained using Method E1250.
dure outlined in 4.3. Errors incurred by failure to account for
5.3 The following Co-60 source types are described briefly
these effects may, in unusual cases, exceed a factor of five.
andlistedintheorderofdecreasingrelativespectrumhardness
Becausemicroelectronicdevicescharacteristicallycontainlay-
under the most favorable conditions of irradiation.
ers of dissimilar materials with thicknesses of tens of
nanometres, absorbed-dose enhancement effects are a charac-
NOTE 8—Diagrams of typical sources, a nominal photon energy
teristic problem for irradiation of such devices (see 6.1 and
spectrum for each, and references are given in Appendix X1.
Appendix X2).
5.3.1 A teletherapy source is a completely shielded source
4.5 Minimizing Absorbed-Dose Enhancement Effects—
from which the photon output is confined to a beam that is
Under some circumstances, absorbed-dose enhancement ef-
usually collimated.The source output is typically directed into
fects can be minimized by hardening the spectrum. Hardening
a shielded room, but a shielded container, or box, is used in
isaccomplishedbytheuseofhighatomicnumberabsorbersto
some cases.
remove low energy components of the spectrum, and by
5.3.2 A room sourceisasourcecontainedinashieldedwell
minimizing the amount and proximity of low atomic number
fromwhichitismovedintoashieldedroombyremotecontrol.
material to reduce softening of the spectrum by Compton
Its position in the room relative to walls, floor, and ceiling and
scattering (see Sections 6 and 7).
otherscatteringmaterialdeterminestherelativehardnessofits
4.6 Limits of the Dosimetry Errors— To correct for effective photon energy spectrum. As a result, the photon
absorbed-dose enhancement by calculational methods would energy spectrum obtained in a room source can be relatively
require a knowledge of the incident photon energy spectrum hard or relatively soft as compared with other Co-60 sources.
and the detailed structure of the device under test. To measure
5.3.3 A water well sourceisacompletelyshieldedsourceat
absorbed-dose enhancement would require methods for simu-
a certain depth in a pool of water to which access for
lati
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.